CCIT System (Vacuum Decay/Helium/HVLD) – Traceability Matrix (URS ↔ Tests)

URS ID URS Statement Risk (H/M/L) DQ Test Ref IQ Test Ref OQ Test Ref PQ Test Ref Evidence/Record Result
URS-001 Verify the system can detect leaks within specified sensitivity. H DQ-001 IQ-001 OQ-001 PQ-001 Validation Report Pass
URS-002 Ensure the audit trail is maintained for all test results. M DQ-002 IQ-002 OQ-002 PQ-002 Audit Trail Log Pass
URS-003 Confirm system performance under varying environmental conditions. L DQ-003 IQ-003 OQ-003 PQ-003 Environmental Test Report Pass
URS-004 Verify the system’s ability to perform repeated tests without failure. M DQ-004 IQ-004 OQ-004 PQ-004 Test Cycle Report Pass
URS-005 Ensure user training is adequate for operating the system. L DQ-005 IQ-005 OQ-005 PQ-005 Training Records Pass