Calibration Documents: Autoclave Temp Sensor (CAL-0150)
| Asset ID | CAL-0150 |
|---|---|
| Equipment | Autoclave Temp Sensor |
| Equipment Type | Temperature Sensor |
| Department | Microbiology |
| Area/Location | Sterilization |
| Process Use | Sterilization |
| Parameter | Temperature |
| Units | °C |
| Range Min | 100 |
| Range Max | 140 |
| Acceptance Criteria | ±0.5°C |
| Calibration Frequency (days) | 180 |
| Calibration Type | External |
| Traceability | ISO17025 |
Table of Contents
Calibration SOP
1. Purpose
The calibration of the Autoclave Temp Sensor is performed to ensure accurate temperature readings during the sterilization process. This is critical for maintaining patient safety, ensuring batch quality, and preserving data integrity. Accurate temperature measurement is essential to prevent under- or over-sterilization, which could compromise product safety and efficacy.
2. Scope
- Asset ID: CAL-0150
- Equipment Name/Type: Autoclave Temp Sensor / Temperature Sensor
- Department/Area: Microbiology / Sterilization
- Process Use: Sterilization
- Parameter: Temperature
- Range: 100°C to 140°C
- Units: °C
- Calibration Frequency: 180 days
3. Definitions
- Calibration: The process of comparing a measurement instrument to a known standard to ensure accuracy.
- Verification: The process of confirming that a measurement instrument is functioning correctly.
- Adjustment: The process of modifying a measurement instrument to bring it into compliance with standards.
- As Found: The condition of the measurement instrument prior to any adjustments.
- As Left: The condition of the measurement instrument after adjustments have been made.
- OOT (Out of Tolerance): A condition where the measurement falls outside the established acceptance criteria.
- Drift: The gradual change in a measurement instrument’s output over time.
- Traceability: The ability to verify the history, location, or application of an item by means of documented recorded identification.
- Measurement Uncertainty: A parameter that characterizes the range of values within which the true value is asserted to lie.
- Intermediate Check: A periodic check of a measurement instrument to ensure it remains within specified limits.
4. References
- Company Metrology SOP [TBD]
- ISO/IEC 17025 traceability via calibration certificates
- ALCOA+ principles
5. Responsibilities
- Technician/Metrology: Perform calibration and document results.
- QC/QA Reviewer: Review calibration results and ensure compliance.
- Owner Department: Ensure equipment is available and maintained.
- External Service Provider: Conduct external calibration as required.
6. Safety & Preconditions
- Ensure environmental controls are in place (temperature, humidity).
- Allow equipment to warm up and stabilize as per manufacturer’s guidelines.
- Maintain cleanliness of the calibration area.
- Check that the equipment is level and stable.
- Verify power supply is stable and within specifications.
7. Standards & Tools
- Reference standards required: [TBD]
- Traceability statement: ISO17025
- Certificate validity check: [TBD]
8. Calibration Strategy
Calibration will follow the As Found and As Left logic. The minimum number of calibration points will be five, including endpoints and midpoints of the specified range (100°C to 140°C). It is recommended to take three replicate readings at each point to ensure accuracy.
9. Step-by-Step Procedure
- Preparation: Gather all necessary tools and documentation.
- Stabilization: Allow the sensor to reach operating conditions.
- As Found checks and data capture: Record initial readings at specified points.
- Adjustment rules: If OOT is found, escalate to QA for further action.
- As Left verification: Confirm that adjustments bring the sensor within acceptance criteria.
- Labeling + documentation completion: Ensure all records are filled out contemporaneously (ALCOA+).
10. Acceptance Criteria & Calculations
Acceptance criteria: ±0.5°C. The error calculation formula is:
Error = Observed Value – True Value
Pass/Fail Logic: If the absolute value of the error is less than or equal to 0.5°C, the calibration passes; otherwise, it fails.
Rounding rule: [TBD]
11. Handling OOT / Failures
- Immediate containment: Tag equipment and stop use.
- Notification & deviation initiation: Notify QA and document the deviation.
- Impact assessment trigger: Review affected results/batches within the impact window.
- Recalibration after repair/adjustment: Ensure recalibration is performed before reuse.
- QA disposition: [TBD]
12. Records & Retention
- Logbook entries: Document all calibration activities.
- Calibration report/certificate filing: Maintain records for audit purposes.
- E-records if applicable: [TBD]
- Label content requirements: Calibrated date, due date, technician name, status.
13. Annexures
Annexure A: Calibration Data Sheet
| Test Points | Reference | Observed | Error | Acceptance | Pass/Fail | As Found | As Left |
|---|---|---|---|---|---|---|---|
| [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] |
Annexure B: Quick Checklist
- Environmental controls verified
- Equipment warm-up completed
- Calibration area cleanliness confirmed
- Power supply checked
Annexure C: Example Label Text
Calibrated Date: [TBD]
Due Date: [TBD]
Calibrated By: [TBD]
Status: [TBD]
Calibration Protocol
Calibration Protocol
| Protocol ID | Asset ID | Equipment Name | Equipment Type | Dept | Area | Process Use | Parameter | Units | Range | Acceptance Criteria | Frequency | Calibration Type | Traceability |
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| [TBD] | CAL-0150 | Autoclave Temp Sensor | Temperature Sensor | Microbiology | Sterilization | Sterilization | Temperature | °C | 100 – 140 | ±0.5°C | 180 days | External | ISO17025 |
Environmental Conditions
Temperature: [TBD] °C
Relative Humidity: [TBD] %
Vibration: [TBD]
Drafts: [TBD]
Test Point Selection
- Minimum: 100°C
- 25% of Range: 110°C
- 50% of Range: 120°C
- 75% of Range: 130°C
- Maximum: 140°C
Results Table
| Test Point | Ref Value | Observed (As Found) | Error | Limit | Pass/Fail | Observed (As Left) | Error | Pass/Fail |
|---|---|---|---|---|---|---|---|---|
| 100°C | [TBD] | [TBD] | [TBD] | ±0.5°C | [TBD] | [TBD] | [TBD] | [TBD] |
| 110°C | [TBD] | [TBD] | [TBD] | ±0.5°C | [TBD] | [TBD] | [TBD] | [TBD] |
| 120°C | [TBD] | [TBD] | [TBD] | ±0.5°C | [TBD] | [TBD] | [TBD] | [TBD] |
| 130°C | [TBD] | [TBD] | [TBD] | ±0.5°C | [TBD] | [TBD] | [TBD] | [TBD] |
| 140°C | [TBD] | [TBD] | [TBD] | ±0.5°C | [TBD] | [TBD] | [TBD] | [TBD] |
Stepwise Instructions
- Preparation: Ensure the autoclave temperature sensor is clean and properly installed.
- Stabilization: Allow the sensor to stabilize at room temperature for [TBD] minutes before calibration.
- Performing Readings:
- Take three replicate readings at each test point.
- Record the observed values in the results table.
- Calculating Error: Calculate the error for each observed value using the formula: Error = Observed – Ref Value.
- Evaluation: Compare the calculated error against the acceptance criteria of ±0.5°C.
- Documentation: Complete the results table and ensure all data is recorded accurately (ALCOA+).
Deviation Instruction for OOT
If any observed value falls outside the acceptance criteria:
- Stop use of the equipment immediately.
- Notify the supervisor and quality assurance (QA) personnel.
- Open a deviation report.
- Quarantine impacted results window: [TBD]
- Await QA decision on further actions.
Approvals
| Prepared By | Reviewed By | Approved By |
|---|---|---|
| [Signature/Date] | [Signature/Date] | [Signature/Date] |
Calibration Checklist
| Asset ID | Equipment | Dept | Area | Parameter | Range | Acceptance | Frequency | Traceability |
|---|---|---|---|---|---|---|---|---|
| CAL-0150 | Autoclave Temp Sensor | Microbiology | Sterilization | Temperature | 100 – 140 °C | ±0.5°C | 180 days | ISO17025 |
Pre-calibration
- [ ] Verify environmental conditions are within specified limits
- [ ] Ensure temperature sensor has warmed up adequately
- [ ] Clean the equipment as per SOP
- [ ] Level the equipment if necessary
- [ ] Check validity of reference standards
- [ ] Ensure calibration logbook is ready for entries
As Found Execution
- [ ] Set the temperature sensor to specified setpoints
- [ ] Perform multiple repeats for accuracy
- [ ] Record all raw data accurately
- [ ] Perform necessary calculations for results
OOT Immediate Actions
- [ ] Tag out the equipment if out of tolerance
- [ ] Stop use of the equipment immediately
- [ ] Notify QA and relevant personnel
- [ ] Document deviation and OOT findings
- [ ] Trigger impact review as per SOP
As Left Verification
- [ ] Conduct post-adjustment checks to confirm accuracy
- [ ] Ensure all readings are within acceptance limits
Documentation & Closure
- [ ] Update equipment label with new calibration date
- [ ] Attach calibration certificate to logbook
- [ ] Obtain reviewer sign-off on calibration records
- [ ] Update next due date in the logbook
Data Integrity (ALCOA+): Ensure all data recorded is Attributable, Legible, Contemporaneous, Original, Accurate, and Complete.
Deviation Impact Assessment
| Event Summary |
|---|
|
Asset: Autoclave Temp Sensor Date: [TBD] Discovered By: [TBD] Document References: [TBD] As Found Results Summary: [TBD] |
Immediate Containment Checklist
- Stop use of the asset
- Label/tag the asset as out of tolerance
- Segregate impacted work
Define Impact Window
| Last Known In-Tolerance Date | Discovery Date |
|---|---|
| [TBD] | [TBD] |
Product/Batch/Test Impact Review
| Batch/Study/Report ID | Results Potentially Impacted | Re-test/Re-weigh Decision |
|---|---|---|
| [TBD] | [TBD] | [TBD] |
Data Integrity Review Prompts (ALCOA+)
- Attributable: [TBD]
- Legible: [TBD]
- Contemporaneous: [TBD]
- Original: [TBD]
- Accurate: [TBD]
- Complete: [TBD]
- Consistent: [TBD]
- Enduring: [TBD]
- Available: [TBD]
Risk Assessment
| Severity | Probability | Detectability | Rationale | Overall Risk |
|---|---|---|---|---|
| [TBD] | [TBD] | [TBD] | [TBD] | [TBD] |
Investigation Summary
Root Cause: [TBD]
CAPA Plan
| Action | Owner | Due Date | Effectiveness Check |
|---|---|---|---|
| [TBD] | [TBD] | [TBD] | [TBD] |
QA Disposition
Release/Hold/Retest/Recall Triggers: [TBD]
Calibration Certificate
Certificate No: [TBD]
Date: [TBD]
Next Due Date: [TBD]
Asset Details
| Asset ID | Equipment Name | Equipment Type | Dept | Area | Make/Model | Serial No | Location | Parameter/Range/Units | Acceptance Criteria |
|---|---|---|---|---|---|---|---|---|---|
| CAL-0150 | Autoclave Temp Sensor | Temperature Sensor | Microbiology | Sterilization | [TBD] | [TBD] | [TBD] | Temperature: 100 – 140 °C | ±0.5°C |
Conditions
Environment: [TBD]
Reference Standards
| Standard Name/ID | Certificate No | Due Date | Traceability Statement |
|---|---|---|---|
| [TBD] | [TBD] | [TBD] | Traceability to ISO17025 |
Results
As Found Results
| Test Point | Measured Value (°C) | Comments |
|---|---|---|
| 100 | [TBD] | [TBD] |
| 110 | [TBD] | [TBD] |
| 120 | [TBD] | [TBD] |
| 130 | [TBD] | [TBD] |
| 140 | [TBD] | [TBD] |
As Left Results
| Test Point | Measured Value (°C) | Comments |
|---|---|---|
| 100 | [TBD] | [TBD] |
| 110 | [TBD] | [TBD] |
| 120 | [TBD] | [TBD] |
| 130 | [TBD] | [TBD] |
| 140 | [TBD] | [TBD] |
Measurement Uncertainty Statement: [TBD]
Statement of Conformity: [Pass/Fail overall]
Remarks: [TBD]
Signatures
Performed by: ______________________ Date: ___________
Checked by: ______________________ Date: ___________
Approved by: ______________________ Date: ___________