Inline Vacuum Transmitter (CAL-0202) – Calibration Documents

Calibration Documents: Inline Vacuum Transmitter (CAL-0202)

Asset ID CAL-0202
Equipment Inline Vacuum Transmitter
Equipment Type Vaccum Transmitter
Department API
Area/Location Dryer
Process Use Vacuum control
Parameter Pressure
Units bar
Range Min -1
Range Max 0
Acceptance Criteria ±1%
Calibration Frequency (days) 365
Calibration Type External
Traceability ISO17025

Calibration SOP

1. Purpose

The calibration of the Inline Vacuum Transmitter is performed to ensure accurate pressure measurements within the specified range. This is critical for maintaining vacuum control during the API production process, thereby safeguarding patient safety, ensuring batch quality, and preserving data integrity.

2. Scope

  • Asset ID: CAL-0202
  • Equipment Name/Type: Inline Vacuum Transmitter / Vacuum Transmitter
  • Department/Area: API / Dryer
  • Process Use: Vacuum control
  • Parameter: Pressure
  • Range: -1 to 0 bar
  • Units: bar
  • Calibration Frequency: 365 days

3. Definitions

  • Calibration: The process of configuring an instrument to provide a result for a sample within an acceptable range.
  • Verification: The process of confirming that an instrument meets specified requirements.
  • Adjustment: The process of modifying an instrument to bring it into compliance with specifications.
  • As Found: The condition of the instrument prior to any adjustments.
  • As Left: The condition of the instrument after adjustments have been made.
  • OOT (Out of Tolerance): A condition where the instrument’s measurements fall outside the specified acceptance criteria.
  • Drift: The gradual change in an instrument’s output over time.
  • Traceability: The ability to verify the history, location, or application of an item by means of documented recorded identification.
  • Measurement Uncertainty: A parameter associated with the result of a measurement that characterizes the range of values within which the true value is asserted to lie.
  • Intermediate Check: A verification process conducted between calibrations to ensure ongoing accuracy.

4. References

  • Company Metrology SOP [TBD]
  • ISO/IEC 17025 traceability via calibration certificates
  • ALCOA+ principles

5. Responsibilities

  • Technician/Metrology: Perform calibration and document results.
  • QC/QA Reviewer: Review calibration documentation and results.
  • Owner Department: Ensure equipment is available and in good condition for calibration.
  • External Service Provider: Conduct calibration as per external calibration type.

6. Safety & Preconditions

  • Ensure environmental controls are in place (temperature, humidity).
  • Allow for warm-up/stabilization of the equipment as per manufacturer specifications.
  • Maintain cleanliness of the equipment and surrounding area.
  • Check leveling of the equipment.
  • Verify power supply is stable and within specifications.
See also  Inline Flow Meter (Solvent) (CAL-0203) - Calibration Documents

7. Standards & Tools

  • Reference standards required: [TBD]
  • Traceability statement: ISO17025
  • Certificate validity check: [TBD]
  • Standard ID placeholders: [TBD]

8. Calibration Strategy

Calibration will follow the As Found then As Left logic. The minimum number of test points will be five, including endpoints and midpoints within the specified range of -1 to 0 bar. It is recommended to take three replicate readings at each point to ensure accuracy.

9. Step-by-Step Procedure

  1. Preparation: Gather all necessary tools and reference standards.
  2. Stabilization: Allow the Inline Vacuum Transmitter to stabilize before taking measurements.
  3. As Found checks and data capture: Record initial measurements and observations contemporaneously (ALCOA+).
  4. Adjustment rules: If OOT is detected, follow escalation procedures as adjustments may not be allowed.
  5. As Left verification: Confirm that the instrument is within specifications after any adjustments.
  6. Labeling + documentation completion: Ensure all calibration records are completed accurately.

10. Acceptance Criteria & Calculations

Acceptance criteria for the Inline Vacuum Transmitter is ±1%. The error calculation formula is:

Error (%) = (Observed Value – True Value) / True Value × 100

Pass/Fail logic: If the calculated error is within ±1%, the calibration passes; otherwise, it fails.

Rounding rule: [TBD]

11. Handling OOT / Failures

  • Immediate containment: Tag equipment and stop use.
  • Notification & deviation initiation: Inform relevant personnel of the OOT condition.
  • Impact assessment trigger: Define impact window and review of affected results/batches.
  • Recalibration after repair/adjustment: Schedule recalibration as necessary.
  • QA disposition placeholders: [TBD]

12. Records & Retention

  • Logbook entries: Document all calibration activities.
  • Calibration report/certificate filing: Maintain records in accordance with company policy.
  • e-Records if applicable: [TBD]
  • Label content requirements: Include calibrated date, due date, technician name, and status.

13. Annexures

Annexure A: Calibration Data Sheet

Test Points Reference Observed Error Acceptance Pass/Fail As Found As Left
[TBD] [TBD] [TBD] [TBD] [TBD] [TBD] [TBD] [TBD]

Annexure B: Quick Checklist

  • Environmental controls verified
  • Equipment warm-up completed
  • Cleanliness checked
  • Leveling confirmed
  • Power supply verified

Annexure C: Example Label Text

Calibration Date: [TBD] | Due Date: [TBD] | Calibrated By: [TBD] | Status: [TBD]

See also  Pycnometer (Density bottle) (CAL-0046) - Calibration Documents

⬆ Back to TOC


Calibration Protocol

Calibration Protocol

Protocol ID Asset ID Equipment Name Equipment Type Dept Area Process Use Parameter Units Range Acceptance Criteria Frequency Calibration Type Traceability
[TBD] CAL-0202 Inline Vacuum Transmitter Vacuum Transmitter API Dryer Vacuum control Pressure bar -1 to 0 ±1% 365 days External ISO17025

Environmental Conditions

Temperature: [TBD]

Relative Humidity: [TBD]

Vibration: [TBD]

Drafts: [TBD]

Test Point Selection

Test points for calibration will be selected based on the following values:

  • Min: -1 bar
  • 25%: -0.75 bar
  • 50%: -0.5 bar
  • 75%: -0.25 bar
  • Max: 0 bar

Results Table

Test Point Ref Value Observed (As Found) Error Limit Pass/Fail Observed (As Left) Error Pass/Fail
-1 bar [TBD] [TBD] [TBD] ±1% [TBD] [TBD] [TBD] [TBD]
-0.75 bar [TBD] [TBD] [TBD] ±1% [TBD] [TBD] [TBD] [TBD]
-0.5 bar [TBD] [TBD] [TBD] ±1% [TBD] [TBD] [TBD] [TBD]
-0.25 bar [TBD] [TBD] [TBD] ±1% [TBD] [TBD] [TBD] [TBD]
0 bar [TBD] [TBD] [TBD] ±1% [TBD] [TBD] [TBD] [TBD]

Stepwise Instructions

  1. Preparation: Ensure the Inline Vacuum Transmitter is clean and properly installed in the system.
  2. Stabilization: Allow the equipment to stabilize for [TBD] minutes before taking measurements.
  3. Performing Readings: Take three replicate readings at each test point to ensure accuracy.
  4. Calculating Error: Calculate the error for each observed value against the reference value.
  5. Evaluation: Compare the calculated error against the acceptance criteria of ±1%.
  6. Documentation: Record all observations, errors, and results in the results table.

Deviation Instruction for OOT

If any observed value is out of tolerance (OOT):

  • Stop use of the equipment immediately.
  • Notify the QA department.
  • Open a deviation report.
  • Quarantine impacted results window: [TBD]
  • Await QA decision on further actions.

Approvals

Prepared Reviewed Approved
[Signature/Date] [Signature/Date] [Signature/Date]

⬆ Back to TOC


Calibration Checklist

Asset ID Equipment Dept Area Parameter Range Acceptance Frequency Traceability
CAL-0202 Inline Vacuum Transmitter API Dryer Pressure -1 to 0 bar ±1% 365 days ISO17025

Pre-calibration

  • [ ] Verify environmental conditions are within specifications
  • [ ] Allow equipment to warm up as per SOP
  • [ ] Clean the equipment as per cleaning procedure
  • [ ] Ensure equipment is leveled
  • [ ] Check validity of reference standards
  • [ ] Ensure logbook is ready for entries

As Found Execution

  • [ ] Set equipment to specified setpoints
  • [ ] Perform multiple repeats for accuracy
  • [ ] Record all raw data accurately
  • [ ] Perform necessary calculations for results
See also  Dry Heat Sterilizer Temp Sensor (CAL-0153) - Calibration Documents

OOT Immediate Actions

  • [ ] Tag out the equipment
  • [ ] Stop use of the equipment immediately
  • [ ] Notify the responsible personnel
  • [ ] Document deviation in the logbook
  • [ ] Trigger impact review process

As Left Verification

  • [ ] Conduct post-adjustment checks
  • [ ] Confirm all results are within acceptance limits

Documentation & Closure

  • [ ] Update equipment label with new calibration date
  • [ ] Attach calibration certificate to documentation
  • [ ] Obtain reviewer sign-off
  • [ ] Update next due date in the logbook

⬆ Back to TOC


Deviation Impact Assessment

Event Summary
Asset: Inline Vacuum Transmitter
Date: [TBD]
Discovered By: [TBD]
Document References: [TBD]
As Found Results Summary: [TBD]

Immediate Containment Checklist

  • Stop use of the asset
  • Label/tag the asset as out of service
  • Segregate impacted work

Define Impact Window

Last known in-tolerance date: [TBD] to discovery date: [TBD]

Product/Batch/Test Impact Review

Batch/Study/Report ID Results Potentially Impacted Re-test/Re-weigh Decision
[TBD] [TBD] [TBD]

Data Integrity Review Prompts (ALCOA+)

  • Attributable: [TBD]
  • Legible: [TBD]
  • Contemporaneous: [TBD]
  • Original: [TBD]
  • Accurate: [TBD]
  • Complete: [TBD]
  • Consistent: [TBD]
  • Enduring: [TBD]
  • Available: [TBD]

Risk Assessment

Severity Probability Detectability Rationale Overall Risk
[TBD] [TBD] [TBD] [TBD] [TBD]

Investigation Summary

Root Cause: [TBD]

CAPA Plan

Action Owner Due Date Effectiveness Check
[TBD] [TBD] [TBD] [TBD]

QA Disposition

Release/Hold/Retest/Recall Triggers: [TBD]

⬆ Back to TOC


Calibration Certificate

Certificate No: [TBD]

Date: [TBD]

Next Due Date: [TBD]

Asset Details

Asset ID Equipment Name Equipment Type Dept Area Make/Model Serial No Location Parameter/Range/Units Acceptance Criteria
CAL-0202 Inline Vacuum Transmitter Vacuum Transmitter API Dryer [TBD] [TBD] [TBD] Pressure: -1 to 0 bar ±1%

Conditions

Environment: [TBD]

Reference Standards

Standard Name/ID Certificate No Due Date Traceability Statement
[TBD] [TBD] [TBD] ISO17025

Results

As Found Results

Test Point Measured Value (bar) Deviation (bar)
Point 1 [TBD] [TBD]
Point 2 [TBD] [TBD]
Point 3 [TBD] [TBD]
Point 4 [TBD] [TBD]
Point 5 [TBD] [TBD]

As Left Results

Test Point Measured Value (bar) Deviation (bar)
Point 1 [TBD] [TBD]
Point 2 [TBD] [TBD]
Point 3 [TBD] [TBD]
Point 4 [TBD] [TBD]
Point 5 [TBD] [TBD]

Measurement Uncertainty Statement: [TBD]

Statement of Conformity: [Pass/Fail overall]

Remarks: [TBD]

Signatures

Performed by: ______________________ Date: ___________

Checked by: ______________________ Date: ___________

Approved by: ______________________ Date: ___________

⬆ Back to TOC