Barcode Scanner (GMP) – PQ Protocol

Performance Qualification Protocol for Barcode Scanner in Solid Dosage Form Equipment Validation

Document Control

Document Number: PQ-OSD-001

Version: 1.0

Effective Date: [Insert Date]

Review Date: [Insert Date]

Objective

The objective of this Performance Qualification (PQ) protocol is to validate the functionality of the Barcode Scanner used for identity verification and traceability in the production and dispensing of solid dosage forms, ensuring compliance with regulatory requirements.

Scope

This protocol applies to the Barcode Scanner utilized in the Production/Dispensing area for solid dosage forms. It covers the validation of the scanner’s performance with respect to scan accuracy data capture and audit trail integrity.

Responsibilities

The following personnel are responsible for executing this protocol:

  • Validation Team: Conducts the PQ tests and compiles results.
  • Quality Assurance: Reviews and approves the PQ protocol and results.
  • IT Support: Ensures the integrity of the data capture system.

Prerequisites

Prior to executing this protocol, the following must be completed:

  • Installation Qualification (IQ) of the Barcode Scanner.
  • Operational Qualification (OQ) of the Barcode Scanner.
  • Training of personnel on the use of the Barcode Scanner.

Equipment Description

The Barcode Scanner is a GMP-compliant instrument used for identity verification and traceability in the production and dispensing of solid dosage forms. It captures barcode data and maintains an audit trail for compliance with regulatory requirements.

Test Plan

Test ID Procedure Acceptance Criteria Evidence
TP-001 Verify scan accuracy by scanning known barcodes. Scan accuracy ≥ 98%. Scan accuracy report.
TP-002 Check audit trail functionality by scanning and logging data. Complete audit trail without data loss. Audit trail log.
See also  Label Printer (GMP) – Validation Summary Report (VSR) Template

Detailed Test Cases

Test Case 1: Scan Accuracy

Objective: To verify the accuracy of the barcode scanning.

Procedure: Scan a set of known barcodes and record the results.

Acceptance Criteria: At least 98% of barcodes must be accurately scanned.

Expected Results: Documented results showing accuracy percentage.

Test Case 2: Audit Trail Verification

Objective: To ensure the audit trail captures all scanning events.

Procedure: Perform a series of scans and review the audit trail log.

Acceptance Criteria: No missing entries in the audit trail.

Expected Results: Complete and accurate audit trail documentation.

Deviations

Any deviations from the acceptance criteria must be documented and assessed for impact on the validation. A root cause analysis should be conducted for significant deviations.

Approvals

Prepared by: [Name, Title, Date]

Reviewed by: [Name, Title, Date]

Approved by: [Name, Title, Date]

Data Integrity Checks

The following data integrity checks will be implemented:

  • Regular audits of the audit trail logs to ensure completeness.
  • Validation of data entry points to prevent manual errors.
  • Secure storage of scanned data to prevent unauthorized access.