Calibration Documents: Inline Conductivity Sensor (WFI) (CAL-0095)
| Asset ID | CAL-0095 |
|---|---|
| Equipment | Inline Conductivity Sensor (WFI) |
| Equipment Type | Inline Conductivity Sensor |
| Department | Utilities |
| Area/Location | WFI Loop |
| Process Use | Conductivity |
| Parameter | Conductivity |
| Units | µS/cm |
| Range Min | 0 |
| Range Max | 2000 |
| Acceptance Criteria | ±1% |
| Calibration Frequency (days) | 180 |
| Calibration Type | External |
| Traceability | ISO17025 |
Table of Contents
Calibration SOP
1. Purpose
The calibration of the Inline Conductivity Sensor (WFI) is performed to ensure accurate measurement of conductivity, which is critical for maintaining the quality of Water for Injection (WFI). This calibration controls risks related to patient safety, batch quality, and data integrity by ensuring that the sensor operates within specified limits.
2. Scope
- Asset ID: CAL-0095
- Equipment Name/Type: Inline Conductivity Sensor (WFI)
- Department/Area: Utilities / WFI Loop
- Process Use: Conductivity
- Parameter: Conductivity
- Range: 0 – 2000 µS/cm
- Units: µS/cm
- Calibration Frequency: 180 days
3. Definitions
- Calibration: The process of comparing a measurement instrument to a known standard to ensure accuracy.
- Verification: The process of confirming that a measurement instrument meets specified requirements.
- Adjustment: The process of modifying a measurement instrument to bring it into compliance with specifications.
- As Found: The condition of the instrument prior to any adjustments.
- As Left: The condition of the instrument after adjustments have been made.
- OOT (Out of Tolerance): A condition where the measurement exceeds specified limits.
- Drift: The gradual change in a measurement instrument’s output over time.
- Traceability: The ability to verify the history, location, or application of an item by means of documented recorded identification.
- Measurement Uncertainty: A parameter that characterizes the range of values within which the true value is asserted to lie.
- Intermediate Check: A check performed between calibrations to ensure continued accuracy.
4. References
- Company Metrology SOP [TBD]
- ISO/IEC 17025 traceability via calibration certificates
- ALCOA+ principles
5. Responsibilities
- Technician/Metrology: Perform calibration and maintain records.
- QC/QA Reviewer: Review calibration results and documentation.
- Owner Department: Ensure equipment is available and in good condition for calibration.
- External Service Provider: Conduct external calibration as required.
6. Safety & Preconditions
- Ensure environmental controls are in place (temperature, humidity).
- Allow for warm-up/stabilization of the sensor as per manufacturer guidelines.
- Maintain cleanliness of the equipment and surrounding area.
- Check leveling of the equipment.
- Verify power supply stability and functionality.
7. Standards & Tools
- Reference Standards: [TBD]
- Traceability Statement: ISO17025
- Certificate Validity Check: Ensure all calibration certificates are current and valid.
8. Calibration Strategy
The calibration will follow an As Found and As Left logic:
- Minimum number of points: At least 5 points including endpoints (0, 2000 µS/cm) and midpoints.
- Replicate readings: Perform 3 repeats at each point as a best practice.
9. Step-by-Step Procedure
- Preparation: Gather all necessary tools and documentation.
- Stabilization: Allow the Inline Conductivity Sensor to stabilize as per manufacturer recommendations.
- As Found checks and data capture: Record initial readings and conditions.
- Adjustment rules: If OOT is found, follow escalation procedures; adjustments are only allowed if specified.
- As Left verification: Confirm that the sensor is within tolerance after adjustments.
- Labeling + documentation completion: Complete all required documentation and label the equipment accordingly.
Record raw observations contemporaneously to ensure data integrity (ALCOA+).
10. Acceptance Criteria & Calculations
Acceptance criteria: ±1%
Error calculation formula: Error (%) = (Measured Value – True Value) / True Value × 100
Pass/Fail Logic: If the error falls within ±1%, the calibration is considered a pass.
Rounding rule: [TBD]
11. Handling OOT / Failures
- Immediate containment: Tag equipment and stop use.
- Notification & deviation initiation: Notify relevant personnel and initiate deviation process.
- Impact assessment trigger: Define impact window and review of affected results/batches.
- Recalibration after repair/adjustment: Ensure recalibration is performed post-repair.
- QA disposition: [TBD]
12. Records & Retention
- Logbook entries: Document all calibration activities.
- Calibration report/certificate filing: Maintain records in accordance with company policy.
- e-Records if applicable: [TBD]
- Label content requirements: Include calibrated date, due date, technician name, and status.
13. Annexures
Annexure A: Calibration Data Sheet
| Test Points | Reference | Observed | Error | Acceptance | Pass/Fail | As Found | As Left |
|---|---|---|---|---|---|---|---|
| [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] |
Annexure B: Quick Checklist
- Environmental controls verified
- Equipment clean and leveled
- Power supply checked
- Calibration standards available
Annexure C: Example Label Text
Calibration Date: [TBD]
Due Date: [TBD]
Calibrated By: [TBD]
Status: [TBD]
Calibration Protocol
Calibration Protocol
| Protocol ID | Asset ID | Equipment Name | Equipment Type | Dept | Area | Process Use | Parameter | Units | Range | Acceptance Criteria | Frequency | Calibration Type | Traceability |
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| [TBD] | CAL-0095 | Inline Conductivity Sensor (WFI) | Inline Conductivity Sensor | Utilities | WFI Loop | Conductivity | Conductivity | µS/cm | 0 – 2000 | ±1% | 180 days | External | ISO17025 |
Environmental Conditions
Temperature: [TBD] °C
Relative Humidity: [TBD] %
Vibration: [TBD]
Drafts: [TBD]
Test Point Selection
- Minimum: 0 µS/cm
- 25% of Range: 500 µS/cm
- 50% of Range: 1000 µS/cm
- 75% of Range: 1500 µS/cm
- Maximum: 2000 µS/cm
Results Table
| Test Point | Ref Value | Observed (As Found) | Error | Limit | Pass/Fail | Observed (As Left) | Error | Pass/Fail |
|---|---|---|---|---|---|---|---|---|
| 0 µS/cm | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] |
| 500 µS/cm | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] |
| 1000 µS/cm | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] |
| 1500 µS/cm | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] |
| 2000 µS/cm | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] |
Stepwise Instructions
- Preparation: Ensure the Inline Conductivity Sensor is clean and properly installed in the WFI Loop.
- Stabilization: Allow the sensor to stabilize for at least [TBD] minutes before taking measurements.
- Performing Readings: Take three replicate readings at each test point.
- Calculating Error: Calculate the error for each reading using the formula: Error = (Observed – Ref Value) / Ref Value * 100.
- Evaluation: Compare the calculated errors against the acceptance criteria of ±1%.
- Documentation: Record all observations in the results table and ensure data integrity (ALCOA+).
Deviation Instruction for OOT
If any reading falls outside the acceptance criteria:
- Stop use of the Inline Conductivity Sensor.
- Notify the supervisor immediately.
- Open a deviation report.
- Quarantine impacted results window: [TBD].
- Await QA decision on further actions.
Approvals
| Prepared | Reviewed | Approved |
|---|---|---|
| [Signature/Date] | [Signature/Date] | [Signature/Date] |
Calibration Checklist
| Asset ID | Equipment | Dept | Area | Parameter | Range | Acceptance | Frequency | Traceability |
|---|---|---|---|---|---|---|---|---|
| CAL-0095 | Inline Conductivity Sensor (WFI) | Utilities | WFI Loop | Conductivity | 0 – 2000 µS/cm | ±1% | 180 days | ISO17025 |
Pre-calibration
- [ ] Verify environmental conditions are within specified limits
- [ ] Ensure sensor has warmed up for the required duration
- [ ] Clean the sensor as per SOP
- [ ] Level the sensor installation if applicable
- [ ] Confirm reference standards are valid and calibrated
- [ ] Ensure calibration logbook is ready for entries
As Found Execution
- [ ] Set the sensor to specified setpoints
- [ ] Perform repeat measurements as required
- [ ] Record all raw data accurately
- [ ] Perform necessary calculations for results
OOT Immediate Actions
- [ ] Tag out the equipment if out of tolerance (OOT)
- [ ] Stop use of the equipment immediately
- [ ] Notify relevant personnel of the OOT condition
- [ ] Document deviation and investigation steps
- [ ] Trigger impact review as per SOP
As Left Verification
- [ ] Conduct post-adjustment checks to confirm accuracy
- [ ] Ensure all measurements are within specified limits
Documentation & Closure
- [ ] Update calibration label with new date
- [ ] Attach calibration certificate to log
- [ ] Obtain reviewer sign-off on calibration documentation
- [ ] Update next due date in the logbook
Data Integrity (ALCOA+): Ensure all data recorded is Attributable, Legible, Contemporaneous, Original, Accurate, and Complete.
Deviation Impact Assessment
| Event Summary | Details |
|---|---|
| Asset | Inline Conductivity Sensor (WFI) |
| Date | [TBD] |
| Discovered By | [TBD] |
| Document References | [TBD] |
| As Found Results Summary | [TBD] |
Immediate Containment Checklist
- Stop use of the asset
- Label/tag the asset as out of tolerance
- Segregate impacted work
Define Impact Window
| Last Known In-Tolerance Date | Discovery Date |
|---|---|
| [TBD] | [TBD] |
Product/Batch/Test Impact Review
| Batch/Study/Report ID | Results Potentially Impacted | Re-test/Re-weigh Decision |
|---|---|---|
| [TBD] | [TBD] | [TBD] |
Data Integrity Review Prompts (ALCOA+)
- Attributable: [TBD]
- Legible: [TBD]
- Contemporaneous: [TBD]
- Original: [TBD]
- Accurate: [TBD]
- Complete: [TBD]
- Consistent: [TBD]
- Enduring: [TBD]
- Available: [TBD]
Risk Assessment
| Severity | Probability | Detectability | Rationale | Overall Risk |
|---|---|---|---|---|
| [TBD] | [TBD] | [TBD] | [TBD] | [TBD] |
Investigation Summary
Root Cause: [TBD]
CAPA Plan
| Action | Owner | Due Date | Effectiveness Check |
|---|---|---|---|
| [TBD] | [TBD] | [TBD] | [TBD] |
QA Disposition
Release/Hold/Retest/Recall Triggers: [TBD]
Calibration Certificate
Certificate No: [TBD]
Date: [TBD]
Next Due Date: [TBD]
Asset Details
| Asset ID | Equipment Name | Equipment Type | Dept | Area | Make/Model | Serial No | Location | Parameter/Range/Units | Acceptance Criteria |
|---|---|---|---|---|---|---|---|---|---|
| CAL-0095 | Inline Conductivity Sensor (WFI) | Inline Conductivity Sensor | Utilities | WFI Loop | [TBD] | [TBD] | [TBD] | Conductivity / 0 – 2000 / µS/cm | ±1% |
Conditions
Environment: [TBD]
Reference Standards
| Standard Name/ID | Certificate No | Due Date | Traceability Statement |
|---|---|---|---|
| [TBD] | [TBD] | [TBD] | ISO17025 |
Results
As Found Results
| Test Point | Measured Value (µS/cm) | Acceptance Criteria | Pass/Fail |
|---|---|---|---|
| 1 | [TBD] | ±1% | [TBD] |
| 2 | [TBD] | ±1% | [TBD] |
| 3 | [TBD] | ±1% | [TBD] |
| 4 | [TBD] | ±1% | [TBD] |
| 5 | [TBD] | ±1% | [TBD] |
As Left Results
| Test Point | Measured Value (µS/cm) | Acceptance Criteria | Pass/Fail |
|---|---|---|---|
| 1 | [TBD] | ±1% | [TBD] |
| 2 | [TBD] | ±1% | [TBD] |
| 3 | [TBD] | ±1% | [TBD] |
| 4 | [TBD] | ±1% | [TBD] |
| 5 | [TBD] | ±1% | [TBD] |
Measurement Uncertainty Statement: [TBD]
Statement of Conformity: [TBD]
Remarks: [TBD]
Signatures
Performed by: ____________________ Date: ____________
Checked by: ______________________ Date: ____________
Approved by: _____________________ Date: ____________