Water System (PW/WFI) – Point of Use for LVP – PQ Protocol

Document Control Number: PQ-WS-001

Version: 1.0

Effective Date: [Insert Date]

Review Date: [Insert Date]

Performance Qualification Protocol for Water System (PW/WFI) – Point of Use for LVP

Meta Description: This document outlines the Performance Qualification protocol for the Water System used in LVP compounding, ensuring compliance with regulatory standards.

Tags: Performance Qualification, Equipment Validation, Water System, IV Infusions, Utilities

Objective

The objective of this Performance Qualification (PQ) protocol is to verify that the Water System (PW/WFI) for point of use in LVP manufacturing meets predefined specifications and operates consistently within the defined critical parameters.

Scope

This protocol applies to the Water System utilized for the preparation of water for compounding and cleaning in place (CIP) processes within the utilities area, impacting the quality of LVP/SVP products directly.

Responsibilities

  • Validation Team: Responsible for executing the PQ protocol and documenting results.
  • Quality Assurance: Responsible for reviewing and approving the PQ documentation.
  • Engineering: Responsible for maintaining the Water System and ensuring it operates within specifications.

Prerequisites

  • Completion of Installation Qualification (IQ) and Operational Qualification (OQ).
  • Availability of necessary equipment and materials for testing.
  • Training of personnel involved in the PQ process.

Equipment Description

The Water System (PW/WFI) serves as a point of use system for providing purified water for compounding LVP/SVP products. It is equipped with monitoring capabilities for critical parameters, including conductivity, total organic carbon (TOC), microbial limits, and temperature.

See also  Plunger Insertion Machine – IQ Protocol

Test Plan

Test ID Procedure Acceptance Criteria Evidence
TP-001 Conductivity Test Conductivity ≤ 1.3 µS/cm Conductivity Log
TP-002 TOC Analysis TOC ≤ 500 ppb TOC Report
TP-003 Microbial Testing No growth in 100 mL sample Microbial Test Report
TP-004 Temperature Monitoring Temperature maintained at 20-25°C Temperature Log

Detailed Test Cases

Test Case 1: Conductivity Test

Procedure: Measure the conductivity of the water at the point of use using a calibrated conductivity meter. Record the values and compare them against acceptance criteria.

Acceptance Criteria: Conductivity must be ≤ 1.3 µS/cm.

Evidence: Documented conductivity log.

Test Case 2: TOC Analysis

Procedure: Collect samples of water and analyze using a TOC analyzer. Ensure proper calibration and maintenance of the analyzer.

Acceptance Criteria: TOC must be ≤ 500 ppb.

Evidence: TOC analysis report.

Test Case 3: Microbial Testing

Procedure: Conduct microbial testing by incubating water samples in appropriate growth media.

Acceptance Criteria: No growth in 100 mL sample.

Evidence: Microbial test report.

Test Case 4: Temperature Monitoring

Procedure: Monitor the water temperature continuously and record the data.

Acceptance Criteria: Temperature must be maintained at 20-25°C.

Evidence: Documented temperature log.

Deviations

Any deviations from the acceptance criteria must be documented and investigated. Root cause analysis should be conducted, and corrective actions must be implemented to ensure compliance.

Approvals

Prepared by: ______________________ Date: ___________

Reviewed by: ______________________ Date: ___________

Approved by: ______________________ Date: ___________