Bioreactor DO Probe (CAL-0302) – Calibration Documents

Calibration Documents: Bioreactor DO Probe (CAL-0302)

Asset ID CAL-0302
Equipment Bioreactor DO Probe
Equipment Type DO Probe
Department Biologics
Area/Location Bioreactor
Process Use DO control
Parameter DO
Units %DO
Range Min 0
Range Max 200
Acceptance Criteria ±2%
Calibration Frequency (days) 180
Calibration Type External
Traceability ISO17025

Calibration SOP

1. Purpose

The calibration of the Bioreactor DO Probe is performed to ensure accurate dissolved oxygen (DO) measurements, which are critical for maintaining optimal conditions in biologics production. Accurate calibration mitigates risks related to patient safety, batch quality, and data integrity, thereby ensuring compliance with Good Manufacturing Practices (GMP).

2. Scope

  • Asset ID: CAL-0302
  • Equipment Name/Type: Bioreactor DO Probe
  • Department/Area: Biologics
  • Process Use: DO control
  • Parameter: DO
  • Range: 0% DO to 200% DO
  • Units: %DO
  • Calibration Frequency: 180 days

3. Definitions

  • Calibration: The process of configuring an instrument to provide a result for a sample within an acceptable range.
  • Verification: The process of confirming that an instrument is functioning correctly.
  • Adjustment: The act of altering the instrument to bring it into compliance with specified standards.
  • As Found: The condition of the instrument prior to any adjustments.
  • As Left: The condition of the instrument after adjustments have been made.
  • OOT (Out of Tolerance): A condition where the measurement exceeds the acceptance criteria.
  • Drift: The gradual change in an instrument’s output over time.
  • Traceability: The ability to verify the history, location, or application of an item by means of documented recorded identification.
  • Measurement Uncertainty: A parameter associated with the result of a measurement that characterizes the dispersion of the values that could reasonably be attributed to the measurement.
  • Intermediate Check: A periodic verification of an instrument’s performance between calibrations.

4. References

  • Company Metrology SOP [TBD]
  • ISO/IEC 17025 traceability via calibration certificates
  • ALCOA+ principles

5. Responsibilities

  • Technician/Metrology: Perform calibration and document results.
  • QC/QA Reviewer: Review calibration documentation for compliance.
  • Owner Department: Ensure equipment is available and properly maintained.
  • External Service Provider: Conduct external calibration as per the defined schedule.

6. Safety & Preconditions

  • Ensure environmental controls are in place (temperature, humidity).
  • Allow equipment warm-up/stabilization time as per manufacturer guidelines.
  • Maintain cleanliness of the equipment and surrounding area.
  • Check leveling of the equipment.
  • Verify power supply stability and functionality.
See also  Rotameter (Gas) (CAL-0084) - Calibration Documents

7. Standards & Tools

  • Reference Standards: [TBD]
  • Traceability Statement: ISO17025
  • Certificate Validity Check: Ensure all certificates are current and valid.

8. Calibration Strategy

The calibration will follow an As Found and As Left logic:

  • Minimum number of points: At least 5 points including endpoints and midpoints across the range of 0% DO to 200% DO.
  • Replicate readings: 3 repeats at each point as a best practice.

9. Step-by-Step Procedure

  1. Preparation: Gather all necessary tools and reference standards.
  2. Stabilization: Allow the Bioreactor DO Probe to stabilize before taking measurements.
  3. As Found checks and data capture: Record initial readings and document them contemporaneously (ALCOA+).
  4. Adjustment rules: If OOT is found, follow the adjustment protocol; otherwise, escalate the issue.
  5. As Left verification: Confirm that the instrument is within tolerance after adjustments.
  6. Labeling + documentation completion: Complete all required documentation and label the equipment with calibration details.

10. Acceptance Criteria & Calculations

Acceptance criteria for the Bioreactor DO Probe is ±2%. The error calculation formula is:

Error (%) = (Observed Value – Reference Value) / Reference Value * 100

Pass/Fail Logic: If the error is within ±2%, the calibration is considered a pass; otherwise, it is a fail.

Rounding rule: [TBD]

11. Handling OOT / Failures

  • Immediate containment: Tag the equipment and stop its use.
  • Notification & deviation initiation: Notify relevant personnel and initiate a deviation report.
  • Impact assessment trigger: Define the impact window and review affected results/batches.
  • Recalibration after repair/adjustment: Ensure recalibration is performed post-adjustment.
  • QA disposition: [TBD]

12. Records & Retention

  • Logbook entries: Document all calibration activities.
  • Calibration report/certificate filing: Maintain records in accordance with company policy.
  • e-Records if applicable: [TBD]
  • Label content requirements: Include calibrated date, due date, technician name, and status.

13. Annexures

Annexure A: Calibration Data Sheet

Test Points Reference Observed Error Acceptance Pass/Fail As Found As Left
[TBD] [TBD] [TBD] [TBD] [TBD] [TBD] [TBD] [TBD]

Annexure B: Quick Checklist

  • Environmental controls verified
  • Equipment warm-up completed
  • Cleanliness checked
  • Leveling confirmed
  • Power supply verified

Annexure C: Example Label Text

Calibration Date: [TBD]

Due Date: [TBD]

Calibrated By: [TBD]

See also  Micropipette 0.1-2 µL (CAL-0020) - Calibration Documents

Status: [TBD]

⬆ Back to TOC


Calibration Protocol

Calibration Protocol

Protocol ID Asset ID Equipment Name Equipment Type Dept Area Process Use Parameter Units Range Acceptance Criteria Frequency Calibration Type Traceability
[TBD] CAL-0302 Bioreactor DO Probe DO Probe Biologics Bioreactor DO control DO %DO 0 – 200 ±2% 180 days External ISO17025

Environmental Conditions

Temperature: [TBD]

Relative Humidity: [TBD]

Vibration: [TBD]

Drafts: [TBD]

Test Point Selection

  • Minimum: 0 %DO
  • 25%: 50 %DO
  • 50%: 100 %DO
  • 75%: 150 %DO
  • Maximum: 200 %DO

Results Table

Test Point Ref Value Observed (As Found) Error Limit Pass/Fail Observed (As Left) Error Pass/Fail
0 %DO [TBD] [TBD] [TBD] ±2% [TBD] [TBD] [TBD] [TBD]
50 %DO [TBD] [TBD] [TBD] ±2% [TBD] [TBD] [TBD] [TBD]
100 %DO [TBD] [TBD] [TBD] ±2% [TBD] [TBD] [TBD] [TBD]
150 %DO [TBD] [TBD] [TBD] ±2% [TBD] [TBD] [TBD] [TBD]
200 %DO [TBD] [TBD] [TBD] ±2% [TBD] [TBD] [TBD] [TBD]

Stepwise Instructions

  1. Preparation: Ensure the Bioreactor DO Probe is clean and free from contaminants.
  2. Stabilization: Allow the probe to stabilize in the calibration environment for at least 30 minutes.
  3. Performing Readings: Take three replicate readings at each test point and record the values.
  4. Calculating Error: Calculate the error for each test point using the formula: Error = Observed – Ref Value.
  5. Evaluation: Compare the calculated error against the acceptance criteria of ±2%.
  6. Documentation: Complete the results table and ensure all data is recorded accurately.

Deviation Instruction for OOT

In the event of an Out of Tolerance (OOT) condition:

  • Stop use of the Bioreactor DO Probe immediately.
  • Notify the QA department.
  • Open a deviation report.
  • Quarantine impacted results window: [TBD]
  • Await QA decision on further actions.

Approvals

Prepared Reviewed Approved
[Signature/Date] [Signature/Date] [Signature/Date]

⬆ Back to TOC


Calibration Checklist

Asset ID Equipment Dept Area Parameter Range Acceptance Frequency Traceability
CAL-0302 Bioreactor DO Probe Biologics Bioreactor DO 0 – 200 %DO ±2% 180 days ISO17025

Pre-calibration

  • [ ] Verify environmental conditions are within specifications
  • [ ] Allow equipment to warm up as per SOP
  • [ ] Clean the probe and surrounding area
  • [ ] Level the equipment if necessary
  • [ ] Check validity of reference standards
  • [ ] Ensure logbook is ready for entries

As Found Execution

  • [ ] Set the probe to specified setpoints
  • [ ] Perform repeat measurements
  • [ ] Record all raw data accurately
  • [ ] Perform necessary calculations
See also  Karl Fischer Titrator (Volumetric) (CAL-0140) - Calibration Documents

OOT Immediate Actions

  • [ ] Tag out the equipment
  • [ ] Stop use of the equipment
  • [ ] Notify relevant personnel
  • [ ] Document deviation
  • [ ] Trigger impact review

As Left Verification

  • [ ] Conduct post-adjustment checks
  • [ ] Confirm all measurements are within acceptance limits

Documentation & Closure

  • [ ] Update equipment label
  • [ ] Attach calibration certificate
  • [ ] Obtain reviewer sign-off
  • [ ] Update next due date in logbook

Data Integrity (ALCOA+): Ensure all data recorded is Attributable, Legible, Contemporaneous, Original, Accurate, and Complete.

⬆ Back to TOC


Deviation Impact Assessment

Event Summary
Asset: Bioreactor DO Probe
Date: [TBD]
Discovered By: [TBD]
Document References: [TBD]
As Found Results Summary: [TBD]

Immediate Containment Checklist

  • Stop use of the asset
  • Label/tag the asset as out of service
  • Segregate impacted work

Define Impact Window

Last known in-tolerance date: [TBD] to discovery date: [TBD]

Product/Batch/Test Impact Review

Batch/Study/Report ID Results Potentially Impacted Re-test/Re-weigh Decision
[TBD] [TBD] [TBD]

Data Integrity Review Prompts (ALCOA+)

  • Attributable: [TBD]
  • Legible: [TBD]
  • Contemporaneous: [TBD]
  • Original: [TBD]
  • Accurate: [TBD]
  • Complete: [TBD]
  • Consistent: [TBD]
  • Enduring: [TBD]
  • Available: [TBD]

Risk Assessment

Severity Probability Detectability Rationale Overall Risk
[TBD] [TBD] [TBD] [TBD] [TBD]

Investigation Summary

Root Cause: [TBD]

CAPA Plan

Action Owner Due Date Effectiveness Check
[TBD] [TBD] [TBD] [TBD]

QA Disposition

Release/Hold/Retest/Recall Triggers: [TBD]

⬆ Back to TOC


Calibration Certificate

Certificate No: [TBD]

Date: [TBD]

Next Due Date: [TBD]

Asset Details

Asset ID Equipment Name Equipment Type Dept Area Make/Model Serial No Location Parameter/Range/Units Acceptance Criteria
CAL-0302 Bioreactor DO Probe DO Probe Biologics Bioreactor [TBD] [TBD] [TBD] DO: 0 – 200 %DO ±2%

Conditions

Environment: [TBD]

Reference Standards

Standard Name/ID Certificate No Due Date Traceability Statement
[TBD] [TBD] [TBD] Traceability to ISO17025

Results

As Found Results

Test Point Measured Value (%DO) Acceptance Criteria Pass/Fail
1 [TBD] ±2% [TBD]
2 [TBD] ±2% [TBD]
3 [TBD] ±2% [TBD]
4 [TBD] ±2% [TBD]
5 [TBD] ±2% [TBD]

As Left Results

Test Point Measured Value (%DO) Acceptance Criteria Pass/Fail
1 [TBD] ±2% [TBD]
2 [TBD] ±2% [TBD]
3 [TBD] ±2% [TBD]
4 [TBD] ±2% [TBD]
5 [TBD] ±2% [TBD]

Measurement Uncertainty Statement: [TBD]

Statement of Conformity: [TBD]

Remarks: [TBD]

Signatures

Performed by: ____________________ Date: __________

Checked by: ____________________ Date: __________

Approved by: ____________________ Date: __________

⬆ Back to TOC