Volumetric Flask 250 mL (CAL-0037) – Calibration Documents

Calibration Documents: Volumetric Flask 250 mL (CAL-0037)

Asset ID CAL-0037
Equipment Volumetric Flask 250 mL
Equipment Type Volumetric Flask
Department QC
Area/Location Glassware Room
Process Use Solution prep
Parameter Volume
Units mL
Range Min 0
Range Max 250
Acceptance Criteria Class A
Calibration Frequency (days) 365
Calibration Type External
Traceability ISO17025

Calibration SOP

1. Purpose

The calibration of the Volumetric Flask 250 mL is performed to ensure accurate volume measurements during solution preparation in the QC department. This calibration controls risks associated with accuracy, patient safety, batch quality, and data integrity, thereby ensuring compliance with regulatory standards.

2. Scope

  • Asset ID: CAL-0037
  • Equipment Name/Type: Volumetric Flask 250 mL / Volumetric Flask
  • Department/Area: QC / Glassware Room
  • Process Use: Solution prep
  • Parameter: Volume
  • Range: 0 – 250 mL
  • Units: mL
  • Calibration Frequency: 365 days

3. Definitions

  • Calibration: The process of comparing a measurement instrument against a standard to determine its accuracy.
  • Verification: The process of confirming that a measurement instrument meets specified requirements.
  • Adjustment: The act of modifying a measurement instrument to bring it into compliance with specifications.
  • As Found: The condition of the equipment at the time of calibration before any adjustments.
  • As Left: The condition of the equipment after calibration and any necessary adjustments.
  • OOT (Out of Tolerance): A condition where the measurement falls outside the specified acceptance criteria.
  • Drift: The gradual change in a measurement instrument’s accuracy over time.
  • Traceability: The ability to verify the history, location, or application of an item by means of documented recorded identification.
  • Measurement Uncertainty: A parameter that characterizes the range of values within which the true value is asserted to lie.
  • Intermediate Check: A verification process conducted between calibrations to ensure continued accuracy.

4. References

  • Company Metrology SOP [TBD]
  • ISO/IEC 17025 traceability via calibration certificates
  • ALCOA+ principles

5. Responsibilities

  • Technician/Metrology: Perform calibration and maintain records.
  • QC/QA Reviewer: Review calibration records and ensure compliance.
  • Owner Department: Ensure equipment is available and in good condition for calibration.
  • External Service Provider: Conduct external calibration as required.

6. Safety & Preconditions

  • Ensure environmental controls are in place (temperature, humidity).
  • Allow equipment to warm up/stabilize as per manufacturer recommendations.
  • Verify cleanliness of the volumetric flask before use.
  • Check leveling of the equipment.
  • Confirm power supply checks are completed [TBD].
See also  Rotameter (Gas) (CAL-0084) - Calibration Documents

7. Standards & Tools

  • Reference Standards Required: Class A volumetric standards.
  • Traceability Statement: Calibration traceable to ISO17025.
  • Certificate Validity Check: Ensure calibration certificates are current and valid.
  • Standard ID Placeholders: [TBD].

8. Calibration Strategy

Calibration will follow the As Found then As Left logic. The minimum number of points for calibration will be 5, including endpoints (0 mL and 250 mL) and midpoints (125 mL). Each point will have 3 replicate readings as a best practice.

9. Step-by-Step Procedure

  1. Preparation: Gather all necessary equipment and reference standards.
  2. Stabilization: Allow the volumetric flask to stabilize in the calibration environment.
  3. As Found checks and data capture: Measure the volume at each defined point and record the data contemporaneously (ALCOA+).
  4. Adjustment rules: If OOT is detected, follow adjustment procedures if allowed; otherwise, escalate to QA.
  5. As Left verification: Measure again to confirm the flask is within tolerance after any adjustments.
  6. Labeling + documentation completion: Complete all required documentation and label the equipment appropriately.

10. Acceptance Criteria & Calculations

The acceptance criteria for the volumetric flask is Class A. The error calculation formula is:

Error (%) = (Observed Value – True Value) / True Value × 100

Pass/Fail logic: If the error is within the acceptance criteria, the calibration is considered a pass; otherwise, it is a fail.

Rounding rule: [TBD].

11. Handling OOT / Failures

  • Immediate containment: Tag the equipment and stop use.
  • Notification & deviation initiation: Notify QA and initiate a deviation report.
  • Impact assessment trigger: Define the impact window and review affected results/batches.
  • Recalibration after repair/adjustment: Ensure recalibration is performed post-adjustment.
  • QA disposition placeholders: [TBD].

12. Records & Retention

  • Logbook entries: Document all calibration activities.
  • Calibration report/certificate filing: Maintain records of calibration certificates.
  • e-Records if applicable: [TBD].
  • Label content requirements: Include calibrated date, due date, by whom, and status.

13. Annexures

Annexure A: Calibration Data Sheet

Test Points Reference Observed Error Acceptance Pass/Fail As Found As Left
[TBD] [TBD] [TBD] [TBD] [TBD] [TBD] [TBD] [TBD]

Annexure B: Quick Checklist

  • Environmental controls verified.
  • Equipment warm-up completed.
  • Cleanliness confirmed.
  • Leveling checked.
  • Power supply verified.

Annexure C: Example Label Text

Calibration Date: [TBD]

Due Date: [TBD]

Calibrated By: [TBD]

Status: [TBD]

⬆ Back to TOC


Calibration Protocol

Calibration Protocol

Protocol ID Asset ID Equipment Name Equipment Type Dept Area Process Use Parameter Units Range Acceptance Criteria Frequency Calibration Type Traceability
[TBD] CAL-0037 Volumetric Flask 250 mL Volumetric Flask QC Glassware Room Solution prep Volume mL 0 – 250 Class A 365 days External ISO17025
See also  Inline Conductivity Sensor (PW) (CAL-0094) - Calibration Documents

Environmental Conditions

Temperature: [TBD]

Relative Humidity: [TBD]

Vibration: [TBD]

Drafts: [TBD]

Test Point Selection

  • Minimum: 0 mL
  • 25%: 62.5 mL
  • 50%: 125 mL
  • 75%: 187.5 mL
  • Maximum: 250 mL

Results Table

Test Point Ref Value Observed (As Found) Error Limit Pass/Fail Observed (As Left) Error Pass/Fail
0 mL [TBD] [TBD] [TBD] [TBD] [TBD] [TBD] [TBD] [TBD]
62.5 mL [TBD] [TBD] [TBD] [TBD] [TBD] [TBD] [TBD] [TBD]
125 mL [TBD] [TBD] [TBD] [TBD] [TBD] [TBD] [TBD] [TBD]
187.5 mL [TBD] [TBD] [TBD] [TBD] [TBD] [TBD] [TBD] [TBD]
250 mL [TBD] [TBD] [TBD] [TBD] [TBD] [TBD] [TBD] [TBD]

Stepwise Instructions

  1. Preparation: Ensure the volumetric flask is clean and free from any contaminants.
  2. Stabilization: Allow the flask to acclimate to the environmental conditions for at least [TBD] minutes.
  3. Performing Readings:
    • Fill the flask to each test point (0 mL, 62.5 mL, 125 mL, 187.5 mL, 250 mL).
    • Take three replicate readings for each test point.
  4. Calculating Error: Calculate the error for each test point using the formula: Error = Observed – Ref Value.
  5. Evaluation: Compare the calculated errors against the acceptance criteria (Class A).
  6. Documentation: Record all observations in the results table and ensure data integrity (ALCOA+).

Deviation Instruction for OOT

  • Stop use of the equipment immediately.
  • Notify the QA department.
  • Open a deviation report.
  • Quarantine impacted results window: [TBD]
  • Await QA decision on further actions.

Approvals

Prepared Reviewed Approved
[Signature/Date] [Signature/Date] [Signature/Date]

⬆ Back to TOC


Calibration Checklist

Asset ID Equipment Dept Area Parameter Range Acceptance Frequency Traceability
CAL-0037 Volumetric Flask 250 mL QC Glassware Room Volume 0 – 250 mL Class A 365 days ISO17025

Pre-calibration

  • [ ] Verify environmental conditions are within specified limits
  • [ ] Allow equipment to warm up as per SOP
  • [ ] Clean the volumetric flask according to cleaning SOP
  • [ ] Level the equipment if applicable
  • [ ] Confirm validity of reference standards
  • [ ] Ensure calibration logbook is ready for entries

As Found Execution

  • [ ] Set the volumetric flask to specified setpoints
  • [ ] Perform repeat measurements as required
  • [ ] Record all raw data accurately
  • [ ] Perform necessary calculations for results

OOT Immediate Actions

  • [ ] Tag out the equipment if out of tolerance (OOT)
  • [ ] Stop use of the equipment immediately
  • [ ] Notify the appropriate personnel
  • [ ] Document deviation from acceptance criteria
  • [ ] Trigger impact review process
See also  Stopwatch (CAL-0262) - Calibration Documents

As Left Verification

  • [ ] Conduct post-adjustment checks to confirm accuracy
  • [ ] Ensure all measurements are within specified limits

Documentation & Closure

  • [ ] Update equipment label with calibration details
  • [ ] Attach calibration certificate to logbook
  • [ ] Obtain reviewer sign-off on calibration documentation
  • [ ] Update next due date for calibration

Acceptance Criteria: Class A standards must be met during calibration.

Data Integrity (ALCOA+): Ensure data is Attributable, Legible, Contemporaneous, Original, Accurate, and Complete.

Traceability: Calibration must adhere to ISO17025 standards.

⬆ Back to TOC


Deviation Impact Assessment

Event Summary
Asset: Volumetric Flask 250 mL
Date: [TBD]
Discovered By: [TBD]
Document References: [TBD]
As Found Results Summary: [TBD]

Immediate Containment Checklist

  • Stop use of the asset
  • Label/Tag the asset as out of tolerance
  • Segregate impacted work

Define Impact Window

Last known in-tolerance date: [TBD] to discovery date: [TBD]

Product/Batch/Test Impact Review

Batch/Study/Report ID Results Potentially Impacted Re-test/Re-weigh Decision
[TBD] [TBD] [TBD]

Data Integrity Review Prompts (ALCOA+)

  • Attributable: [TBD]
  • Legible: [TBD]
  • Contemporaneous: [TBD]
  • Original: [TBD]
  • Accurate: [TBD]
  • Complete: [TBD]
  • Consistent: [TBD]
  • Enduring: [TBD]
  • Available: [TBD]

Risk Assessment

Severity Probability Detectability Rationale Overall Risk
[TBD] [TBD] [TBD] [TBD] [TBD]

Investigation Summary

Root Cause: [TBD]

CAPA Plan

Action Owner Due Date Effectiveness Check
[TBD] [TBD] [TBD] [TBD]

QA Disposition

Release/Hold/Retest/Recall Triggers: [TBD]

⬆ Back to TOC


Calibration Certificate

Certificate No: [TBD]

Date: [TBD]

Next Due Date: [TBD]

Asset Details

Asset ID Equipment Name Equipment Type Dept Area Make/Model Serial No Location Parameter/Range/Units Acceptance Criteria
CAL-0037 Volumetric Flask 250 mL Volumetric Flask QC Glassware Room [TBD] [TBD] [TBD] Volume: 0 – 250 mL Class A

Conditions

Environment: [TBD]

Reference Standards

Standard Name/ID Certificate No Due Date Traceability Statement
[TBD] [TBD] [TBD] Traceability: ISO17025

Results

As Found Results

Test Point Measured Value (mL) Acceptance Criteria Pass/Fail
0 mL [TBD] Class A [TBD]
50 mL [TBD] Class A [TBD]
100 mL [TBD] Class A [TBD]
150 mL [TBD] Class A [TBD]
250 mL [TBD] Class A [TBD]

As Left Results

Test Point Measured Value (mL) Acceptance Criteria Pass/Fail
0 mL [TBD] Class A [TBD]
50 mL [TBD] Class A [TBD]
100 mL [TBD] Class A [TBD]
150 mL [TBD] Class A [TBD]
250 mL [TBD] Class A [TBD]

Measurement Uncertainty Statement: [TBD]

Statement of Conformity: [Pass/Fail overall]

Remarks: [TBD]

Signatures

Performed by: ____________________ Date: ___________

Checked by: ____________________ Date: ___________

Approved by: ____________________ Date: ___________

⬆ Back to TOC