Calibration Documents: Burette 25 mL (CAL-0029)
| Asset ID | CAL-0029 |
|---|---|
| Equipment | Burette 25 mL |
| Equipment Type | Burette |
| Department | QC |
| Area/Location | Chemistry Lab |
| Process Use | Titration |
| Parameter | Volume |
| Units | mL |
| Range Min | 0 |
| Range Max | 25 |
| Acceptance Criteria | Class A |
| Calibration Frequency (days) | 365 |
| Calibration Type | External |
| Traceability | ISO17025 |
Table of Contents
Calibration SOP
1. Purpose
The calibration of the Burette 25 mL is performed to ensure accurate volume measurements during titration processes in the QC department. This calibration controls risks related to accuracy, patient safety, batch quality, and data integrity, thereby ensuring compliance with regulatory standards.
2. Scope
- Asset ID: CAL-0029
- Equipment Name/Type: Burette 25 mL / Burette
- Department/Area: QC / Chemistry Lab
- Process Use: Titration
- Parameter: Volume
- Range: 0 – 25 mL
- Units: mL
- Calibration Frequency: 365 days
3. Definitions
- Calibration: The process of verifying and adjusting the accuracy of an instrument against a known standard.
- Verification: The process of confirming that the instrument meets specified requirements.
- Adjustment: The act of modifying an instrument to bring it into compliance with specifications.
- As Found: The condition of the instrument prior to any adjustments.
- As Left: The condition of the instrument after adjustments have been made.
- OOT (Out of Tolerance): A condition where the instrument measurements fall outside specified acceptance criteria.
- Drift: The gradual change in the measurement accuracy of an instrument over time.
- Traceability: The ability to verify the history, location, or application of an item by means of documented recorded identification.
- Measurement Uncertainty: A parameter that characterizes the range of values within which the true value is asserted to lie.
- Intermediate Check: A periodic verification of the instrument’s performance between calibrations.
4. References
- Company Metrology SOP [TBD]
- ISO/IEC 17025 traceability via calibration certificates
- ALCOA+ principles
5. Responsibilities
- Technician/Metrology: Perform calibration and document results.
- QC/QA Reviewer: Review calibration documentation and results.
- Owner Department: Ensure equipment is available and maintained.
- External Service Provider: Conduct external calibrations as required.
6. Safety & Preconditions
- Ensure environmental controls are in place (temperature, humidity).
- Allow equipment to warm up and stabilize as per manufacturer guidelines.
- Maintain cleanliness of the equipment and surrounding area.
- Check that the burette is leveled before use.
- Verify power supply is stable and within specifications.
7. Standards & Tools
- Reference Standards Required: Class A volumetric standards.
- Traceability Statement: Calibration traceable to ISO17025.
- Certificate Validity Check: Ensure calibration certificates are current and valid.
- Standard ID Placeholders: [TBD]
8. Calibration Strategy
Calibration will follow the As Found and As Left logic. A minimum of 5 measurement points will be taken, including endpoints (0 mL and 25 mL) and midpoints (12.5 mL). Each point will have 3 replicate readings to ensure accuracy.
9. Step-by-Step Procedure
- Preparation: Gather all necessary tools and standards.
- Stabilization: Allow the burette to stabilize in the environment.
- As Found checks and data capture: Measure and record the volume at each point.
- Adjustment rules: If OOT is detected, follow adjustment procedures or escalate as necessary.
- As Left verification: Confirm that the instrument is within tolerance after adjustments.
- Labeling + documentation completion: Complete all required documentation and label the equipment.
Record raw observations contemporaneously to ensure data integrity (ALCOA+).
10. Acceptance Criteria & Calculations
Acceptance criteria for the Burette 25 mL is Class A. The error calculation formula is as follows:
Error (%) = (Observed Value – True Value) / True Value x 100
Pass/Fail logic: If the error falls within the acceptance criteria, the calibration is deemed a pass; otherwise, it is a fail.
Rounding rule: [TBD]
11. Handling OOT / Failures
- Immediate containment: Tag equipment and cease use.
- Notification & deviation initiation: Notify QA and initiate deviation process.
- Impact assessment trigger: Define impact window and review affected results/batches.
- Recalibration after repair/adjustment: Ensure recalibration is performed post-adjustment.
- QA disposition placeholders: [TBD]
12. Records & Retention
- Logbook entries: Document all calibration activities.
- Calibration report/certificate filing: Maintain records of calibration certificates.
- e-Records if applicable: [TBD]
- Label content requirements: Include calibrated date, due date, technician name, and status.
13. Annexures
Annexure A: Calibration Data Sheet
| Test Points | Reference | Observed | Error | Acceptance | Pass/Fail | As Found | As Left |
|---|---|---|---|---|---|---|---|
| [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] |
Annexure B: Quick Checklist
- Environmental controls verified
- Equipment cleaned and leveled
- Calibration standards available
- Documentation ready
Annexure C: Example Label Text
Calibration Date: [TBD]
Due Date: [TBD]
Calibrated By: [TBD]
Status: [TBD]
Calibration Protocol
Calibration Protocol
| Protocol ID | Asset ID | Equipment Name | Equipment Type | Dept | Area | Process Use | Parameter | Units | Range | Acceptance Criteria | Frequency | Calibration Type | Traceability |
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| [TBD] | CAL-0029 | Burette 25 mL | Burette | QC | Chemistry Lab | Titration | Volume | mL | 0 – 25 | Class A | 365 days | External | ISO17025 |
Environmental Conditions
Temperature: [TBD]
Relative Humidity: [TBD]
Vibration: [TBD]
Drafts: [TBD]
Test Point Selection
Test points for calibration will be selected as follows:
- Minimum (0 mL)
- 25% of Range (6.25 mL)
- 50% of Range (12.5 mL)
- 75% of Range (18.75 mL)
- Maximum (25 mL)
Results Table
| Test Point | Ref Value | Observed (As Found) | Error | Limit | Pass/Fail | Observed (As Left) | Error | Pass/Fail |
|---|---|---|---|---|---|---|---|---|
| 0 mL | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] |
| 6.25 mL | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] |
| 12.5 mL | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] |
| 18.75 mL | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] |
| 25 mL | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] |
Stepwise Instructions
- Preparation: Ensure the burette is clean and free from any contaminants.
- Stabilization: Allow the burette to stabilize at room temperature for at least 30 minutes.
- Performing Readings:
- Fill the burette with the calibration standard.
- Take three replicate readings at each test point.
- Calculating Error: Calculate the error for each observed value against the reference value.
- Evaluation: Compare the calculated errors against the acceptance criteria (Class A).
- Documentation: Record all observations in the results table and ensure data integrity (ALCOA+).
Deviation Instruction for OOT
If any observed value is out of tolerance (OOT):
- Stop use of the equipment immediately.
- Notify the supervisor and quality assurance (QA) department.
- Open a deviation report.
- Quarantine impacted results window: [TBD]
- Await QA decision on further actions.
Approvals
| Prepared | Reviewed | Approved |
|---|---|---|
| [Signature/Date] | [Signature/Date] | [Signature/Date] |
Calibration Checklist
| Asset ID | Equipment | Dept | Area | Parameter | Range | Acceptance | Frequency | Traceability |
|---|---|---|---|---|---|---|---|---|
| CAL-0029 | Burette 25 mL | QC | Chemistry Lab | Volume | 0 – 25 mL | Class A | 365 days | ISO17025 |
1) Pre-calibration
- [ ] Verify environmental conditions are within specified limits
- [ ] Ensure equipment has warmed up adequately
- [ ] Clean the burette as per SOP
- [ ] Level the burette properly
- [ ] Check validity of reference standards
- [ ] Ensure logbook is ready for documentation
2) As Found Execution
- [ ] Set initial calibration setpoints
- [ ] Perform repeat measurements as required
- [ ] Record all raw data accurately
- [ ] Perform necessary calculations for results
3) OOT Immediate Actions
- [ ] Tag out the equipment if out of tolerance (OOT)
- [ ] Stop use of the equipment immediately
- [ ] Notify the appropriate personnel
- [ ] Document deviation in the logbook
- [ ] Trigger impact review process
4) As Left Verification
- [ ] Conduct post-adjustment checks
- [ ] Confirm that all measurements are within specified limits
5) Documentation & Closure
- [ ] Update equipment label with calibration details
- [ ] Attach calibration certificate to logbook
- [ ] Obtain reviewer sign-off on calibration report
- [ ] Update next due date for calibration
Acceptance Criteria: Class A
Data Integrity (ALCOA+): Ensure all data is Attributable, Legible, Contemporaneous, Original, Accurate, and Complete.
As Found/As Left: Document the condition of the equipment before and after calibration.
Deviation Impact Assessment
| Event Summary |
|---|
|
Asset: Burette 25 mL Date: [TBD] Discovered By: [TBD] Document References: [TBD] As Found Results Summary: [TBD] |
Immediate Containment Checklist
- Stop use of the asset
- Label/tag the asset as out of tolerance
- Segregate impacted work
Define Impact Window
Last known in-tolerance date: [TBD] to discovery date: [TBD]
Product/Batch/Test Impact Review
| Batch/Study/Report ID | Results Potentially Impacted | Re-test/Re-weigh Decision |
|---|---|---|
| [TBD] | [TBD] | [TBD] |
Data Integrity Review Prompts (ALCOA+)
- Attributable: [TBD]
- Legible: [TBD]
- Contemporaneous: [TBD]
- Original: [TBD]
- Accurate: [TBD]
- Complete: [TBD]
- Consistent: [TBD]
- Enduring: [TBD]
- Available: [TBD]
Risk Assessment
| Severity | Probability | Detectability | Rationale | Overall Risk |
|---|---|---|---|---|
| [TBD] | [TBD] | [TBD] | [TBD] | [TBD] |
Investigation Summary
Root Cause: [TBD]
CAPA Plan
| Action | Owner | Due Date | Effectiveness Check |
|---|---|---|---|
| [TBD] | [TBD] | [TBD] | [TBD] |
QA Disposition
Release/Hold/Retest/Recall Triggers: [TBD]
Calibration Certificate
Certificate No: [TBD]
Date: [TBD]
Next Due Date: [TBD]
Asset Details
| Asset ID | Equipment Name | Equipment Type | Dept | Area | Make/Model | Serial No | Location | Parameter/Range/Units | Acceptance Criteria |
|---|---|---|---|---|---|---|---|---|---|
| CAL-0029 | Burette 25 mL | Burette | QC | Chemistry Lab | [TBD] | [TBD] | [TBD] | Volume: 0 – 25 mL | Class A |
Conditions
Environment: [TBD]
Reference Standards
| Standard Name/ID | Certificate No | Due Date | Traceability Statement |
|---|---|---|---|
| [TBD] | [TBD] | [TBD] | Traceable to ISO17025 |
Results
As Found Results
| Test Point | Measured Value (mL) | Acceptance Criteria | Pass/Fail |
|---|---|---|---|
| 0 mL | [TBD] | Class A | [TBD] |
| 5 mL | [TBD] | Class A | [TBD] |
| 10 mL | [TBD] | Class A | [TBD] |
| 15 mL | [TBD] | Class A | [TBD] |
| 20 mL | [TBD] | Class A | [TBD] |
| 25 mL | [TBD] | Class A | [TBD] |
As Left Results
| Test Point | Measured Value (mL) | Acceptance Criteria | Pass/Fail |
|---|---|---|---|
| 0 mL | [TBD] | Class A | [TBD] |
| 5 mL | [TBD] | Class A | [TBD] |
| 10 mL | [TBD] | Class A | [TBD] |
| 15 mL | [TBD] | Class A | [TBD] |
| 20 mL | [TBD] | Class A | [TBD] |
| 25 mL | [TBD] | Class A | [TBD] |
Measurement Uncertainty Statement: [TBD]
Statement of Conformity: [TBD]
Remarks: [TBD]
Signatures
Performed by: ____________________ Date: ___________
Checked by: ______________________ Date: ___________
Approved by: _____________________ Date: ___________