Disintegration Tester (CAL-0131) – Calibration Documents

Calibration Documents: Disintegration Tester (CAL-0131)

Asset ID CAL-0131
Equipment Disintegration Tester
Equipment Type Disintegration Tester
Department QC
Area/Location Instrument Room
Process Use Disintegration
Parameter Temperature
Units °C
Range Min 20
Range Max 50
Acceptance Criteria ±0.5°C
Calibration Frequency (days) 365
Calibration Type External
Traceability ISO17025

Calibration SOP

1. Purpose

The calibration of the Disintegration Tester (Asset ID: CAL-0131) is performed to ensure accurate measurement of temperature during the disintegration process. This is critical for maintaining patient safety, ensuring batch quality, and upholding data integrity. Accurate temperature readings help prevent deviations that could affect product efficacy and safety.

2. Scope

Asset ID: CAL-0131
Equipment Name/Type: Disintegration Tester
Department/Area: QC, Instrument Room
Process Use: Disintegration
Parameter: Temperature
Range: 20°C to 50°C
Units: °C
Calibration Frequency: 365 days

3. Definitions

  • Calibration: The process of comparing a measurement instrument against a known standard to determine its accuracy.
  • Verification: The process of confirming that a measurement instrument meets specified requirements.
  • Adjustment: The process of modifying a measurement instrument to bring it into compliance with specifications.
  • As Found: The condition of the instrument prior to any adjustments.
  • As Left: The condition of the instrument after adjustments have been made.
  • OOT (Out of Tolerance): A condition where a measurement falls outside the specified acceptance criteria.
  • Drift: The gradual change in the output of a measurement instrument over time.
  • Traceability: The ability to verify the history, location, or application of an item by means of documented recorded identification.
  • Measurement Uncertainty: A parameter that characterizes the range of values within which the true value is asserted to lie.
  • Intermediate Check: A periodic check of the instrument’s performance between calibrations.

4. References

  • Company Metrology SOP [TBD]
  • ISO/IEC 17025 traceability via calibration certificates
  • ALCOA+ principles

5. Responsibilities

  • Technician/Metrology: Perform calibration and document results.
  • QC/QA Reviewer: Review calibration results and ensure compliance.
  • Owner Department: Ensure equipment is available and maintained.
  • External Service Provider: Conduct external calibration as per schedule.

6. Safety & Preconditions

  • Ensure environmental controls are in place (temperature, humidity).
  • Allow equipment to warm up and stabilize as per manufacturer’s instructions.
  • Maintain cleanliness of the instrument and surrounding area.
  • Check that the instrument is level and properly positioned.
  • Verify power supply is stable and within specifications.
See also  Balance Calibration Weight 1000 g (CAL-0339) - Calibration Documents

7. Standards & Tools

Reference Standards Required: [TBD]
Traceability Statement: ISO17025
Certificate Validity Check: Ensure calibration certificates are current and valid.

8. Calibration Strategy

Calibration will follow the As Found and As Left logic. The minimum number of test points will be five, including the endpoints (20°C and 50°C) and midpoints. Each point should have three replicate readings to ensure accuracy and reliability.

9. Step-by-Step Procedure

  1. Preparation: Gather all necessary tools and documentation.
  2. Stabilization: Allow the Disintegration Tester to stabilize at the operating temperature range.
  3. As Found checks and data capture: Measure temperature at specified points and record results contemporaneously.
  4. Adjustment rules: If any readings are OOT, follow the adjustment protocol; if adjustments are not permissible, escalate to QA.
  5. As Left verification: Re-measure to confirm that the instrument is within acceptance criteria after adjustments.
  6. Labeling + documentation completion: Complete all required documentation and label the instrument with calibration details.

10. Acceptance Criteria & Calculations

Acceptance Criteria: ±0.5°C
Error Calculation Formula: Error = Observed Value – True Value
Pass/Fail Logic: If the error falls within the acceptance criteria, the calibration passes; otherwise, it fails.
Rounding Rule: [TBD]

11. Handling OOT / Failures

  • Immediate containment: Tag the equipment and cease use.
  • Notification & deviation initiation: Notify QA and initiate deviation process.
  • Impact assessment trigger: Define the impact window and review all affected results/batches.
  • Recalibration after repair/adjustment: Ensure recalibration is performed post-adjustment.
  • QA disposition placeholders: [TBD]

12. Records & Retention

Maintain logbook entries, calibration report/certificate filing, and e-records if applicable [TBD].
Label Content Requirements: Calibrated date, due date, technician name, and calibration status.

13. Annexures

Annexure A: Calibration Data Sheet (test points, reference, observed, error, acceptance, pass/fail; include As Found + As Left columns)
Annexure B: Quick checklist
Annexure C: Example label text (with [TBD] placeholders)

⬆ Back to TOC


Calibration Protocol

Calibration Protocol

Protocol ID Asset ID Equipment Name Equipment Type Dept Area Process Use Parameter Units Range Acceptance Criteria Frequency Calibration Type Traceability
[TBD] CAL-0131 Disintegration Tester Disintegration Tester QC Instrument Room Disintegration Temperature °C 20 – 50 ±0.5°C 365 days External ISO17025
See also  Capsule Filler Dosing Disc Calibration (CAL-0212) - Calibration Documents

Environmental Conditions

Temperature: [TBD] °C

Relative Humidity: [TBD]

Vibration: [TBD]

Drafts: [TBD]

Test Point Selection

  • Minimum: 20°C
  • 25%: 27.5°C
  • 50%: 35°C
  • 75%: 42.5°C
  • Maximum: 50°C

Results Table

Test Point Ref Value Observed (As Found) Error Limit Pass/Fail Observed (As Left) Error Pass/Fail
20°C 20°C [TBD] [TBD] ±0.5°C [TBD] [TBD] [TBD] [TBD]
27.5°C 27.5°C [TBD] [TBD] ±0.5°C [TBD] [TBD] [TBD] [TBD]
35°C 35°C [TBD] [TBD] ±0.5°C [TBD] [TBD] [TBD] [TBD]
42.5°C 42.5°C [TBD] [TBD] ±0.5°C [TBD] [TBD] [TBD] [TBD]
50°C 50°C [TBD] [TBD] ±0.5°C [TBD] [TBD] [TBD] [TBD]

Stepwise Instructions

  1. Preparation: Ensure the Disintegration Tester is clean and calibrated prior to use.
  2. Stabilization: Allow the equipment to stabilize at the ambient temperature for at least 30 minutes.
  3. Performing Readings:
    • Set the Disintegration Tester to the first test point.
    • Take three replicate readings for each test point.
  4. Calculating Error: Calculate the error for each observed value against the reference value.
  5. Evaluation: Compare the calculated error against the acceptance criteria.
  6. Documentation: Record all observations in the results table and ensure data integrity (ALCOA+).

Deviation Instruction for OOT

If any observed value falls outside the acceptance criteria:

  • Stop use of the equipment immediately.
  • Notify the Quality Assurance (QA) department.
  • Open a deviation report.
  • Quarantine impacted results window: [TBD]
  • Await QA decision on further actions.

Approvals

Prepared By Reviewed By Approved By
[Signature/Date] [Signature/Date] [Signature/Date]

⬆ Back to TOC


Calibration Checklist

Asset ID Equipment Dept Area Parameter Range Acceptance Frequency Traceability
CAL-0131 Disintegration Tester QC Instrument Room Temperature 20 – 50 °C ±0.5°C 365 days ISO17025

Pre-calibration

  • [ ] Verify environmental conditions are within specified limits
  • [ ] Ensure equipment has warmed up adequately
  • [ ] Clean the equipment as per SOP
  • [ ] Level the equipment if necessary
  • [ ] Confirm reference standards are valid and calibrated
  • [ ] Ensure logbook is ready for documentation

As Found Execution

  • [ ] Set temperature setpoints
  • [ ] Perform repeat measurements as required
  • [ ] Record raw data accurately
  • [ ] Perform necessary calculations for results
See also  ICP-MS (CAL-0117) - Calibration Documents

OOT Immediate Actions

  • [ ] Tag out the equipment
  • [ ] Stop use of the equipment
  • [ ] Notify the appropriate personnel
  • [ ] Document deviation from acceptance criteria
  • [ ] Trigger impact review as necessary

As Left Verification

  • [ ] Conduct post-adjustment checks
  • [ ] Confirm all measurements are within specified limits

Documentation & Closure

  • [ ] Update equipment label with calibration details
  • [ ] Attach calibration certificate to documentation
  • [ ] Obtain reviewer sign-off
  • [ ] Update next due date for calibration

⬆ Back to TOC


Deviation Impact Assessment

Event Summary
Asset: Disintegration Tester
Date: [TBD]
Discovered By: [TBD]
Document References: [TBD]
As Found Results Summary: [TBD]

Immediate Containment Checklist

  • Stop use of the equipment
  • Label/tag the equipment as out of service
  • Segregate impacted work

Define Impact Window

Last Known In-Tolerance Date Discovery Date
[TBD] [TBD]

Product/Batch/Test Impact Review

Batch/Study/Report ID Results Potentially Impacted Re-Test/Re-Weigh Decision
[TBD] [TBD] [TBD]

Data Integrity Review Prompts (ALCOA+)

  • Attributable: [TBD]
  • Legible: [TBD]
  • Contemporaneous: [TBD]
  • Original: [TBD]
  • Accurate: [TBD]
  • Complete: [TBD]
  • Consistent: [TBD]
  • Enduring: [TBD]
  • Available: [TBD]

Risk Assessment

Severity Probability Detectability Rationale Overall Risk
[TBD] [TBD] [TBD] [TBD] [TBD]

Investigation Summary

Root Cause: [TBD]

CAPA Plan

Action Owner Due Date Effectiveness Check
[TBD] [TBD] [TBD] [TBD]

QA Disposition

Release/Hold/Retest/Recall Triggers: [TBD]

⬆ Back to TOC


Calibration Certificate

Certificate No: [TBD]

Date: [TBD]

Next Due Date: [TBD]

Asset Details

Asset ID Equipment Name Equipment Type Dept Area Make/Model Serial No Location Parameter/Range/Units Acceptance Criteria
CAL-0131 Disintegration Tester Disintegration Tester QC Instrument Room [TBD] [TBD] [TBD] Temperature: 20 – 50 °C ±0.5°C

Conditions

Environment: [TBD]

Reference Standards

Standard Name/ID Certificate No Due Date Traceability Statement
[TBD] [TBD] [TBD] ISO17025

Results

As Found Results

Test Point Measurement (°C) Pass/Fail
1 [TBD] [TBD]
2 [TBD] [TBD]
3 [TBD] [TBD]
4 [TBD] [TBD]
5 [TBD] [TBD]

As Left Results

Test Point Measurement (°C) Pass/Fail
1 [TBD] [TBD]
2 [TBD] [TBD]
3 [TBD] [TBD]
4 [TBD] [TBD]
5 [TBD] [TBD]

Measurement Uncertainty Statement: [TBD]

Statement of Conformity: [Pass/Fail overall]

Remarks: [TBD]

Signatures

Performed by: ____________________ Date: __________

Checked by: ____________________ Date: __________

Approved by: ____________________ Date: __________

⬆ Back to TOC