Calibration Documents: Friability Tester (CAL-0134)
| Asset ID | CAL-0134 |
|---|---|
| Equipment | Friability Tester |
| Equipment Type | Friability Tester |
| Department | QC |
| Area/Location | Instrument Room |
| Process Use | Friability |
| Parameter | Speed |
| Units | RPM |
| Range Min | 0 |
| Range Max | 100 |
| Acceptance Criteria | ±1% |
| Calibration Frequency (days) | 365 |
| Calibration Type | External |
| Traceability | ISO17025 |
Table of Contents
Calibration SOP
1. Purpose
The calibration of the Friability Tester (Asset ID: CAL-0134) is performed to ensure accurate measurement of the friability parameter, which is critical for maintaining batch quality and patient safety. Accurate calibration mitigates risks associated with inaccurate readings that could lead to compromised product integrity and data integrity.
2. Scope
- Asset ID: CAL-0134
- Equipment Name/Type: Friability Tester
- Department/Area: QC, Instrument Room
- Process Use: Friability
- Parameter: Speed
- Range: 0 – 100 RPM
- Units: RPM
- Calibration Frequency: 365 days
3. Definitions
- Calibration: The process of comparing a measurement instrument against a known standard to determine its accuracy.
- Verification: The process of confirming that a measurement instrument meets specified requirements.
- Adjustment: The process of modifying a measurement instrument to bring it into compliance with specifications.
- As Found: The condition of the instrument prior to any adjustments being made.
- As Left: The condition of the instrument after adjustments have been made.
- OOT (Out of Tolerance): A condition where the measurement falls outside the specified acceptance criteria.
- Drift: The gradual change in a measurement instrument’s output over time.
- Traceability: The ability to verify the history, location, or application of an item by means of documented recorded identification.
- Measurement Uncertainty: A parameter that characterizes the range of values within which the true value of a measurement lies.
- Intermediate Check: A periodic check of the instrument’s performance between calibrations.
4. References
- Company Metrology SOP [TBD]
- ISO/IEC 17025 traceability via calibration certificates
- ALCOA+ principles
5. Responsibilities
- Technician/Metrology: Perform calibration and document results.
- QC/QA Reviewer: Review calibration results and ensure compliance.
- Owner Department: Ensure equipment is available and maintained.
- External Service Provider: Conduct external calibration as per requirements.
6. Safety & Preconditions
- Ensure environmental controls are in place (temperature, humidity).
- Allow for warm-up/stabilization of the equipment as per manufacturer guidelines.
- Maintain cleanliness of the instrument and surrounding area.
- Check leveling of the equipment.
- Verify power supply stability and functionality.
7. Standards & Tools
- Reference Standards: [TBD]
- Traceability Statement: ISO17025
- Certificate Validity Check: Ensure all certificates are current and valid.
8. Calibration Strategy
The calibration will follow the As Found and As Left logic:
- Minimum number of test points: 5 points including endpoints (0 RPM, 100 RPM) and midpoints (50 RPM).
- Replicate readings at each point: 3 repeats as a best practice.
9. Step-by-Step Procedure
- Preparation: Gather all necessary tools and reference standards.
- Stabilization: Allow the Friability Tester to warm up for [TBD] minutes.
- As Found checks and data capture: Record initial readings at all test points.
- Adjustment rules: If OOT is identified, follow the adjustment protocol; if not allowed, escalate to QA.
- As Left verification: Record final readings after adjustments.
- Labeling + documentation completion: Ensure all records are filled out and raw observations are recorded contemporaneously (ALCOA+).
10. Acceptance Criteria & Calculations
Acceptance criteria for the Friability Tester is ±1% of the set point. The error calculation formula is:
Error (%) = (Observed Value – True Value) / True Value × 100
Pass/Fail Logic: If the error is within ±1%, the test point passes; otherwise, it fails.
Rounding rule: [TBD]
11. Handling OOT / Failures
- Immediate containment: Tag equipment and stop use.
- Notification & deviation initiation: Notify QA and document the deviation.
- Impact assessment trigger: Define impact window and review of affected results/batches.
- Recalibration after repair/adjustment: Schedule recalibration as necessary.
- QA disposition placeholders: [TBD]
12. Records & Retention
- Logbook entries: Document all calibration activities.
- Calibration report/certificate filing: Retain all calibration certificates.
- e-Records if applicable: [TBD]
- Label content requirements: Calibrated date, due date, by whom, status.
13. Annexures
Annexure A: Calibration Data Sheet
| Test Points | Reference | Observed | Error | Acceptance | Pass/Fail | As Found | As Left |
|---|---|---|---|---|---|---|---|
| [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] |
Annexure B: Quick Checklist
- Environmental controls verified
- Equipment warm-up completed
- Cleanliness checked
- Leveling confirmed
- Power supply verified
Annexure C: Example Label Text
Calibration Date: [TBD]
Due Date: [TBD]
Calibrated By: [TBD]
Status: [TBD]
Calibration Protocol
Calibration Protocol
| Protocol ID | Asset ID | Equipment Name | Equipment Type | Dept | Area | Process Use | Parameter | Units | Range | Acceptance Criteria | Frequency | Calibration Type | Traceability |
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| [TBD] | CAL-0134 | Friability Tester | Friability Tester | QC | Instrument Room | Friability | Speed | RPM | 0 – 100 | ±1% | 365 days | External | ISO17025 |
Environmental Conditions
Temperature: [TBD]
Relative Humidity: [TBD]
Vibration: [TBD]
Drafts: [TBD]
Test Point Selection
- Minimum: 0 RPM
- 25% of Range: 25 RPM
- 50% of Range: 50 RPM
- 75% of Range: 75 RPM
- Maximum: 100 RPM
Results Table
| Test Point | Ref Value | Observed (As Found) | Error | Limit | Pass/Fail | Observed (As Left) | Error | Pass/Fail |
|---|---|---|---|---|---|---|---|---|
| 0 RPM | [TBD] | [TBD] | [TBD] | ±1% | [TBD] | [TBD] | [TBD] | [TBD] |
| 25 RPM | [TBD] | [TBD] | [TBD] | ±1% | [TBD] | [TBD] | [TBD] | [TBD] |
| 50 RPM | [TBD] | [TBD] | [TBD] | ±1% | [TBD] | [TBD] | [TBD] | [TBD] |
| 75 RPM | [TBD] | [TBD] | [TBD] | ±1% | [TBD] | [TBD] | [TBD] | [TBD] |
| 100 RPM | [TBD] | [TBD] | [TBD] | ±1% | [TBD] | [TBD] | [TBD] | [TBD] |
Stepwise Instructions
- Preparation: Ensure the Friability Tester is clean and free from any previous calibration residues.
- Stabilization: Allow the equipment to stabilize for [TBD] minutes before calibration.
- Performing Readings: Take three replicate readings at each test point.
- Calculating Error: Calculate the error for each reading using the formula: Error = (Observed – Ref Value) / Ref Value * 100.
- Evaluation: Compare the calculated error against the acceptance criteria of ±1%.
- Documentation: Record all observations in the results table and ensure data integrity (ALCOA+).
Deviation Instruction for OOT
If any reading is outside the acceptance criteria:
- Stop use of the equipment immediately.
- Notify the supervisor and QA.
- Open a deviation report.
- Quarantine impacted results window: [TBD].
- Await QA decision on further actions.
Approvals
| Prepared | Reviewed | Approved |
|---|---|---|
| [Signature/Date] | [Signature/Date] | [Signature/Date] |
Calibration Checklist
| Asset ID | Equipment | Dept | Area | Parameter | Range | Acceptance | Frequency | Traceability |
|---|---|---|---|---|---|---|---|---|
| CAL-0134 | Friability Tester | QC | Instrument Room | Speed | 0 – 100 RPM | ±1% | 365 days | ISO17025 |
Pre-calibration
- [ ] Verify environmental conditions are within specified limits
- [ ] Allow equipment to warm up as per SOP
- [ ] Clean the equipment as per cleaning procedure
- [ ] Ensure equipment is leveled
- [ ] Check validity of reference standards
- [ ] Ensure logbook is ready for entries
As Found Execution
- [ ] Set the equipment to specified setpoints
- [ ] Perform repeat measurements as required
- [ ] Record raw data accurately
- [ ] Perform necessary calculations
OOT Immediate Actions
- [ ] Tag out the equipment
- [ ] Stop use of the equipment
- [ ] Notify the appropriate personnel
- [ ] Document deviation
- [ ] Trigger impact review
As Left Verification
- [ ] Perform post-adjustment checks
- [ ] Confirm all parameters are within specified limits
Documentation & Closure
- [ ] Update equipment label
- [ ] Attach calibration certificate
- [ ] Obtain reviewer sign-off
- [ ] Update next due date in logbook
Data Integrity (ALCOA+): Ensure all data recorded is Attributable, Legible, Contemporaneous, Original, and Accurate, with additional focus on Completeness and Consistency.
Deviation Impact Assessment
| Event Summary | Details |
|---|---|
| Asset | Friability Tester (CAL-0134) |
| Date | [TBD] |
| Discovered By | [TBD] |
| Document References | [TBD] |
| As Found Results Summary | [TBD] |
Immediate Containment Checklist
- Stop use of the asset
- Label/Tag the asset as out of tolerance
- Segregate impacted work
Define Impact Window
| Last Known In-Tolerance Date | Discovery Date |
|---|---|
| [TBD] | [TBD] |
Product/Batch/Test Impact Review
| Batch/Study/Report ID | Results Potentially Impacted | Re-Test/Re-Weigh Decision |
|---|---|---|
| [TBD] | [TBD] | [TBD] |
Data Integrity Review Prompts (ALCOA+)
- Attributable: [TBD]
- Legible: [TBD]
- Contemporaneous: [TBD]
- Original: [TBD]
- Accurate: [TBD]
- Complete: [TBD]
- Consistent: [TBD]
- Enduring: [TBD]
- Available: [TBD]
Risk Assessment
| Severity | Probability | Detectability | Rationale | Overall Risk |
|---|---|---|---|---|
| [TBD] | [TBD] | [TBD] | [TBD] | [TBD] |
Investigation Summary
Root Cause: [TBD]
CAPA Plan
| Action | Owner | Due Date | Effectiveness Check |
|---|---|---|---|
| [TBD] | [TBD] | [TBD] | [TBD] |
QA Disposition
Release/Hold/Retest/Recall Triggers: [TBD]
Calibration Certificate
Certificate No: [TBD]
Date: [TBD]
Next Due Date: [TBD]
Asset Details
| Asset ID | Equipment Name | Equipment Type | Dept | Area | Make/Model | Serial No | Location | Parameter/Range/Units | Acceptance Criteria |
|---|---|---|---|---|---|---|---|---|---|
| CAL-0134 | Friability Tester | Friability Tester | QC | Instrument Room | [TBD] | [TBD] | [TBD] | Speed: 0 – 100 RPM | ±1% |
Conditions
Environment: [TBD]
Reference Standards
| Standard Name/ID | Certificate No | Due Date | Traceability Statement |
|---|---|---|---|
| [TBD] | [TBD] | [TBD] | Traceability to ISO17025 |
Results
As Found Results
| Test Point | Measured Value (RPM) | Acceptance Criteria | Conformance |
|---|---|---|---|
| 1 | [TBD] | ±1% | [TBD] |
| 2 | [TBD] | ±1% | [TBD] |
| 3 | [TBD] | ±1% | [TBD] |
| 4 | [TBD] | ±1% | [TBD] |
| 5 | [TBD] | ±1% | [TBD] |
As Left Results
| Test Point | Measured Value (RPM) | Acceptance Criteria | Conformance |
|---|---|---|---|
| 1 | [TBD] | ±1% | [TBD] |
| 2 | [TBD] | ±1% | [TBD] |
| 3 | [TBD] | ±1% | [TBD] |
| 4 | [TBD] | ±1% | [TBD] |
| 5 | [TBD] | ±1% | [TBD] |
Measurement Uncertainty Statement: [TBD]
Statement of Conformity: [TBD]
Remarks: [TBD]
Signatures
Performed by: ______________________ Date: ___________
Checked by: ______________________ Date: ___________
Approved by: ______________________ Date: ___________