CO2 Analyzer (Incubator) (CAL-0099) – Calibration Documents

Calibration Documents: CO2 Analyzer (Incubator) (CAL-0099)

Asset ID CAL-0099
Equipment CO2 Analyzer (Incubator)
Equipment Type CO2 Analyzer
Department Biologics
Area/Location Cell Culture
Process Use CO2 control
Parameter CO2
Units %CO2
Range Min 0
Range Max 20
Acceptance Criteria ±0.5%
Calibration Frequency (days) 180
Calibration Type External
Traceability ISO17025

Calibration SOP

1. Purpose

The calibration of the CO2 Analyzer (Incubator) is performed to ensure accurate CO2 control within the incubator, which is critical for maintaining optimal cell culture conditions. This calibration mitigates risks associated with inaccurate CO2 levels that could compromise patient safety, batch quality, and data integrity.

2. Scope

  • Asset ID: CAL-0099
  • Equipment Name/Type: CO2 Analyzer (Incubator)
  • Department/Area: Biologics / Cell Culture
  • Process Use: CO2 control
  • Parameter: CO2
  • Range: 0%CO2 to 20%CO2
  • Units: %CO2
  • Calibration Frequency: 180 days

3. Definitions

  • Calibration: The process of verifying and adjusting the accuracy of an instrument.
  • Verification: The confirmation that an instrument meets specified requirements.
  • Adjustment: The process of modifying an instrument to bring it within specified limits.
  • As Found: The condition of the instrument prior to any adjustments.
  • As Left: The condition of the instrument after adjustments have been made.
  • OOT (Out of Tolerance): A condition where the instrument’s measurements fall outside the specified acceptance criteria.
  • Drift: The gradual change in an instrument’s output over time.
  • Traceability: The ability to verify the history, location, or application of an item by means of documented recorded identification.
  • Measurement Uncertainty: A parameter that characterizes the range of values within which the true value is expected to lie.
  • Intermediate Check: A periodic assessment of an instrument’s performance between calibrations.

4. References

  • Company Metrology SOP [TBD]
  • ISO/IEC 17025 traceability via calibration certificates
  • ALCOA+ principles

5. Responsibilities

  • Technician/Metrology: Perform calibration and document results.
  • QC/QA Reviewer: Review calibration documentation and results.
  • Owner Department: Ensure equipment is available and maintained.
  • External Service Provider: Conduct external calibration as required.

6. Safety & Preconditions

  • Ensure environmental controls are in place (temperature, humidity).
  • Allow equipment to warm up and stabilize as per manufacturer’s instructions.
  • Maintain cleanliness of the equipment and surrounding area.
  • Check leveling of the equipment.
  • Verify power supply stability.
See also  Load Cell (Bin blender) (CAL-0012) - Calibration Documents

7. Standards & Tools

  • Reference Standards Required: [TBD]
  • Traceability Statement: ISO17025
  • Certificate Validity Check: Ensure all reference standards are within their validity period.

8. Calibration Strategy

Calibration will follow an As Found and As Left logic:

  • Minimum number of points: At least 5 points including endpoints and midpoints within the range of 0%CO2 to 20%CO2.
  • Replicate readings at each point: 3 repeats as a best practice.

9. Step-by-Step Procedure

  1. Preparation: Gather all necessary tools and reference standards.
  2. Stabilization: Allow the CO2 Analyzer to stabilize in the operating environment.
  3. As Found checks and data capture: Measure CO2 levels at defined points and record results contemporaneously.
  4. Adjustment rules: If OOT is detected, follow adjustment protocols; otherwise, escalate to QA.
  5. As Left verification: Confirm that all adjustments bring the instrument within acceptance criteria.
  6. Labeling + documentation completion: Complete all required documentation and label the equipment appropriately.

10. Acceptance Criteria & Calculations

The acceptance criteria for the CO2 Analyzer is ±0.5%. The error calculation formula is:

Error (%) = (Observed Value – True Value) / True Value × 100

Pass/Fail Logic: If the error falls within ±0.5%, the calibration is considered a pass; otherwise, it is a fail.

Rounding rule: [TBD]

11. Handling OOT / Failures

  • Immediate containment: Tag the equipment and stop use.
  • Notification & deviation initiation: Notify QA and initiate deviation procedures.
  • Impact assessment trigger: Define impact window and review of affected results/batches.
  • Recalibration after repair/adjustment: Ensure recalibration is performed post-adjustment.
  • QA disposition placeholders: [TBD]

12. Records & Retention

  • Logbook entries: Document all calibration activities.
  • Calibration report/certificate filing: Maintain records of calibration certificates.
  • e-Records if applicable: [TBD]
  • Label content requirements: Include calibrated date, due date, technician name, and status.

13. Annexures

Annexure A: Calibration Data Sheet

Test Points Reference Observed Error Acceptance Pass/Fail As Found As Left
[TBD] [TBD] [TBD] [TBD] [TBD] [TBD] [TBD] [TBD]

Annexure B: Quick Checklist

  • Environmental controls verified
  • Equipment stabilized
  • Calibration tools prepared
  • Documentation ready

Annexure C: Example Label Text

Calibration Date: [TBD]

Due Date: [TBD]

Calibrated By: [TBD]

Status: [TBD]

⬆ Back to TOC


Calibration Protocol

Calibration Protocol

Protocol ID Asset ID Equipment Name Equipment Type Dept Area Process Use Parameter Units Range Acceptance Criteria Frequency Calibration Type Traceability
[TBD] CAL-0099 CO2 Analyzer (Incubator) CO2 Analyzer Biologics Cell Culture CO2 control CO2 %CO2 0 – 20 ±0.5% 180 days External ISO17025
See also  Stability Chamber Controller Display Temp (CAL-0182) - Calibration Documents

Environmental Conditions

Temperature: [TBD]

Relative Humidity: [TBD]

Vibration: [TBD]

Drafts: [TBD]

Test Point Selection

  • Min: 0% CO2
  • 25%: 5% CO2
  • 50%: 10% CO2
  • 75%: 15% CO2
  • Max: 20% CO2

Results Table

Test Point Ref Value Observed (As Found) Error Limit Pass/Fail Observed (As Left) Error Pass/Fail
0% CO2 0% CO2 [TBD] [TBD] ±0.5% [TBD] [TBD] [TBD] [TBD]
5% CO2 5% CO2 [TBD] [TBD] ±0.5% [TBD] [TBD] [TBD] [TBD]
10% CO2 10% CO2 [TBD] [TBD] ±0.5% [TBD] [TBD] [TBD] [TBD]
15% CO2 15% CO2 [TBD] [TBD] ±0.5% [TBD] [TBD] [TBD] [TBD]
20% CO2 20% CO2 [TBD] [TBD] ±0.5% [TBD] [TBD] [TBD] [TBD]

Stepwise Instructions

  1. Preparation: Ensure the CO2 Analyzer is clean and properly calibrated prior to use.
  2. Stabilization: Allow the analyzer to stabilize for at least [TBD] minutes before taking readings.
  3. Performing Readings: Take three replicate readings at each test point.
  4. Calculating Error: Calculate the error for each observed value against the reference value.
  5. Evaluation: Compare the calculated errors against the acceptance criteria.
  6. Documentation: Record all observations, errors, and results in the results table.

Deviation Instruction for OOT

If any observed value falls outside the acceptance criteria:

  • Stop use of the equipment immediately.
  • Notify the supervisor and QA.
  • Open a deviation report.
  • Quarantine the impacted results window: [TBD].
  • Await QA decision on further actions.

Approvals

Prepared By Reviewed By Approved By
[Signature/Date] [Signature/Date] [Signature/Date]

⬆ Back to TOC


Calibration Checklist

Asset ID Equipment Dept Area Parameter Range Acceptance Frequency Traceability
CAL-0099 CO2 Analyzer (Incubator) Biologics Cell Culture CO2 0 – 20 %CO2 ±0.5% 180 days ISO17025

Pre-calibration

  • [ ] Verify environmental conditions are within specifications
  • [ ] Ensure equipment has warmed up for the required duration
  • [ ] Clean the equipment as per SOP
  • [ ] Level the equipment if necessary
  • [ ] Check validity of reference standards
  • [ ] Ensure logbook is ready for documentation

As Found Execution

  • [ ] Set the initial CO2 setpoints
  • [ ] Perform repeat measurements as required
  • [ ] Record all raw data accurately
  • [ ] Perform necessary calculations for results
See also  Dew Point Meter (CAL-0062) - Calibration Documents

OOT Immediate Actions

  • [ ] Tag out the equipment
  • [ ] Stop use of the equipment
  • [ ] Notify the appropriate personnel
  • [ ] Document the deviation
  • [ ] Trigger impact review process

As Left Verification

  • [ ] Conduct post-adjustment checks
  • [ ] Confirm all measurements are within acceptance limits

Documentation & Closure

  • [ ] Update equipment label
  • [ ] Attach calibration certificate
  • [ ] Obtain reviewer sign-off
  • [ ] Update next due date in logbook

Data Integrity (ALCOA+): Ensure all data recorded is Attributable, Legible, Contemporaneous, Original, Accurate, and Complete.

⬆ Back to TOC


Deviation Impact Assessment

Event Summary Details
Asset CO2 Analyzer (Incubator)
Date [TBD]
Discovered By [TBD]
Document References [TBD]
As Found Results Summary [TBD]

Immediate Containment Checklist

  • Stop use of the equipment
  • Label/Tag the equipment as out of service
  • Segregate impacted work

Define Impact Window

Last Known In-Tolerance Date Discovery Date
[TBD] [TBD]

Product/Batch/Test Impact Review

Batch/Study/Report ID Results Potentially Impacted Re-Test/Re-Weigh Decision
[TBD] [TBD] [TBD]

Data Integrity Review Prompts (ALCOA+)

  • Attributable: [TBD]
  • Legible: [TBD]
  • Contemporaneous: [TBD]
  • Original: [TBD]
  • Accurate: [TBD]
  • Complete: [TBD]
  • Consistent: [TBD]
  • Enduring: [TBD]
  • Available: [TBD]

Risk Assessment

Severity Probability Detectability Rationale Overall Risk
[TBD] [TBD] [TBD] [TBD] [TBD]

Investigation Summary

Root Cause: [TBD]

CAPA Plan

Action Owner Due Date Effectiveness Check
[TBD] [TBD] [TBD] [TBD]

QA Disposition

Release/Hold/Retest/Recall Triggers: [TBD]

⬆ Back to TOC


Calibration Certificate

Certificate No: [TBD]

Date: [TBD]

Next Due Date: [TBD]

Asset Details

Asset ID Equipment Name Equipment Type Dept Area Make/Model Serial No Location Parameter/Range/Units Acceptance Criteria
CAL-0099 CO2 Analyzer (Incubator) CO2 Analyzer Biologics Cell Culture [TBD] [TBD] [TBD] CO2: 0 – 20 %CO2 ±0.5%

Conditions

Environment: [TBD]

Reference Standards

Standard Name/ID Certificate No Due Date Traceability Statement
[TBD] [TBD] [TBD] Traceability: ISO17025

Results

As Found Results

Test Point Measured Value (%CO2) Acceptance Criteria Pass/Fail
1 [TBD] ±0.5% [TBD]
2 [TBD] ±0.5% [TBD]
3 [TBD] ±0.5% [TBD]
4 [TBD] ±0.5% [TBD]
5 [TBD] ±0.5% [TBD]

As Left Results

Test Point Measured Value (%CO2) Acceptance Criteria Pass/Fail
1 [TBD] ±0.5% [TBD]
2 [TBD] ±0.5% [TBD]
3 [TBD] ±0.5% [TBD]
4 [TBD] ±0.5% [TBD]
5 [TBD] ±0.5% [TBD]

Measurement Uncertainty Statement: [TBD]

Statement of Conformity: [TBD]

Remarks: [TBD]

Signatures

Performed by: ____________________ Date: ___________

Checked by: ______________________ Date: ___________

Approved by: _____________________ Date: ___________

⬆ Back to TOC