Calibration Documents: Burette 50 mL (CAL-0030)
| Asset ID | CAL-0030 |
|---|---|
| Equipment | Burette 50 mL |
| Equipment Type | Burette |
| Department | QC |
| Area/Location | Chemistry Lab |
| Process Use | Titration |
| Parameter | Volume |
| Units | mL |
| Range Min | 0 |
| Range Max | 50 |
| Acceptance Criteria | Class A |
| Calibration Frequency (days) | 365 |
| Calibration Type | External |
| Traceability | ISO17025 |
Table of Contents
Calibration SOP
1. Purpose
The calibration of the Burette 50 mL is performed to ensure accurate volume measurements during titration processes in the QC department. This calibration controls risks related to accuracy, patient safety, batch quality, and data integrity, thereby ensuring compliance with regulatory standards.
2. Scope
Asset ID: CAL-0030
Equipment Name/Type: Burette 50 mL / Burette
Department/Area: QC / Chemistry Lab
Process Use: Titration
Parameter: Volume
Range: 0 – 50 mL
Units: mL
Calibration Frequency: 365 days
3. Definitions
- Calibration: The process of verifying and adjusting the accuracy of a measurement instrument.
- Verification: The process of confirming that an instrument meets specified requirements.
- Adjustment: The act of modifying an instrument to bring it into compliance with specified standards.
- As Found: The condition of the instrument prior to any adjustments.
- As Left: The condition of the instrument after adjustments have been made.
- OOT (Out of Tolerance): A condition where the instrument’s measurements fall outside specified limits.
- Drift: The gradual change in an instrument’s measurement accuracy over time.
- Traceability: The ability to verify the history, location, or application of an item by means of documented recorded identification.
- Measurement Uncertainty: A parameter that characterizes the range of values within which the true value is asserted to lie.
- Intermediate Check: A periodic check of an instrument’s performance between full calibrations.
4. References
- Company Metrology SOP [TBD]
- ISO/IEC 17025 traceability via calibration certificates
- ALCOA+ principles
5. Responsibilities
- Technician/Metrology: Perform calibration and maintain records.
- QC/QA Reviewer: Review calibration results and documentation.
- Owner Department: Ensure equipment is available and in good condition for calibration.
- External Service Provider: Conduct external calibrations as required.
6. Safety & Preconditions
- Ensure environmental controls are in place (temperature, humidity).
- Allow equipment to warm up and stabilize as per manufacturer guidelines.
- Maintain cleanliness of the equipment and surrounding area.
- Check that the burette is leveled properly before use.
- Verify power supply is stable and within required specifications.
7. Standards & Tools
Reference Standards Required: [TBD]
Traceability Statement: Calibration traceable to ISO17025.
Certificate Validity Check: Ensure all calibration certificates are current and valid.
8. Calibration Strategy
Calibration will follow the As Found and As Left logic. A minimum of 5 measurement points will be taken, including endpoints (0 mL and 50 mL) and midpoints. Each point will have 3 replicate readings to ensure accuracy.
9. Step-by-Step Procedure
- Preparation: Gather all necessary tools and standards.
- Stabilization: Allow the burette to stabilize in the environment.
- As Found checks and data capture: Measure and record the volume at each point.
- Adjustment rules: If OOT is found, escalate for adjustment as per SOP.
- As Left verification: Confirm that the instrument is within tolerance after adjustments.
- Labeling + documentation completion: Complete all required documentation and label the equipment.
Record all raw observations contemporaneously to ensure data integrity (ALCOA+).
10. Acceptance Criteria & Calculations
Acceptance Criteria: Class A
Error Calculation Formula: Error (%) = (Observed Value – True Value) / True Value * 100
Pass/Fail Logic: If error is within acceptance criteria, pass; otherwise, fail.
Rounding Rule: [TBD]
11. Handling OOT / Failures
- Immediate containment: Tag equipment and stop use.
- Notification & deviation initiation: Notify relevant personnel.
- Impact assessment trigger: Define impact window and review of affected results/batches.
- Recalibration after repair/adjustment: Ensure recalibration is performed.
- QA disposition placeholders: [TBD]
12. Records & Retention
- Logbook entries: Document all calibration activities.
- Calibration report/certificate filing: Maintain records for audit purposes.
- e-Records if applicable: [TBD]
- Label content requirements: Calibrated date, due date, by whom, status.
13. Annexures
Annexure A: Calibration Data Sheet (test points, ref, observed, error, acceptance, pass/fail; include As Found + As Left columns)
Annexure B: Quick checklist
Annexure C: Example label text (with [TBD] placeholders)
Calibration Protocol
Calibration Protocol
| Protocol ID | Asset ID | Equipment Name | Equipment Type | Dept | Area | Process Use | Parameter | Units | Range | Acceptance Criteria | Frequency | Calibration Type | Traceability |
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| [TBD] | CAL-0030 | Burette 50 mL | Burette | QC | Chemistry Lab | Titration | Volume | mL | 0 – 50 | Class A | 365 days | External | ISO17025 |
Environmental Conditions
Temperature: [TBD]
Relative Humidity: [TBD]
Vibration: [TBD]
Drafts: [TBD]
Test Point Selection
Test points will be selected based on the following values:
- Min: 0 mL
- 25%: 12.5 mL
- 50%: 25 mL
- 75%: 37.5 mL
- Max: 50 mL
Results Table
| Test Point | Ref Value | Observed (As Found) | Error | Limit | Pass/Fail | Observed (As Left) | Error | Pass/Fail |
|---|---|---|---|---|---|---|---|---|
| 0 mL | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] |
| 12.5 mL | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] |
| 25 mL | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] |
| 37.5 mL | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] |
| 50 mL | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] |
Stepwise Instructions
- Preparation: Ensure the burette is clean and free from any contaminants.
- Stabilization: Allow the burette to stabilize at room temperature for at least 30 minutes.
- Performing Readings:
- Fill the burette with the calibration standard.
- Take three replicate readings at each test point.
- Calculating Error: Calculate the error for each observed value against the reference value.
- Evaluation: Compare the errors against the acceptance criteria.
- Documentation: Record all observations and calculations in the results table.
Deviation Instruction for OOT
If any observed value is out of tolerance (OOT):
- Stop use of the equipment immediately.
- Notify the QA department.
- Open a deviation report.
- Quarantine impacted results window: [TBD]
- Await QA decision on further actions.
Approvals
| Prepared | Reviewed | Approved |
|---|---|---|
| [Signature/Date] | [Signature/Date] | [Signature/Date] |
Calibration Checklist
| Asset ID | Equipment | Dept | Area | Parameter | Range | Acceptance | Frequency | Traceability |
|---|---|---|---|---|---|---|---|---|
| CAL-0030 | Burette 50 mL | QC | Chemistry Lab | Volume | 0 – 50 mL | Class A | 365 days | ISO17025 |
Pre-calibration
- [ ] Verify environmental conditions are within specified limits
- [ ] Allow equipment to warm up as per SOP
- [ ] Clean the burette thoroughly
- [ ] Ensure the burette is level
- [ ] Check validity of reference standards
- [ ] Ensure logbook is ready for entries
As Found Execution
- [ ] Set initial measurement setpoints
- [ ] Perform repeat measurements as required
- [ ] Record all raw data accurately
- [ ] Perform necessary calculations
OOT Immediate Actions
- [ ] Tag out the equipment
- [ ] Stop use of the equipment immediately
- [ ] Notify the appropriate personnel
- [ ] Document deviation
- [ ] Trigger impact review process
As Left Verification
- [ ] Conduct post-adjustment checks
- [ ] Confirm all measurements are within specified limits
Documentation & Closure
- [ ] Update equipment label
- [ ] Attach calibration certificate
- [ ] Obtain reviewer sign-off
- [ ] Update next due date in logbook
Acceptance Criteria: All measurements must meet Class A standards.
Data Integrity (ALCOA+): Ensure all data is Attributable, Legible, Contemporaneous, Original, Accurate, and Complete.
Traceability: Calibration must be traceable to ISO17025 standards.
Deviation Impact Assessment
| Event Summary |
|---|
|
Asset: Burette 50 mL Date: [TBD] Discovered By: [TBD] Document References: [TBD] As Found Results Summary: [TBD] |
Immediate Containment Checklist
- Stop use of the asset
- Label/Tag the asset as out of tolerance
- Segregate impacted work
Define Impact Window
Last known in-tolerance date: [TBD] to discovery date: [TBD]
Product/Batch/Test Impact Review
| Batch/Study/Report ID | Results Potentially Impacted | Re-test/Re-weigh Decision |
|---|---|---|
| [TBD] | [TBD] | [TBD] |
Data Integrity Review Prompts (ALCOA+)
- Attributable: [TBD]
- Legible: [TBD]
- Contemporaneous: [TBD]
- Original: [TBD]
- Accurate: [TBD]
- Complete: [TBD]
- Consistent: [TBD]
- Enduring: [TBD]
- Available: [TBD]
Risk Assessment
| Severity | Probability | Detectability | Rationale | Overall Risk |
|---|---|---|---|---|
| [TBD] | [TBD] | [TBD] | [TBD] | [TBD] |
Investigation Summary
Root Cause: [TBD]
CAPA Plan
| Action | Owner | Due Date | Effectiveness Check |
|---|---|---|---|
| [TBD] | [TBD] | [TBD] | [TBD] |
QA Disposition
Release/Hold/Retest/Recall Triggers: [TBD]
Calibration Certificate
Certificate No: [TBD]
Date: [TBD]
Next Due Date: [TBD]
Asset Details
| Asset ID | Equipment Name | Equipment Type | Dept | Area | Make/Model | Serial No | Location | Parameter/Range/Units | Acceptance Criteria |
|---|---|---|---|---|---|---|---|---|---|
| CAL-0030 | Burette 50 mL | Burette | QC | Chemistry Lab | [TBD] | [TBD] | [TBD] | Volume: 0 – 50 mL | Class A |
Conditions
Environment: [TBD]
Reference Standards
| Standard Name/ID | Certificate No | Due Date | Traceability Statement |
|---|---|---|---|
| [TBD] | [TBD] | [TBD] | Traceability: ISO17025 |
Results
As Found Results
| Test Point | Measured Value (mL) | Acceptance Criteria | Pass/Fail |
|---|---|---|---|
| 0 mL | [TBD] | Class A | [TBD] |
| 10 mL | [TBD] | Class A | [TBD] |
| 20 mL | [TBD] | Class A | [TBD] |
| 30 mL | [TBD] | Class A | [TBD] |
| 40 mL | [TBD] | Class A | [TBD] |
As Left Results
| Test Point | Measured Value (mL) | Acceptance Criteria | Pass/Fail |
|---|---|---|---|
| 0 mL | [TBD] | Class A | [TBD] |
| 10 mL | [TBD] | Class A | [TBD] |
| 20 mL | [TBD] | Class A | [TBD] |
| 30 mL | [TBD] | Class A | [TBD] |
| 40 mL | [TBD] | Class A | [TBD] |
Measurement Uncertainty Statement: [TBD]
Statement of Conformity: [Pass/Fail overall]
Remarks: [TBD]
Signatures
Performed by: ______________________ Date: ____________
Checked by: ______________________ Date: ____________
Approved by: ______________________ Date: ____________