Micropipette 1-5 mL (CAL-0024) – Calibration Documents

Calibration Documents: Micropipette 1-5 mL (CAL-0024)

Asset ID CAL-0024
Equipment Micropipette 1-5 mL
Equipment Type Micropipette
Department QC
Area/Location Chemistry Lab
Process Use Sample prep
Parameter Volume
Units mL
Range Min 1
Range Max 5
Acceptance Criteria ±1%
Calibration Frequency (days) 180
Calibration Type External
Traceability ISO17025

Calibration SOP

1. Purpose

The calibration of the Micropipette 1-5 mL is performed to ensure accurate volume measurements during sample preparation in the QC Chemistry Lab. This calibration controls risks associated with accuracy, patient safety, batch quality, and data integrity, thereby ensuring compliance with regulatory standards.

2. Scope

  • Asset ID: CAL-0024
  • Equipment Name/Type: Micropipette 1-5 mL / Micropipette
  • Department/Area: QC / Chemistry Lab
  • Process Use: Sample prep
  • Parameter: Volume
  • Range: 1 mL to 5 mL
  • Units: mL
  • Calibration Frequency: Every 180 days

3. Definitions

  • Calibration: The process of verifying and adjusting the accuracy of an instrument.
  • Verification: The confirmation that the instrument meets specified requirements.
  • Adjustment: The process of correcting an instrument to ensure it meets the required specifications.
  • As Found: The condition of the instrument before any adjustments are made.
  • As Left: The condition of the instrument after adjustments have been made.
  • OOT (Out of Tolerance): A condition where the measurement falls outside the specified acceptance criteria.
  • Drift: The gradual change in an instrument’s output over time.
  • Traceability: The ability to verify the history, location, or application of an item by means of documented recorded identification.
  • Measurement Uncertainty: A parameter that characterizes the range of values within which the true value is asserted to lie.
  • Intermediate Check: A verification process performed between calibrations to ensure continued accuracy.

4. References

  • Company Metrology SOP [TBD]
  • ISO/IEC 17025 traceability via calibration certificates
  • ALCOA+ principles

5. Responsibilities

  • Technician/Metrology: Perform calibration and document results.
  • QC/QA Reviewer: Review calibration documentation and results.
  • Owner Department: Ensure equipment is available and maintained.
  • External Service Provider: Conduct external calibration as per schedule.

6. Safety & Preconditions

  • Ensure environmental controls are in place (temperature, humidity).
  • Allow warm-up/stabilization period for the micropipette.
  • Maintain cleanliness of the equipment and workspace.
  • Check leveling of the micropipette.
  • Verify power supply checks are completed.
See also  Inline Pressure Transmitter (Reactor) (CAL-0201) - Calibration Documents

7. Standards & Tools

  • Reference Standards Required: [TBD]
  • Traceability Statement: ISO17025
  • Certificate Validity Check: Ensure calibration certificates are current.
  • Standard ID: [TBD]

8. Calibration Strategy

Calibration will follow the As Found and As Left logic. A minimum of 5 calibration points will be taken, including endpoints (1 mL and 5 mL) and midpoints (3 mL). Each point will be measured in triplicate as a best practice.

9. Step-by-Step Procedure

  1. Preparation: Gather all necessary equipment and standards.
  2. Stabilization: Allow the micropipette to stabilize as per manufacturer guidelines.
  3. As Found checks and data capture: Measure and record the volume at each calibration point.
  4. Adjustment rules: If OOT is found, adjust the micropipette according to manufacturer specifications; if not allowed, escalate to QA.
  5. As Left verification: Re-measure the adjusted points to confirm compliance.
  6. Labeling + documentation completion: Complete all required documentation and label the equipment with calibration details.

Record raw observations contemporaneously to ensure data integrity (ALCOA+).

10. Acceptance Criteria & Calculations

Acceptance criteria for the micropipette calibration is ±1%. The error calculation formula is as follows:

Error (%) = ((Observed Value – True Value) / True Value) x 100

Pass/Fail logic: If the error is within ±1%, the calibration passes; otherwise, it fails.

Rounding rule: [TBD]

11. Handling OOT / Failures

  • Immediate containment: Tag equipment and stop use.
  • Notification & deviation initiation: Notify QA and initiate deviation process.
  • Impact assessment trigger: Define impact window and review of affected results/batches.
  • Recalibration after repair/adjustment: Schedule recalibration post-adjustment.
  • QA disposition: [TBD]

12. Records & Retention

  • Logbook entries: Document all calibration activities.
  • Calibration report/certificate filing: Maintain records as per company policy.
  • e-Records if applicable: [TBD]
  • Label content requirements: Include calibrated date, due date, technician name, and status.

13. Annexures

Annexure A: Calibration Data Sheet

Test Points Reference Observed Error Acceptance Pass/Fail As Found As Left
[TBD] [TBD] [TBD] [TBD] [TBD] [TBD] [TBD] [TBD]

Annexure B: Quick Checklist

  • Environmental controls in place
  • Equipment warm-up completed
  • Workspace cleanliness verified
  • Power supply checked

Annexure C: Example Label Text

Calibration Date: [TBD] | Due Date: [TBD] | Calibrated By: [TBD] | Status: [TBD]

⬆ Back to TOC


Calibration Protocol

Calibration Protocol

Protocol ID Asset ID Equipment Name Equipment Type Dept Area Process Use Parameter Units Range Acceptance Criteria Frequency Calibration Type Traceability
[TBD] CAL-0024 Micropipette 1-5 mL Micropipette QC Chemistry Lab Sample prep Volume mL 1 – 5 ±1% 180 days External ISO17025
See also  Skid Conductivity Sensor (CAL-0313) - Calibration Documents

Environmental Conditions

Temperature: [TBD]

Relative Humidity: [TBD]

Vibration: [TBD]

Drafts: [TBD]

Test Point Selection

Test points will be selected based on the specified range of 1 to 5 mL:

  • Min: 1 mL
  • 25%: 2 mL
  • 50%: 3 mL
  • 75%: 4 mL
  • Max: 5 mL

Results Table

Test Point Ref Value Observed (As Found) Error Limit Pass/Fail Observed (As Left) Error Pass/Fail
1 mL [TBD] [TBD] [TBD] ±1% [TBD] [TBD] [TBD] [TBD]
2 mL [TBD] [TBD] [TBD] ±1% [TBD] [TBD] [TBD] [TBD]
3 mL [TBD] [TBD] [TBD] ±1% [TBD] [TBD] [TBD] [TBD]
4 mL [TBD] [TBD] [TBD] ±1% [TBD] [TBD] [TBD] [TBD]
5 mL [TBD] [TBD] [TBD] ±1% [TBD] [TBD] [TBD] [TBD]

Stepwise Instructions

  1. Preparation: Ensure the micropipette is clean and free from any contaminants.
  2. Stabilization: Allow the micropipette to stabilize at room temperature for at least 30 minutes.
  3. Performing Readings:
    • Calibrate at each test point (1 mL, 2 mL, 3 mL, 4 mL, 5 mL).
    • Take three replicate measurements at each test point.
  4. Calculating Error: Calculate the error for each test point using the formula: Error = (Observed – Ref Value) / Ref Value * 100.
  5. Evaluation: Compare the calculated error against the acceptance criteria of ±1%.
  6. Documentation: Record all observations in the results table and ensure data integrity (ALCOA+).

Deviation Instruction for OOT

If any results fall outside the acceptance criteria:

  • Stop use of the micropipette.
  • Notify the supervisor immediately.
  • Open a deviation report.
  • Quarantine impacted results window: [TBD].
  • Await QA decision on further actions.

Approvals

Prepared By Reviewed By Approved By
[Signature/Date] [Signature/Date] [Signature/Date]

⬆ Back to TOC


Calibration Checklist

Asset ID Equipment Dept Area Parameter Range Acceptance Frequency Traceability
CAL-0024 Micropipette 1-5 mL QC Chemistry Lab Volume 1 – 5 mL ±1% 180 days ISO17025

Pre-calibration

  • [ ] Verify environmental conditions are within specified limits
  • [ ] Allow micropipette to warm up as per manufacturer guidelines
  • [ ] Clean the micropipette as per SOP
  • [ ] Level the micropipette if applicable
  • [ ] Check validity of reference standards
  • [ ] Ensure logbook is ready for recording

As Found Execution

  • [ ] Set initial setpoints for calibration
  • [ ] Perform multiple repeats for accuracy
  • [ ] Record raw data accurately
  • [ ] Perform necessary calculations
See also  Graduated Cylinder 50 mL (CAL-0042) - Calibration Documents

OOT Immediate Actions

  • [ ] Tag out the micropipette
  • [ ] Stop use immediately
  • [ ] Notify QA and relevant personnel
  • [ ] Document deviation
  • [ ] Trigger impact review

As Left Verification

  • [ ] Conduct post-adjustment checks
  • [ ] Confirm results are within acceptance limits

Documentation & Closure

  • [ ] Update equipment label
  • [ ] Attach calibration certificate
  • [ ] Obtain reviewer sign-off
  • [ ] Update next due date in logbook

Data Integrity (ALCOA+): Ensure all data recorded is Attributable, Legible, Contemporaneous, Original, Accurate, and Complete.

⬆ Back to TOC


Deviation Impact Assessment

Event Summary
Asset: Micropipette 1-5 mL
Date: [TBD]
Discovered By: [TBD]
Document References: [TBD]
As Found Results Summary: [TBD]

Immediate Containment Checklist

  • Stop use of the asset
  • Label/tag the asset as out of tolerance
  • Segregate impacted work

Define Impact Window

Last Known In-Tolerance Date Discovery Date
[TBD] [TBD]

Product/Batch/Test Impact Review

Batch/Study/Report ID Results Potentially Impacted Re-test/Re-weigh Decision
[TBD] [TBD] [TBD]

Data Integrity Review Prompts (ALCOA+)

  • Attributable: [TBD]
  • Legible: [TBD]
  • Contemporaneous: [TBD]
  • Original: [TBD]
  • Accurate: [TBD]
  • Complete: [TBD]
  • Consistent: [TBD]
  • Enduring: [TBD]
  • Available: [TBD]

Risk Assessment

Severity Probability Detectability Rationale Overall Risk
[TBD] [TBD] [TBD] [TBD] [TBD]

Investigation Summary

Root Cause: [TBD]

CAPA Plan

Action Owner Due Date Effectiveness Check
[TBD] [TBD] [TBD] [TBD]

QA Disposition

Release/Hold/Retest/Recall Triggers: [TBD]

⬆ Back to TOC


Calibration Certificate

Certificate No: [TBD]

Date: [TBD]

Next Due Date: [TBD]

Asset Details

Asset ID Equipment Name Equipment Type Dept Area Make/Model Serial No Location Parameter/Range/Units Acceptance Criteria
CAL-0024 Micropipette 1-5 mL Micropipette QC Chemistry Lab [TBD] [TBD] [TBD] Volume (1-5 mL) ±1%

Conditions

Environment: [TBD]

Reference Standards

Standard Name/ID Certificate No Due Date Traceability Statement
[TBD] [TBD] [TBD] ISO17025

Results

As Found Results

Test Point Measured Value (mL) Acceptance Criteria Pass/Fail
1 mL [TBD] ±1% [TBD]
2 mL [TBD] ±1% [TBD]
3 mL [TBD] ±1% [TBD]
4 mL [TBD] ±1% [TBD]
5 mL [TBD] ±1% [TBD]

As Left Results

Test Point Measured Value (mL) Acceptance Criteria Pass/Fail
1 mL [TBD] ±1% [TBD]
2 mL [TBD] ±1% [TBD]
3 mL [TBD] ±1% [TBD]
4 mL [TBD] ±1% [TBD]
5 mL [TBD] ±1% [TBD]

Measurement Uncertainty Statement: [TBD]

Statement of Conformity: [Pass/Fail overall]

Remarks: [TBD]

Signatures

Performed by: ____________________ Date: ___________

Checked by: ____________________ Date: ___________

Approved by: ____________________ Date: ___________

⬆ Back to TOC