Inhalation Flow Resistance Validation in Dry Powder Inhalers (DPI) Manufacturing

Inhalation Flow Resistance Validation in Dry Powder Inhalers (DPI) Manufacturing

Inhalation Flow Resistance Validation for Ensuring Quality in Dry Powder Inhalers Manufacturing

All equipment used in this process validation must be duly qualified and validated for its intended use and performance specifications. Equipment qualification (IQ/OQ/PQ) is assumed to be completed prior to this process validation.

Introduction to Inhalation Flow Resistance in DPI Manufacturing

Inhalation flow resistance is a critical parameter in dry powder inhaler (DPI) manufacturing. It governs the ease with which a patient can draw air through the device, directly influencing the delivered dose performance and pulmonary drug deposition. Correct validation of inhalation flow resistance assures that the DPI meets designed airflow specifications, supporting therapeutic efficacy and patient adherence.

This process validation aims to establish documented evidence that inhalation flow resistance consistently meets predetermined acceptance criteria aligning with the device’s quality target product profile (QTPP), which encapsulates safety, efficacy, and usability considerations under cGMP principles.

Role of Inhalation Flow Resistance in cGMP and Product Consistency

Maintaining control and validation of flow resistance is integral to current Good Manufacturing Practices (cGMP) to guarantee product uniformity batch-to-batch. It ensures the DPI’s airflow pathways are manufactured without deviation from specifications that could lead to suboptimal drug delivery or patient usability issues. Consistency in this parameter supports regulatory expectations for robust process control and reduces risks of batch failure.

Validation documentation provides quality assurance teams with confidence that the production process is stable, capable, and well-understood. It also supports ongoing process verification and lifecycle management of the DPI product.

Understanding the Quality Target Product Profile (QTPP) Related to Flow Resistance

Step 1: Define the QTPP attributes pertinent to inhalation flow resistance. These include the maximum and minimum allowable resistance limits expressed as pressure drop (typically measured in kPa or cm H2O) at standardized flow rates (e.g., 30 L/min or 60 L/min). The QTPP should align with the DPI’s intended patient population and inhalation technique.

Step 2: Establish QTPP criteria ensuring that the inhalation flow resistance is balanced—neither too high to cause patient difficulty nor too low to risk inadequate powder deaggregation and dispersion. This balance ensures dose uniformity and pharmacokinetic performance.

Desired Attributes for DPI Inhalation Flow Resistance

Step 3: Specify desired flow resistance attributes based on device design and clinical requirements:

  • Consistent resistance across manufactured lots
  • Resistance within defined upper and lower specification limits
  • Correlation of flow resistance with delivered dose uniformity and aerodynamic particle size distribution
  • Stability of resistance over the product shelf life

Step 4: For validation planning, incorporate the impact of inhalation flow resistance on critical device performance factors such as ease of use, dose reproducibility, and emitted dose profile.

Impact of Flow Resistance on QTPP and Critical Quality Attributes (CQAs)

Step 5: Map the relationship between inhalation flow resistance and CQAs:

  • Dose Consistency: Resistance affects airflow velocity and powder dispersion, impacting uniformity of delivered dose per actuation.
  • Particle Size Distribution: Adequate flow resistance supports deagglomeration dynamics – crucial for pulmonary deposition.
  • Patient Usability: Resistance influences patient inhalation effort and could affect adherence and treatment outcomes.

Step 6: Define acceptance criteria for inhalation flow resistance based on clinical data, regulatory guidance, and device design limits to ensure CQAs remain within approved ranges.

Critical Properties and Key Parameters for Validation

Step 7: Identify critical properties that influence inhalation flow resistance:

  • Device Geometry: Airflow path dimensions, mesh screens, and orifices
  • Material Properties: Surface roughness and wear of airflow channels over time
  • Assembly Integrity: Seals and component fit affecting airtightness
  • Measurement Conditions: Standardized flow rates, temperature, and humidity during testing

Step 8: Establish key validation parameters:

  • Specified airflow rates for resistance testing consistent with patient inhalation profiles
  • Test reproducibility and repeatability of measuring instruments
  • Defined sample size and acceptance criteria for statistical confidence
  • Frequency of testing (e.g., during pilot, validation batches, and routine production)

Inhalation Flow Resistance Validation in Dry Powder Inhalers Manufacturing

Inhalation Flow Resistance Validation in Dry Powder Inhalers Manufacturing

All equipment used in this process validation must be duly qualified and validated for its intended use and performance specifications. Equipment qualification (IQ/OQ/PQ) is assumed to be completed prior to this process validation.

Desired Attributes of Inhalation Flow Resistance in DPI

The desired attributes for inhalation flow resistance in dry powder inhalers encompass an optimal balance that facilitates patient-friendly airflow while ensuring reliable aerosolization of the drug powder. Specifically, the DPI should exhibit resistance values within a defined range that neither burdens the patient’s inspiratory effort nor compromises the dispersion of the powder into fine particles suitable for lung deposition. This range must be supported by empirical data correlating flow resistance with clinical performance parameters.

Attributes include:

  • Consistent pressure drop at specified flow rates, representative of typical patient inhalation profiles.
  • Repeatability of resistance values across manufactured batches to ensure uniform device function.
  • Robustness against minor variations in manufacturing affecting airflow passageways.
  • Compatibility with the targeted patient demographic’s inspiratory capacity, such as pediatric or elderly populations.

Impact of Inhalation Flow Resistance on the Quality Target Product Profile (QTPP)

The inhalation flow resistance directly influences several QTPP elements, including dose uniformity, delivered dose accuracy, and patient usability. Deviations outside established resistance thresholds can alter the inspiratory flow dynamics, resulting in inconsistent drug delivery efficiency and potentially diminished therapeutic outcomes.

Specifically, increased resistance may:

  • Reduce the inspiratory volume, decreasing lung deposition.
  • Cause incomplete aerosolization, leading to variable dose delivery.
  • Impact patient compliance due to increased inhalation effort.

Conversely, resistance values too low may not induce adequate powder deaggregation, also impacting delivered dose performance. Hence, thorough validation ensures that inhalation flow resistance supports all critical QTPP characteristics for product safety and efficacy.

See also  Homogeneity Validation in Emulsion-based Injectables Manufacturing

Critical Quality Attributes (CQAs) Related to Inhalation Flow Resistance

Identifying and controlling CQAs related to inhalation flow resistance is paramount in DPI manufacturing. These attributes include:

  • Pressure Drop Uniformity: Consistency of measured pressure difference across devices at standard flow rates.
  • Flow Rate Consistency: The ability of the device to maintain designed airflow resistance over the intended operational lifespan.
  • Device Mechanical Integrity: Structural design parameters influencing airflow pathway resistance.
  • Powder Dispersion Efficiency: Indirect impact of flow resistance on aerosol particle size distribution.

Monitoring these CQAs during validation and routine manufacturing ensures product quality adherence and regulatory compliance.

Key Properties Influencing Inhalation Flow Resistance

The following key properties must be monitored and controlled as part of flow resistance validation:

  • Device Geometry: Internal dimensions, channel design, and air inlet size that determine airflow restriction.
  • Surface Roughness: Influences air turbulence and resistance within the DPI.
  • Material Properties: Material elasticity or deformation potential that may alter flow resistance over time.
  • Device Assembly Consistency: Alignment and sealing of internal components contributing to resistance variability.

Control of these properties through robust manufacturing processes supports validated inhalation flow resistance parameters and assures product performance.

Introduction to Inhalation Flow Resistance Validation in DPI Manufacturing

Inhalation Flow Resistance validation is critical in the manufacturing of Dry Powder Inhalers (DPIs) to ensure consistent patient dosing and device performance. This validation confirms that the resistance encountered during patient inhalation falls within predefined limits, which guarantees proper aerosolization and drug delivery. The process validation should be conducted after all equipment qualifications (IQ/OQ/PQ) are completed and focuses on key process parameters that affect flow resistance.

Risk Assessment and Failure Modes Effects Analysis (FMEA)

Begin with a thorough risk assessment identifying all potential failure modes impacting flow resistance. Use FMEA methodology to document each failure mode along with its severity, occurrence, and detectability:

  • Severity (S): Rate the impact of incorrect flow resistance on patient performance and drug delivery efficacy.
  • Occurrence (O): Estimate the frequency of deviations due to manufacturing variability or device defects.
  • Detectability (D): Evaluate the likelihood of detection during routine testing and quality control.

Calculate Risk Priority Numbers (RPN = S x O x D) and prioritize failure modes such as manufacturing defects in the resistance orifice, blockages, or component distortion.

Identification of Critical Process Parameters (CPPs)

Identify CPPs directly influencing inhalation flow resistance through process knowledge, prior studies, and design of experiments (DoE). Typical CPPs include:

  • Orifice dimensions and tolerance limits
  • Assembly torque of components affecting flow pathway
  • Powder bulk density and formulation flow properties
  • Device airflow path integrity and sealing

Map these parameters to critical quality attributes (CQAs) like target resistance range (measured in kPa/(L/min)), ensuring tight control for reproducible patient experience.

Design of Experiments (DoE) for CPP Optimization

Design and execute a DoE to understand the relationships between identified CPPs and flow resistance response. Use factorial or fractional factorial designs to efficiently study multiple parameters simultaneously. Follow these steps:

  1. Define factor levels based on engineering specifications and past data.
  2. Conduct experiments systematically varying CPPs and measure the inhalation flow resistance using standardized test setups (e.g., calibrated flow meters).
  3. Analyze results using statistical software to identify significant factors and interaction effects.
  4. Define acceptable ranges of CPPs that yield desired flow resistance values within product specifications.

Control Strategy and Acceptance Criteria

Develop a control strategy focused on maintaining CPPs within established ranges to ensure flow resistance consistency:

  • In-process controls: Monitor component dimensions and assembly parameters.
  • Critical in-line inspections: Check orifice integrity and pathway continuity.
  • Finished product testing: Conduct inhalation resistance testing on representative samples from every batch.

The acceptance criteria for flow resistance should be set based on clinical performance requirements and regulatory guidance, typically within a range expressed as a specific pressure drop (e.g., 0.02 to 0.04 kPa/(L/min)), aligned with the inhalation profile of target patients.

Process Flow and Stepwise Validation Workflow

Follow a clearly defined stepwise procedure for inhalation flow resistance validation:

  1. Preparation: Confirm equipment qualification and calibration status of flow measurement instruments.
  2. Batch Manufacturing: Produce three consecutive validation batches using controlled CPP settings derived from DoE.
  3. Sampling: Collect representative samples from each batch at defined points – typically, 10% of units or a minimum of 10 devices per batch.
  4. Resistance Testing: Measure inhalation flow resistance under standardized laboratory conditions at set flow rates reflective of patient use.
  5. Data Analysis: Evaluate intra-batch and inter-batch variability. Confirm results fall within pre-established acceptance criteria.
  6. Documentation and Reporting: Compile validation protocols, raw data, statistical analyses, and final conclusions.

Sampling and Decision Points

Define sampling strategies considering device variability and batch size:

  • Select samples at multiple production points for thorough representation (beginning, middle, and end of batch).
  • Apply acceptance criteria evaluating individual measurements and aggregate batch data.
  • Utilize statistical tools like capability indices (Cp, Cpk) to determine process capability and control.
  • Decision points should trigger investigations if samples fall outside acceptable flow resistance ranges or show trends indicating process drift.

Performance Qualification (PPQ) Batch Execution and Evaluation

Execute PPQ batches with full process parameters in intended production environments:

  1. Produce a minimum of three commercial-scale batches following qualified procedures.
  2. Implement full sampling and testing as per control strategy.
  3. Record any deviations, non-conformances, or incidents impacting flow resistance.
  4. Perform comprehensive data trending and statistical analysis to ensure robust control.
  5. Review PPQ data against acceptance criteria; confirm consistent flow resistance within defined limits.
  6. Upon successful completion, approve the DPI manufacturing process for commercial production.

Protocol Design for Inhalation Flow Resistance Validation

Draft a detailed validation protocol including these essential elements:

  • Objective: Define the goal to verify flow resistance meets product specifications.
  • Scope: Describe affected equipment, process steps, and product variants.
  • Responsibilities: Assign roles for execution, sampling, testing, and data analysis.
  • Materials & Equipment: List all instruments, calibration standards, and materials used.
  • Procedures: Outline batch manufacturing, sampling plans, test methods, and acceptance criteria.
  • Data Analysis: Specify statistical methods and decision rules.
  • Reporting: Define format and content of validation report.
  • Change Control: State how protocol amendments are managed.
See also  Compatibility/Emulsion Stability Validation in Parenteral Nutrition (TPN) Manufacturing

Monitoring and Continuous Improvement

Post-validation, establish routine monitoring to sustain process control:

  • Incorporate flow resistance tests in regular quality assurance testing.
  • Apply statistical process control (SPC) charts to detect trends or shifts.
  • Schedule periodic re-validation or verification in response to process changes or equipment maintenance.
  • Use complaint data and clinical feedback to reassess risk and control strategies periodically.

Conclusion

Validating inhalation flow resistance in DPI manufacturing is a complex but essential task to ensure therapeutic efficacy and patient safety. By applying rigorous risk assessment, CPP identification, DoE, controlled execution, and robust statistical evaluation, manufacturers can demonstrate a validated, reliable process for flow resistance control. This stepwise approach supports regulatory compliance and continuous product quality improvements.

Control Strategy Development

Develop a control strategy based on the DoE results and risk assessment to maintain inhalation flow resistance within acceptable ranges. The control strategy should include:

  • Specification limits for CPPs linked to flow resistance (e.g., orifice dimensions, assembly torque)
  • Routine in-process checks for critical assembly steps impacting airflow
  • Calibration and periodic verification of flow resistance measurement equipment
  • Preventive maintenance schedule for equipment influencing resistance pathways

Implement statistical process control (SPC) tools to monitor CPPs and detect trends that could lead to deviations.

Establishment of Acceptable Ranges and Specifications

Establish scientifically justified acceptable ranges for inhalation flow resistance and related CPPs. These should be based on:

  • Patient usability data and clinical inhalation flow profiles
  • Results from pre-validation DoE and robustness studies
  • Regulatory guidelines for DPI performance characteristics

Typical inhalation flow resistance is expressed in kPa/(L/min) and should reflect a target range ensuring optimal powder aerosolization without excessive patient effort.

Sampling Plan and Decision Points

Design a detailed sampling plan for validation batches:

  • Define representative sampling locations and quantities per batch to capture variability
  • Include time-based or batch-based sampling intervals for continuous process verification
  • Incorporate acceptance criteria and decision rules for batch release based on resistance testing

Implement a strategy to escalate investigation or halt manufacturing if samples fall outside acceptable ranges.

Process Performance Qualification (PPQ) Batch Execution and Evaluation

Execute multiple consecutive PPQ batches under normal operating conditions to demonstrate consistent flow resistance within specification:

  1. Perform pre-run equipment checks and confirm calibration of measurement devices
  2. Document all CPP settings and process conditions thoroughly
  3. Collect inhalation flow resistance data per sampling plan during production
  4. Statistically analyze batch data assessing process capability (Cp, Cpk)
  5. Evaluate compliance with predetermined acceptance criteria
  6. Investigate and document any deviations or out-of-trend results immediately

Upon successful PPQ completion, finalize and approve the process validation report confirming robust inhalation flow resistance control.

Monitoring and Ongoing Control

Establish a long-term monitoring plan post-PPQ, including:

  • Routine testing of inhalation flow resistance during manufacturing using validated methods
  • Trend analysis to identify potential drift or shifts in process behavior
  • Periodic revalidation of critical equipment and CPPs affected by wear or design changes
  • Continual review of risk assessments incorporating new data or observed deviations

This ensures sustained product quality and patient safety throughout commercial production.

Process Flow and Stepwise Workflow for Inhalation Flow Resistance Validation

  1. Complete equipment qualification (IQ/OQ/PQ).
  2. Conduct initial risk assessment and FMEA.
  3. Identify CPPs and CQAs related to inhalation flow resistance.
  4. Design and execute DoE for CPP optimization.
  5. Develop control strategy and establish acceptance criteria.
  6. Design sampling plan and define decision points.
  7. Execute PPQ batches with comprehensive monitoring.
  8. Analyze data and generate validation report.
  9. Implement ongoing monitoring and control in routine manufacturing.

Introduction to Inhalation Flow Resistance Validation in DPI Manufacturing

Inhalation flow resistance is a critical parameter impacting the performance, dose delivery, and patient compliance of Dry Powder Inhalers (DPIs). Validating this parameter ensures product consistency and efficacy, complying with regulatory standards. This guide outlines a stepwise procedural framework for comprehensive inhalation flow resistance validation in DPI manufacturing.

Preparation and Equipment Qualification Verification

Confirm all measuring instruments intended for inhalation flow resistance validation have completed IQ (Installation Qualification), OQ (Operational Qualification), and PQ (Performance Qualification). Calibration certificates and maintenance logs must be current and traceable.

  • Verify flow resistance test setups, including flow meters, pressure differential gauges, and test fixtures, are qualified.
  • Ensure software systems for data acquisition and analysis are validated.
  • Document all qualification records in a dedicated equipment file.

Defining Validation Protocol and Acceptance Criteria

Develop a detailed Validation Protocol encompassing methodology, sampling plan, acceptance criteria, and statistical tools. This protocol should reference product-specific inhalation resistance limits derived from development studies and pharmacopoeial requirements.

  • Define sampling size, typically three consecutive production batches for initial validation.
  • Set acceptance criteria for flow resistance values based on validated design space; common range is ±10% of target flow resistance.
  • Specify test conditions, including airflow rates (e.g., 30 L/min, 60 L/min), test temperature and humidity.
  • Outline documentation strategy for raw data, calculations, and batch-specific reports.

Sample Collection and Pre-Test Checks

Collect samples representative of normal manufacturing operation. Select fully manufactured, packaged DPI units from three consecutive batches following routine production.

  • Label and track samples meticulously to prevent lot confusion.
  • Perform environmental conditioning as per protocol before tests.
  • Verify equipment calibration immediately prior to testing.

Performing Inhalation Flow Resistance Tests

Execute flow resistance measurements in accordance with the validation protocol.

  1. Set up the DPI unit on the designed test rig ensuring airtight seals.
  2. Apply airflow at specified test rates; typically conduct tests at multiple flow rates to simulate patient use.
  3. Record pressure drop (∆P) caused by the DPI at each flow rate to calculate flow resistance using the formula:

    Resistance (R) = ∆P / Flow rate.
  4. Perform triplicate measurements on multiple units per batch to capture intra-batch variability.
  5. Record all raw data including environmental conditions and equipment settings.
See also  Base Consistency Validation in Pastes Manufacturing

Documentation and Compilation of Validation Results

Validation Result Tabulation Table (3 Batches)
Batch No. Flow Rate (L/min) Pressure Drop (Pa) Calculated Resistance (Pa/L/min) Mean Resistance (Pa/L/min) Standard Deviation RSD (%)
Batch 1 30 120 4.00 4.05 0.05 1.23
Batch 2 30 122 4.07
Batch 3 30 123 4.08

Comparative Summary and Statistical Analysis

Comparative Summary Table Across Batches
Parameter Batch 1 Batch 2 Batch 3 Overall Mean Compliance Status
Mean Flow Resistance (Pa/L/min) 4.00 4.07 4.08 4.05 Compliant (±10% target 4.0)
Relative Standard Deviation (RSD %) 1.5 1.2 1.1 1.27 Acceptable (≤5%)

The relative standard deviation (RSD) values under 5% across batches demonstrate consistency and precision in the manufacturing process. Mean resistance values within ±10% of the expected target confirm compliance with the product design specifications.

Validation Conclusion and Recommendations

Summarize findings and confirm that inhalation flow resistance parameter is validated successfully for the tested DPI batches within defined acceptance criteria. If deviations occur, investigate root causes before proceeding.

  • Recommend routine monitoring of inhalation flow resistance as part of Critical Process Validation (CPV) post-validation phases.
  • Integrate validated measurement methodology into annual product quality review and monitoring (APQR) for trending analysis.
  • Ensure retention of complete raw data, analytical reports, and validation documentation for regulatory audits.

Continuous Process Verification (CPV) and Routine Monitoring

Establish a controlled monitoring program to track inhalation flow resistance through the product lifecycle.

  • Conduct periodic testing on routine batches using identical validated methods.
  • Use trending tools such as control charts to detect shifts or trends beyond normal variability.
  • Set alert and action limits aligned with validated acceptance criteria.

Annual Product Quality Review (APQR) Integration and Trending

Incorporate inhalation flow resistance data trends into the APQR documentation to identify potential process changes or drift over time.

  • Compare yearly batch data statistically to ensure ongoing compliance.
  • Highlight out-of-trend or out-of-specification data for root cause analysis and corrective actions.

Annexure Templates for Documentation

Include the following annexure templates to standardize documentation:

  • Annexure I: Flow Resistance Validation Protocol Template – outlines objective, scope, sampling, equipment, methods, and acceptance criteria.
  • Annexure II: Equipment Qualification Summary Template – records IQ/OQ/PQ status and calibration details of flow resistance measurement equipment.
  • Annexure III: Raw Data Recording Sheet – structured format to capture batch number, sample ID, airflow rates, pressure drop readings, and calculated resistance.
  • Annexure IV: Validation Results and Statistical Analysis Template – includes tables for mean, standard deviation, RSD%, and compliance evaluation.
  • Annexure V: Validation Summary and Approval Sheet – documents final conclusion, deviations (if any), corrective actions, and sign-off by Quality Assurance and Production Heads.

All templates should be adapted to site-specific documentation systems and electronic or paper record practices.

Validation Result Tabulation and Data Analysis

Compile all measured inhalation flow resistance data systematically to facilitate comprehensive analysis and interpretation. Use tabulated formats for clarity and ease of comparison.

Table 1: Validation Results Across Three Consecutive Batches
Batch Number Sample ID Flow Resistance at 30 L/min (kPa L/min) Flow Resistance at 60 L/min (kPa L/min) Comments
Batch 1 Sample 1 XX.XX XX.XX Within acceptable range
Batch 1 Sample 2 XX.XX XX.XX Within acceptable range
Batch 1 Sample 3 XX.XX XX.XX Within acceptable range
Batch 2 Sample 1 XX.XX XX.XX Within acceptable range
Batch 2 Sample 2 XX.XX XX.XX Within acceptable range
Batch 2 Sample 3 XX.XX XX.XX Within acceptable range
Batch 3 Sample 1 XX.XX XX.XX Within acceptable range
Batch 3 Sample 2 XX.XX XX.XX Within acceptable range
Batch 3 Sample 3 XX.XX XX.XX Within acceptable range

Comparative Summary and Statistical Evaluation

Perform a comparative summary to evaluate batch-to-batch consistency and compliance with acceptance criteria using statistical parameters such as Mean, Standard Deviation (SD), and Relative Standard Deviation (RSD, %).

Table 2: Comparative Summary of Inhalation Flow Resistance
Batch Mean Flow Resistance at 30 L/min (kPa L/min) SD RSD (%) Mean Flow Resistance at 60 L/min (kPa L/min) SD RSD (%) Compliance Status
Batch 1 XX.XX XX.XX XX.X XX.XX XX.XX XX.X Pass/Fail
Batch 2 XX.XX XX.XX XX.X XX.XX XX.XX XX.X Pass/Fail
Batch 3 XX.XX XX.XX XX.X XX.XX XX.XX XX.X Pass/Fail
  • Assess RSD values to confirm consistency; typically, RSD ≤ 10% indicates acceptable variability.
  • Validate if batch results comply within ±10% of target inhalation flow resistance.
  • If any out-of-specification (OOS) results are observed, initiate investigation and root cause analysis.

Continued Process Verification (CPV) and Routine Monitoring

Establish a CPV program to monitor inhalation flow resistance in routine manufacturing to ensure continued product quality post-validation.

  1. Define a sampling frequency, e.g., monthly testing of three random batches post-validation.
  2. Utilize the same validated instruments, test methods, and environmental conditions as in the initial validation.
  3. Document all results in CPV reports, flagging any trends or deviations.
  4. Perform trend analysis quarterly or annually as part of the Annual Product Quality Review (APQR).
  5. Implement corrective actions promptly in case of any observed deviations or drift.

Documentation and Reporting

Complete thorough documentation to demonstrate compliance with regulatory expectations and support quality management systems.

  • Compile a comprehensive validation report summarizing objectives, methodology, results, and conclusions.
  • Include raw data, calibration certificates, and equipment qualification records as annexures.
  • Submit the validation report for approval by quality assurance (QA) and signatories.
  • Archive all documentation systematically for audit readiness and regulatory inspection.

Annexures: Templates for Standardized Documentation

Use the following templates to standardize documentation for inhalation flow resistance validation and monitoring:

  1. Annexure I: Equipment Qualification Checklist (IQ/OQ/PQ)
  2. Annexure II: Validation Protocol Template
  3. Annexure III: Sample Collection and Labeling Log
  4. Annexure IV: Raw Data Recording Sheet for Flow Resistance Tests
  5. Annexure V: Validation Summary Report and Approval Form