Micropipette 100-1000 µL (CAL-0023) – Calibration Documents

Calibration Documents: Micropipette 100-1000 µL (CAL-0023)

Asset ID CAL-0023
Equipment Micropipette 100-1000 µL
Equipment Type Micropipette
Department QC
Area/Location Chemistry Lab
Process Use Sample prep
Parameter Volume
Units µL
Range Min 100
Range Max 1000
Acceptance Criteria ±1%
Calibration Frequency (days) 180
Calibration Type External
Traceability ISO17025

Calibration SOP

1. Purpose

The calibration of the Micropipette 100-1000 µL is performed to ensure accurate volume measurements during sample preparation in the Chemistry Lab. This calibration controls risks related to accuracy, patient safety, batch quality, and data integrity, thereby ensuring compliance with Good Manufacturing Practices (GMP).

2. Scope

  • Asset ID: CAL-0023
  • Equipment Name/Type: Micropipette 100-1000 µL / Micropipette
  • Department/Area: QC / Chemistry Lab
  • Process Use: Sample prep
  • Parameter: Volume
  • Range: 100 – 1000 µL
  • Units: µL
  • Calibration Frequency: 180 days

3. Definitions

  • Calibration: The process of comparing a measurement instrument to a known standard to ensure accuracy.
  • Verification: The process of confirming that a measurement instrument meets specified requirements.
  • Adjustment: The act of modifying a measurement instrument to bring it into compliance with specifications.
  • As Found: The condition of the instrument prior to any adjustments being made.
  • As Left: The condition of the instrument after adjustments have been made.
  • OOT (Out of Tolerance): A condition where the measurement falls outside the specified acceptance criteria.
  • Drift: The gradual change in the measurement output of an instrument over time.
  • Traceability: The ability to verify the history, location, or application of an item by means of documented recorded identification.
  • Measurement Uncertainty: A parameter that characterizes the dispersion of the values attributed to a measured quantity.
  • Intermediate Check: A verification process conducted between calibrations to ensure ongoing accuracy.

4. References

  • Company Metrology SOP [TBD]
  • ISO/IEC 17025 traceability via calibration certificates
  • ALCOA+ principles

5. Responsibilities

  • Technician/Metrology: Perform calibration and maintain records.
  • QC/QA Reviewer: Review calibration results and documentation.
  • Owner Department: Ensure equipment is available and in good condition for calibration.
  • External Service Provider: Conduct external calibration as required.

6. Safety & Preconditions

  • Ensure environmental controls are in place (temperature, humidity).
  • Allow equipment to warm-up/stabilize as per manufacturer recommendations.
  • Maintain cleanliness of the equipment and workspace.
  • Check leveling of the equipment.
  • Verify power supply is stable and within specifications.
See also  Microscope Stage Micrometer (Calibration) (CAL-0240) - Calibration Documents

7. Standards & Tools

  • Reference Standards Required: [TBD]
  • Traceability Statement: ISO17025
  • Certificate Validity Check: Ensure all reference standards are within their validity period.
  • Standard ID: [TBD]

8. Calibration Strategy

Calibration will follow the As Found and As Left logic:

  • Minimum number of points: At least 5 points including endpoints (100 µL, 1000 µL) and midpoints (e.g., 500 µL).
  • Replicate readings at each point: 3 repeats are recommended for accuracy.

9. Step-by-Step Procedure

  1. Preparation: Gather all necessary tools and reference standards.
  2. Stabilization: Allow the micropipette to stabilize in the environment for at least 30 minutes.
  3. As Found checks and data capture: Measure the volume at each defined point and record observations contemporaneously (ALCOA+).
  4. Adjustment rules: If OOT is detected, follow the adjustment protocol; if not allowed, escalate to QA.
  5. As Left verification: Re-measure to confirm adjustments are within specifications.
  6. Labeling + documentation completion: Complete all required documentation and label the equipment with calibration details.

10. Acceptance Criteria & Calculations

Acceptance criteria: ±1% of the measured value.

Error calculation formula: Error (%) = (Observed Value – True Value) / True Value × 100

Pass/Fail Logic: If the error is within ±1%, the calibration passes; otherwise, it fails.

Rounding rule: [TBD]

11. Handling OOT / Failures

  • Immediate containment: Tag equipment and stop use.
  • Notification & deviation initiation: Notify QA and initiate deviation process.
  • Impact assessment trigger: Define impact window and review of affected results/batches.
  • Recalibration after repair/adjustment: Ensure recalibration is performed post-adjustment.
  • QA disposition: [TBD]

12. Records & Retention

  • Logbook entries: Document all calibration activities.
  • Calibration report/certificate filing: Maintain records for audit purposes.
  • e-Records if applicable: [TBD]
  • Label content requirements: Include calibrated date, due date, by whom, and status.

13. Annexures

Annexure A: Calibration Data Sheet

Test Points Reference Observed Error Acceptance Pass/Fail As Found As Left
[TBD] [TBD] [TBD] [TBD] [TBD] [TBD] [TBD] [TBD]

Annexure B: Quick Checklist

  • Equipment clean and ready for calibration.
  • All necessary tools and standards available.
  • Environmental conditions verified.
  • Calibration documentation prepared.

Annexure C: Example Label Text

Calibration Date: [TBD]

Due Date: [TBD]

Calibrated By: [TBD]

Status: [TBD]

⬆ Back to TOC


Calibration Protocol

Calibration Protocol

Protocol ID Asset ID Equipment Name Equipment Type Dept Area Process Use Parameter Units Range Acceptance Criteria Frequency Calibration Type Traceability
[TBD] CAL-0023 Micropipette 100-1000 µL Micropipette QC Chemistry Lab Sample prep Volume µL 100 – 1000 ±1% 180 days External ISO17025
See also  Sieve Shaker Timer (CAL-0135) - Calibration Documents

Environmental Conditions

Temperature: [TBD]

Relative Humidity: [TBD]

Vibration: [TBD]

Drafts: [TBD]

Test Point Selection

  • Minimum: 100 µL
  • 25%: 250 µL
  • 50%: 500 µL
  • 75%: 750 µL
  • Maximum: 1000 µL

Results Table

Test Point Ref Value Observed (As Found) Error Limit Pass/Fail Observed (As Left) Error Pass/Fail
100 µL [TBD] [TBD] [TBD] [TBD] [TBD] [TBD] [TBD] [TBD]
250 µL [TBD] [TBD] [TBD] [TBD] [TBD] [TBD] [TBD] [TBD]
500 µL [TBD] [TBD] [TBD] [TBD] [TBD] [TBD] [TBD] [TBD]
750 µL [TBD] [TBD] [TBD] [TBD] [TBD] [TBD] [TBD] [TBD]
1000 µL [TBD] [TBD] [TBD] [TBD] [TBD] [TBD] [TBD] [TBD]

Stepwise Instructions

  1. Preparation: Ensure the micropipette is clean and calibrated equipment is ready.
  2. Stabilization: Allow the micropipette to stabilize at room temperature for [TBD] minutes.
  3. Performing Readings:
    • Set the micropipette to the first test point.
    • Perform three replicates for each test point.
  4. Calculating Error: Calculate the error for each replicate using the formula: Error = (Observed – Ref Value) / Ref Value * 100.
  5. Evaluation: Compare the calculated errors against the acceptance criteria of ±1%.
  6. Documentation: Record all observations in the results table.

Deviation Instruction for OOT

If any observed value falls outside the acceptance criteria:

  • Stop use of the micropipette.
  • Notify the supervisor immediately.
  • Open a deviation report.
  • Quarantine impacted results window: [TBD].
  • Await QA decision on further actions.

Approvals

Prepared By Reviewed By Approved By
[Signature/Date] [Signature/Date] [Signature/Date]

⬆ Back to TOC


Calibration Checklist

Asset ID Equipment Dept Area Parameter Range Acceptance Frequency Traceability
CAL-0023 Micropipette 100-1000 µL QC Chemistry Lab Volume 100 – 1000 µL ±1% 180 days ISO17025

Pre-calibration

  • [ ] Verify environmental conditions are within specified limits
  • [ ] Allow equipment to warm up as per SOP
  • [ ] Clean the micropipette according to cleaning procedures
  • [ ] Level the micropipette if applicable
  • [ ] Check validity of reference standards
  • [ ] Ensure calibration logbook is ready for entries

As Found Execution

  • [ ] Set initial setpoints for calibration
  • [ ] Perform repeat measurements as per protocol
  • [ ] Record all raw data accurately
  • [ ] Perform necessary calculations for results

OOT Immediate Actions

  • [ ] Tag out the micropipette if out of tolerance
  • [ ] Stop use of the micropipette immediately
  • [ ] Notify the supervisor of the OOT condition
  • [ ] Document deviation in the logbook
  • [ ] Trigger impact review process as necessary
See also  Load Cell (Loss-in-weight feeder) (CAL-0013) - Calibration Documents

As Left Verification

  • [ ] Conduct post-adjustment checks to confirm accuracy
  • [ ] Ensure all measurements are within specified limits

Documentation & Closure

  • [ ] Update equipment label with new calibration date
  • [ ] Attach calibration certificate to logbook
  • [ ] Obtain reviewer sign-off on calibration records
  • [ ] Update next due date for calibration

Data Integrity (ALCOA+): Ensure all data is attributable, legible, contemporaneous, original, and accurate. Maintain records in compliance with GMP standards.

⬆ Back to TOC


Deviation Impact Assessment

Event Summary
Asset: Micropipette 100-1000 µL
Date: [TBD]
Discovered By: [TBD]
Document References: [TBD]
As Found Results Summary: [TBD]

Immediate Containment Checklist

  • Stop use of the asset
  • Label/tag the asset as out of tolerance
  • Segregate impacted work

Define Impact Window

Last known in-tolerance date: [TBD] to discovery date: [TBD]

Product/Batch/Test Impact Review

Batch/Study/Report ID Results Potentially Impacted Re-test/Re-weigh Decision
[TBD] [TBD] [TBD]

Data Integrity Review Prompts (ALCOA+)

  • Attributable: [TBD]
  • Legible: [TBD]
  • Contemporaneous: [TBD]
  • Original: [TBD]
  • Accurate: [TBD]
  • Complete: [TBD]
  • Consistent: [TBD]
  • Enduring: [TBD]
  • Available: [TBD]

Risk Assessment

Severity Probability Detectability Rationale Overall Risk
[TBD] [TBD] [TBD] [TBD] [TBD]

Investigation Summary

Root Cause: [TBD]

CAPA Plan

Action Owner Due Date Effectiveness Check
[TBD] [TBD] [TBD] [TBD]

QA Disposition

Release/Hold/Retest/Recall Triggers: [TBD]

⬆ Back to TOC


Calibration Certificate

Certificate No: [TBD]

Date: [TBD]

Next Due Date: [TBD]

Asset Details

Asset ID Equipment Name Equipment Type Dept Area Make/Model Serial No Location Parameter/Range/Units Acceptance Criteria
CAL-0023 Micropipette 100-1000 µL Micropipette QC Chemistry Lab [TBD] [TBD] [TBD] Volume / 100-1000 / µL ±1%

Conditions

Environment: [TBD]

Reference Standards

Standard Name/ID Certificate No Due Date Traceability Statement
[TBD] [TBD] [TBD] ISO17025

Results

As Found Results

Test Point Measured Value (µL) Acceptance Criteria Pass/Fail
100 [TBD] ±1% [TBD]
200 [TBD] ±1% [TBD]
500 [TBD] ±1% [TBD]
800 [TBD] ±1% [TBD]
1000 [TBD] ±1% [TBD]

As Left Results

Test Point Measured Value (µL) Acceptance Criteria Pass/Fail
100 [TBD] ±1% [TBD]
200 [TBD] ±1% [TBD]
500 [TBD] ±1% [TBD]
800 [TBD] ±1% [TBD]
1000 [TBD] ±1% [TBD]

Measurement Uncertainty Statement: [TBD]

Statement of Conformity: [TBD]

Remarks: [TBD]

Signatures

Performed by: ____________________ Date: ___________

Checked by: ____________________ Date: ___________

Approved by: ____________________ Date: ___________

⬆ Back to TOC