Device Inspection System (Vision/Microscopy) – OQ Protocol

Document Control

Protocol Number: OQ-DEV-INS-001

Version: 1.0

Effective Date: [Insert Date]

Review Date: [Insert Date]

Prepared By: [Name]

Approved By: [Name]

Operational Qualification Protocol for Device Inspection System

Objective: To validate the Device Inspection System for inspecting coating defects and dimensions of drug-eluting stents, ensuring compliance with established acceptance criteria.

Scope: This protocol applies to the Operational Qualification of the Device Inspection System used in the QC/R&D area for drug-eluting stents and coated devices.

Responsibilities:

  • Validation Team: Execute the OQ protocol and document results.
  • Quality Assurance: Review and approve the OQ protocol and results.
  • Equipment Owner: Ensure equipment is maintained and calibrated prior to OQ execution.

Prerequisites:

  • Device Inspection System is installed and operational.
  • Calibration of the Device Inspection System is completed and documented.
  • All personnel involved are trained on the use of the Device Inspection System.

Equipment Description:

The Device Inspection System (Vision/Microscopy) is designed to inspect the coating defects and dimensions of drug-eluting stents. It utilizes advanced imaging technology to ensure high detection accuracy.

Test ID Procedure Acceptance Criteria Evidence
OQ-01 Calibration Verification Calibration data must show accuracy within ±5% of known standards. Calibration report
OQ-02 Image Quality Assessment Images must be clear, with no distortion or artifacts. Sample images
OQ-03 Defect Detection Test Must correctly identify at least 95% of defects. Test results report
OQ-04 Data Integrity Check Data must be retrievable and unchanged post-analysis. Data integrity report
See also  Nasal Spray Filling Machine – Deviation Impact Assessment

Detailed Test Cases:

Test Case OQ-01: Calibration Verification

  • Procedure: Verify calibration using standard reference materials.
  • Expected Result: Calibration data shows accuracy within the specified limits.

Test Case OQ-02: Image Quality Assessment

  • Procedure: Capture images of the device under standard conditions.
  • Expected Result: Images are clear and free from distortions.

Test Case OQ-03: Defect Detection Test

  • Procedure: Test system on known defective and non-defective samples.
  • Expected Result: System identifies at least 95% of defects correctly.

Test Case OQ-04: Data Integrity Check

  • Procedure: Perform checks to ensure data is intact and retrievable.
  • Expected Result: Data remains unchanged and accessible.

Deviations:

Any deviations from the acceptance criteria must be documented and investigated. A corrective action plan must be developed and approved by the Quality Assurance team.

Approvals:

__________________________
Prepared By: [Name]
Date: [Insert Date]

__________________________
Approved By: [Name]
Date: [Insert Date]