Calibration Documents: Dissolved Oxygen Sensor (Bioreactor) (CAL-0098)
| Asset ID | CAL-0098 |
|---|---|
| Equipment | Dissolved Oxygen Sensor (Bioreactor) |
| Equipment Type | DO Sensor |
| Department | Biologics |
| Area/Location | Bioreactor |
| Process Use | DO control |
| Parameter | Dissolved Oxygen |
| Units | %DO |
| Range Min | 0 |
| Range Max | 200 |
| Acceptance Criteria | ±2% |
| Calibration Frequency (days) | 180 |
| Calibration Type | External |
| Traceability | ISO17025 |
Table of Contents
Calibration SOP
1. Purpose
The calibration of the Dissolved Oxygen Sensor (Bioreactor) is performed to ensure accurate measurement of dissolved oxygen levels, which is critical for maintaining optimal conditions in bioprocessing. This calibration controls risks associated with inaccurate readings that could compromise patient safety, batch quality, and data integrity.
2. Scope
- Asset ID: CAL-0098
- Equipment Name/Type: Dissolved Oxygen Sensor (DO Sensor)
- Department/Area: Biologics/Bioreactor
- Process Use: DO control
- Parameter: Dissolved Oxygen
- Range: 0 – 200 %DO
- Units: %DO
- Calibration Frequency: 180 days
3. Definitions
- Calibration: The process of comparing a measurement instrument against a known standard to determine its accuracy.
- Verification: The process of confirming that a measurement instrument is functioning correctly.
- Adjustment: The act of correcting a measurement instrument to ensure it meets specified standards.
- As Found: The condition of the instrument prior to any adjustments.
- As Left: The condition of the instrument after adjustments have been made.
- OOT (Out of Tolerance): A condition where the measurement falls outside the specified acceptance criteria.
- Drift: The gradual change in the output of a measurement instrument over time.
- Traceability: The ability to verify the history, location, or application of an item by means of documented recorded identification.
- Measurement Uncertainty: A parameter that characterizes the range of values within which the true value is expected to lie.
- Intermediate Check: A periodic check performed to ensure the instrument remains within specified limits between calibrations.
4. References
- Company Metrology SOP [TBD]
- ISO/IEC 17025 traceability via calibration certificates
- ALCOA+ principles
5. Responsibilities
- Technician/Metrology: Perform calibration and document results.
- QC/QA Reviewer: Review calibration documentation and results.
- Owner Department: Ensure equipment is available and maintained.
- External Service Provider: Conduct external calibration as required.
6. Safety & Preconditions
- Ensure environmental controls are in place (temperature, humidity).
- Allow for warm-up/stabilization of the sensor as per manufacturer guidelines.
- Maintain cleanliness of the equipment and surrounding area.
- Check leveling of the equipment.
- Verify power supply is stable and within specifications.
7. Standards & Tools
- Reference standards required: [TBD]
- Traceability statement: ISO17025
- Certificate validity check: Ensure all certificates are current and valid.
- Standard ID placeholders: [TBD]
8. Calibration Strategy
Calibration will follow the As Found then As Left logic. The minimum number of calibration points will be five, including endpoints and midpoints. Each point will have three replicate readings to ensure accuracy.
9. Step-by-Step Procedure
- Preparation: Gather all necessary tools and standards.
- Stabilization: Allow the sensor to stabilize as per manufacturer instructions.
- As Found checks and data capture: Record initial readings before any adjustments.
- Adjustment rules: If OOT is found, adjustments may be made if permitted; otherwise, escalate to QA.
- As Left verification: Confirm that the sensor is within tolerance after adjustments.
- Labeling + documentation completion: Ensure all records are filled out accurately and contemporaneously (ALCOA+).
10. Acceptance Criteria & Calculations
Acceptance criteria: ±2% from the true value. The error calculation formula is:
Error (%) = (Observed Value – True Value) / True Value * 100
Pass/Fail logic: If the error is within ±2%, the calibration passes; otherwise, it fails.
Rounding rule: [TBD]
11. Handling OOT / Failures
- Immediate containment: Tag equipment and stop use.
- Notification & deviation initiation: Notify QA and initiate deviation process.
- Impact assessment trigger: Define impact window and review affected results/batches.
- Recalibration after repair/adjustment: Ensure recalibration is performed post-adjustment.
- QA disposition placeholders: [TBD]
12. Records & Retention
- Logbook entries: Document all calibration activities.
- Calibration report/certificate filing: Maintain records as per company policy.
- e-Records if applicable: [TBD]
- Label content requirements: Include calibrated date, due date, technician name, and status.
13. Annexures
Annexure A: Calibration Data Sheet
| Test Points | Reference | Observed | Error | Acceptance | Pass/Fail | As Found | As Left |
|---|---|---|---|---|---|---|---|
| [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] |
Annexure B: Quick Checklist
- Environmental controls verified
- Equipment clean and leveled
- Power supply checked
- Calibration standards available
Annexure C: Example Label Text
Calibration Date: [TBD]
Due Date: [TBD]
Calibrated By: [TBD]
Status: [TBD]
Calibration Protocol
Calibration Protocol
| Protocol ID | Asset ID | Equipment Name | Equipment Type | Dept | Area | Process Use | Parameter | Units | Range | Acceptance Criteria | Frequency | Calibration Type | Traceability |
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| [TBD] | CAL-0098 | Dissolved Oxygen Sensor (Bioreactor) | DO Sensor | Biologics | Bioreactor | DO control | Dissolved Oxygen | %DO | 0 – 200 | ±2% | 180 days | External | ISO17025 |
Environmental Conditions
Temperature: [TBD]
Relative Humidity: [TBD]
Vibration: [TBD]
Drafts: [TBD]
Test Point Selection
- Minimum: 0 %DO
- 25% of Range: 50 %DO
- 50% of Range: 100 %DO
- 75% of Range: 150 %DO
- Maximum: 200 %DO
Results Table
| Test Point | Ref Value | Observed (As Found) | Error | Limit | Pass/Fail | Observed (As Left) | Error | Pass/Fail |
|---|---|---|---|---|---|---|---|---|
| 0 %DO | [TBD] | [TBD] | [TBD] | ±2% | [TBD] | [TBD] | [TBD] | [TBD] |
| 50 %DO | [TBD] | [TBD] | [TBD] | ±2% | [TBD] | [TBD] | [TBD] | [TBD] |
| 100 %DO | [TBD] | [TBD] | [TBD] | ±2% | [TBD] | [TBD] | [TBD] | [TBD] |
| 150 %DO | [TBD] | [TBD] | [TBD] | ±2% | [TBD] | [TBD] | [TBD] | [TBD] |
| 200 %DO | [TBD] | [TBD] | [TBD] | ±2% | [TBD] | [TBD] | [TBD] | [TBD] |
Stepwise Instructions
- Preparation: Ensure the Dissolved Oxygen Sensor is clean and calibrated according to the manufacturer’s specifications.
- Stabilization: Allow the sensor to stabilize in the calibration environment for at least [TBD] minutes.
- Performing Readings: Take three replicate readings at each test point to ensure accuracy.
- Calculating Error: Calculate the error for each observed value using the formula: Error = Observed – Ref Value.
- Evaluation: Compare the calculated error against the acceptance criteria of ±2%.
- Documentation: Record all observations, errors, and pass/fail results in the results table.
Deviation Instruction for OOT
If any observed value falls outside the acceptance criteria:
- Stop use of the equipment immediately.
- Notify the QA department.
- Open a deviation report.
- Quarantine the impacted results window: [TBD].
- Await QA decision before proceeding.
Approvals
| Prepared | Reviewed | Approved |
|---|---|---|
| [Signature/Date] | [Signature/Date] | [Signature/Date] |
Calibration Checklist
| Asset ID | Equipment | Dept | Area | Parameter | Range | Acceptance | Frequency | Traceability |
|---|---|---|---|---|---|---|---|---|
| CAL-0098 | Dissolved Oxygen Sensor (Bioreactor) | Biologics | Bioreactor | Dissolved Oxygen | 0 – 200 %DO | ±2% | 180 days | ISO17025 |
Pre-calibration
- [ ] Verify environmental conditions are within specified limits
- [ ] Allow sensor to warm up for the recommended time
- [ ] Clean the sensor as per SOP
- [ ] Level the sensor if applicable
- [ ] Check validity of reference standards
- [ ] Ensure logbook is ready for documentation
As Found Execution
- [ ] Set the sensor to specified setpoints
- [ ] Perform repeat measurements as required
- [ ] Record all raw data accurately
- [ ] Perform necessary calculations for results
OOT Immediate Actions
- [ ] Tag out the equipment
- [ ] Stop use of the sensor
- [ ] Notify relevant personnel
- [ ] Document deviation
- [ ] Trigger impact review
As Left Verification
- [ ] Conduct post-adjustment checks
- [ ] Confirm all results are within acceptance limits
Documentation & Closure
- [ ] Update equipment label
- [ ] Attach calibration certificate
- [ ] Obtain reviewer sign-off
- [ ] Update next due date in logbook
Deviation Impact Assessment
| Event Summary |
|---|
|
Asset: Dissolved Oxygen Sensor (Bioreactor) Date: [TBD] Discovered By: [TBD] Document References: [TBD] As Found Results Summary: [TBD] |
Immediate Containment Checklist
- Stop use of the asset
- Label/tag the asset as out of tolerance
- Segregate impacted work
Define Impact Window
Last known in-tolerance date: [TBD] to discovery date: [TBD]
Product/Batch/Test Impact Review
| Batch/Study/Report ID | Results Potentially Impacted | Re-test/Re-weigh Decision |
|---|---|---|
| [TBD] | [TBD] | [TBD] |
Data Integrity Review Prompts (ALCOA+)
- Attributable: [TBD]
- Legible: [TBD]
- Contemporaneous: [TBD]
- Original: [TBD]
- Accurate: [TBD]
- Complete: [TBD]
- Consistent: [TBD]
- Enduring: [TBD]
- Available: [TBD]
Risk Assessment
| Severity | Probability | Detectability | Rationale | Overall Risk |
|---|---|---|---|---|
| [TBD] | [TBD] | [TBD] | [TBD] | [TBD] |
Investigation Summary
Root Cause: [TBD]
CAPA Plan
| Action | Owner | Due Date | Effectiveness Check |
|---|---|---|---|
| [TBD] | [TBD] | [TBD] | [TBD] |
QA Disposition
Release/Hold/Retest/Recall Triggers: [TBD]
Calibration Certificate
Certificate No: [TBD]
Date: [TBD]
Next Due Date: [TBD]
Asset Details
| Asset ID | Equipment Name | Equipment Type | Dept | Area | Make/Model | Serial No | Location | Parameter/Range/Units | Acceptance Criteria |
|---|---|---|---|---|---|---|---|---|---|
| CAL-0098 | Dissolved Oxygen Sensor (Bioreactor) | DO Sensor | Biologics | Bioreactor | [TBD] | [TBD] | [TBD] | Dissolved Oxygen / 0-200 %DO | ±2% |
Conditions
Environment: [TBD]
Reference Standards
| Standard Name/ID | Certificate No | Due Date | Traceability Statement |
|---|---|---|---|
| [TBD] | [TBD] | [TBD] | Traceability: ISO17025 |
Results
As Found Results
| Test Point | Measured Value (%DO) | Pass/Fail |
|---|---|---|
| 1 | [TBD] | [TBD] |
| 2 | [TBD] | [TBD] |
| 3 | [TBD] | [TBD] |
| 4 | [TBD] | [TBD] |
| 5 | [TBD] | [TBD] |
As Left Results
| Test Point | Measured Value (%DO) | Pass/Fail |
|---|---|---|
| 1 | [TBD] | [TBD] |
| 2 | [TBD] | [TBD] |
| 3 | [TBD] | [TBD] |
| 4 | [TBD] | [TBD] |
| 5 | [TBD] | [TBD] |
Measurement Uncertainty Statement: [TBD]
Statement of Conformity: [Pass/Fail overall]
Remarks: [TBD]
Signatures
Performed by: ______________________ Date: ___________
Checked by: ______________________ Date: ___________
Approved by: ______________________ Date: ___________