GC System (CAL-0122) – Calibration Documents

Calibration Documents: GC System (CAL-0122)

Asset ID CAL-0122
Equipment GC System
Equipment Type GC
Department QC
Area/Location Instrument Room
Process Use Chromatography
Parameter Temperature
Units °C
Range Min 40
Range Max 400
Acceptance Criteria ±1°C
Calibration Frequency (days) 365
Calibration Type External
Traceability ISO17025

Calibration SOP

1. Purpose

The calibration of the GC System is performed to ensure accurate temperature measurements during chromatography processes. This calibration controls risks related to data integrity, patient safety, and batch quality by ensuring that the equipment operates within specified limits.

2. Scope

  • Asset ID: CAL-0122
  • Equipment Name/Type: GC System / GC
  • Department/Area: QC / Instrument Room
  • Process Use: Chromatography
  • Parameter: Temperature
  • Range: 40°C to 400°C
  • Units: °C
  • Calibration Frequency: 365 days

3. Definitions

  • Calibration: The process of configuring an instrument to provide a result for a sample within an acceptable range.
  • Verification: The process of confirming that the calibration of an instrument is still valid.
  • Adjustment: The process of modifying an instrument to bring it into compliance with specified standards.
  • As Found: The condition of the instrument prior to any adjustments.
  • As Left: The condition of the instrument after adjustments have been made.
  • OOT (Out of Tolerance): A condition where the measurement exceeds the acceptance criteria.
  • Drift: The gradual change in an instrument’s output over time.
  • Traceability: The ability to verify the history, location, or application of an item by means of documented recorded identification.
  • Measurement Uncertainty: A parameter that characterizes the range of values within which the true value is expected to lie.
  • Intermediate Check: A periodic check performed between calibrations to ensure continued accuracy.

4. References

  • Company Metrology SOP [TBD]
  • ISO/IEC 17025 traceability via calibration certificates
  • ALCOA+ principles

5. Responsibilities

  • Technician/Metrology: Perform calibration and document results.
  • QC/QA Reviewer: Review calibration documentation and results.
  • Owner Department: Ensure equipment is available and in good condition for calibration.
  • External Service Provider: Conduct external calibration as required.

6. Safety & Preconditions

  • Ensure environmental controls are in place (temperature, humidity).
  • Allow equipment to warm up and stabilize before calibration.
  • Maintain cleanliness of the instrument and surrounding area.
  • Check that the instrument is level and properly positioned.
  • Verify power supply is stable and within specifications.
See also  Micropipette 1-5 mL (CAL-0024) - Calibration Documents

7. Standards & Tools

  • Reference standards required: [TBD]
  • Traceability statement: ISO17025
  • Certificate validity check: [TBD]
  • Standard ID placeholders: [TBD]

8. Calibration Strategy

Calibration will follow the As Found then As Left logic. The minimum number of points for calibration will be five, including endpoints and midpoints. It is recommended to take three replicate readings at each point to ensure accuracy.

9. Step-by-Step Procedure

  1. Preparation: Gather all necessary tools and standards.
  2. Stabilization: Allow the GC System to reach operational temperature.
  3. As Found checks and data capture: Measure temperature at defined points and record results contemporaneously.
  4. Adjustment rules: If OOT is detected, escalate for adjustment as per SOP.
  5. As Left verification: Confirm that the instrument is within tolerance after adjustments.
  6. Labeling + documentation completion: Complete all required documentation and label the equipment.

10. Acceptance Criteria & Calculations

Acceptance criteria: ±1°C. The error calculation formula is: Error = Observed Value – True Value. Pass/Fail logic: If the absolute value of the error is within the acceptance criteria, the calibration passes; otherwise, it fails. Rounding rule: [TBD].

11. Handling OOT / Failures

  • Immediate containment: Tag equipment and stop use.
  • Notification & deviation initiation: Notify relevant personnel.
  • Impact assessment trigger: Define impact window and review of affected results/batches.
  • Recalibration after repair/adjustment: Ensure recalibration is performed post-adjustment.
  • QA disposition placeholders: [TBD].

12. Records & Retention

  • Logbook entries: Document all calibration activities.
  • Calibration report/certificate filing: Maintain records as per company policy.
  • e-records if applicable: [TBD].
  • Label content requirements: Calibrated date, due date, by whom, status.

13. Annexures

Annexure A: Calibration Data Sheet

Test Points Reference Observed Error Acceptance Pass/Fail As Found As Left
[TBD] [TBD] [TBD] [TBD] [TBD] [TBD] [TBD] [TBD]

Annexure B: Quick Checklist

  • Environmental controls verified
  • Equipment warmed up
  • Cleanliness checked
  • Leveling confirmed
  • Power supply stable

Annexure C: Example Label Text

Calibration Date: [TBD] | Due Date: [TBD] | Calibrated By: [TBD] | Status: [TBD]

⬆ Back to TOC


Calibration Protocol

Calibration Protocol

Protocol ID Asset ID Equipment Name Equipment Type Dept Area Process Use Parameter Units Range Acceptance Criteria Frequency Calibration Type Traceability
[TBD] CAL-0122 GC System GC QC Instrument Room Chromatography Temperature °C 40 – 400 ±1°C 365 days External ISO17025
See also  IR Thermometer (CAL-0054) - Calibration Documents

Environmental Conditions

Temperature: [TBD] °C

Relative Humidity: [TBD] %

Vibration: [TBD]

Drafts: [TBD]

Test Point Selection

  • Minimum: 40 °C
  • 25% of Range: 135 °C
  • 50% of Range: 220 °C
  • 75% of Range: 305 °C
  • Maximum: 400 °C

Results Table

Test Point Ref Value Observed (As Found) Error Limit Pass/Fail Observed (As Left) Error Pass/Fail
40 °C [TBD] [TBD] [TBD] ±1°C [TBD] [TBD] [TBD] [TBD]
135 °C [TBD] [TBD] [TBD] ±1°C [TBD] [TBD] [TBD] [TBD]
220 °C [TBD] [TBD] [TBD] ±1°C [TBD] [TBD] [TBD] [TBD]
305 °C [TBD] [TBD] [TBD] ±1°C [TBD] [TBD] [TBD] [TBD]
400 °C [TBD] [TBD] [TBD] ±1°C [TBD] [TBD] [TBD] [TBD]

Stepwise Instructions

  1. Preparation: Ensure the GC System is clean and properly set up in the Instrument Room.
  2. Stabilization: Allow the system to stabilize for at least [TBD] minutes before taking measurements.
  3. Performing Readings: Take three replicate readings at each test point to ensure accuracy.
  4. Calculating Error: Calculate the error for each test point using the formula: Error = Observed – Ref Value.
  5. Evaluation: Compare the calculated error against the acceptance criteria of ±1°C.
  6. Documentation: Record all observations and calculations in the results table.

Deviation Instruction for OOT

In the event of an Out of Tolerance (OOT) condition:

  • Stop use of the GC System immediately.
  • Notify the Quality Assurance (QA) department.
  • Open a deviation report.
  • Quarantine impacted results window: [TBD]
  • Await QA decision on further actions.

Approvals

Prepared By Reviewed By Approved By
[Signature/Date] [Signature/Date] [Signature/Date]

⬆ Back to TOC


Calibration Checklist

Asset ID Equipment Dept Area Parameter Range Acceptance Frequency Traceability
CAL-0122 GC System QC Instrument Room Temperature 40 – 400 °C ±1°C 365 days ISO17025

Pre-calibration

  • [ ] Verify environmental conditions are within specified limits
  • [ ] Allow GC System to warm up for the recommended duration
  • [ ] Clean the equipment as per SOP
  • [ ] Level the GC System if required
  • [ ] Check validity of reference standards
  • [ ] Ensure logbook is ready for documentation

As Found Execution

  • [ ] Record initial setpoints
  • [ ] Perform repeat measurements
  • [ ] Record raw data accurately
  • [ ] Perform necessary calculations
See also  CO2 Incubator CO2 Sensor (CAL-0305) - Calibration Documents

OOT Immediate Actions

  • [ ] Tag out the equipment
  • [ ] Stop use of the GC System
  • [ ] Notify the supervisor
  • [ ] Document deviation
  • [ ] Trigger impact review

As Left Verification

  • [ ] Conduct post-adjustment checks
  • [ ] Confirm all parameters are within acceptance limits

Documentation & Closure

  • [ ] Update equipment label
  • [ ] Attach calibration certificate
  • [ ] Obtain reviewer sign-off
  • [ ] Update next due date in logbook

Data Integrity (ALCOA+): Ensure all data is attributable, legible, contemporaneous, original, and accurate. Maintain records in compliance with GMP standards.

⬆ Back to TOC


Deviation Impact Assessment

Event Summary
Asset: GC System
Date: [TBD]
Discovered By: [TBD]
Document References: [TBD]
As Found Results Summary: [TBD]

Immediate Containment Checklist

  • Stop use of the equipment
  • Label/tag the equipment as out of service
  • Segregate impacted work

Define Impact Window

Last Known In-Tolerance Date Discovery Date
[TBD] [TBD]

Product/Batch/Test Impact Review

Batch/Study/Report ID Results Potentially Impacted Re-Test/Re-Weigh Decision
[TBD] [TBD] [TBD]

Data Integrity Review Prompts (ALCOA+)

  • Attributable: [TBD]
  • Legible: [TBD]
  • Contemporaneous: [TBD]
  • Original: [TBD]
  • Accurate: [TBD]
  • Complete: [TBD]
  • Consistent: [TBD]
  • Enduring: [TBD]
  • Available: [TBD]

Risk Assessment

Severity Probability Detectability Rationale Overall Risk
[TBD] [TBD] [TBD] [TBD] [TBD]

Investigation Summary

Root Cause: [TBD]

CAPA Plan

Action Owner Due Date Effectiveness Check
[TBD] [TBD] [TBD] [TBD]

QA Disposition

Release/Hold/Retest/Recall Triggers: [TBD]

⬆ Back to TOC


Calibration Certificate

Certificate No: [TBD]

Date: [TBD]

Next Due Date: [TBD]

Asset Details

Asset ID Equipment Name Equipment Type Dept Area Make/Model Serial No Location Parameter/Range/Units Acceptance Criteria
CAL-0122 GC System GC QC Instrument Room [TBD] [TBD] [TBD] Temperature: 40 – 400 °C ±1°C

Conditions

Environment: [TBD]

Reference Standards

Standard Name/ID Certificate No Due Date Traceability Statement
[TBD] [TBD] [TBD] Traceability: ISO17025

Results

As Found Results

Test Point Measured Value (°C) Pass/Fail
1 [TBD] [TBD]
2 [TBD] [TBD]
3 [TBD] [TBD]
4 [TBD] [TBD]
5 [TBD] [TBD]

As Left Results

Test Point Measured Value (°C) Pass/Fail
1 [TBD] [TBD]
2 [TBD] [TBD]
3 [TBD] [TBD]
4 [TBD] [TBD]
5 [TBD] [TBD]

Measurement Uncertainty Statement: [TBD]

Statement of Conformity: [Pass/Fail overall]

Remarks: [TBD]

Signatures

Performed by: ____________________ Date: ___________

Checked by: ____________________ Date: ___________

Approved by: ____________________ Date: ___________

⬆ Back to TOC