Tablet Checkweigher (Inline) – PQ Protocol

Document Control

Document ID: PQ-OSD-TCW-001

Version: 1.0

Effective Date: [Insert Date]

Reviewed By: [Insert Name]

Approval Date: [Insert Date]

Performance Qualification Protocol for Inline Tablet Checkweigher

Meta Description: This document outlines the Performance Qualification protocol for the Inline Tablet Checkweigher used in Solid Dosage Form production, focusing on weight verification and critical parameters.

Tags: Equipment Validation, Performance Qualification, Tablet Checkweigher, Solid Dosage Form

Objective

The objective of this protocol is to verify the performance of the Inline Tablet Checkweigher to ensure it meets the specified acceptance criteria for weight verification in the Solid Dosage Form production area.

Scope

This protocol applies to the Inline Tablet Checkweigher used in the Production/Compression area for weight verification of solid dosage forms. It encompasses all critical parameters as outlined in the User Requirement Specification (URS) Annex 11.

Responsibilities

  • Validation Team: Responsible for executing the PQ protocol and documenting results.
  • Quality Assurance: Responsible for reviewing and approving the protocol and results.
  • Production Personnel: Responsible for ensuring the equipment is maintained in a qualified state during the PQ process.

Prerequisites

  • Completion of Installation Qualification (IQ) and Operational Qualification (OQ).
  • Training of personnel on the operation of the Inline Tablet Checkweigher.
  • Availability of necessary calibration and verification tools.

Equipment Description

The Inline Tablet Checkweigher is designed for real-time weight verification of tablets during the compression process. It features an accuracy reject mechanism and an audit trail for compliance tracking.

See also  Split Butterfly Valve (SBV) Transfer System – Equipment Validation SOP

Test Plan

Test ID Procedure Acceptance Evidence
PQ-TCW-001 Verify accuracy of weight measurement using calibrated weights. Weight measurement must be within ±5% of target weight. Calibration records, test results.
PQ-TCW-002 Audit reject mechanism to ensure proper operation. Reject mechanism must activate for out-of-spec weights. Audit trail logs, observation records.

Detailed Test Cases

Test Case 1: Weight Measurement Accuracy

Objective: To verify the accuracy of the Inline Tablet Checkweigher.

Procedure: Use calibrated weights to test the checkweigher at various weight points (e.g., 100mg, 200mg, 300mg).

Acceptance Criteria: All measurements must fall within ±5% of the target weight.

Data Integrity Checks: Ensure that all weight measurements are logged in a secure database and that the data is retrievable for audit purposes.

Test Case 2: Reject Mechanism Audit

Objective: To ensure the reject mechanism functions correctly.

Procedure: Introduce weights outside of the acceptable range and verify that the reject mechanism activates.

Acceptance Criteria: The checkweigher must reject all out-of-spec weights and log the event in the audit trail.

Data Integrity Checks: Review the audit trail to confirm that all reject events are documented correctly.

Deviations

Any deviations from the acceptance criteria must be documented and investigated. A root cause analysis will be performed to determine the impact on product quality.