Calibration Documents: Refrigerator Temp Probe (2-8°C) (CAL-0190)
| Asset ID | CAL-0190 |
|---|---|
| Equipment | Refrigerator Temp Probe (2-8°C) |
| Equipment Type | Temperature Probe |
| Department | Warehouse/QA |
| Area/Location | Cold Storage |
| Process Use | GDP storage |
| Parameter | Temperature |
| Units | °C |
| Range Min | 2 |
| Range Max | 8 |
| Acceptance Criteria | ±0.5°C |
| Calibration Frequency (days) | 180 |
| Calibration Type | External |
| Traceability | ISO17025 |
Table of Contents
Calibration SOP
1. Purpose
The calibration of the Refrigerator Temp Probe (2-8°C) is performed to ensure accurate temperature readings within the specified range, thereby controlling risks associated with product integrity, patient safety, and batch quality. Accurate temperature monitoring is critical for maintaining Good Distribution Practice (GDP) storage conditions, ensuring that products remain within acceptable limits to prevent degradation and ensure compliance with regulatory standards.
2. Scope
- Asset ID: CAL-0190
- Equipment Name/Type: Refrigerator Temp Probe (2-8°C) / Temperature Probe
- Department/Area: Warehouse/QA / Cold Storage
- Process Use: GDP storage
- Parameter: Temperature
- Range: 2 – 8 °C
- Units: °C
- Calibration Frequency: 180 days
3. Definitions
- Calibration: The process of comparing a measurement device to a known standard to determine its accuracy.
- Verification: The process of confirming that a measurement device operates within specified limits.
- Adjustment: The act of modifying a measurement device to bring it within specified limits.
- As Found: The condition of the equipment prior to any adjustments being made.
- As Left: The condition of the equipment after adjustments have been made.
- OOT (Out of Tolerance): A condition where the measurement exceeds the defined acceptance criteria.
- Drift: The gradual change in a measurement device’s output over time.
- Traceability: The ability to verify the history, location, or application of an item by means of documented recorded identification.
- Measurement Uncertainty: A parameter that characterizes the range of values within which the true value is estimated to lie.
- Intermediate Check: A verification process performed between calibrations to ensure continued accuracy.
4. References
- Company Metrology SOP [TBD]
- ISO/IEC 17025 traceability via calibration certificates
- ALCOA+ principles
5. Responsibilities
- Technician/Metrology: Responsible for performing the calibration and documenting results.
- QC/QA Reviewer: Responsible for reviewing calibration documentation and ensuring compliance.
- Owner Department: Responsible for ensuring equipment is available and maintained.
- External Service Provider: Responsible for conducting external calibrations as per the calibration type.
6. Safety & Preconditions
- Ensure environmental controls are in place to maintain the specified temperature range.
- Allow the equipment to warm up/stabilize for at least [TBD] minutes prior to calibration.
- Ensure cleanliness of the equipment and surrounding area.
- Verify that the equipment is level and properly positioned.
- Check power supply and functionality of the equipment prior to calibration.
7. Standards & Tools
- Reference standards required: [TBD]
- Traceability statement: ISO17025
- Certificate validity check: Ensure all calibration certificates are current and valid.
- Standard ID placeholders: [TBD]
8. Calibration Strategy
Calibration will follow the As Found and As Left logic:
- Minimum number of points: At least 5 points including endpoints (2°C, 8°C) and midpoints (5°C, 3°C, 7°C).
- Replicate readings: Perform 3 repeats at each point as a best practice.
9. Step-by-Step Procedure
- Preparation: Gather all necessary tools and reference standards.
- Stabilization: Allow the temperature probe to stabilize at the operating environment.
- As Found checks and data capture: Record initial temperature readings at defined points.
- Adjustment rules: If OOT is detected, follow adjustment procedures if allowed; otherwise, escalate to QA.
- As Left verification: After adjustments, recheck and record temperature readings.
- Labeling + documentation completion: Ensure all records are filled out contemporaneously (ALCOA+).
10. Acceptance Criteria & Calculations
Acceptance criteria: ±0.5°C from the target range.
Error calculation formula: Error = Observed Value – True Value.
Pass/Fail Logic: If the absolute value of Error is less than or equal to 0.5°C, then Pass; otherwise, Fail.
Rounding rule placeholder: [TBD]
11. Handling OOT / Failures
- Immediate containment: Tag equipment and stop use until resolved.
- Notification & deviation initiation: Notify QA and initiate deviation process.
- Impact assessment trigger: Define impact window and review of affected results/batches.
- Recalibration after repair/adjustment: Ensure recalibration is performed post-adjustment.
- QA disposition placeholders: [TBD]
12. Records & Retention
- Logbook entries: Document all calibration activities in the logbook.
- Calibration report/certificate filing: Maintain records of calibration certificates.
- e-records if applicable: [TBD]
- Label content requirements: Include calibrated date, due date, technician name, and status.
13. Annexures
Annexure A: Calibration Data Sheet
| Test Points | Reference | Observed | Error | Acceptance | Pass/Fail | As Found | As Left |
|---|---|---|---|---|---|---|---|
| [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] |
Annexure B: Quick Checklist
- Environmental controls verified
- Equipment warm-up completed
- Cleanliness checked
- Leveling confirmed
- Power supply checked
Annexure C: Example Label Text
Calibration Date: [TBD]
Due Date: [TBD]
Calibrated By: [TBD]
Status: [TBD]
Calibration Protocol
Calibration Protocol
| Protocol ID | Asset ID | Equipment Name | Equipment Type | Dept | Area | Process Use | Parameter | Units | Range | Acceptance Criteria | Frequency | Calibration Type | Traceability |
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| [TBD] | CAL-0190 | Refrigerator Temp Probe (2-8°C) | Temperature Probe | Warehouse/QA | Cold Storage | GDP storage | Temperature | °C | 2 – 8 | ±0.5°C | 180 days | External | ISO17025 |
Environmental Conditions
Temperature: [TBD] °C
Relative Humidity: [TBD] %
Vibration: [TBD]
Drafts: [TBD]
Test Point Selection
- Minimum: 2°C
- 25% of Range: 3°C
- 50% of Range: 5°C
- 75% of Range: 7°C
- Maximum: 8°C
Results Table
| Test Point | Ref Value | Observed (As Found) | Error | Limit | Pass/Fail | Observed (As Left) | Error | Pass/Fail |
|---|---|---|---|---|---|---|---|---|
| 2°C | 2°C | [TBD] | [TBD] | ±0.5°C | [TBD] | [TBD] | [TBD] | [TBD] |
| 3°C | 3°C | [TBD] | [TBD] | ±0.5°C | [TBD] | [TBD] | [TBD] | [TBD] |
| 5°C | 5°C | [TBD] | [TBD] | ±0.5°C | [TBD] | [TBD] | [TBD] | [TBD] |
| 7°C | 7°C | [TBD] | [TBD] | ±0.5°C | [TBD] | [TBD] | [TBD] | [TBD] |
| 8°C | 8°C | [TBD] | [TBD] | ±0.5°C | [TBD] | [TBD] | [TBD] | [TBD] |
Stepwise Instructions
- Preparation: Ensure the temperature probe is clean and calibrated within the last 180 days.
- Stabilization: Allow the probe to stabilize in the cold storage environment for at least 30 minutes before taking readings.
- Performing Readings: Take three replicate readings at each test point to ensure accuracy.
- Calculating Error: Calculate the error by subtracting the reference value from the observed value.
- Evaluation: Compare the error against the acceptance criteria of ±0.5°C.
- Documentation: Record all observed values, errors, and pass/fail results in the results table.
Deviation Instruction for OOT
- Stop use of the equipment immediately.
- Notify the QA department.
- Open a deviation report.
- Quarantine impacted results window: [TBD]
- Await QA decision on further actions.
Approvals
| Prepared By | Reviewed By | Approved By |
|---|---|---|
| [Signature/Date] | [Signature/Date] | [Signature/Date] |
Calibration Checklist
| Asset ID | Equipment | Dept | Area | Parameter | Range | Acceptance | Frequency | Traceability |
|---|---|---|---|---|---|---|---|---|
| CAL-0190 | Refrigerator Temp Probe (2-8°C) | Warehouse/QA | Cold Storage | Temperature | 2 – 8 °C | ±0.5°C | 180 days | ISO17025 |
Pre-calibration
- [ ] Verify environmental conditions are within specified limits
- [ ] Allow equipment to warm up as per SOP
- [ ] Clean the temperature probe and surrounding area
- [ ] Ensure the probe is level and properly positioned
- [ ] Confirm validity of reference standards used for calibration
- [ ] Ensure calibration logbook is ready for documentation
As Found Execution
- [ ] Set temperature setpoints for calibration
- [ ] Perform repeat measurements as per SOP
- [ ] Record all raw data accurately
- [ ] Perform necessary calculations for results
OOT Immediate Actions
- [ ] Tag out the equipment to prevent further use
- [ ] Stop use of the temperature probe immediately
- [ ] Notify QA and relevant personnel of the OOT condition
- [ ] Document deviation from acceptance criteria
- [ ] Trigger impact review as per SOP
As Left Verification
- [ ] Conduct post-adjustment checks to ensure accuracy
- [ ] Confirm that all measurements are within specified limits
Documentation & Closure
- [ ] Update equipment label with calibration details
- [ ] Attach calibration certificate to logbook
- [ ] Obtain reviewer sign-off on calibration documentation
- [ ] Update next due date for calibration in logbook
Deviation Impact Assessment
| Event Summary |
|---|
|
Asset: Refrigerator Temp Probe (2-8°C) Date: [TBD] Discovered By: [TBD] Document References: [TBD] As Found Results Summary: [TBD] |
Immediate Containment Checklist
- Stop use of the affected equipment
- Label/tag the equipment as out of service
- Segregate impacted work areas
Define Impact Window
| Last Known In-Tolerance Date | Discovery Date |
|---|---|
| [TBD] | [TBD] |
Product/Batch/Test Impact Review
| Batch/Study/Report ID | Results Potentially Impacted | Re-test/Re-weigh Decision |
|---|---|---|
| [TBD] | [TBD] | [TBD] |
Data Integrity Review Prompts (ALCOA+)
- Attributable: [TBD]
- Legible: [TBD]
- Contemporaneous: [TBD]
- Original: [TBD]
- Accurate: [TBD]
- Complete: [TBD]
- Consistent: [TBD]
- Enduring: [TBD]
- Available: [TBD]
Risk Assessment
| Severity | Probability | Detectability | Rationale | Overall Risk |
|---|---|---|---|---|
| [TBD] | [TBD] | [TBD] | [TBD] | [TBD] |
Investigation Summary
Root Cause: [TBD]
CAPA Plan
| Action | Owner | Due Date | Effectiveness Check |
|---|---|---|---|
| [TBD] | [TBD] | [TBD] | [TBD] |
QA Disposition
Release/Hold/Retest/Recall Triggers: [TBD]
Calibration Certificate
Certificate No: [TBD]
Date: [TBD]
Next Due Date: [TBD]
Asset Details
| Asset ID | Equipment Name | Equipment Type | Dept | Area | Make/Model | Serial No | Location | Parameter/Range/Units | Acceptance Criteria |
|---|---|---|---|---|---|---|---|---|---|
| CAL-0190 | Refrigerator Temp Probe (2-8°C) | Temperature Probe | Warehouse/QA | Cold Storage | [TBD] | [TBD] | [TBD] | Temperature (2-8°C) | ±0.5°C |
Conditions
Environment: [TBD]
Reference Standards
| Standard Name/ID | Certificate No | Due Date | Traceability Statement |
|---|---|---|---|
| [TBD] | [TBD] | [TBD] | ISO17025 |
Results
As Found Results
| Test Point | Measured Value (°C) | Pass/Fail |
|---|---|---|
| 1 | [TBD] | [TBD] |
| 2 | [TBD] | [TBD] |
| 3 | [TBD] | [TBD] |
| 4 | [TBD] | [TBD] |
| 5 | [TBD] | [TBD] |
As Left Results
| Test Point | Measured Value (°C) | Pass/Fail |
|---|---|---|
| 1 | [TBD] | [TBD] |
| 2 | [TBD] | [TBD] |
| 3 | [TBD] | [TBD] |
| 4 | [TBD] | [TBD] |
| 5 | [TBD] | [TBD] |
Measurement Uncertainty Statement: [TBD]
Statement of Conformity: [Pass/Fail overall]
Remarks: [TBD]
Signatures
Performed by: ______________________ Date: __________
Checked by: ______________________ Date: __________
Approved by: ______________________ Date: __________