Calibration Documents: Bioreactor Temperature Probe (CAL-0300)
| Asset ID | CAL-0300 |
|---|---|
| Equipment | Bioreactor Temperature Probe |
| Equipment Type | Temperature Probe |
| Department | Biologics |
| Area/Location | Bioreactor |
| Process Use | Temp control |
| Parameter | Temperature |
| Units | °C |
| Range Min | 0 |
| Range Max | 60 |
| Acceptance Criteria | ±0.5°C |
| Calibration Frequency (days) | 180 |
| Calibration Type | External |
| Traceability | ISO17025 |
Table of Contents
Calibration SOP
1. Purpose
The calibration of the Bioreactor Temperature Probe is performed to ensure accurate temperature control within the bioreactor. This calibration is critical for maintaining product quality, ensuring patient safety, and preserving data integrity during biologics manufacturing processes. Accurate temperature measurements are essential to prevent deviations that could affect batch quality and compliance with regulatory standards.
2. Scope
- Asset ID: CAL-0300
- Equipment Name/Type: Bioreactor Temperature Probe / Temperature Probe
- Department/Area: Biologics / Bioreactor
- Process Use: Temperature control
- Parameter: Temperature
- Range: 0°C to 60°C
- Units: °C
- Calibration Frequency: 180 days
3. Definitions
- Calibration: The process of comparing a measurement device against a known standard to determine its accuracy.
- Verification: The process of confirming that a measurement device meets specified requirements.
- Adjustment: The process of modifying a measurement device to bring it into compliance with specified standards.
- As Found: The condition of the measurement device prior to any adjustments.
- As Left: The condition of the measurement device after adjustments have been made.
- OOT (Out of Tolerance): A condition where the measurement device does not meet specified acceptance criteria.
- Drift: The gradual change in a measurement device’s output over time.
- Traceability: The ability to verify the history, location, or application of an item by means of documented recorded identification.
- Measurement Uncertainty: A parameter that characterizes the range of values within which the true value is expected to lie.
- Intermediate Check: A periodic check performed between calibrations to ensure continued accuracy.
4. References
- Company Metrology SOP [TBD]
- ISO/IEC 17025 traceability via calibration certificates
- ALCOA+ principles
5. Responsibilities
- Technician/Metrology: Perform calibration and document results.
- QC/QA Reviewer: Review calibration documentation for compliance.
- Owner Department: Ensure equipment is available and maintained.
- External Service Provider: Conduct external calibration as required.
6. Safety & Preconditions
- Ensure environmental controls are in place (temperature, humidity).
- Allow equipment to warm up and stabilize as per manufacturer guidelines.
- Maintain cleanliness of the calibration area and equipment.
- Check leveling of the equipment prior to calibration.
- Verify power supply stability and functionality.
7. Standards & Tools
- Reference Standards: [TBD]
- Traceability Statement: Calibration traceable to ISO17025.
- Certificate Validity Check: Ensure all calibration certificates are current.
- Standard ID: [TBD]
8. Calibration Strategy
Calibration will follow the As Found and As Left logic:
- Minimum number of points: At least 5 points including endpoints (0°C, 60°C) and midpoints (30°C).
- Replicate readings: Perform 3 repeats at each point as a best practice.
9. Step-by-Step Procedure
- Preparation: Gather all necessary tools and standards.
- Stabilization: Allow the temperature probe to stabilize at the calibration environment.
- As Found checks and data capture: Record initial temperature readings at specified points.
- Adjustment rules: If OOT is found, follow escalation procedures; adjustments are only permitted if allowed.
- As Left verification: Confirm that the probe is within acceptance criteria after adjustments.
- Labeling + documentation completion: Complete all required documentation and label the equipment accordingly.
Record all raw observations contemporaneously to ensure ALCOA+ compliance.
10. Acceptance Criteria & Calculations
Acceptance criteria: ±0.5°C.
Error calculation formula: Error = Observed Value – True Value.
Pass/Fail Logic: If the absolute value of Error is within acceptance criteria, mark as Pass; otherwise, mark as Fail.
Rounding rule: [TBD]
11. Handling OOT / Failures
- Immediate containment: Tag equipment and stop use until resolved.
- Notification & deviation initiation: Notify QA and initiate deviation process.
- Impact assessment trigger: Define impact window and review affected results/batches.
- Recalibration after repair/adjustment: Ensure recalibration is performed post-adjustment.
- QA disposition placeholders: [TBD]
12. Records & Retention
- Logbook entries: Document all calibration activities.
- Calibration report/certificate filing: Maintain records in accordance with company policy.
- e-Records if applicable: [TBD]
- Label content requirements: Include calibrated date, due date, technician name, and status.
13. Annexures
Annexure A: Calibration Data Sheet
| Test Points | Reference | Observed | Error | Acceptance | Pass/Fail | As Found | As Left |
|---|---|---|---|---|---|---|---|
| [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] |
Annexure B: Quick Checklist
- Environmental controls verified
- Equipment warm-up completed
- Calibration tools prepared
- Documentation ready
Annexure C: Example Label Text
Calibration Date: [TBD]
Due Date: [TBD]
Calibrated By: [TBD]
Status: [TBD]
Calibration Protocol
Calibration Protocol
| Protocol ID | Asset ID | Equipment Name | Equipment Type | Dept | Area | Process Use | Parameter | Units | Range | Acceptance Criteria | Frequency | Calibration Type | Traceability |
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| [TBD] | CAL-0300 | Bioreactor Temperature Probe | Temperature Probe | Biologics | Bioreactor | Temp control | Temperature | °C | 0 – 60 | ±0.5°C | 180 days | External | ISO17025 |
Environmental Conditions
Temperature: [TBD] °C
Relative Humidity: [TBD] %
Vibration: [TBD]
Drafts: [TBD]
Test Point Selection
- Minimum: 0°C
- 25% of Range: 15°C
- 50% of Range: 30°C
- 75% of Range: 45°C
- Maximum: 60°C
Results Table
| Test Point | Ref Value | Observed (As Found) | Error | Limit | Pass/Fail | Observed (As Left) | Error | Pass/Fail |
|---|---|---|---|---|---|---|---|---|
| 0°C | [TBD] | [TBD] | [TBD] | ±0.5°C | [TBD] | [TBD] | [TBD] | [TBD] |
| 15°C | [TBD] | [TBD] | [TBD] | ±0.5°C | [TBD] | [TBD] | [TBD] | [TBD] |
| 30°C | [TBD] | [TBD] | [TBD] | ±0.5°C | [TBD] | [TBD] | [TBD] | [TBD] |
| 45°C | [TBD] | [TBD] | [TBD] | ±0.5°C | [TBD] | [TBD] | [TBD] | [TBD] |
| 60°C | [TBD] | [TBD] | [TBD] | ±0.5°C | [TBD] | [TBD] | [TBD] | [TBD] |
Stepwise Instructions
- Preparation: Ensure the bioreactor temperature probe is clean and properly connected.
- Stabilization: Allow the probe to stabilize at room temperature for at least 30 minutes.
- Performing Readings: Take three replicate readings at each test point to ensure accuracy.
- Calculating Error: Calculate the error as the difference between the observed value and the reference value.
- Evaluation: Compare the calculated error against the acceptance criteria of ±0.5°C.
- Documentation: Record all observations, errors, and results in the results table.
Deviation Instruction for OOT
If any reading is outside the acceptance criteria:
- Stop use of the equipment immediately.
- Notify the QA department.
- Open a deviation report.
- Quarantine impacted results window: [TBD]
- Await QA decision on further actions.
Approvals
| Prepared | Reviewed | Approved |
|---|---|---|
| [Signature/Date] | [Signature/Date] | [Signature/Date] |
Calibration Checklist
| Asset ID | Equipment | Dept | Area | Parameter | Range | Acceptance | Frequency | Traceability |
|---|---|---|---|---|---|---|---|---|
| CAL-0300 | Bioreactor Temperature Probe | Biologics | Bioreactor | Temperature | 0 to 60 °C | ±0.5°C | 180 days | ISO17025 |
Pre-calibration
- [ ] Verify environmental conditions are within specified limits
- [ ] Allow equipment to warm up as per SOP
- [ ] Clean the equipment as per cleaning procedure
- [ ] Level the equipment if necessary
- [ ] Check validity of reference standards
- [ ] Ensure logbook is ready for documentation
As Found Execution
- [ ] Set the temperature probe to specified setpoints
- [ ] Perform repeat measurements as required
- [ ] Record raw data accurately
- [ ] Perform necessary calculations for results
OOT Immediate Actions
- [ ] Tag out the equipment if out of tolerance
- [ ] Stop use of the equipment immediately
- [ ] Notify relevant personnel of the OOT condition
- [ ] Document deviation in the logbook
- [ ] Trigger impact review as per SOP
As Left Verification
- [ ] Perform post-adjustment checks to confirm accuracy
- [ ] Ensure all measurements are within acceptance limits
Documentation & Closure
- [ ] Update equipment label with calibration details
- [ ] Attach calibration certificate to logbook
- [ ] Obtain reviewer sign-off on calibration record
- [ ] Update next due date for calibration
Data Integrity (ALCOA+): Ensure all data recorded is Attributable, Legible, Contemporaneous, Original, Accurate, and Complete.
Deviation Impact Assessment
| Event Summary |
|---|
|
Asset: Bioreactor Temperature Probe Date: [TBD] Discovered By: [TBD] Document References: [TBD] As Found Results Summary: [TBD] |
Immediate Containment Checklist
- Stop use of the asset
- Label/Tag the asset as out of service
- Segregate impacted work
Define Impact Window
| Last Known In-Tolerance Date | Discovery Date |
|---|---|
| [TBD] | [TBD] |
Product/Batch/Test Impact Review
| Batch/Study/Report ID | Results Potentially Impacted | Re-test/Re-weigh Decision |
|---|---|---|
| [TBD] | [TBD] | [TBD] |
Data Integrity Review Prompts (ALCOA+)
- Attributable: [TBD]
- Legible: [TBD]
- Contemporaneous: [TBD]
- Original: [TBD]
- Accurate: [TBD]
- Complete: [TBD]
- Consistent: [TBD]
- Enduring: [TBD]
- Available: [TBD]
Risk Assessment
| Severity | Probability | Detectability | Rationale | Overall Risk |
|---|---|---|---|---|
| [TBD] | [TBD] | [TBD] | [TBD] | [TBD] |
Investigation Summary
Root Cause: [TBD]
CAPA Plan
| Action | Owner | Due Date | Effectiveness Check |
|---|---|---|---|
| [TBD] | [TBD] | [TBD] | [TBD] |
QA Disposition
Release/Hold/Retest/Recall Triggers: [TBD]
Calibration Certificate
Certificate No: [TBD]
Date: [TBD]
Next Due Date: [TBD]
Asset Details
| Asset ID | Equipment Name | Equipment Type | Dept | Area | Make/Model | Serial No | Location | Parameter/Range/Units | Acceptance Criteria |
|---|---|---|---|---|---|---|---|---|---|
| CAL-0300 | Bioreactor Temperature Probe | Temperature Probe | Biologics | Bioreactor | [TBD] | [TBD] | [TBD] | Temperature: 0 – 60 °C | ±0.5°C |
Conditions
Environment: [TBD]
Reference Standards
| Standard Name/ID | Certificate No | Due Date | Traceability Statement |
|---|---|---|---|
| [TBD] | [TBD] | [TBD] | Traceability to ISO17025 |
Results
As Found Results
| Test Point | Measured Value (°C) | Pass/Fail |
|---|---|---|
| 0 | [TBD] | [TBD] |
| 15 | [TBD] | [TBD] |
| 30 | [TBD] | [TBD] |
| 45 | [TBD] | [TBD] |
| 60 | [TBD] | [TBD] |
As Left Results
| Test Point | Measured Value (°C) | Pass/Fail |
|---|---|---|
| 0 | [TBD] | [TBD] |
| 15 | [TBD] | [TBD] |
| 30 | [TBD] | [TBD] |
| 45 | [TBD] | [TBD] |
| 60 | [TBD] | [TBD] |
Measurement Uncertainty Statement: [TBD]
Statement of Conformity: [Pass/Fail overall]
Remarks: [TBD]
Signatures
Performed by: ______________________ Date: ___________
Checked by: ______________________ Date: ___________
Approved by: ______________________ Date: ___________