Calibration Documents: Dissolution Tester Apparatus 1/2 (CAL-0130)
| Asset ID | CAL-0130 |
|---|---|
| Equipment | Dissolution Tester Apparatus 1/2 |
| Equipment Type | Dissolution Tester |
| Department | QC |
| Area/Location | Instrument Room |
| Process Use | Dissolution |
| Parameter | Temperature |
| Units | °C |
| Range Min | 20 |
| Range Max | 50 |
| Acceptance Criteria | ±0.5°C |
| Calibration Frequency (days) | 365 |
| Calibration Type | External |
| Traceability | ISO17025 |
Table of Contents
Calibration SOP
1. Purpose
The calibration of the Dissolution Tester Apparatus 1/2 is performed to ensure accurate temperature measurements during dissolution testing. This is critical for maintaining patient safety, ensuring batch quality, and preserving data integrity. Accurate temperature control is essential to meet regulatory requirements and to avoid any potential risks associated with inaccurate dissolution results.
2. Scope
- Asset ID: CAL-0130
- Equipment Name/Type: Dissolution Tester Apparatus 1/2 / Dissolution Tester
- Department/Area: QC / Instrument Room
- Process Use: Dissolution
- Parameter: Temperature
- Range: 20°C to 50°C
- Units: °C
- Calibration Frequency: 365 days
3. Definitions
- Calibration: The process of verifying and adjusting the accuracy of a measurement instrument.
- Verification: The confirmation that a measurement instrument meets specified requirements.
- Adjustment: The process of modifying a measurement instrument to bring it into compliance with specified standards.
- As Found: The condition of the instrument before any adjustments are made.
- As Left: The condition of the instrument after adjustments have been made.
- OOT (Out of Tolerance): A condition where a measurement falls outside the specified acceptance criteria.
- Drift: The gradual change in a measurement instrument’s output over time.
- Traceability: The ability to verify the history, location, or application of an item by means of documented recorded identification.
- Measurement Uncertainty: A parameter that characterizes the range of values within which the true value of a measurement lies.
- Intermediate Check: A periodic verification of an instrument’s performance between full calibrations.
4. References
- Company Metrology SOP [TBD]
- ISO/IEC 17025 traceability via calibration certificates
- ALCOA+ principles
5. Responsibilities
- Technician/Metrology: Perform calibration and maintain records.
- QC/QA Reviewer: Review calibration results and documentation.
- Owner Department: Ensure equipment is available and in good condition for calibration.
- External Service Provider: Conduct external calibration as required.
6. Safety & Preconditions
- Ensure environmental controls are in place (temperature, humidity).
- Allow for warm-up/stabilization of the equipment as per manufacturer guidelines.
- Maintain cleanliness of the instrument and surrounding area.
- Check that the instrument is level before calibration.
- Verify power supply is stable and within specifications.
7. Standards & Tools
- Reference standards required: [TBD]
- Traceability statement: ISO17025
- Certificate validity check: [TBD]
- Standard ID placeholders: [TBD]
8. Calibration Strategy
The calibration will follow the As Found and As Left logic. A minimum of 5 points will be taken across the specified range of 20°C to 50°C, including endpoints and midpoints. It is recommended to take 3 replicate readings at each point to ensure accuracy, unless otherwise specified.
9. Step-by-Step Procedure
- Preparation: Gather all necessary tools and reference standards.
- Stabilization: Allow the instrument to stabilize at the operating temperature.
- As Found checks and data capture: Record the temperature readings at the defined points.
- Adjustment rules: If any readings are OOT, escalate for adjustment as per SOP.
- As Left verification: Confirm that the instrument is within the acceptance criteria after adjustments.
- Labeling + documentation completion: Ensure all calibration records are filled out accurately.
Record all raw observations contemporaneously to maintain ALCOA+ compliance.
10. Acceptance Criteria & Calculations
The acceptance criteria for the temperature parameter is ±0.5°C. The error calculation formula is:
Error = Observed Value – True Value
Pass/Fail Logic: If the absolute value of the error is within the acceptance criteria, the calibration passes; otherwise, it fails.
Rounding rule: [TBD]
11. Handling OOT / Failures
- Immediate containment: Tag the equipment and stop its use.
- Notification & deviation initiation: Notify the relevant personnel.
- Impact assessment trigger: Define the impact window and review affected results/batches.
- Recalibration after repair/adjustment: Ensure recalibration is performed post-adjustment.
- QA disposition placeholders: [TBD]
12. Records & Retention
- Logbook entries: Maintain detailed records of all calibration activities.
- Calibration report/certificate filing: Store all calibration certificates securely.
- e-Records if applicable: [TBD]
- Label content requirements: Include calibrated date, due date, technician name, and status.
13. Annexures
Annexure A: Calibration Data Sheet
| Test Points | Reference | Observed | Error | Acceptance | Pass/Fail | As Found | As Left |
|---|---|---|---|---|---|---|---|
| [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] |
Annexure B: Quick Checklist
- Environmental controls verified
- Instrument warm-up completed
- Cleanliness checked
- Leveling confirmed
- Power supply stable
Annexure C: Example Label Text
Calibration Date: [TBD]
Due Date: [TBD]
Calibrated By: [TBD]
Status: [TBD]
Calibration Protocol
Calibration Protocol
| Protocol ID | Asset ID | Equipment Name | Equipment Type | Dept | Area | Process Use | Parameter | Units | Range | Acceptance Criteria | Frequency | Calibration Type | Traceability |
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| [TBD] | CAL-0130 | Dissolution Tester Apparatus 1/2 | Dissolution Tester | QC | Instrument Room | Dissolution | Temperature | °C | 20 – 50 | ±0.5°C | 365 days | External | ISO17025 |
Environmental Conditions
Temperature: [TBD]
Relative Humidity: [TBD]
Vibration: [TBD]
Drafts: [TBD]
Test Point Selection
- Minimum: 20°C
- 25%: 27.5°C
- 50%: 35°C
- 75%: 42.5°C
- Maximum: 50°C
Results Table
| Test Point | Ref Value | Observed (As Found) | Error | Limit | Pass/Fail | Observed (As Left) | Error | Pass/Fail |
|---|---|---|---|---|---|---|---|---|
| 20°C | [TBD] | [TBD] | [TBD] | ±0.5°C | [TBD] | [TBD] | [TBD] | [TBD] |
| 27.5°C | [TBD] | [TBD] | [TBD] | ±0.5°C | [TBD] | [TBD] | [TBD] | [TBD] |
| 35°C | [TBD] | [TBD] | [TBD] | ±0.5°C | [TBD] | [TBD] | [TBD] | [TBD] |
| 42.5°C | [TBD] | [TBD] | [TBD] | ±0.5°C | [TBD] | [TBD] | [TBD] | [TBD] |
| 50°C | [TBD] | [TBD] | [TBD] | ±0.5°C | [TBD] | [TBD] | [TBD] | [TBD] |
Stepwise Instructions
- Preparation: Ensure the Dissolution Tester is clean and calibrated prior to use.
- Stabilization: Allow the equipment to stabilize at the ambient temperature for at least 30 minutes.
- Performing Readings:
- Set the temperature to each test point.
- Allow the temperature to stabilize for 10 minutes.
- Take three replicate readings at each test point.
- Calculating Error: Calculate the error for each observed value against the reference value.
- Evaluation: Compare the calculated error to the acceptance criteria.
- Documentation: Record all observations in the results table and ensure data integrity (ALCOA+).
Deviation Instruction for OOT
If any observed value is outside the acceptance criteria:
- Stop use of the equipment immediately.
- Notify the supervisor and quality assurance.
- Open a deviation report.
- Quarantine impacted results window: [TBD]
- Await QA decision on further actions.
Approvals
| Prepared | Reviewed | Approved |
|---|---|---|
| [Signature/Date] | [Signature/Date] | [Signature/Date] |
Calibration Checklist
| Asset ID | Equipment | Dept | Area | Parameter | Range | Acceptance | Frequency | Traceability |
|---|---|---|---|---|---|---|---|---|
| CAL-0130 | Dissolution Tester Apparatus 1/2 | QC | Instrument Room | Temperature | 20 – 50 °C | ±0.5°C | 365 days | ISO17025 |
Pre-calibration
- [ ] Verify environmental conditions are within specified limits
- [ ] Ensure equipment has warmed up adequately
- [ ] Clean the apparatus as per SOP
- [ ] Level the equipment
- [ ] Confirm validity of reference standards
- [ ] Logbook is ready for entries
As Found Execution
- [ ] Set the temperature to specified setpoints
- [ ] Perform necessary repeats for accuracy
- [ ] Record all raw data accurately
- [ ] Perform calculations as required
OOT Immediate Actions
- [ ] Tag out the equipment
- [ ] Stop use of the apparatus
- [ ] Notify relevant personnel
- [ ] Document deviation
- [ ] Trigger impact review
As Left Verification
- [ ] Conduct post-adjustment checks
- [ ] Confirm all parameters are within limits
Documentation & Closure
- [ ] Update equipment label
- [ ] Attach calibration certificate
- [ ] Obtain reviewer sign-off
- [ ] Update next due date in logbook
Data Integrity (ALCOA+): Ensure all data recorded is Attributable, Legible, Contemporaneous, Original, Accurate, and Complete.
Deviation Impact Assessment
| Event Summary |
|---|
|
Asset: Dissolution Tester Apparatus 1/2 Date: [TBD] Discovered By: [TBD] Document References: [TBD] As Found Results Summary: [TBD] |
Immediate Containment Checklist
- Stop use of the affected equipment
- Label/tag the equipment as out of service
- Segregate impacted work
Define Impact Window
| Last Known In-Tolerance Date | Discovery Date |
|---|---|
| [TBD] | [TBD] |
Product/Batch/Test Impact Review
| Batch/Study/Report ID | Results Potentially Impacted | Re-Test/Re-Weigh Decision |
|---|---|---|
| [TBD] | [TBD] | [TBD] |
Data Integrity Review Prompts (ALCOA+)
- Attributable: [TBD]
- Legible: [TBD]
- Contemporaneous: [TBD]
- Original: [TBD]
- Accurate: [TBD]
- Complete: [TBD]
- Consistent: [TBD]
- Enduring: [TBD]
- Available: [TBD]
Risk Assessment
| Severity | Probability | Detectability | Rationale | Overall Risk |
|---|---|---|---|---|
| [TBD] | [TBD] | [TBD] | [TBD] | [TBD] |
Investigation Summary
Root Cause: [TBD]
CAPA Plan
| Action | Owner | Due Date | Effectiveness Check |
|---|---|---|---|
| [TBD] | [TBD] | [TBD] | [TBD] |
QA Disposition
Release/Hold/Re-Test/Recall Triggers: [TBD]
Calibration Certificate
Certificate No: [TBD]
Date: [TBD]
Next Due Date: [TBD]
Asset Details
| Asset ID | Equipment Name | Equipment Type | Dept | Area | Make/Model | Serial No | Location | Parameter/Range/Units | Acceptance Criteria |
|---|---|---|---|---|---|---|---|---|---|
| CAL-0130 | Dissolution Tester Apparatus 1/2 | Dissolution Tester | QC | Instrument Room | [TBD] | [TBD] | [TBD] | Temperature: 20 – 50 °C | ±0.5°C |
Conditions
Environment: [TBD]
Reference Standards
| Standard Name/ID | Certificate No | Due Date | Traceability Statement |
|---|---|---|---|
| [TBD] | [TBD] | [TBD] | Traceability: ISO17025 |
Results
As Found Results
| Test Point | Measured Value (°C) | Acceptance Criteria |
|---|---|---|
| 1 | [TBD] | ±0.5°C |
| 2 | [TBD] | ±0.5°C |
| 3 | [TBD] | ±0.5°C |
| 4 | [TBD] | ±0.5°C |
| 5 | [TBD] | ±0.5°C |
As Left Results
| Test Point | Measured Value (°C) | Acceptance Criteria |
|---|---|---|
| 1 | [TBD] | ±0.5°C |
| 2 | [TBD] | ±0.5°C |
| 3 | [TBD] | ±0.5°C |
| 4 | [TBD] | ±0.5°C |
| 5 | [TBD] | ±0.5°C |
Measurement Uncertainty Statement: [TBD]
Statement of Conformity: [Pass/Fail overall]
Remarks: [TBD]
Signatures
Performed by: ____________________ Date: ___________
Checked by: ______________________ Date: ___________
Approved by: _____________________ Date: ___________