Calibration Documents: Freezer Temp Probe (-20°C) (CAL-0191)
| Asset ID | CAL-0191 |
|---|---|
| Equipment | Freezer Temp Probe (-20°C) |
| Equipment Type | Temperature Probe |
| Department | Warehouse/QA |
| Area/Location | Cold Storage |
| Process Use | GDP storage |
| Parameter | Temperature |
| Units | °C |
| Range Min | -30 |
| Range Max | 0 |
| Acceptance Criteria | ±0.5°C |
| Calibration Frequency (days) | 180 |
| Calibration Type | External |
| Traceability | ISO17025 |
Table of Contents
Calibration SOP
1. Purpose
The calibration of the Freezer Temp Probe (-20°C) is performed to ensure accurate temperature measurements within the cold storage area, which is critical for maintaining Good Distribution Practice (GDP) storage conditions. Accurate temperature readings are essential for patient safety, batch quality, and data integrity, as deviations may compromise product stability and efficacy.
2. Scope
- Asset ID: CAL-0191
- Equipment Name/Type: Freezer Temp Probe (-20°C) / Temperature Probe
- Department/Area: Warehouse/QA / Cold Storage
- Process Use: GDP storage
- Parameter: Temperature
- Range: -30°C to 0°C
- Units: °C
- Calibration Frequency: 180 days
3. Definitions
- Calibration: The process of configuring an instrument to provide a result for a sample within an acceptable range.
- Verification: The process of confirming that an instrument is functioning correctly.
- Adjustment: The act of modifying an instrument to bring it within specified limits.
- As Found: The condition of the equipment prior to any adjustments made during calibration.
- As Left: The condition of the equipment after adjustments have been made.
- OOT (Out of Tolerance): A condition where the measurement falls outside the specified acceptance criteria.
- Drift: The gradual change in the output of a measurement instrument over time.
- Traceability: The ability to verify the history, location, or application of an item by means of documented recorded identification.
- Measurement Uncertainty: A parameter associated with the result of a measurement that characterizes the dispersion of the values that could reasonably be attributed to the measurement.
- Intermediate Check: A check performed between calibrations to ensure continued accuracy of the instrument.
4. References
- Company Metrology SOP [TBD]
- ISO/IEC 17025 traceability via calibration certificates
- ALCOA+ principles
5. Responsibilities
- Technician/Metrology: Perform calibration and document results.
- QC/QA Reviewer: Review calibration documentation and results.
- Owner Department: Ensure equipment is available and accessible for calibration.
- External Service Provider: Conduct external calibration as per the defined schedule.
6. Safety & Preconditions
- Ensure environmental controls are in place (temperature, humidity).
- Allow equipment to warm-up/stabilize as per manufacturer guidelines.
- Maintain cleanliness of the equipment and surrounding area.
- Check that the equipment is level and stable.
- Verify power supply is functioning correctly.
7. Standards & Tools
- Reference standards required: [TBD]
- Traceability statement: ISO17025
- Certificate validity check: [TBD]
- Standard ID placeholders: [TBD]
8. Calibration Strategy
Calibration will follow the As Found then As Left logic. The minimum number of points for calibration will include at least 5 points across the specified range of -30°C to 0°C, including endpoints and midpoints. It is recommended to take 3 replicate readings at each point to ensure accuracy and reliability.
9. Step-by-Step Procedure
- Preparation: Gather all necessary tools and reference standards.
- Stabilization: Allow the temperature probe to stabilize in the calibration environment.
- As Found checks and data capture: Record initial temperature readings at defined points.
- Adjustment rules: If OOT is found, escalate for adjustment as per SOP; otherwise, proceed to As Left verification.
- As Left verification: Record final readings after any adjustments.
- Labeling + documentation completion: Ensure all calibration records are filled out accurately and contemporaneously (ALCOA+).
10. Acceptance Criteria & Calculations
Acceptance criteria for the Freezer Temp Probe are ±0.5°C. The error calculation formula is:
Error = Observed Value – Reference Value
Pass/Fail logic: If the absolute value of the error is within the acceptance criteria, the calibration passes; otherwise, it fails.
Rounding rule: [TBD]
11. Handling OOT / Failures
- Immediate containment: Tag equipment and stop use if OOT is detected.
- Notification & deviation initiation: Notify relevant personnel and initiate deviation procedures.
- Impact assessment trigger: Define impact window and review of affected results/batches.
- Recalibration after repair/adjustment: Ensure recalibration is performed post-adjustment.
- QA disposition placeholders: [TBD]
12. Records & Retention
- Logbook entries: Document all calibration activities.
- Calibration report/certificate filing: Maintain records in accordance with company policy.
- e-records if applicable: [TBD]
- Label content requirements: Include calibrated date, due date, technician name, and status.
13. Annexures
Annexure A: Calibration Data Sheet
| Test Points | Reference | Observed | Error | Acceptance | Pass/Fail | As Found | As Left |
|---|---|---|---|---|---|---|---|
| [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] |
Annexure B: Quick Checklist
- Environmental controls verified
- Equipment warm-up completed
- Cleanliness checked
- Leveling confirmed
- Power supply verified
Annexure C: Example Label Text
Calibration Date: [TBD] | Due Date: [TBD] | Calibrated By: [TBD] | Status: [TBD]
Calibration Protocol
Calibration Protocol
| Protocol ID | Asset ID | Equipment Name | Equipment Type | Dept | Area | Process Use | Parameter | Units | Range | Acceptance Criteria | Frequency | Calibration Type | Traceability |
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| [TBD] | CAL-0191 | Freezer Temp Probe (-20°C) | Temperature Probe | Warehouse/QA | Cold Storage | GDP storage | Temperature | °C | -30 to 0 | ±0.5°C | 180 days | External | ISO17025 |
Environmental Conditions
Temperature: [TBD] °C
Relative Humidity: [TBD] %
Vibration: [TBD]
Drafts: [TBD]
Test Point Selection
- Minimum: -30°C
- 25% of Range: -22.5°C
- 50% of Range: -15°C
- 75% of Range: -7.5°C
- Maximum: 0°C
Results Table
| Test Point | Ref Value | Observed (As Found) | Error | Limit | Pass/Fail | Observed (As Left) | Error | Pass/Fail |
|---|---|---|---|---|---|---|---|---|
| -30°C | [TBD] | [TBD] | [TBD] | ±0.5°C | [TBD] | [TBD] | [TBD] | [TBD] |
| -22.5°C | [TBD] | [TBD] | [TBD] | ±0.5°C | [TBD] | [TBD] | [TBD] | [TBD] |
| -15°C | [TBD] | [TBD] | [TBD] | ±0.5°C | [TBD] | [TBD] | [TBD] | [TBD] |
| -7.5°C | [TBD] | [TBD] | [TBD] | ±0.5°C | [TBD] | [TBD] | [TBD] | [TBD] |
| 0°C | [TBD] | [TBD] | [TBD] | ±0.5°C | [TBD] | [TBD] | [TBD] | [TBD] |
Stepwise Instructions
- Preparation: Ensure the temperature probe is clean and properly calibrated prior to use.
- Stabilization: Allow the probe to stabilize in the cold storage environment for at least 30 minutes.
- Performing Readings: Take three replicate readings at each test point to ensure accuracy.
- Calculating Error: Calculate the error for each test point by subtracting the reference value from the observed value.
- Evaluation: Compare the calculated error against the acceptance criteria of ±0.5°C.
- Documentation: Record all observations and calculations in the results table.
Deviation Instruction for OOT
If any reading is outside the acceptance criteria:
- Stop use of the equipment immediately.
- Notify the QA department.
- Open a deviation report.
- Quarantine impacted results window: [TBD]
- Await QA decision on further actions.
Approvals
| Prepared | Reviewed | Approved |
|---|---|---|
| [Signature/Date] | [Signature/Date] | [Signature/Date] |
Calibration Checklist
| Asset ID | Equipment | Dept | Area | Parameter | Range | Acceptance | Frequency | Traceability |
|---|---|---|---|---|---|---|---|---|
| CAL-0191 | Freezer Temp Probe (-20°C) | Warehouse/QA | Cold Storage | Temperature | -30°C to 0°C | ±0.5°C | 180 days | ISO17025 |
Pre-calibration
- [ ] Verify environment conditions are stable
- [ ] Allow equipment to warm up as per SOP
- [ ] Clean the probe as necessary
- [ ] Ensure the probe is leveled
- [ ] Check validity of reference standards
- [ ] Confirm logbook is ready for entries
As Found Execution
- [ ] Set the temperature probe to specified setpoints
- [ ] Perform multiple repeats for accuracy
- [ ] Record all raw data accurately
- [ ] Perform necessary calculations for results
OOT Immediate Actions
- [ ] Tag out the equipment
- [ ] Stop use of the equipment immediately
- [ ] Notify relevant personnel
- [ ] Document deviation from acceptance criteria
- [ ] Trigger impact review as per SOP
As Left Verification
- [ ] Conduct post-adjustment checks
- [ ] Confirm all readings are within acceptance limits
Documentation & Closure
- [ ] Update equipment label with calibration details
- [ ] Attach calibration certificate to log
- [ ] Obtain reviewer sign-off
- [ ] Update next due date in logbook
Deviation Impact Assessment
| Event Summary |
|---|
|
Asset: Freezer Temp Probe (-20°C) Date: [TBD] Discovered By: [TBD] Document References: [TBD] As Found Results Summary: [TBD] |
Immediate Containment Checklist
- Stop use of the affected equipment
- Label/tag the equipment as out of service
- Segregate impacted work areas
Define Impact Window
Last known in-tolerance date: [TBD] to discovery date: [TBD]
Product/Batch/Test Impact Review
| Batch/Study/Report ID | Results Potentially Impacted | Re-test/Re-weigh Decision |
|---|---|---|
| [TBD] | [TBD] | [TBD] |
Data Integrity Review Prompts (ALCOA+)
- Attributable: [TBD]
- Legible: [TBD]
- Contemporaneous: [TBD]
- Original: [TBD]
- Accurate: [TBD]
- Complete: [TBD]
- Consistent: [TBD]
- Enduring: [TBD]
- Available: [TBD]
Risk Assessment
| Severity | Probability | Detectability | Rationale | Overall Risk |
|---|---|---|---|---|
| [TBD] | [TBD] | [TBD] | [TBD] | [TBD] |
Investigation Summary
Root Cause: [TBD]
CAPA Plan
| Action | Owner | Due Date | Effectiveness Check |
|---|---|---|---|
| [TBD] | [TBD] | [TBD] | [TBD] |
QA Disposition
Release/Hold/Retest/Recall Triggers: [TBD]
Calibration Certificate
Certificate No: [TBD]
Date: [TBD]
Next Due Date: [TBD]
Asset Details
| Asset ID | Equipment Name | Equipment Type | Dept | Area | Make/Model | Serial No | Location | Parameter/Range/Units | Acceptance Criteria |
|---|---|---|---|---|---|---|---|---|---|
| CAL-0191 | Freezer Temp Probe (-20°C) | Temperature Probe | Warehouse/QA | Cold Storage | [TBD] | [TBD] | [TBD] | Temperature: -30°C to 0°C (°C) | ±0.5°C |
Conditions
Environment: [TBD]
Reference Standards
| Standard Name/ID | Certificate No | Due Date | Traceability Statement |
|---|---|---|---|
| [TBD] | [TBD] | [TBD] | ISO17025 |
Results
As Found Results
| Test Point | Measured Value (°C) | Pass/Fail |
|---|---|---|
| Point 1 | [TBD] | [TBD] |
| Point 2 | [TBD] | [TBD] |
| Point 3 | [TBD] | [TBD] |
| Point 4 | [TBD] | [TBD] |
| Point 5 | [TBD] | [TBD] |
As Left Results
| Test Point | Measured Value (°C) | Pass/Fail |
|---|---|---|
| Point 1 | [TBD] | [TBD] |
| Point 2 | [TBD] | [TBD] |
| Point 3 | [TBD] | [TBD] |
| Point 4 | [TBD] | [TBD] |
| Point 5 | [TBD] | [TBD] |
Measurement Uncertainty Statement
[TBD]
Statement of Conformity
[TBD]
Remarks
[TBD]
Signatures
Performed by: ____________________ Date: __________
Checked by: ______________________ Date: __________
Approved by: _____________________ Date: __________