NIR Spectrometer (CAL-0113) – Calibration Documents

Calibration Documents: NIR Spectrometer (CAL-0113)

Asset ID CAL-0113
Equipment NIR Spectrometer
Equipment Type NIR
Department QC/Production
Area/Location PAT
Process Use Blend monitoring
Parameter Wavelength
Units nm
Range Min 800
Range Max 2500
Acceptance Criteria ±2 nm
Calibration Frequency (days) 365
Calibration Type External
Traceability ISO17025

Calibration SOP

1. Purpose

The calibration of the NIR Spectrometer (CAL-0113) is performed to ensure accurate wavelength measurements for blend monitoring in the QC/Production department. This calibration controls risks related to accuracy, patient safety, batch quality, and data integrity, thereby ensuring compliance with Good Manufacturing Practices (GMP).

2. Scope

  • Asset ID: CAL-0113
  • Equipment Name/Type: NIR Spectrometer / NIR
  • Department/Area: QC/Production / PAT
  • Process Use: Blend monitoring
  • Parameter: Wavelength
  • Range: 800 nm to 2500 nm
  • Units: nm
  • Calibration Frequency: 365 days

3. Definitions

  • Calibration: The process of verifying and adjusting the accuracy of an instrument.
  • Verification: The confirmation that an instrument meets specified requirements.
  • Adjustment: The process of modifying an instrument to bring it into compliance with specifications.
  • As Found: The condition of the instrument prior to any adjustments.
  • As Left: The condition of the instrument after adjustments have been made.
  • OOT (Out of Tolerance): A condition where the instrument’s measurements fall outside the specified acceptance criteria.
  • Drift: The gradual change in an instrument’s performance over time.
  • Traceability: The ability to verify the history, location, or application of an item by means of documented recorded identification.
  • Measurement Uncertainty: A parameter that characterizes the range of values within which the true value is asserted to lie.
  • Intermediate Check: A periodic check of an instrument’s performance between calibrations.

4. References

  • Company Metrology SOP [TBD]
  • ISO/IEC 17025 traceability via calibration certificates
  • ALCOA+ principles

5. Responsibilities

  • Technician/Metrology: Perform calibration and maintain records.
  • QC/QA Reviewer: Review calibration results and documentation.
  • Owner Department: Ensure equipment is available and in good condition for calibration.
  • External Service Provider: Conduct calibration as per external calibration type.

6. Safety & Preconditions

  • Ensure environmental controls are in place (temperature, humidity).
  • Allow equipment warm-up/stabilization as per manufacturer guidelines.
  • Maintain cleanliness of the equipment and surrounding area.
  • Check leveling of the instrument.
  • Verify power supply stability and functionality.

7. Standards & Tools

  • Reference Standards: [TBD]
  • Traceability Statement: ISO17025
  • Certificate Validity Check: Ensure all certificates are current and valid.
  • Standard ID: [TBD]
See also  Tablet Thickness Gauge (CAL-0133) - Calibration Documents

8. Calibration Strategy

Calibration will follow the As Found and As Left logic. The minimum number of test points will be five, including endpoints and midpoints of the specified range (800 nm to 2500 nm). It is recommended to take three replicate readings at each point to ensure accuracy.

9. Step-by-Step Procedure

  1. Preparation: Gather all necessary tools and documentation.
  2. Stabilization: Allow the NIR Spectrometer to warm up for the specified time.
  3. As Found checks and data capture: Record initial measurements and observations contemporaneously (ALCOA+).
  4. Adjustment rules: If any readings are OOT, follow the adjustment protocol; if adjustments are not allowed, escalate to QA.
  5. As Left verification: Confirm that all adjustments have been made and recheck measurements.
  6. Labeling + documentation completion: Ensure all calibration records are filled out accurately and labels are affixed to the equipment.

10. Acceptance Criteria & Calculations

The acceptance criteria for the NIR Spectrometer calibration is ±2 nm. The error calculation formula is:

Error = Observed Value – True Value

Pass/Fail logic: If the absolute value of the error is within the acceptance criteria, the calibration passes; otherwise, it fails.

Rounding rule: [TBD]

11. Handling OOT / Failures

  • Immediate containment: Tag the equipment and stop use.
  • Notification & deviation initiation: Notify relevant personnel of the OOT condition.
  • Impact assessment trigger: Define the impact window and review affected results/batches.
  • Recalibration after repair/adjustment: Schedule recalibration as necessary.
  • QA disposition placeholders: [TBD]

12. Records & Retention

  • Logbook entries: Document all calibration activities.
  • Calibration report/certificate filing: Maintain a file of all calibration certificates.
  • e-Records if applicable: [TBD]
  • Label content requirements: Include calibrated date, due date, technician name, and status.

13. Annexures

Annexure A: Calibration Data Sheet

Test Points Reference Observed Error Acceptance Pass/Fail As Found As Left
[TBD] [TBD] [TBD] [TBD] [TBD] [TBD] [TBD] [TBD]

Annexure B: Quick Checklist

  • Environmental controls verified
  • Equipment warm-up completed
  • Cleanliness checked
  • Leveling confirmed
  • Power supply verified

Annexure C: Example Label Text

Calibration Date: [TBD]

Due Date: [TBD]

Calibrated By: [TBD]

Status: [TBD]

⬆ Back to TOC


Calibration Protocol

Calibration Protocol

Protocol ID Asset ID Equipment Name Equipment Type Dept Area Process Use Parameter Units Range Acceptance Criteria Frequency Calibration Type Traceability
[TBD] CAL-0113 NIR Spectrometer NIR QC/Production PAT Blend monitoring Wavelength nm 800 – 2500 ±2 nm 365 days External ISO17025
See also  Volumetric Flask 500 mL (CAL-0038) - Calibration Documents

Environmental Conditions

Temperature: [TBD]

Relative Humidity: [TBD]

Vibration: [TBD]

Drafts: [TBD]

Test Point Selection

Test points will be selected as follows:

  • Minimum: 800 nm
  • 25%: 1100 nm
  • 50%: 1300 nm
  • 75%: 1750 nm
  • Maximum: 2500 nm

Results Table

Test Point Ref Value Observed (As Found) Error Limit Pass/Fail Observed (As Left) Error Pass/Fail
800 nm [TBD] [TBD] [TBD] ±2 nm [TBD] [TBD] [TBD] [TBD]
1100 nm [TBD] [TBD] [TBD] ±2 nm [TBD] [TBD] [TBD] [TBD]
1300 nm [TBD] [TBD] [TBD] ±2 nm [TBD] [TBD] [TBD] [TBD]
1750 nm [TBD] [TBD] [TBD] ±2 nm [TBD] [TBD] [TBD] [TBD]
2500 nm [TBD] [TBD] [TBD] ±2 nm [TBD] [TBD] [TBD] [TBD]

Stepwise Instructions

  1. Preparation: Ensure the NIR Spectrometer is clean and calibrated according to the manufacturer’s specifications.
  2. Stabilization: Allow the equipment to stabilize for [TBD] minutes before taking measurements.
  3. Performing Readings: Take three replicate readings at each test point to ensure accuracy.
  4. Calculating Error: Calculate the error for each observed value against the reference value.
  5. Evaluation: Compare the calculated error against the acceptance criteria of ±2 nm.
  6. Documentation: Record all observations in the results table and ensure data integrity (ALCOA+).

Deviation Instruction for OOT

If any observed value falls outside the acceptance criteria:

  • Stop use of the equipment immediately.
  • Notify the supervisor and quality assurance.
  • Open a deviation report.
  • Quarantine impacted results window: [TBD]
  • Await QA decision on further actions.

Approvals

Prepared By Reviewed By Approved By
[Signature/Date] [Signature/Date] [Signature/Date]

⬆ Back to TOC


Calibration Checklist

Asset ID Equipment Dept Area Parameter Range Acceptance Frequency Traceability
CAL-0113 NIR Spectrometer QC/Production PAT Wavelength 800 – 2500 nm ±2 nm 365 days ISO17025

Pre-calibration

  • [ ] Verify environmental conditions are within specified limits
  • [ ] Allow equipment to warm up as per manufacturer’s guidelines
  • [ ] Clean the equipment as per SOP
  • [ ] Ensure equipment is leveled properly
  • [ ] Confirm validity of reference standards
  • [ ] Ensure logbook is ready for documentation

As Found Execution

  • [ ] Set initial measurement setpoints
  • [ ] Perform repeat measurements
  • [ ] Record all raw data accurately
  • [ ] Perform necessary calculations for results

OOT Immediate Actions

  • [ ] Tag out the equipment
  • [ ] Stop use of the equipment immediately
  • [ ] Notify the appropriate personnel
  • [ ] Document deviation in the logbook
  • [ ] Trigger impact review as per SOP
See also  Manometer (U-tube/Digital) (CAL-0078) - Calibration Documents

As Left Verification

  • [ ] Conduct post-adjustment checks
  • [ ] Confirm all measurements are within acceptance limits

Documentation & Closure

  • [ ] Update equipment label with calibration details
  • [ ] Attach calibration certificate to documentation
  • [ ] Obtain reviewer sign-off
  • [ ] Update next due date in logbook

Data Integrity (ALCOA+): Ensure all data is attributable, legible, contemporaneous, original, and accurate, with appropriate documentation for all steps.

Out of Tolerance (OOT) Handling: Follow established procedures for addressing OOT results, including documentation and corrective actions.

⬆ Back to TOC


Deviation Impact Assessment

Event Summary Details
Asset NIR Spectrometer
Date [TBD]
Discovered By [TBD]
Document References [TBD]
As Found Results Summary [TBD]

Immediate Containment Checklist

  • Stop use of the equipment
  • Label/tag the equipment as out of service
  • Segregate impacted work

Define Impact Window

Last Known In-Tolerance Date Discovery Date
[TBD] [TBD]

Product/Batch/Test Impact Review

Batch/Study/Report ID Results Potentially Impacted Re-Test/Re-Weigh Decision
[TBD] [TBD] [TBD]

Data Integrity Review Prompts (ALCOA+)

  • Attributable: [TBD]
  • Legible: [TBD]
  • Contemporaneous: [TBD]
  • Original: [TBD]
  • Accurate: [TBD]
  • Complete: [TBD]
  • Consistent: [TBD]
  • Enduring: [TBD]
  • Available: [TBD]

Risk Assessment

Severity Probability Detectability Rationale Overall Risk
[TBD] [TBD] [TBD] [TBD] [TBD]

Investigation Summary

Root Cause: [TBD]

CAPA Plan

Action Owner Due Date Effectiveness Check
[TBD] [TBD] [TBD] [TBD]

QA Disposition

Release/Hold/Retest/Recall Triggers: [TBD]

⬆ Back to TOC


Calibration Certificate

Certificate No: [TBD]

Date: [TBD]

Next Due Date: [TBD]

Asset Details

Asset ID Equipment Name Equipment Type Dept Area Make/Model Serial No Location Parameter/Range/Units Acceptance Criteria
CAL-0113 NIR Spectrometer NIR QC/Production PAT [TBD] [TBD] [TBD] Wavelength: 800 – 2500 nm ±2 nm

Conditions

Environment: [TBD]

Reference Standards

Standard Name/ID Certificate No Due Date Traceability Statement
[TBD] [TBD] [TBD] ISO17025

Results

As Found Results

Test Point Measured Value (nm) Acceptance Criteria Pass/Fail
Point 1 [TBD] ±2 nm [TBD]
Point 2 [TBD] ±2 nm [TBD]
Point 3 [TBD] ±2 nm [TBD]
Point 4 [TBD] ±2 nm [TBD]
Point 5 [TBD] ±2 nm [TBD]

As Left Results

Test Point Measured Value (nm) Acceptance Criteria Pass/Fail
Point 1 [TBD] ±2 nm [TBD]
Point 2 [TBD] ±2 nm [TBD]
Point 3 [TBD] ±2 nm [TBD]
Point 4 [TBD] ±2 nm [TBD]
Point 5 [TBD] ±2 nm [TBD]

Measurement Uncertainty Statement: [TBD]

Statement of Conformity: [TBD]

Remarks: [TBD]

Signatures

Performed by: ______________________ Date: ___________

Checked by: ______________________ Date: ___________

Approved by: ______________________ Date: ___________

⬆ Back to TOC