Dynamic Light Scattering (DLS) Analyzer – DQ Protocol

Document ID: DQ-001

Version: 1.0

Effective Date: 2023-10-01

Reviewed By: [Name]

Approval Date: [Date]

Design Qualification Protocol for Dynamic Light Scattering (DLS) Analyzer

This document outlines the Design Qualification Protocol for the Dynamic Light Scattering (DLS) Analyzer used in QC/R&D for particle size and PDI measurement.

#EquipmentValidation #NDDS #DLSAnalyzer #QualityControl #ResearchAndDevelopment

Objective

The objective of this Design Qualification (DQ) protocol is to ensure that the Dynamic Light Scattering (DLS) Analyzer is properly designed and meets the specified requirements for particle size and PDI measurement.

Scope

This protocol applies to the installation and qualification of the DLS Analyzer used in the QC/R&D area for the measurement of critical quality attributes (CQA) and critical process parameters (CPP).

Responsibilities

  • Validation Team: Responsible for executing the DQ protocol and documenting results.
  • Quality Assurance: Responsible for reviewing and approving the DQ protocol.
  • Equipment Owner: Responsible for ensuring the DLS Analyzer is available and maintained.

Prerequisites

  • Availability of the DLS Analyzer.
  • Completion of User Requirement Specification (URS) as per Annex 11.
  • Training of personnel on DLS Analyzer operation.

Equipment Description

The Dynamic Light Scattering (DLS) Analyzer is an instrument designed for measuring particle size and polydispersity index (PDI) of samples in a liquid medium. It utilizes laser light scattering techniques to provide accurate size measurements.

Test Plan

Test ID Procedure Acceptance Criteria Evidence
DQ-01 Verify size accuracy Size measurements within ±5% of expected values Calibration report
DQ-02 Assess PDI repeatability PDI values within ±0.1 for repeat measurements Test run data
DQ-03 Check audit trail functionality Complete audit trail for all measurements System log review
See also  Lyophilizer (Freeze Dryer) - NDDS – Traceability Matrix (URS ↔ Tests)

Detailed Test Cases

Test Case DQ-01: Size Accuracy

Procedure: Measure size of standard particles and compare to expected sizes.

Acceptance Criteria: Size measurements must be within ±5% of expected values.

Evidence: Calibration report demonstrating compliance.

Test Case DQ-02: PDI Repeatability

Procedure: Conduct multiple measurements of a standard sample and calculate the PDI.

Acceptance Criteria: PDI values must be within ±0.1 for all measurements.

Evidence: Test run data with statistical analysis.

Test Case DQ-03: Audit Trail

Procedure: Review system logs to ensure all measurements are recorded.

Acceptance Criteria: Complete and accessible audit trail for all operations.

Evidence: System log review documentation.

Deviations

Any deviations from the acceptance criteria must be documented and justified. A corrective action plan should be developed for any critical failures observed during testing.

Approvals

Approved By: [Name, Title]

Date of Approval: [Date]