Bottle Top Dispenser 5-50 mL (CAL-0027) – Calibration Documents

Calibration Documents: Bottle Top Dispenser 5-50 mL (CAL-0027)

Asset ID CAL-0027
Equipment Bottle Top Dispenser 5-50 mL
Equipment Type Bottle Top Dispenser
Department QC
Area/Location Chemistry Lab
Process Use Reagent dispensing
Parameter Volume
Units mL
Range Min 5
Range Max 50
Acceptance Criteria ±1%
Calibration Frequency (days) 365
Calibration Type External
Traceability ISO17025

Calibration SOP

1. Purpose

The calibration of the Bottle Top Dispenser 5-50 mL is performed to ensure accurate reagent dispensing within the specified volume range. This calibration controls risks related to accuracy, patient safety, batch quality, and data integrity, thereby ensuring compliance with Good Manufacturing Practices (GMP).

2. Scope

  • Asset ID: CAL-0027
  • Equipment Name/Type: Bottle Top Dispenser 5-50 mL / Bottle Top Dispenser
  • Department/Area: QC / Chemistry Lab
  • Process Use: Reagent dispensing
  • Parameter: Volume
  • Range: 5 – 50 mL
  • Units: mL
  • Calibration Frequency: 365 days

3. Definitions

  • Calibration: The process of verifying and adjusting the accuracy of measurement equipment against a known standard.
  • Verification: The process of confirming that the equipment meets specified requirements.
  • Adjustment: The act of correcting the measurement output of the equipment.
  • As Found: The condition of the equipment prior to any adjustments.
  • As Left: The condition of the equipment after adjustments have been made.
  • OOT (Out of Tolerance): A condition where the measurement falls outside the defined acceptance criteria.
  • Drift: The gradual change in the measurement output over time.
  • Traceability: The ability to verify the history, location, or application of an item by means of documented recorded identification.
  • Measurement Uncertainty: A parameter that characterizes the range of values within which the true value is expected to lie.
  • Intermediate Check: A periodic check to ensure that the equipment remains within specified limits between full calibrations.

4. References

  • Company Metrology SOP [TBD]
  • ISO/IEC 17025 traceability via calibration certificates
  • ALCOA+ principles

5. Responsibilities

  • Technician/Metrology: Perform calibration and maintain records.
  • QC/QA Reviewer: Review calibration results and documentation.
  • Owner Department: Ensure equipment is available and in good condition for calibration.
  • External Service Provider: Conduct external calibration as required.

6. Safety & Preconditions

  • Ensure environmental controls are in place (temperature, humidity).
  • Allow for warm-up/stabilization of the equipment as per manufacturer guidelines.
  • Maintain cleanliness of the equipment and surrounding area.
  • Check leveling of the equipment.
  • Verify power supply checks are completed.
See also  Volumetric Flask 250 mL (CAL-0037) - Calibration Documents

7. Standards & Tools

  • Reference standards required: [TBD]
  • Traceability statement: ISO17025
  • Certificate validity check: [TBD]
  • Standard ID placeholders: [TBD]

8. Calibration Strategy

The calibration will follow the As Found and As Left logic:

  • Minimum number of points: At least 5 points including endpoints (5 mL, 50 mL) and midpoints (e.g., 10 mL, 25 mL, 40 mL).
  • Replicate readings: Perform 3 repeats at each point as a best practice.

9. Step-by-Step Procedure

  1. Preparation: Gather all necessary tools and standards.
  2. Stabilization: Allow the equipment to stabilize as per manufacturer instructions.
  3. As Found checks and data capture: Measure and record the volume dispensed at each test point.
  4. Adjustment rules: If OOT is found, follow adjustment protocols; if not allowed, escalate to QA.
  5. As Left verification: Measure and record the volume dispensed again after adjustments.
  6. Labeling + documentation completion: Complete all required documentation and label the equipment accordingly.

Record raw observations contemporaneously to ensure data integrity (ALCOA+).

10. Acceptance Criteria & Calculations

Acceptance criteria: ±1%

Error calculation formula: Error (%) = (Observed Value – True Value) / True Value * 100

Pass/Fail logic: If the error falls within the acceptance criteria, the calibration passes; otherwise, it fails.

Rounding rule placeholder: [TBD]

11. Handling OOT / Failures

  • Immediate containment: Tag equipment and stop use.
  • Notification & deviation initiation: Notify relevant personnel and initiate deviation procedures.
  • Impact assessment trigger: Define impact window and review of affected results/batches.
  • Recalibration after repair/adjustment: Ensure recalibration is performed post-adjustment.
  • QA disposition placeholders: [TBD]

12. Records & Retention

  • Logbook entries: Maintain a logbook of all calibration activities.
  • Calibration report/certificate filing: File all calibration reports and certificates appropriately.
  • e-Records if applicable: [TBD]
  • Label content requirements: Include calibrated date, due date, by whom, and status.

13. Annexures

Annexure A: Calibration Data Sheet

Test Points Reference Observed Error Acceptance Pass/Fail As Found As Left
[TBD] [TBD] [TBD] [TBD] [TBD] [TBD] [TBD] [TBD]

Annexure B: Quick Checklist

  • Environmental controls verified
  • Equipment stabilized
  • Calibration standards checked
  • Documentation prepared

Annexure C: Example Label Text

Calibration Date: [TBD]

Due Date: [TBD]

Calibrated By: [TBD]

Status: [TBD]

⬆ Back to TOC


Calibration Protocol

Calibration Protocol

Protocol ID Asset ID Equipment Name Equipment Type Dept Area Process Use Parameter Units Range Acceptance Criteria Frequency Calibration Type Traceability
[TBD] CAL-0027 Bottle Top Dispenser 5-50 mL Bottle Top Dispenser QC Chemistry Lab Reagent dispensing Volume mL 5 – 50 ±1% 365 days External ISO17025
See also  Graduated Cylinder 25 mL (CAL-0041) - Calibration Documents

Environmental Conditions

Temperature: [TBD] °C

Relative Humidity: [TBD] %

Vibration: [TBD]

Drafts: [TBD]

Test Point Selection

  • Minimum: 5 mL
  • 25%: 15 mL
  • 50%: 25 mL
  • 75%: 35 mL
  • Maximum: 50 mL

Results Table

Test Point Ref Value Observed (As Found) Error Limit Pass/Fail Observed (As Left) Error Pass/Fail
5 mL [TBD] [TBD] [TBD] [TBD] [TBD] [TBD] [TBD] [TBD]
15 mL [TBD] [TBD] [TBD] [TBD] [TBD] [TBD] [TBD] [TBD]
25 mL [TBD] [TBD] [TBD] [TBD] [TBD] [TBD] [TBD] [TBD]
35 mL [TBD] [TBD] [TBD] [TBD] [TBD] [TBD] [TBD] [TBD]
50 mL [TBD] [TBD] [TBD] [TBD] [TBD] [TBD] [TBD] [TBD]

Stepwise Instructions

  1. Preparation: Ensure the Bottle Top Dispenser is clean and calibrated equipment is available.
  2. Stabilization: Allow the equipment to stabilize in the environmental conditions for at least [TBD] minutes.
  3. Performing Readings: Dispense each test point volume in replicates (minimum of 3) and record the observed values.
  4. Calculating Error: Calculate error using the formula: Error = (Observed – Ref Value) / Ref Value * 100.
  5. Evaluation: Compare the calculated error against the acceptance criteria of ±1%.
  6. Documentation: Complete the results table and ensure all data is recorded accurately following ALCOA+ principles.

Deviation Instruction for OOT

If any observed value is outside the acceptance criteria:

  • Stop use of the equipment immediately.
  • Notify the supervisor and Quality Assurance (QA) department.
  • Open a deviation report.
  • Quarantine impacted results window: [TBD]
  • Await QA decision on further actions.

Approvals

Prepared By Reviewed By Approved By
[Signature/Date] [Signature/Date] [Signature/Date]

⬆ Back to TOC


Calibration Checklist

Asset ID Equipment Dept Area Parameter Range Acceptance Frequency Traceability
CAL-0027 Bottle Top Dispenser 5-50 mL QC Chemistry Lab Volume 5 – 50 mL ±1% 365 days ISO17025

Pre-calibration

  • [ ] Verify environmental conditions are within specified limits
  • [ ] Allow equipment to warm up as per SOP
  • [ ] Clean the equipment according to the cleaning procedure
  • [ ] Ensure the equipment is leveled
  • [ ] Check validity of reference standards
  • [ ] Ensure logbook is ready for documentation

As Found Execution

  • [ ] Set initial measurement setpoints
  • [ ] Perform repeat measurements as per protocol
  • [ ] Record raw data accurately
  • [ ] Perform necessary calculations for results
See also  Inline Temp Sensor (Reactor Jacket) (CAL-0200) - Calibration Documents

OOT Immediate Actions

  • [ ] Tag out the equipment
  • [ ] Stop use of the equipment immediately
  • [ ] Notify the supervisor/QA
  • [ ] Document deviation in the logbook
  • [ ] Trigger impact review as per SOP

As Left Verification

  • [ ] Conduct post-adjustment checks
  • [ ] Confirm results are within acceptance limits

Documentation & Closure

  • [ ] Update equipment label with calibration details
  • [ ] Attach calibration certificate to logbook
  • [ ] Obtain reviewer sign-off
  • [ ] Update next due date in the system

Data Integrity (ALCOA+): Ensure all data is Attributable, Legible, Contemporaneous, Original, Accurate, and Complete.

⬆ Back to TOC


Deviation Impact Assessment

Event Summary
Asset: Bottle Top Dispenser 5-50 mL
Date: [TBD]
Discovered By: [TBD]
Document References: [TBD]
As Found Results Summary: [TBD]

Immediate Containment Checklist

  • Stop use of the asset
  • Label/tag the asset as out of tolerance
  • Segregate impacted work

Define Impact Window

Last Known In-Tolerance Date Discovery Date
[TBD] [TBD]

Product/Batch/Test Impact Review

Batch/Study/Report ID Results Potentially Impacted Re-test/Re-weigh Decision
[TBD] [TBD] [TBD]

Data Integrity Review Prompts (ALCOA+)

  • Attributable: [TBD]
  • Legible: [TBD]
  • Contemporaneous: [TBD]
  • Original: [TBD]
  • Accurate: [TBD]
  • Complete: [TBD]
  • Consistent: [TBD]
  • Enduring: [TBD]
  • Available: [TBD]

Risk Assessment

Severity Probability Detectability Rationale Overall Risk
[TBD] [TBD] [TBD] [TBD] [TBD]

Investigation Summary

Root Cause: [TBD]

CAPA Plan

Action Owner Due Date Effectiveness Check
[TBD] [TBD] [TBD] [TBD]

QA Disposition

Release/Hold/Retest/Recall Triggers: [TBD]

⬆ Back to TOC


Calibration Certificate

Certificate No: [TBD]

Date: [TBD]

Next Due Date: [TBD]

Asset Details

Asset ID Equipment Name Equipment Type Dept Area Make/Model Serial No Location Parameter/Range/Units Acceptance Criteria
CAL-0027 Bottle Top Dispenser 5-50 mL Bottle Top Dispenser QC Chemistry Lab [TBD] [TBD] [TBD] Volume (5-50 mL) ±1%

Conditions

Environment: [TBD]

Reference Standards

Standard Name/ID Certificate No Due Date Traceability Statement
[TBD] [TBD] [TBD] Traceability: ISO17025

Results

As Found Results

Test Point Measured Value (mL) Pass/Fail
5 mL [TBD] [TBD]
10 mL [TBD] [TBD]
20 mL [TBD] [TBD]
35 mL [TBD] [TBD]
50 mL [TBD] [TBD]

As Left Results

Test Point Measured Value (mL) Pass/Fail
5 mL [TBD] [TBD]
10 mL [TBD] [TBD]
20 mL [TBD] [TBD]
35 mL [TBD] [TBD]
50 mL [TBD] [TBD]

Measurement Uncertainty Statement: [TBD]

Statement of Conformity: [Pass/Fail overall]

Remarks: [TBD]

Signatures

Performed by: ______________________ Date: ____________

Checked by: ______________________ Date: ____________

Approved by: ______________________ Date: ____________

⬆ Back to TOC