Calibration Documents: Load Cell (Loss-in-weight feeder) (CAL-0013)
| Asset ID | CAL-0013 |
|---|---|
| Equipment | Load Cell (Loss-in-weight feeder) |
| Equipment Type | Load Cell |
| Department | Production |
| Area/Location | Granulation |
| Process Use | Feed rate control |
| Parameter | Weight |
| Units | kg |
| Range Min | 0 |
| Range Max | 50 |
| Acceptance Criteria | ±0.1% |
| Calibration Frequency (days) | 365 |
| Calibration Type | External |
| Traceability | ISO17025 |
Table of Contents
Calibration SOP
1. Purpose
The calibration of the Load Cell (Loss-in-weight feeder) is performed to ensure accurate weight measurements for feed rate control in the granulation process. This calibration is critical to maintaining batch quality, ensuring patient safety, and preserving data integrity. Inaccurate weight measurements can lead to improper dosing, affecting product quality and compliance with Good Manufacturing Practices (GMP).
2. Scope
- Asset ID: CAL-0013
- Equipment Name/Type: Load Cell (Loss-in-weight feeder)
- Department/Area: Production / Granulation
- Process Use: Feed rate control
- Parameter: Weight
- Range: 0 kg to 50 kg
- Units: kg
- Calibration Frequency: 365 days
3. Definitions
- Calibration: The process of verifying and adjusting the accuracy of an instrument against a known standard.
- Verification: Confirming that an instrument meets specified requirements.
- Adjustment: The process of altering an instrument to bring it into compliance with specifications.
- As Found: The condition of the instrument prior to any adjustments or corrections.
- As Left: The condition of the instrument after adjustments have been made.
- OOT (Out of Tolerance): A condition where the instrument’s measurement falls outside the defined acceptance criteria.
- Drift: The gradual change in an instrument’s output over time.
- Traceability: The ability to verify the history, location, or application of an item by means of documented recorded identification.
- Measurement Uncertainty: A parameter that characterizes the range of values within which the true value is asserted to lie.
- Intermediate Check: A periodic verification of an instrument’s performance between scheduled calibrations.
4. References
- Company Metrology SOP [TBD]
- ISO/IEC 17025 traceability via calibration certificates
- ALCOA+ principles
5. Responsibilities
- Technician/Metrology: Perform calibration and maintain records.
- QC/QA Reviewer: Review calibration records and ensure compliance.
- Owner Department: Ensure equipment is available for calibration.
- External Service Provider: Conduct external calibrations as required.
6. Safety & Preconditions
- Ensure environmental controls are in place (temperature, humidity).
- Allow equipment to warm up and stabilize as per manufacturer specifications.
- Maintain cleanliness of the equipment and surrounding area.
- Check that the load cell is level before calibration.
- Verify power supply is stable and within specifications.
7. Standards & Tools
- Reference standards required: [TBD]
- Traceability statement: ISO17025
- Certificate validity check: Ensure all certificates are current and valid.
8. Calibration Strategy
Calibration will follow an As Found and As Left logic:
- Minimum number of points: At least 5 points including endpoints (0 kg and 50 kg) and midpoints (e.g., 25 kg).
- Replicate readings: Perform 3 repeats at each point as a best practice.
9. Step-by-Step Procedure
- Preparation: Gather all necessary tools and reference standards.
- Stabilization: Allow the load cell to stabilize in the calibration environment.
- As Found checks and data capture: Record initial measurements at defined points.
- Adjustment rules: If OOT is detected, follow the adjustment protocol; if not allowed, escalate to QA.
- As Left verification: Confirm that the instrument is within acceptance criteria after adjustments.
- Labeling + documentation completion: Complete all required documentation and label the equipment appropriately.
Record raw observations contemporaneously to ensure data integrity (ALCO+).
10. Acceptance Criteria & Calculations
Acceptance criteria: ±0.1%
Error calculation formula: Error (%) = (Observed Value – True Value) / True Value × 100
Pass/Fail logic: If the error falls within the acceptance criteria, the calibration passes; otherwise, it fails.
Rounding rule: [TBD]
11. Handling OOT / Failures
- Immediate containment: Tag equipment and stop use until resolved.
- Notification & deviation initiation: Notify QA and initiate deviation report.
- Impact assessment trigger: Define impact window and review affected results/batches.
- Recalibration after repair/adjustment: Ensure recalibration is performed post-adjustment.
- QA disposition: [TBD]
12. Records & Retention
- Logbook entries: Document all calibration activities.
- Calibration report/certificate filing: Maintain records in accordance with company policy.
- e-Records if applicable: [TBD]
- Label content requirements: Include calibrated date, due date, technician name, and status.
13. Annexures
Annexure A: Calibration Data Sheet
| Test Points | Reference | Observed | Error | Acceptance | Pass/Fail | As Found | As Left |
|---|---|---|---|---|---|---|---|
| [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] |
Annexure B: Quick Checklist
- Environmental controls verified
- Equipment warm-up completed
- Cleanliness checked
- Leveling confirmed
- Power supply verified
Annexure C: Example Label Text
Calibration Date: [TBD]
Due Date: [TBD]
Calibrated By: [TBD]
Status: [TBD]
Calibration Protocol
Calibration Protocol
| Protocol ID | Asset ID | Equipment Name | Equipment Type | Dept | Area | Process Use | Parameter | Units | Range | Acceptance Criteria | Frequency | Calibration Type | Traceability |
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| [TBD] | CAL-0013 | Load Cell (Loss-in-weight feeder) | Load Cell | Production | Granulation | Feed rate control | Weight | kg | 0 – 50 | ±0.1% | 365 days | External | ISO17025 |
Environmental Conditions
Temperature: [TBD] °C
Relative Humidity: [TBD] %
Vibration: [TBD]
Drafts: [TBD]
Test Point Selection
Test points will be selected based on the following values:
- Minimum: 0 kg
- 25% of Range: 12.5 kg
- 50% of Range: 25 kg
- 75% of Range: 37.5 kg
- Maximum: 50 kg
Results Table
| Test Point | Ref Value | Observed (As Found) | Error | Limit | Pass/Fail | Observed (As Left) | Error | Pass/Fail |
|---|---|---|---|---|---|---|---|---|
| 0 kg | [TBD] | [TBD] | [TBD] | ±0.1% | [TBD] | [TBD] | [TBD] | [TBD] |
| 12.5 kg | [TBD] | [TBD] | [TBD] | ±0.1% | [TBD] | [TBD] | [TBD] | [TBD] |
| 25 kg | [TBD] | [TBD] | [TBD] | ±0.1% | [TBD] | [TBD] | [TBD] | [TBD] |
| 37.5 kg | [TBD] | [TBD] | [TBD] | ±0.1% | [TBD] | [TBD] | [TBD] | [TBD] |
| 50 kg | [TBD] | [TBD] | [TBD] | ±0.1% | [TBD] | [TBD] | [TBD] | [TBD] |
Stepwise Instructions
- Preparation: Ensure the load cell is clean and free from any obstructions.
- Stabilization: Allow the equipment to stabilize for [TBD] minutes before taking measurements.
- Performing Readings: Take at least three replicate measurements at each test point.
- Calculating Error: Calculate the error for each test point using the formula: Error = (Observed – Ref Value) / Ref Value * 100.
- Evaluation: Compare the calculated error against the acceptance criteria of ±0.1%.
- Documentation: Record all observations and calculations in the results table.
Deviation Instruction for OOT
If any observed values fall outside the acceptance criteria:
- Stop use of the equipment immediately.
- Notify the supervisor and quality assurance (QA) team.
- Open a deviation report.
- Quarantine the impacted results window: [TBD]
- Await QA decision on further actions.
Approvals
| Prepared | Reviewed | Approved |
|---|---|---|
| [Signature/Date] | [Signature/Date] | [Signature/Date] |
Calibration Checklist
| Asset ID | Equipment | Dept | Area | Parameter | Range | Acceptance | Frequency | Traceability |
|---|---|---|---|---|---|---|---|---|
| CAL-0013 | Load Cell (Loss-in-weight feeder) | Production | Granulation | Weight | 0 – 50 kg | ±0.1% | 365 days | ISO17025 |
Pre-calibration
- [ ] Verify environmental conditions are within specified limits
- [ ] Ensure equipment has been warmed up for the required duration
- [ ] Clean the equipment and surrounding area
- [ ] Level the load cell properly
- [ ] Check validity of reference standards used for calibration
- [ ] Ensure calibration logbook is ready for documentation
As Found Execution
- [ ] Set the load cell to specified setpoints
- [ ] Perform multiple repeats for accuracy
- [ ] Record all raw data accurately
- [ ] Perform necessary calculations to determine deviations
OOT Immediate Actions
- [ ] Tag out the equipment if out of tolerance (OOT)
- [ ] Stop use of the load cell immediately
- [ ] Notify relevant personnel about the OOT condition
- [ ] Document the deviation in the logbook
- [ ] Trigger impact review process as necessary
As Left Verification
- [ ] Conduct post-adjustment checks to confirm accuracy
- [ ] Ensure all measurements are within acceptance limits
Documentation & Closure
- [ ] Update equipment label with calibration details
- [ ] Attach calibration certificate to the logbook
- [ ] Obtain reviewer sign-off on calibration documentation
- [ ] Update next due date for calibration in the logbook
Data Integrity (ALCOA+): Ensure all data recorded is Attributable, Legible, Contemporaneous, Original, Accurate, and Complete.
Deviation Impact Assessment
| Event Summary |
|---|
|
Asset: Load Cell (Loss-in-weight feeder) Date: [TBD] Discovered By: [TBD] Document References: [TBD] As Found Results Summary: [TBD] |
Immediate Containment Checklist
- Stop use of the affected equipment
- Label/tag the equipment as out of service
- Segregate impacted work areas
Define Impact Window
Last known in-tolerance date: [TBD] to discovery date: [TBD]
Product/Batch/Test Impact Review
| Batch/Study/Report ID | Results Potentially Impacted | Re-test/Re-weigh Decision |
|---|---|---|
| [TBD] | [TBD] | [TBD] |
Data Integrity Review Prompts (ALCOA+)
- Attributable: [TBD]
- Legible: [TBD]
- Contemporaneous: [TBD]
- Original: [TBD]
- Accurate: [TBD]
- Complete: [TBD]
- Consistent: [TBD]
- Enduring: [TBD]
- Available: [TBD]
Risk Assessment
| Severity | Probability | Detectability | Rationale | Overall Risk |
|---|---|---|---|---|
| [TBD] | [TBD] | [TBD] | [TBD] | [TBD] |
Investigation Summary
Root Cause: [TBD]
CAPA Plan
| Action | Owner | Due Date | Effectiveness Check |
|---|---|---|---|
| [TBD] | [TBD] | [TBD] | [TBD] |
QA Disposition
Release/Hold/Retest/Recall Triggers: [TBD]
Calibration Certificate
Certificate No: [TBD]
Date: [TBD]
Next Due Date: [TBD]
Asset Details
| Asset ID | Equipment Name | Equipment Type | Dept | Area | Make/Model | Serial No | Location | Parameter/Range/Units | Acceptance Criteria |
|---|---|---|---|---|---|---|---|---|---|
| CAL-0013 | Load Cell (Loss-in-weight feeder) | Load Cell | Production | Granulation | [TBD] | [TBD] | [TBD] | Weight (0 – 50 kg) | ±0.1% |
Conditions
Environment: [TBD]
Reference Standards
| Standard Name/ID | Certificate No | Due Date | Traceability Statement |
|---|---|---|---|
| [TBD] | [TBD] | [TBD] | ISO17025 |
Results
As Found Results
| Test Point | Measured Value (kg) | Acceptance Criteria | Pass/Fail |
|---|---|---|---|
| 1 | [TBD] | ±0.1% | [TBD] |
| 2 | [TBD] | ±0.1% | [TBD] |
| 3 | [TBD] | ±0.1% | [TBD] |
| 4 | [TBD] | ±0.1% | [TBD] |
| 5 | [TBD] | ±0.1% | [TBD] |
As Left Results
| Test Point | Measured Value (kg) | Acceptance Criteria | Pass/Fail |
|---|---|---|---|
| 1 | [TBD] | ±0.1% | [TBD] |
| 2 | [TBD] | ±0.1% | [TBD] |
| 3 | [TBD] | ±0.1% | [TBD] |
| 4 | [TBD] | ±0.1% | [TBD] |
| 5 | [TBD] | ±0.1% | [TBD] |
Measurement Uncertainty Statement: [TBD]
Statement of Conformity: [TBD]
Remarks: [TBD]
Signatures
Performed by: ____________________ Date: ____________
Checked by: ______________________ Date: ____________
Approved by: _____________________ Date: ____________