Balance Calibration Weight 50 g (CAL-0335) – Calibration Documents

Calibration Documents: Balance Calibration Weight 50 g (CAL-0335)

Asset ID CAL-0335
Equipment Balance Calibration Weight 50 g
Equipment Type Standard Weight
Department Calibration Lab
Area/Location Metrology
Process Use Mass standard
Parameter Weight
Units g
Range Min 50
Range Max 50
Acceptance Criteria Class F1/E2
Calibration Frequency (days) 365
Calibration Type External
Traceability ISO17025

Calibration SOP

1. Purpose

The calibration of the Balance Calibration Weight 50 g is performed to ensure the accuracy of mass measurements, which is critical for maintaining patient safety, ensuring batch quality, and preserving data integrity. Accurate weight measurements directly impact the formulation and dosing of pharmaceutical products, thereby controlling risks associated with incorrect dosages.

2. Scope

  • Asset ID: CAL-0335
  • Equipment Name/Type: Balance Calibration Weight 50 g / Standard Weight
  • Department/Area: Calibration Lab / Metrology
  • Process Use: Mass standard
  • Parameter: Weight
  • Range: 50 g
  • Units: g
  • Calibration Frequency: 365 days

3. Definitions

  • Calibration: The process of verifying and adjusting the accuracy of a measurement instrument against a known standard.
  • Verification: The process of confirming that an instrument meets specified requirements.
  • Adjustment: The act of modifying a measurement instrument to bring it into compliance with specified standards.
  • As Found: The condition of the instrument prior to any adjustments being made.
  • As Left: The condition of the instrument after adjustments have been made.
  • OOT (Out of Tolerance): A condition where the measurement falls outside the specified acceptance criteria.
  • Drift: The gradual change in the measurement output of an instrument over time.
  • Traceability: The ability to verify the history, location, or application of an item by means of documented recorded identification.
  • Measurement Uncertainty: A parameter that characterizes the range of values within which the true value is expected to lie.
  • Intermediate Check: A periodic assessment of an instrument’s performance between calibrations.

4. References

  • Company Metrology SOP [TBD]
  • ISO/IEC 17025 traceability via calibration certificates
  • ALCOA+ principles

5. Responsibilities

  • Technician/Metrology: Perform calibration and maintain records.
  • QC/QA Reviewer: Review calibration results and documentation.
  • Owner Department: Ensure compliance with calibration schedules.
  • External Service Provider: Conduct external calibrations as required.

6. Safety & Preconditions

  • Ensure environmental controls are in place (temperature, humidity).
  • Allow for warm-up/stabilization of the equipment as per manufacturer guidelines.
  • Maintain cleanliness of the calibration area and equipment.
  • Check leveling of the balance before use.
  • Verify power supply stability and functionality.
See also  Disintegration Tester (CAL-0131) - Calibration Documents

7. Standards & Tools

  • Reference Standards Required: ISO17025 certified calibration weights.
  • Traceability Statement: Calibration is traceable to ISO17025.
  • Certificate Validity Check: Ensure calibration certificates are current and valid.
  • Standard ID: [TBD]

8. Calibration Strategy

Calibration will follow the As Found and As Left logic:

  • Minimum number of points: 5 points including endpoints and midpoints.
  • Replicate readings at each point: 3 repeats as a best practice.

9. Step-by-Step Procedure

  1. Preparation: Gather all necessary tools and reference standards.
  2. Stabilization: Allow the balance to stabilize before calibration.
  3. As Found checks and data capture: Record initial measurements before any adjustments.
  4. Adjustment rules: If OOT is found, escalate for further investigation; adjustments should only be made if allowed.
  5. As Left verification: Confirm that the instrument is within tolerance after adjustments.
  6. Labeling + documentation completion: Complete all required documentation and label the equipment accordingly.

Record raw observations contemporaneously to ensure data integrity (ALCOA+).

10. Acceptance Criteria & Calculations

Acceptance criteria: Class F1/E2.

Error calculation formula: Error (%) = (Observed Value – True Value) / True Value × 100

Pass/Fail logic: If the error falls within the acceptance criteria, the calibration passes; otherwise, it fails.

Rounding rule: [TBD]

11. Handling OOT / Failures

  • Immediate containment: Tag equipment and stop use.
  • Notification & deviation initiation: Notify relevant personnel of the OOT condition.
  • Impact assessment trigger: Define impact window and review of affected results/batches.
  • Recalibration after repair/adjustment: Ensure recalibration is performed post-adjustment.
  • QA disposition placeholders: [TBD]

12. Records & Retention

  • Logbook entries: Document all calibration activities.
  • Calibration report/certificate filing: Maintain records of calibration certificates.
  • e-Records if applicable: [TBD]
  • Label content requirements: Calibrated date, due date, by whom, status.

13. Annexures

Annexure A: Calibration Data Sheet

Test Points Reference Observed Error Acceptance Pass/Fail As Found As Left
[TBD] [TBD] [TBD] [TBD] [TBD] [TBD] [TBD] [TBD]

Annexure B: Quick Checklist

  • Environmental controls checked
  • Equipment warm-up completed
  • Cleanliness verified
  • Leveling confirmed
  • Power supply checked

Annexure C: Example Label Text

Calibration Date: [TBD]

Due Date: [TBD]

Calibrated By: [TBD]

See also  Tachometer (CAL-0263) - Calibration Documents

Status: [TBD]

⬆ Back to TOC


Calibration Protocol

Calibration Protocol

Protocol ID Asset ID Equipment Name Equipment Type Dept Area Process Use Parameter Units Range Acceptance Criteria Frequency Calibration Type Traceability
[TBD] CAL-0335 Balance Calibration Weight 50 g Standard Weight Calibration Lab Metrology Mass standard Weight g 50 – 50 Class F1/E2 365 days External ISO17025

Environmental Conditions

Temperature: [TBD]

Relative Humidity: [TBD]

Vibration: [TBD]

Drafts: [TBD]

Test Point Selection

  • Minimum: 50 g
  • 25%: 50 g
  • 50%: 50 g
  • 75%: 50 g
  • Maximum: 50 g

Results Table

Test Point Ref Value Observed (As Found) Error Limit Pass/Fail Observed (As Left) Error Pass/Fail
50 g 50 g [TBD] [TBD] [TBD] [TBD] [TBD] [TBD] [TBD]

Stepwise Instructions

  1. Preparation: Ensure the calibration weight is clean and free from contaminants.
  2. Stabilization: Allow the calibration weight to acclimate to the environmental conditions for at least [TBD] minutes.
  3. Performing Readings: Take three replicate readings at each test point to ensure accuracy.
  4. Calculating Error: Calculate the error by subtracting the reference value from the observed value.
  5. Evaluation: Compare the calculated error against the acceptance criteria.
  6. Documentation: Record all observations, calculations, and results in the calibration log.

Deviation Instruction for OOT

  • Stop use of the asset immediately.
  • Notify the QA department.
  • Open a deviation report.
  • Quarantine impacted results window: [TBD]
  • Await QA decision on further actions.

Approvals

Prepared By Reviewed By Approved By
[Signature/Date] [Signature/Date] [Signature/Date]

⬆ Back to TOC


Calibration Checklist

Asset ID Equipment Dept Area Parameter Range Acceptance Frequency Traceability
CAL-0335 Balance Calibration Weight 50 g Calibration Lab Metrology Weight 50 g Class F1/E2 365 days ISO17025

Pre-calibration

  • [ ] Verify environmental conditions are within specified limits
  • [ ] Allow equipment to warm up as per SOP
  • [ ] Clean the calibration weight and surrounding area
  • [ ] Level the balance as per manufacturer instructions
  • [ ] Confirm validity of reference standards
  • [ ] Ensure calibration logbook is ready for entries

As Found Execution

  • [ ] Set the calibration weight to specified setpoints
  • [ ] Perform repeat measurements as required
  • [ ] Record all raw data accurately
  • [ ] Perform necessary calculations for results

OOT Immediate Actions

  • [ ] Tag out the equipment if out of tolerance (OOT)
  • [ ] Stop use of the equipment immediately
  • [ ] Notify the appropriate personnel
  • [ ] Document deviation in the logbook
  • [ ] Trigger impact review process
See also  Photo-stability Chamber Lux (CAL-0184) - Calibration Documents

As Left Verification

  • [ ] Conduct post-adjustment checks
  • [ ] Confirm that results are within specified limits

Documentation & Closure

  • [ ] Update equipment label with calibration details
  • [ ] Attach calibration certificate to logbook
  • [ ] Obtain reviewer sign-off
  • [ ] Update next due date in the logbook

Acceptance Criteria: Class F1/E2

Data Integrity (ALCOA+): Ensure all data is attributable, legible, contemporaneous, original, and accurate.

Traceability: ISO17025

As Found/As Left: Document conditions before and after calibration.

⬆ Back to TOC


Deviation Impact Assessment

Event Summary
Asset: Balance Calibration Weight 50 g
Date: [TBD]
Discovered By: [TBD]
Document References: [TBD]
As Found Results Summary: [TBD]

Immediate Containment Checklist

  • Stop use of the asset
  • Label/tag the asset as out of tolerance
  • Segregate impacted work

Define Impact Window

Last known in-tolerance date: [TBD] to discovery date: [TBD]

Product/Batch/Test Impact Review

Batch/Study/Report ID Results Potentially Impacted Re-test/Re-weigh Decision
[TBD] [TBD] [TBD]

Data Integrity Review Prompts (ALCOA+)

  • Attributable: [TBD]
  • Legible: [TBD]
  • Contemporaneous: [TBD]
  • Original: [TBD]
  • Accurate: [TBD]
  • Complete: [TBD]
  • Consistent: [TBD]
  • Enduring: [TBD]
  • Available: [TBD]

Risk Assessment

Severity Probability Detectability Rationale Overall Risk
[TBD] [TBD] [TBD] [TBD] [TBD]

Investigation Summary

Root Cause: [TBD]

CAPA Plan

Action Owner Due Date Effectiveness Check
[TBD] [TBD] [TBD] [TBD]

QA Disposition

Release/Hold/Retest/Recall Triggers: [TBD]

⬆ Back to TOC


Calibration Certificate

Certificate No: [TBD]

Date: [TBD]

Next Due Date: [TBD]

Asset Details

Asset ID Equipment Name Equipment Type Dept Area Make/Model Serial No Location Parameter/Range/Units Acceptance Criteria
CAL-0335 Balance Calibration Weight 50 g Standard Weight Calibration Lab Metrology [TBD] [TBD] [TBD] Weight / 50 g / g Class F1/E2

Conditions

Environment: [TBD]

Reference Standards

Standard Name/ID Certificate No Due Date Traceability Statement
[TBD] [TBD] [TBD] ISO17025

Results

As Found Results

Test Point Measured Value (g) Acceptance Criteria Pass/Fail
1 [TBD] Class F1/E2 [TBD]
2 [TBD] Class F1/E2 [TBD]
3 [TBD] Class F1/E2 [TBD]
4 [TBD] Class F1/E2 [TBD]
5 [TBD] Class F1/E2 [TBD]

As Left Results

Test Point Measured Value (g) Acceptance Criteria Pass/Fail
1 [TBD] Class F1/E2 [TBD]
2 [TBD] Class F1/E2 [TBD]
3 [TBD] Class F1/E2 [TBD]
4 [TBD] Class F1/E2 [TBD]
5 [TBD] Class F1/E2 [TBD]

Measurement Uncertainty Statement: [TBD]

Statement of Conformity: [TBD]

Remarks: [TBD]

Signatures

Performed by: ______________________ Date: ____________

Checked by: ______________________ Date: ____________

Approved by: ______________________ Date: ____________

⬆ Back to TOC