Calibration Documents: CO2 Incubator Temp Sensor (CAL-0306)
| Asset ID | CAL-0306 |
|---|---|
| Equipment | CO2 Incubator Temp Sensor |
| Equipment Type | Temperature Sensor |
| Department | Biologics |
| Area/Location | Cell Culture |
| Process Use | Temperature |
| Parameter | Temperature |
| Units | °C |
| Range Min | 20 |
| Range Max | 45 |
| Acceptance Criteria | ±0.5°C |
| Calibration Frequency (days) | 180 |
| Calibration Type | External |
| Traceability | ISO17025 |
Table of Contents
Calibration SOP
1. Purpose
The calibration of the CO2 Incubator Temperature Sensor is performed to ensure accurate temperature readings within the specified range, which is critical for maintaining optimal conditions in cell culture processes. This calibration controls risks related to patient safety, batch quality, and data integrity by ensuring that the temperature measurements are reliable and consistent.
2. Scope
- Asset ID: CAL-0306
- Equipment Name/Type: CO2 Incubator Temp Sensor / Temperature Sensor
- Department/Area: Biologics / Cell Culture
- Process Use: Temperature
- Parameter: Temperature
- Range: 20°C to 45°C
- Units: °C
- Calibration Frequency: Every 180 days
3. Definitions
- Calibration: The process of comparing a measurement instrument to a known standard to determine its accuracy.
- Verification: The process of confirming that a measurement instrument is functioning correctly.
- Adjustment: The act of modifying a measurement instrument to bring it into alignment with a standard.
- As Found: The condition of the equipment prior to any adjustments being made.
- As Left: The condition of the equipment after adjustments have been made.
- OOT (Out of Tolerance): A condition where the measurement falls outside the defined acceptance criteria.
- Drift: The gradual deviation of a measurement instrument from its true value over time.
- Traceability: The ability to verify the history, location, or application of an item by means of documented recorded identification.
- Measurement Uncertainty: A parameter that characterizes the range of values within which the true value of a measurement lies.
- Intermediate Check: A verification process conducted between calibrations to ensure ongoing accuracy.
4. References
- Company Metrology SOP [TBD]
- ISO/IEC 17025 traceability via calibration certificates
- ALCOA+ principles
5. Responsibilities
- Technician/Metrology: Perform calibration and document results.
- QC/QA Reviewer: Review calibration documentation and ensure compliance.
- Owner Department: Ensure equipment is available and maintained.
- External Service Provider: Conduct external calibration as per schedule.
6. Safety & Preconditions
- Ensure environmental controls are in place (temperature, humidity).
- Allow equipment to warm-up and stabilize as per manufacturer’s recommendations.
- Maintain cleanliness of the equipment and surrounding area.
- Verify that the equipment is level and stable.
- Check power supply and connections prior to calibration.
7. Standards & Tools
- Reference Standards: [TBD]
- Traceability Statement: ISO17025
- Certificate Validity Check: Ensure all calibration certificates are current and valid.
8. Calibration Strategy
Calibration will follow the As Found and As Left logic:
- Minimum number of points: At least 5 points including endpoints (20°C and 45°C) and midpoints.
- Replicate readings: 3 repeats at each point as a best practice.
9. Step-by-Step Procedure
- Preparation: Gather all necessary tools and reference standards.
- Stabilization: Allow the temperature sensor to stabilize within the operational range.
- As Found checks and data capture: Record initial readings at defined points.
- Adjustment rules: If OOT is detected, follow adjustment protocols; if not allowed, escalate to QA.
- As Left verification: Confirm that the sensor readings are within acceptance criteria after adjustments.
- Labeling + documentation completion: Complete all required documentation and label the equipment accordingly.
Record raw observations contemporaneously to ensure data integrity (ALCOA+).
10. Acceptance Criteria & Calculations
Acceptance criteria: ±0.5°C
Error calculation formula: Error = Observed Value – True Value
Pass/Fail Logic: If the error falls within the acceptance criteria, the calibration passes; otherwise, it fails.
Rounding rule: [TBD]
11. Handling OOT / Failures
- Immediate containment: Tag equipment and stop use until resolved.
- Notification & deviation initiation: Notify QA and initiate a deviation report.
- Impact assessment trigger: Define impact window and review affected results/batches.
- Recalibration after repair/adjustment: Perform recalibration post-adjustment.
- QA disposition placeholders: [TBD]
12. Records & Retention
- Logbook entries: Document all calibration activities.
- Calibration report/certificate filing: Maintain records of calibration certificates.
- e-Records if applicable: [TBD]
- Label content requirements: Include calibrated date, due date, technician name, and status.
13. Annexures
Annexure A: Calibration Data Sheet
| Test Points | Reference | Observed | Error | Acceptance | Pass/Fail | As Found | As Left |
|---|---|---|---|---|---|---|---|
| [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] |
Annexure B: Quick Checklist
- Equipment clean and operational
- Environmental conditions verified
- Calibration standards available
- Documentation ready
Annexure C: Example Label Text
Calibration Date: [TBD]
Due Date: [TBD]
Calibrated By: [TBD]
Status: [TBD]
Calibration Protocol
Calibration Protocol
| Protocol ID | Asset ID | Equipment Name | Equipment Type | Dept | Area | Process Use | Parameter | Units | Range | Acceptance Criteria | Frequency | Calibration Type | Traceability |
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| [TBD] | CAL-0306 | CO2 Incubator Temp Sensor | Temperature Sensor | Biologics | Cell Culture | Temperature | Temperature | °C | 20 – 45 | ±0.5°C | 180 days | External | ISO17025 |
Environmental Conditions
Temperature: [TBD] °C
Relative Humidity: [TBD] %
Vibration: [TBD]
Drafts: [TBD]
Test Point Selection
- Minimum: 20°C
- 25%: 26.25°C
- 50%: 32.5°C
- 75%: 38.75°C
- Maximum: 45°C
Results Table
| Test Point | Ref Value | Observed (As Found) | Error | Limit | Pass/Fail | Observed (As Left) | Error | Pass/Fail |
|---|---|---|---|---|---|---|---|---|
| 20°C | [TBD] | [TBD] | [TBD] | ±0.5°C | [TBD] | [TBD] | [TBD] | [TBD] |
| 26.25°C | [TBD] | [TBD] | [TBD] | ±0.5°C | [TBD] | [TBD] | [TBD] | [TBD] |
| 32.5°C | [TBD] | [TBD] | [TBD] | ±0.5°C | [TBD] | [TBD] | [TBD] | [TBD] |
| 38.75°C | [TBD] | [TBD] | [TBD] | ±0.5°C | [TBD] | [TBD] | [TBD] | [TBD] |
| 45°C | [TBD] | [TBD] | [TBD] | ±0.5°C | [TBD] | [TBD] | [TBD] | [TBD] |
Stepwise Instructions
- Preparation: Ensure the CO2 incubator is operational and at the desired temperature range.
- Stabilization: Allow the incubator to stabilize for a minimum of [TBD] hours before calibration.
- Performing Readings: Take at least three replicate readings at each test point.
- Calculating Error: Calculate the error for each observed value against the reference value.
- Evaluation: Compare the calculated errors to the acceptance criteria of ±0.5°C.
- Documentation: Record all observations and calculations in the results table.
Deviation Instruction for OOT
If any reading is outside the acceptance criteria:
- Stop use of the equipment immediately.
- Notify the supervisor and quality assurance.
- Open a deviation report.
- Quarantine impacted results window: [TBD]
- Await QA decision on further actions.
Approvals
| Prepared | Reviewed | Approved |
|---|---|---|
| [Signature/Date] | [Signature/Date] | [Signature/Date] |
Calibration Checklist
| Asset ID | Equipment | Dept | Area | Parameter | Range | Acceptance | Frequency | Traceability |
|---|---|---|---|---|---|---|---|---|
| CAL-0306 | CO2 Incubator Temp Sensor | Biologics | Cell Culture | Temperature | 20 – 45 °C | ±0.5°C | 180 days | ISO17025 |
Pre-calibration
- [ ] Verify environmental conditions are within specified limits
- [ ] Allow equipment to warm up as per SOP
- [ ] Clean the equipment as per cleaning procedure
- [ ] Level the equipment if necessary
- [ ] Check validity of reference standards
- [ ] Ensure calibration logbook is ready for documentation
As Found Execution
- [ ] Set temperature setpoints as per calibration protocol
- [ ] Perform repeat measurements as required
- [ ] Record raw data accurately
- [ ] Perform necessary calculations for results
OOT Immediate Actions
- [ ] Tag out the equipment if out of tolerance (OOT)
- [ ] Stop use of the equipment immediately
- [ ] Notify the supervisor and QA
- [ ] Document deviation in the logbook
- [ ] Trigger impact review process
As Left Verification
- [ ] Conduct post-adjustment checks to confirm accuracy
- [ ] Ensure all measurements are within acceptance limits
Documentation & Closure
- [ ] Update equipment label with new calibration date
- [ ] Attach calibration certificate to logbook
- [ ] Obtain reviewer sign-off on calibration results
- [ ] Update next due date for calibration
Data Integrity (ALCOA+): Ensure all data recorded is Attributable, Legible, Contemporaneous, Original, and Accurate, with complete traceability.
Deviation Impact Assessment
| Event Summary |
|---|
|
Asset: CO2 Incubator Temp Sensor Date: [TBD] Discovered By: [TBD] Document References: [TBD] As Found Results Summary: [TBD] |
Immediate Containment Checklist
- Stop use of the affected equipment
- Label/tag the equipment as out of service
- Segregate impacted work
Define Impact Window
| Last Known In-Tolerance Date | Discovery Date |
|---|---|
| [TBD] | [TBD] |
Product/Batch/Test Impact Review
| Batch/Study/Report ID | Results Potentially Impacted | Re-test/Re-weigh Decision |
|---|---|---|
| [TBD] | [TBD] | [TBD] |
Data Integrity Review Prompts (ALCOA+)
- Attributable: [TBD]
- Legible: [TBD]
- Contemporaneous: [TBD]
- Original: [TBD]
- Accurate: [TBD]
- Complete: [TBD]
- Consistent: [TBD]
- Enduring: [TBD]
- Available: [TBD]
Risk Assessment
| Severity | Probability | Detectability | Rationale | Overall Risk |
|---|---|---|---|---|
| [TBD] | [TBD] | [TBD] | [TBD] | [TBD] |
Investigation Summary
Root Cause: [TBD]
CAPA Plan
| Action | Owner | Due Date | Effectiveness Check |
|---|---|---|---|
| [TBD] | [TBD] | [TBD] | [TBD] |
QA Disposition
Release/Hold/Retest/Recall Triggers: [TBD]
Calibration Certificate
Certificate No: [TBD]
Date: [TBD]
Next Due Date: [TBD]
Asset Details
| Asset ID | Equipment Name | Equipment Type | Dept | Area | Make/Model | Serial No | Location | Parameter/Range/Units | Acceptance Criteria |
|---|---|---|---|---|---|---|---|---|---|
| CAL-0306 | CO2 Incubator Temp Sensor | Temperature Sensor | Biologics | Cell Culture | [TBD] | [TBD] | [TBD] | Temperature: 20 – 45 °C | ±0.5°C |
Conditions
Environment: [TBD]
Reference Standards
| Standard Name/ID | Certificate No | Due Date | Traceability Statement |
|---|---|---|---|
| [TBD] | [TBD] | [TBD] | Traceability to ISO17025 |
Results
As Found Results
| Test Point | Measured Value (°C) | Acceptance Criteria | Pass/Fail |
|---|---|---|---|
| 1 | [TBD] | ±0.5°C | [TBD] |
| 2 | [TBD] | ±0.5°C | [TBD] |
| 3 | [TBD] | ±0.5°C | [TBD] |
| 4 | [TBD] | ±0.5°C | [TBD] |
| 5 | [TBD] | ±0.5°C | [TBD] |
As Left Results
| Test Point | Measured Value (°C) | Acceptance Criteria | Pass/Fail |
|---|---|---|---|
| 1 | [TBD] | ±0.5°C | [TBD] |
| 2 | [TBD] | ±0.5°C | [TBD] |
| 3 | [TBD] | ±0.5°C | [TBD] |
| 4 | [TBD] | ±0.5°C | [TBD] |
| 5 | [TBD] | ±0.5°C | [TBD] |
Measurement Uncertainty Statement: [TBD]
Statement of Conformity: [Pass/Fail overall]
Remarks: [TBD]
Signatures
Performed by: ______________________ Date: ___________
Checked by: ______________________ Date: ___________
Approved by: ______________________ Date: ___________