Magnetic Stirrer Hotplate Temp (CAL-0160) – Calibration Documents

Calibration Documents: Magnetic Stirrer Hotplate Temp (CAL-0160)

Asset ID CAL-0160
Equipment Magnetic Stirrer Hotplate Temp
Equipment Type Hotplate
Department QC
Area/Location Chemistry Lab
Process Use Heating
Parameter Temperature
Units °C
Range Min 20
Range Max 300
Acceptance Criteria ±2°C
Calibration Frequency (days) 365
Calibration Type External
Traceability ISO17025

Calibration SOP

1. Purpose

The calibration of the Magnetic Stirrer Hotplate Temp is performed to ensure accurate temperature control during heating processes in the Chemistry Lab. This calibration mitigates risks associated with inaccurate temperature readings that could affect batch quality and data integrity, ultimately safeguarding patient safety.

2. Scope

  • Asset ID: CAL-0160
  • Equipment Name/Type: Magnetic Stirrer Hotplate Temp / Hotplate
  • Department/Area: QC / Chemistry Lab
  • Process Use: Heating
  • Parameter: Temperature
  • Range: 20°C to 300°C
  • Units: °C
  • Calibration Frequency: 365 days

3. Definitions

  • Calibration: The process of verifying and adjusting the accuracy of an instrument.
  • Verification: The confirmation that an instrument meets specified requirements.
  • Adjustment: The act of correcting an instrument to ensure it operates within specified limits.
  • As Found: The condition of the instrument prior to any adjustments.
  • As Left: The condition of the instrument after adjustments have been made.
  • OOT (Out of Tolerance): A condition where the instrument readings fall outside the specified acceptance criteria.
  • Drift: The gradual change in an instrument’s accuracy over time.
  • Traceability: The ability to trace the measurement back to national or international standards.
  • Measurement Uncertainty: A parameter that characterizes the range of values within which the true value is expected to lie.
  • Intermediate Check: A verification process performed between calibrations to ensure ongoing accuracy.

4. References

  • Company Metrology SOP [TBD]
  • ISO/IEC 17025 traceability via calibration certificates
  • ALCOA+ principles

5. Responsibilities

  • Technician/Metrology: Perform calibration and record data.
  • QC/QA Reviewer: Review calibration records and ensure compliance.
  • Owner Department: Ensure equipment is available and in good condition for calibration.
  • External Service Provider: Conduct external calibration as per the calibration type.

6. Safety & Preconditions

  • Ensure the environment is controlled for temperature and humidity.
  • Allow the equipment to warm up and stabilize before calibration.
  • Maintain cleanliness of the equipment and surrounding area.
  • Check that the equipment is level and stable.
  • Verify power supply is stable and within specifications.
See also  IR Thermometer (CAL-0054) - Calibration Documents

7. Standards & Tools

  • Reference standards required: [TBD]
  • Traceability Statement: Calibration is traceable to ISO17025.
  • Certificate validity check: [TBD]
  • Standard ID placeholders: [TBD]

8. Calibration Strategy

Calibration will follow the As Found and As Left logic. A minimum of 5 calibration points will be taken, including endpoints (20°C and 300°C) and midpoints. Each point will have 3 replicate readings to ensure accuracy.

9. Step-by-Step Procedure

  1. Preparation: Gather all necessary tools and standards.
  2. Stabilization: Allow the equipment to reach operating conditions.
  3. As Found checks and data capture: Measure temperature at defined points and record data contemporaneously.
  4. Adjustment rules: If OOT is detected, adjust the equipment if permitted; otherwise, escalate the issue.
  5. As Left verification: Measure again to confirm adjustments are within acceptance criteria.
  6. Labeling + documentation completion: Ensure all records are completed and labeled appropriately.

10. Acceptance Criteria & Calculations

Acceptance criteria for the Magnetic Stirrer Hotplate Temp is ±2°C. The error calculation formula is:

Error = Observed Value – True Value

Pass/Fail Logic: If the absolute value of the error is within ±2°C, the calibration passes; otherwise, it fails.

Rounding rule: [TBD]

11. Handling OOT / Failures

  • Immediate containment: Tag equipment and cease use.
  • Notification & deviation initiation: Notify relevant personnel of the OOT condition.
  • Impact assessment trigger: Review affected results/batches within the defined impact window.
  • Recalibration after repair/adjustment: Ensure recalibration is performed post-adjustment.
  • QA disposition placeholders: [TBD]

12. Records & Retention

  • Logbook entries must be made for each calibration.
  • Calibration report/certificate filing must be maintained.
  • E-records if applicable: [TBD]
  • Label content requirements: calibrated date, due date, by whom, status.

13. Annexures

Annexure A: Calibration Data Sheet

Test Points Reference Observed Error Acceptance Pass/Fail As Found As Left
[TBD] [TBD] [TBD] [TBD] [TBD] [TBD] [TBD] [TBD]

Annexure B: Quick Checklist

  • Environmental controls verified
  • Equipment warmed up
  • Cleanliness checked
  • Leveling confirmed
  • Power supply verified

Annexure C: Example Label Text

Calibration Date: [TBD]

Due Date: [TBD]

Calibrated By: [TBD]

Status: [TBD]

⬆ Back to TOC


Calibration Protocol

Calibration Protocol

Protocol ID Asset ID Equipment Name Equipment Type Dept Area Process Use Parameter Units Range Acceptance Criteria Frequency Calibration Type Traceability
[TBD] CAL-0160 Magnetic Stirrer Hotplate Temp Hotplate QC Chemistry Lab Heating Temperature °C 20 – 300 ±2°C 365 days External ISO17025
See also  Autoclave Temp Sensor (CAL-0150) - Calibration Documents

Environmental Conditions

Temperature: [TBD] °C

Relative Humidity: [TBD] %

Vibration: [TBD]

Drafts: [TBD]

Test Point Selection

Test points will be selected based on the following range:

  • Min: 20°C
  • 25%: 85°C
  • 50%: 160°C
  • 75%: 235°C
  • Max: 300°C

Results Table

Test Point Ref Value Observed (As Found) Error Limit Pass/Fail Observed (As Left) Error Pass/Fail
20°C [TBD] [TBD] [TBD] ±2°C [TBD] [TBD] [TBD] [TBD]
85°C [TBD] [TBD] [TBD] ±2°C [TBD] [TBD] [TBD] [TBD]
160°C [TBD] [TBD] [TBD] ±2°C [TBD] [TBD] [TBD] [TBD]
235°C [TBD] [TBD] [TBD] ±2°C [TBD] [TBD] [TBD] [TBD]
300°C [TBD] [TBD] [TBD] ±2°C [TBD] [TBD] [TBD] [TBD]

Stepwise Instructions

  1. Preparation: Ensure the Magnetic Stirrer Hotplate is clean and free from any previous residues.
  2. Stabilization: Turn on the equipment and allow it to stabilize for at least 30 minutes before taking measurements.
  3. Performing Readings: Measure the temperature at each test point. Take three replicates for each point and record the values.
  4. Calculating Error: For each test point, calculate the error as follows: Error = Observed Value – Ref Value.
  5. Evaluation: Compare the calculated error against the acceptance criteria of ±2°C.
  6. Documentation: Complete the results table with As Found and As Left values, including any errors and pass/fail status.

Deviation Instruction for OOT

If any observed value falls outside the acceptance criteria:

  • Stop use of the equipment immediately.
  • Notify the supervisor and quality assurance (QA) personnel.
  • Open a deviation report.
  • Quarantine the impacted results window: [TBD].
  • Await QA decision on further actions.

Approvals

Prepared Reviewed Approved
[Signature/Date] [Signature/Date] [Signature/Date]

⬆ Back to TOC


Calibration Checklist

Asset ID Equipment Dept Area Parameter Range Acceptance Frequency Traceability
CAL-0160 Magnetic Stirrer Hotplate Temp QC Chemistry Lab Temperature 20 – 300 °C ±2°C 365 days ISO17025

Pre-calibration

  • [ ] Verify environmental conditions are within specified limits
  • [ ] Ensure equipment has warmed up adequately
  • [ ] Clean the equipment as per SOP
  • [ ] Level the equipment if necessary
  • [ ] Check validity of reference standards
  • [ ] Ensure logbook is ready for entries

As Found Execution

  • [ ] Set equipment to specified setpoints
  • [ ] Perform multiple repeats for accuracy
  • [ ] Record all raw data accurately
  • [ ] Perform necessary calculations for results

OOT Immediate Actions

  • [ ] Tag out the equipment
  • [ ] Stop use of the equipment immediately
  • [ ] Notify relevant personnel
  • [ ] Document deviation
  • [ ] Trigger impact review process
See also  Platform Scale 300 kg (CAL-0005) - Calibration Documents

As Left Verification

  • [ ] Conduct post-adjustment checks
  • [ ] Confirm all readings are within acceptance limits

Documentation & Closure

  • [ ] Update equipment label
  • [ ] Attach calibration certificate
  • [ ] Obtain reviewer sign-off
  • [ ] Update next due date in logbook

Data Integrity (ALCOA+): Ensure all data recorded is Attributable, Legible, Contemporaneous, Original, and Accurate.

Out of Tolerance (OOT) Handling: Follow established procedures for OOT situations to ensure compliance and safety.

⬆ Back to TOC


Deviation Impact Assessment

Event Summary
Asset: Magnetic Stirrer Hotplate Temp
Date: [TBD]
Discovered By: [TBD]
Document References: [TBD]
As Found Results Summary: [TBD]

Immediate Containment Checklist

  • Stop use of the asset
  • Label/tag the asset as out of tolerance
  • Segregate impacted work

Define Impact Window

Last known in-tolerance date: [TBD] to discovery date: [TBD]

Product/Batch/Test Impact Review

Batch/Study/Report ID Results Potentially Impacted Re-test/Re-weigh Decision
[TBD] [TBD] [TBD]

Data Integrity Review Prompts (ALCOA+)

  • Attributable: [TBD]
  • Legible: [TBD]
  • Contemporaneous: [TBD]
  • Original: [TBD]
  • Accurate: [TBD]
  • Complete: [TBD]
  • Consistent: [TBD]
  • Enduring: [TBD]
  • Available: [TBD]

Risk Assessment

Severity Probability Detectability Rationale Overall Risk
[TBD] [TBD] [TBD] [TBD] [TBD]

Investigation Summary

Investigation Summary Placeholders: [TBD]

Root Cause: [TBD]

CAPA Plan

Action Owner Due Date Effectiveness Check
[TBD] [TBD] [TBD] [TBD]

QA Disposition

Release/Hold/Retest/Recall Triggers: [TBD]

⬆ Back to TOC


Calibration Certificate

Certificate No: [TBD]

Date: [TBD]

Next Due Date: [TBD]

Asset Details

Asset ID Equipment Name Equipment Type Dept Area Make/Model Serial No Location Parameter/Range/Units Acceptance Criteria
CAL-0160 Magnetic Stirrer Hotplate Temp Hotplate QC Chemistry Lab [TBD] [TBD] [TBD] Temperature: 20 – 300 °C ±2°C

Conditions

Environment: [TBD]

Reference Standards

Standard Name/ID Certificate No Due Date Traceability Statement
[TBD] [TBD] [TBD] Traceability: ISO17025

Results

As Found Results

Test Point Measured Value (°C) Acceptance Criteria Pass/Fail
20 [TBD] ±2°C [TBD]
100 [TBD] ±2°C [TBD]
150 [TBD] ±2°C [TBD]
200 [TBD] ±2°C [TBD]
300 [TBD] ±2°C [TBD]

As Left Results

Test Point Measured Value (°C) Acceptance Criteria Pass/Fail
20 [TBD] ±2°C [TBD]
100 [TBD] ±2°C [TBD]
150 [TBD] ±2°C [TBD]
200 [TBD] ±2°C [TBD]
300 [TBD] ±2°C [TBD]

Measurement Uncertainty Statement: [TBD]

Statement of Conformity: [Pass/Fail overall]

Remarks: [TBD]

Signatures

Performed by: ____________________ Date: ___________

Checked by: ______________________ Date: ___________

Approved by: _____________________ Date: ___________

⬆ Back to TOC