Bioreactor Pressure Sensor (CAL-0303) – Calibration Documents

Calibration Documents: Bioreactor Pressure Sensor (CAL-0303)

Asset ID CAL-0303
Equipment Bioreactor Pressure Sensor
Equipment Type Pressure Sensor
Department Biologics
Area/Location Bioreactor
Process Use Pressure
Parameter Pressure
Units bar
Range Min 0
Range Max 3
Acceptance Criteria ±1%
Calibration Frequency (days) 365
Calibration Type External
Traceability ISO17025

Calibration SOP

1. Purpose

The calibration of the Bioreactor Pressure Sensor is performed to ensure accurate pressure measurements within the bioreactor process. This calibration is critical for maintaining patient safety, ensuring batch quality, and preserving data integrity. Accurate pressure readings are essential to prevent deviations that could compromise product quality and safety.

2. Scope

  • Asset ID: CAL-0303
  • Equipment Name/Type: Bioreactor Pressure Sensor / Pressure Sensor
  • Department/Area: Biologics / Bioreactor
  • Process Use: Pressure
  • Parameter: Pressure
  • Range: 0 to 3 bar
  • Units: bar
  • Calibration Frequency: 365 days

3. Definitions

  • Calibration: The process of comparing a measurement instrument against a known standard to determine its accuracy.
  • Verification: The process of confirming that a measurement instrument meets specified requirements.
  • Adjustment: The process of modifying a measurement instrument to bring it into compliance with specified standards.
  • As Found: The condition of the instrument prior to any adjustments made during calibration.
  • As Left: The condition of the instrument after adjustments have been made and verified.
  • OOT (Out of Tolerance): A condition where the measurement results fall outside the specified acceptance criteria.
  • Drift: The gradual change in a measurement instrument’s output over time.
  • Traceability: The ability to verify the history, location, or application of an item by means of documented recorded identification.
  • Measurement Uncertainty: A parameter that characterizes the range of values within which the true value of a measurement lies.
  • Intermediate Check: A verification process performed between calibrations to ensure ongoing accuracy.

4. References

  • Company Metrology SOP [TBD]
  • ISO/IEC 17025 traceability via calibration certificates
  • ALCOA+ principles

5. Responsibilities

  • Technician/Metrology: Perform calibration and document results.
  • QC/QA Reviewer: Review calibration documentation and results for compliance.
  • Owner Department: Ensure equipment is available and in good condition for calibration.
  • External Service Provider: Conduct calibration as per external calibration type requirements.

6. Safety & Preconditions

  • Ensure environmental controls are in place (temperature, humidity).
  • Allow for warm-up/stabilization of the equipment as per manufacturer specifications.
  • Maintain cleanliness of the equipment and surrounding area.
  • Check leveling of the equipment to ensure accurate readings.
  • Verify power supply stability before calibration.
See also  Volumetric Flask 250 mL (CAL-0037) - Calibration Documents

7. Standards & Tools

  • Reference Standards Required: [TBD]
  • Traceability Statement: ISO17025
  • Certificate Validity Check: Ensure all reference standards are within their validity period.
  • Standard ID Placeholders: [TBD]

8. Calibration Strategy

Calibration will follow the As Found then As Left logic. The minimum number of test points will be five, including endpoints (0 bar and 3 bar) and midpoints (1.5 bar). Each point will have three replicate readings taken as a best practice.

9. Step-by-Step Procedure

  1. Preparation: Gather all necessary tools and reference standards.
  2. Stabilization: Allow the Bioreactor Pressure Sensor to stabilize as per manufacturer guidelines.
  3. As Found checks and data capture: Measure pressure at defined points and record raw observations contemporaneously (ALCOA+).
  4. Adjustment rules: If OOT is detected, follow adjustment protocols as allowed; otherwise, escalate the issue.
  5. As Left verification: Re-measure to confirm that adjustments have brought the sensor within tolerance.
  6. Labeling + documentation completion: Complete all required documentation and label the equipment with calibration details.

10. Acceptance Criteria & Calculations

Acceptance criteria for the Bioreactor Pressure Sensor is ±1%. The error calculation formula is:

Error (%) = (Measured Value – True Value) / True Value × 100

Pass/Fail logic: If the error is within ±1%, the calibration is considered a pass; otherwise, it is a fail.

Rounding rule placeholder: [TBD]

11. Handling OOT / Failures

  • Immediate containment: Tag equipment and stop use until resolved.
  • Notification & deviation initiation: Notify relevant personnel and initiate a deviation report.
  • Impact assessment trigger: Define impact window and review of affected results/batches.
  • Recalibration after repair/adjustment: Ensure recalibration is performed post-adjustment.
  • QA disposition placeholders: [TBD]

12. Records & Retention

  • Logbook entries: Document all calibration activities.
  • Calibration report/certificate filing: Maintain records in accordance with company policy.
  • e-Records if applicable: [TBD]
  • Label content requirements: Include calibrated date, due date, technician name, and status.

13. Annexures

Annexure A: Calibration Data Sheet

Test Points Reference Observed Error Acceptance Pass/Fail As Found As Left
[TBD] [TBD] [TBD] [TBD] [TBD] [TBD] [TBD] [TBD]

Annexure B: Quick Checklist

  • Environmental controls verified
  • Equipment stabilized
  • Calibration tools prepared
  • Documentation ready

Annexure C: Example Label Text

Calibration Date: [TBD] | Due Date: [TBD] | Calibrated By: [TBD] | Status: [TBD]

See also  Heating Mantle Controller Temp (CAL-0159) - Calibration Documents

⬆ Back to TOC


Calibration Protocol

Calibration Protocol

Protocol ID Asset ID Equipment Name Equipment Type Dept Area Process Use Parameter Units Range Acceptance Criteria Frequency Calibration Type Traceability
[TBD] CAL-0303 Bioreactor Pressure Sensor Pressure Sensor Biologics Bioreactor Pressure Pressure bar 0 – 3 ±1% 365 days External ISO17025

Environmental Conditions

Temperature: [TBD]

Relative Humidity: [TBD]

Vibration: [TBD]

Drafts: [TBD]

Test Point Selection

  • Minimum: 0 bar
  • 25% of Range: 0.75 bar
  • 50% of Range: 1.5 bar
  • 75% of Range: 2.25 bar
  • Maximum: 3 bar

Results Table

Test Point Ref Value Observed (As Found) Error Limit Pass/Fail Observed (As Left) Error Pass/Fail
0 bar [TBD] [TBD] [TBD] ±0.03 bar [TBD] [TBD] [TBD] [TBD]
0.75 bar [TBD] [TBD] [TBD] ±0.03 bar [TBD] [TBD] [TBD] [TBD]
1.5 bar [TBD] [TBD] [TBD] ±0.03 bar [TBD] [TBD] [TBD] [TBD]
2.25 bar [TBD] [TBD] [TBD] ±0.03 bar [TBD] [TBD] [TBD] [TBD]
3 bar [TBD] [TBD] [TBD] ±0.03 bar [TBD] [TBD] [TBD] [TBD]

Stepwise Instructions

  1. Preparation: Ensure the bioreactor pressure sensor is clean and free from any contaminants.
  2. Stabilization: Allow the sensor to stabilize for [TBD] minutes before taking measurements.
  3. Performing Readings: Take three replicate readings at each test point.
  4. Calculating Error: Calculate the error for each observed value against the reference value.
  5. Evaluation: Compare the calculated errors against the acceptance criteria of ±1%.
  6. Documentation: Record all findings in the results table, ensuring ALCOA+ principles are followed.

Deviation Instruction for OOT

If any observed value falls outside the acceptance criteria:

  • Stop use of the equipment.
  • Notify the QA department immediately.
  • Open a deviation report.
  • Quarantine impacted results window: [TBD]
  • Await QA decision on further actions.

Approvals

Prepared By Reviewed By Approved By
[Signature/Date] [Signature/Date] [Signature/Date]

⬆ Back to TOC


Calibration Checklist

Asset ID Equipment Dept Area Parameter Range Acceptance Frequency Traceability
CAL-0303 Bioreactor Pressure Sensor Biologics Bioreactor Pressure 0 – 3 bar ±1% 365 days ISO17025

Pre-calibration

  • [ ] Verify environmental conditions are within specified limits
  • [ ] Ensure equipment has warmed up adequately
  • [ ] Clean the equipment as per SOP
  • [ ] Level the equipment if necessary
  • [ ] Check validity of reference standards
  • [ ] Ensure calibration logbook is ready for entries

As Found Execution

  • [ ] Record initial setpoints
  • [ ] Perform repeat measurements
  • [ ] Record raw data accurately
  • [ ] Perform necessary calculations
See also  Micropipette 1-5 mL (CAL-0024) - Calibration Documents

OOT Immediate Actions

  • [ ] Tag out the equipment
  • [ ] Stop use of the equipment
  • [ ] Notify relevant personnel
  • [ ] Document deviation
  • [ ] Trigger impact review

As Left Verification

  • [ ] Conduct post-adjustment checks
  • [ ] Confirm all measurements are within acceptance limits

Documentation & Closure

  • [ ] Update equipment label
  • [ ] Attach calibration certificate
  • [ ] Obtain reviewer sign-off
  • [ ] Update next due date in logbook

Data Integrity (ALCOA+): Ensure all data recorded is Attributable, Legible, Contemporaneous, Original, Accurate, and Complete.

⬆ Back to TOC


Deviation Impact Assessment

Event Summary
Asset: Bioreactor Pressure Sensor
Date: [TBD]
Discovered By: [TBD]
Document References: [TBD]
As Found Results Summary: [TBD]

Immediate Containment Checklist

  • Stop use of the asset
  • Label/Tag the asset as out of tolerance
  • Segregate impacted work

Define Impact Window

Last Known In-Tolerance Date Discovery Date
[TBD] [TBD]

Product/Batch/Test Impact Review

Batch/Study/Report ID Results Potentially Impacted Re-test/Re-weigh Decision
[TBD] [TBD] [TBD]

Data Integrity Review Prompts (ALCOA+)

  • Attributable: [TBD]
  • Legible: [TBD]
  • Contemporaneous: [TBD]
  • Original: [TBD]
  • Accurate: [TBD]
  • Complete: [TBD]
  • Consistent: [TBD]
  • Enduring: [TBD]
  • Available: [TBD]

Risk Assessment

Severity Probability Detectability Rationale Overall Risk
[TBD] [TBD] [TBD] [TBD] [TBD]

Investigation Summary

Investigation Summary Placeholders: [TBD]

Root Cause: [TBD]

CAPA Plan

Action Owner Due Date Effectiveness Check
[TBD] [TBD] [TBD] [TBD]

QA Disposition

Release/Hold/Retest/Recall Triggers: [TBD]

⬆ Back to TOC


Calibration Certificate

Certificate No: [TBD]

Date: [TBD]

Next Due Date: [TBD]

Asset Details

Asset ID Equipment Name Equipment Type Dept Area Make/Model Serial No Location Parameter/Range/Units Acceptance Criteria
CAL-0303 Bioreactor Pressure Sensor Pressure Sensor Biologics Bioreactor [TBD] [TBD] [TBD] Pressure: 0 – 3 bar ±1%

Conditions

Environment: [TBD]

Reference Standards

Standard Name/ID Certificate No Due Date Traceability Statement
[TBD] [TBD] [TBD] Traceability to ISO17025

Results

As Found Results

Test Point Measured Value (bar) Acceptance Criteria Pass/Fail
0.0 [TBD] ±1% [TBD]
1.0 [TBD] ±1% [TBD]
2.0 [TBD] ±1% [TBD]
3.0 [TBD] ±1% [TBD]
2.5 [TBD] ±1% [TBD]

As Left Results

Test Point Measured Value (bar) Acceptance Criteria Pass/Fail
0.0 [TBD] ±1% [TBD]
1.0 [TBD] ±1% [TBD]
2.0 [TBD] ±1% [TBD]
3.0 [TBD] ±1% [TBD]
2.5 [TBD] ±1% [TBD]

Measurement Uncertainty Statement: [TBD]

Statement of Conformity: [TBD]

Remarks: [TBD]

Signatures

Performed by: ____________________ Date: ___________

Checked by: ______________________ Date: ___________

Approved by: _____________________ Date: ___________

⬆ Back to TOC