Calibration Documents: Stability Chamber Controller Display Temp (CAL-0182)
| Asset ID | CAL-0182 |
|---|---|
| Equipment | Stability Chamber Controller Display Temp |
| Equipment Type | Temp Controller |
| Department | Stability |
| Area/Location | Stability Room |
| Process Use | ICH stability |
| Parameter | Temperature |
| Units | °C |
| Range Min | 0 |
| Range Max | 60 |
| Acceptance Criteria | ±0.5°C |
| Calibration Frequency (days) | 365 |
| Calibration Type | External |
| Traceability | ISO17025 |
Table of Contents
Calibration SOP
1. Purpose
The calibration of the Stability Chamber Controller Display Temperature is performed to ensure accurate temperature readings within the stability room, which is critical for maintaining the integrity of stability studies as per ICH guidelines. This calibration controls risks related to accuracy, patient safety, batch quality, and data integrity.
2. Scope
- Asset ID: CAL-0182
- Equipment Name/Type: Stability Chamber Controller Display Temp / Temp Controller
- Department/Area: Stability / Stability Room
- Process Use: ICH stability
- Parameter: Temperature
- Range: 0°C to 60°C
- Units: °C
- Calibration Frequency: 365 days
3. Definitions
- Calibration: The process of comparing a measurement instrument to a known standard to ensure accuracy.
- Verification: The process of confirming that a measurement instrument is functioning correctly.
- Adjustment: The act of correcting a measurement instrument to ensure it provides accurate readings.
- As Found: The condition of the instrument prior to any adjustments.
- As Left: The condition of the instrument after adjustments have been made.
- OOT (Out of Tolerance): A condition where the measurement falls outside the defined acceptance criteria.
- Drift: The gradual change in the accuracy of a measurement instrument over time.
- Traceability: The ability to verify the history, location, or application of an item by means of documented recorded identification.
- Measurement Uncertainty: A parameter that characterizes the range of values within which the true value of a measurement lies.
- Intermediate Check: A quick assessment of the instrument’s performance between full calibrations.
4. References
- Company Metrology SOP [TBD]
- ISO/IEC 17025 traceability via calibration certificates
- ALCOA+ principles
5. Responsibilities
- Technician/Metrology: Perform calibration and maintain records.
- QC/QA Reviewer: Review calibration records and ensure compliance.
- Owner Department: Ensure equipment is available and in good condition for calibration.
- External Service Provider: Conduct external calibrations as required.
6. Safety & Preconditions
- Ensure environmental controls are in place (temperature, humidity).
- Allow equipment to warm-up and stabilize as per manufacturer specifications.
- Maintain cleanliness of the calibration area and equipment.
- Check leveling of the equipment to ensure accurate readings.
- Verify power supply is stable and within specifications.
7. Standards & Tools
- Reference Standards Required: [TBD]
- Traceability Statement: ISO17025
- Certificate Validity Check: Ensure all calibration certificates are current and valid.
- Standard ID: [TBD]
8. Calibration Strategy
Calibration will follow the As Found and As Left logic. The minimum number of points for calibration will be 5, including endpoints (0°C and 60°C) and midpoints (15°C, 30°C, and 45°C). Each point will have 3 replicate readings to ensure accuracy.
9. Step-by-Step Procedure
- Preparation: Gather all necessary tools and reference standards.
- Stabilization: Allow the stability chamber to reach the desired temperature range.
- As Found checks and data capture: Measure the temperature at each defined point and record the data contemporaneously.
- Adjustment rules: If OOT is found, escalate to the QA department for further action.
- As Left verification: Confirm that the instrument is within acceptance criteria after any adjustments.
- Labeling + documentation completion: Complete all required documentation and label the equipment with calibration details.
10. Acceptance Criteria & Calculations
The acceptance criteria for this calibration is ±0.5°C. The error calculation formula is:
Error = Observed Value – True Value
Pass/Fail logic: If the absolute value of the error is within the acceptance criteria, the calibration passes; otherwise, it fails.
Rounding rule: [TBD]
11. Handling OOT / Failures
- Immediate containment: Tag the equipment and stop its use.
- Notification & deviation initiation: Notify QA and initiate a deviation report.
- Impact assessment trigger: Define the impact window and review affected results/batches.
- Recalibration after repair/adjustment: Ensure recalibration is performed post-adjustment.
- QA disposition placeholders: [TBD]
12. Records & Retention
- Logbook entries: Document all calibration activities in the logbook.
- Calibration report/certificate filing: Maintain a file of all calibration reports and certificates.
- e-Records if applicable: [TBD]
- Label content requirements: Include calibrated date, due date, technician name, and status.
13. Annexures
Annexure A: Calibration Data Sheet
| Test Points | Reference | Observed | Error | Acceptance | Pass/Fail | As Found | As Left |
|---|---|---|---|---|---|---|---|
| [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] |
Annexure B: Quick Checklist
- Environmental controls verified
- Equipment warmed up
- Cleanliness checked
- Leveling confirmed
- Power supply stable
Annexure C: Example Label Text
Calibration Date: [TBD]
Due Date: [TBD]
Calibrated By: [TBD]
Status: [TBD]
Calibration Protocol
Calibration Protocol
| Protocol ID | Asset ID | Equipment Name | Equipment Type | Dept | Area | Process Use | Parameter | Units | Range | Acceptance Criteria | Frequency | Calibration Type | Traceability |
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| [TBD] | CAL-0182 | Stability Chamber Controller Display Temp | Temp Controller | Stability | Stability Room | ICH stability | Temperature | °C | 0 – 60 | ±0.5°C | 365 days | External | ISO17025 |
Environmental Conditions
Temperature: [TBD]
Relative Humidity: [TBD]
Vibration: [TBD]
Drafts: [TBD]
Test Point Selection
- Minimum: 0°C
- 25% of Range: 15°C
- 50% of Range: 30°C
- 75% of Range: 45°C
- Maximum: 60°C
Results Table
| Test Point | Ref Value | Observed (As Found) | Error | Limit | Pass/Fail | Observed (As Left) | Error | Pass/Fail |
|---|---|---|---|---|---|---|---|---|
| 0°C | [TBD] | [TBD] | [TBD] | ±0.5°C | [TBD] | [TBD] | [TBD] | [TBD] |
| 15°C | [TBD] | [TBD] | [TBD] | ±0.5°C | [TBD] | [TBD] | [TBD] | [TBD] |
| 30°C | [TBD] | [TBD] | [TBD] | ±0.5°C | [TBD] | [TBD] | [TBD] | [TBD] |
| 45°C | [TBD] | [TBD] | [TBD] | ±0.5°C | [TBD] | [TBD] | [TBD] | [TBD] |
| 60°C | [TBD] | [TBD] | [TBD] | ±0.5°C | [TBD] | [TBD] | [TBD] | [TBD] |
Stepwise Instructions
- Preparation: Ensure the stability chamber is clean and operational. Verify the calibration equipment is within its calibration date.
- Stabilization: Allow the stability chamber to stabilize at each test point for a minimum of [TBD] minutes.
- Performing Readings: Take at least three replicate readings at each test point to ensure accuracy.
- Calculating Error: Calculate the error for each observed value using the formula: Error = Observed – Ref Value.
- Evaluation: Compare the calculated error against the acceptance criteria of ±0.5°C.
- Documentation: Record all observations, calculations, and results in the results table.
Deviation Instruction for OOT
- Stop use of the equipment immediately.
- Notify the QA department.
- Open a deviation report.
- Quarantine impacted results window: [TBD]
- Await QA decision on further actions.
Approvals
| Prepared | Reviewed | Approved |
|---|---|---|
| [Signature/Date] | [Signature/Date] | [Signature/Date] |
Calibration Checklist
| Asset ID | Equipment | Dept | Area | Parameter | Range | Acceptance | Frequency | Traceability |
|---|---|---|---|---|---|---|---|---|
| CAL-0182 | Stability Chamber Controller Display Temp | Stability | Stability Room | Temperature | 0 to 60 °C | ±0.5°C | 365 days | ISO17025 |
Pre-calibration
- [ ] Verify environmental conditions are within acceptable limits
- [ ] Ensure equipment has been powered on for the required warm-up period
- [ ] Clean the equipment as per SOP
- [ ] Level the equipment if necessary
- [ ] Check validity of reference standards
- [ ] Ensure calibration logbook is ready for documentation
As Found Execution
- [ ] Record initial setpoints of the temperature controller
- [ ] Perform repeat measurements to ensure consistency
- [ ] Record all raw data accurately
- [ ] Perform necessary calculations for data analysis
OOT Immediate Actions
- [ ] Tag out the equipment if out of tolerance (OOT)
- [ ] Stop use of the equipment immediately
- [ ] Notify the appropriate personnel
- [ ] Document the deviation
- [ ] Trigger impact review process
As Left Verification
- [ ] Conduct post-adjustment checks to confirm accuracy
- [ ] Ensure all measurements are within specified limits
Documentation & Closure
- [ ] Update equipment label with calibration details
- [ ] Attach calibration certificate to the logbook
- [ ] Obtain reviewer sign-off on calibration documentation
- [ ] Update next due date for calibration
Data Integrity (ALCOA+): Ensure all data recorded is attributable, legible, contemporaneous, original, and accurate. Maintain compliance with all applicable regulations.
Deviation Impact Assessment
| Event Summary |
|---|
|
Asset: Stability Chamber Controller Display Temp Date: [TBD] Discovered By: [TBD] Document References: [TBD] As Found Results Summary: [TBD] |
Immediate Containment Checklist
- Stop use of the asset
- Label/Tag the asset as out of service
- Segregate impacted work
Define Impact Window
Last known in-tolerance date: [TBD] to discovery date: [TBD]
Product/Batch/Test Impact Review
| Batch/Study/Report ID | Results Potentially Impacted | Re-test/Re-weigh Decision |
|---|---|---|
| [TBD] | [TBD] | [TBD] |
Data Integrity Review Prompts (ALCOA+)
- Attributable: [TBD]
- Legible: [TBD]
- Contemporaneous: [TBD]
- Original: [TBD]
- Accurate: [TBD]
- Complete: [TBD]
- Consistent: [TBD]
- Enduring: [TBD]
- Available: [TBD]
Risk Assessment
| Severity | Probability | Detectability | Rationale | Overall Risk |
|---|---|---|---|---|
| [TBD] | [TBD] | [TBD] | [TBD] | [TBD] |
Investigation Summary
Root Cause: [TBD]
CAPA Plan
| Action | Owner | Due Date | Effectiveness Check |
|---|---|---|---|
| [TBD] | [TBD] | [TBD] | [TBD] |
QA Disposition
Release/Hold/Retest/Recall Triggers: [TBD]
Calibration Certificate
Certificate No: [TBD]
Date: [TBD]
Next Due Date: [TBD]
Asset Details
| Asset ID | Equipment Name | Equipment Type | Dept | Area | Make/Model | Serial No | Location | Parameter/Range/Units | Acceptance Criteria |
|---|---|---|---|---|---|---|---|---|---|
| CAL-0182 | Stability Chamber Controller Display Temp | Temp Controller | Stability | Stability Room | [TBD] | [TBD] | [TBD] | Temperature: 0 – 60 °C | ±0.5°C |
Conditions
Environment: [TBD]
Reference Standards
| Standard Name/ID | Certificate No | Due Date | Traceability Statement |
|---|---|---|---|
| [TBD] | [TBD] | [TBD] | Traceability to ISO17025 |
Results
As Found Results
| Test Point | Measured Value (°C) | Pass/Fail |
|---|---|---|
| 1 | [TBD] | [TBD] |
| 2 | [TBD] | [TBD] |
| 3 | [TBD] | [TBD] |
| 4 | [TBD] | [TBD] |
| 5 | [TBD] | [TBD] |
As Left Results
| Test Point | Measured Value (°C) | Pass/Fail |
|---|---|---|
| 1 | [TBD] | [TBD] |
| 2 | [TBD] | [TBD] |
| 3 | [TBD] | [TBD] |
| 4 | [TBD] | [TBD] |
| 5 | [TBD] | [TBD] |
Measurement Uncertainty Statement: [TBD]
Statement of Conformity: [Pass/Fail overall]
Remarks: [TBD]
Signatures
Performed by: ______________________ Date: ___________
Checked by: ______________________ Date: ___________
Approved by: ______________________ Date: ___________