Conductivity Meter – DQ Protocol

Document Control:

Document Number: DQ-QC-001

Version: 1.0

Effective Date: YYYY-MM-DD

Reviewed By: [Name]

Approval Date: YYYY-MM-DD

Design Qualification Protocol for Conductivity Meter in QC Laboratory

Objective: To ensure that the Conductivity Meter used in the QC Laboratory meets all specified requirements and performs reliably for conductivity measurement.

Scope: This protocol applies to the Design Qualification of the Conductivity Meter situated in the QC Laboratory, addressing its criticality and impact on product quality.

Responsibilities:

  • Validation Team: Responsible for executing the DQ protocol and documenting results.
  • Quality Assurance: Ensures compliance with regulatory standards and reviews validation documentation.
  • Laboratory Personnel: Operates the Conductivity Meter and assists in testing.

Prerequisites:

  • Completion of User Requirements Specification (URS).
  • Installation Qualification (IQ) must be completed.
  • Training of personnel on the operation of the Conductivity Meter.

Equipment Description:

The Conductivity Meter is an instrument used for measuring the electrical conductivity of solutions in the QC Laboratory. It is critical for ensuring product quality and compliance with specifications.

Test ID Procedure Acceptance Criteria Evidence
T1 Perform accuracy calibration using standard solutions. Results must fall within ±0.5 μS/cm of the standard. Calibration report and audit trail.
T2 Verify audit trail functionality. Audit trail must document all changes and user actions. Audit trail log review.

Detailed Test Cases:

  1. Test Case 1: Accuracy Calibration
    • Prepare standard conductivity solutions.
    • Calibrate the Conductivity Meter.
    • Document results and compare against acceptance criteria.
  2. Test Case 2: Audit Trail Verification
    • Simulate user actions on the Conductivity Meter.
    • Review the generated audit trail.
    • Ensure all actions are logged correctly.
See also  Blow-Fill-Seal (BFS) Machine – OQ Protocol

Deviations: Any deviations from the acceptance criteria must be documented and assessed for impact on product quality. Corrective actions should be initiated as necessary.

Approvals:

____________________

Validation Team Lead

____________________

Quality Assurance Manager

“`