Homogenizer (Inline / High Shear) – PQ Protocol

Performance Qualification Protocol for Inline/High Shear Homogenizer

Document Number: PQ-001

Version: 1.0

Effective Date: YYYY-MM-DD

Review Date: YYYY-MM-DD

Objective

The objective of this Performance Qualification (PQ) protocol is to ensure that the Inline/High Shear Homogenizer operates within the defined parameters to achieve uniform API dispersion in the production of suppositories and implants.

Scope

This protocol applies to the Inline/High Shear Homogenizer used in the production area for the manufacturing of suppositories and implants. It covers the validation of critical parameters affecting product quality.

Responsibilities

  • Validation Team: Responsible for executing the PQ protocol and documenting results.
  • Quality Assurance: Responsible for reviewing and approving the PQ documentation.
  • Production Team: Responsible for providing access to the equipment and supporting the validation activities.

Prerequisites

  • Completion of Installation Qualification (IQ) and Operational Qualification (OQ).
  • Availability of all necessary documentation, including User Requirement Specifications (URS) and Annex 11 compliance.
  • Training of personnel on the use of the homogenizer.

Equipment Description

The Inline/High Shear Homogenizer is designed to ensure uniform dispersion of active pharmaceutical ingredients (API) in a liquid medium. The equipment operates by applying high shear forces to the product, enhancing mixing and dispersion quality.

Test Plan

Test ID Procedure Acceptance Criteria Evidence
PQ-001 Verify speed settings at 1000, 2000, and 3000 RPM. Speed logs must match set points. PLC logs, calibration records.
PQ-002 Measure temperature during operation. Temperature should remain within 20°C – 30°C. Temperature logs, audit trail.
PQ-003 Assess shear force applied during operation. Shear force must be within specified range. Shear force logs, calibration records.
See also  Solution Preparation Tank – Equipment Validation SOP

Detailed Test Cases

Test Case: PQ-001

Objective: Verify the speed settings of the homogenizer.

Procedure: Set the homogenizer to 1000, 2000, and 3000 RPM. Record the speed from the PLC logs.

Acceptance Criteria: Recorded speeds must match the set points.

Evidence: PLC logs and confirmation from validation team.

Test Case: PQ-002

Objective: Measure the temperature during operation.

Procedure: Operate the homogenizer and monitor the temperature using the integrated sensors.

Acceptance Criteria: Temperature must remain within 20°C – 30°C.

Evidence: Temperature logs and audit trail documentation.

Test Case: PQ-003

Objective: Assess the shear force applied during operation.

Procedure: Operate the homogenizer and record the shear force readings from the PLC.

Acceptance Criteria: Shear force must be within the specified range.

Evidence: Shear force logs and calibration records.

Deviations

Any deviations from the acceptance criteria must be documented and assessed for impact on product quality. A corrective action plan should be initiated as necessary.

Approvals

Prepared by: ______________________ (Validation Team)

Approved by: ______________________ (Quality Assurance)

Date: ______________________

Data Integrity Checks

  • Regular audits of PLC logs to ensure accuracy and completeness.
  • Backup of all electronic records in accordance with data integrity policies.