Calibration Documents: Stability Chamber Temp Probe (CAL-0180)
| Asset ID | CAL-0180 |
|---|---|
| Equipment | Stability Chamber Temp Probe |
| Equipment Type | Temperature Probe |
| Department | Stability |
| Area/Location | Stability Room |
| Process Use | ICH stability |
| Parameter | Temperature |
| Units | °C |
| Range Min | 0 |
| Range Max | 60 |
| Acceptance Criteria | ±0.5°C |
| Calibration Frequency (days) | 180 |
| Calibration Type | External |
| Traceability | ISO17025 |
Table of Contents
Calibration SOP
1. Purpose
The calibration of the Stability Chamber Temperature Probe (CAL-0180) is performed to ensure accurate temperature measurements within the stability room. This calibration controls risks associated with data integrity, patient safety, and batch quality by ensuring that the temperature remains within specified limits during ICH stability testing.
2. Scope
- Asset ID: CAL-0180
- Equipment Name/Type: Stability Chamber Temp Probe / Temperature Probe
- Department/Area: Stability / Stability Room
- Process Use: ICH stability
- Parameter: Temperature
- Range: 0°C to 60°C
- Units: °C
- Calibration Frequency: 180 days
3. Definitions
- Calibration: The process of comparing a measurement device to a known standard to ensure accuracy.
- Verification: The process of confirming that a measurement device meets specified requirements.
- Adjustment: The process of modifying a measurement device to bring it into compliance with standards.
- As Found: The condition of the measurement device prior to any adjustments.
- As Left: The condition of the measurement device after adjustments have been made.
- OOT (Out of Tolerance): A condition where the measurement falls outside the specified acceptance criteria.
- Drift: The gradual change in a measurement device’s performance over time.
- Traceability: The ability to verify the history, location, or application of an item by means of documented recorded identification.
- Measurement Uncertainty: A parameter that characterizes the range of values within which the true value of a measurement lies.
- Intermediate Check: A verification process conducted between scheduled calibrations to ensure continued accuracy.
4. References
- Company Metrology SOP [TBD]
- ISO/IEC 17025 traceability via calibration certificates
- ALCOA+ principles
5. Responsibilities
- Technician/Metrology: Perform calibration and document results.
- QC/QA Reviewer: Review calibration documentation for compliance.
- Owner Department: Ensure equipment is available and accessible for calibration.
- External Service Provider: Conduct external calibration as per schedule.
6. Safety & Preconditions
- Ensure environmental controls are in place (temperature, humidity).
- Allow equipment to warm up and stabilize as per manufacturer specifications.
- Maintain cleanliness in the calibration area.
- Check that the equipment is level and properly positioned.
- Verify power supply is stable and within specifications.
7. Standards & Tools
- Reference Standards: [TBD]
- Traceability Statement: ISO17025
- Certificate Validity Check: Ensure calibration certificates are current and valid.
8. Calibration Strategy
Calibration will follow the As Found and As Left logic:
- Minimum number of points: At least 5 points including endpoints (0°C and 60°C) and midpoints (15°C, 30°C, 45°C).
- Replicate readings at each point: 3 repeats as a best practice.
9. Step-by-Step Procedure
- Preparation: Gather all necessary tools and reference standards.
- Stabilization: Allow the temperature probe to stabilize in the environment.
- As Found checks and data capture: Record initial temperature readings at specified points.
- Adjustment rules: If OOT is found, escalate for adjustment as per SOP.
- As Left verification: Record final readings after any adjustments.
- Labeling + documentation completion: Ensure all records are filled out contemporaneously (ALCOA+).
10. Acceptance Criteria & Calculations
Acceptance criteria: ±0.5°C
Error calculation formula: Error = Observed Value – True Value
Pass/Fail Logic: If the absolute value of Error ≤ 0.5°C, then Pass; otherwise, Fail.
Rounding rule: [TBD]
11. Handling OOT / Failures
- Immediate containment: Tag equipment and stop use.
- Notification & deviation initiation: Notify QA and document deviation.
- Impact assessment trigger: Review affected results/batches within the impact window.
- Recalibration after repair/adjustment: Schedule recalibration as necessary.
- QA disposition: [TBD]
12. Records & Retention
- Logbook entries: Document all calibration activities.
- Calibration report/certificate filing: Maintain records as per company policy.
- e-Records if applicable: [TBD]
- Label content requirements: Include calibrated date, due date, technician name, and status.
13. Annexures
Annexure A: Calibration Data Sheet
| Test Points | Reference | Observed | Error | Acceptance | Pass/Fail | As Found | As Left |
|---|---|---|---|---|---|---|---|
| [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] |
Annexure B: Quick Checklist
- Environmental controls verified
- Equipment warm-up completed
- Calibration tools prepared
- Documentation ready
Annexure C: Example Label Text
Calibration Date: [TBD]
Due Date: [TBD]
Calibrated By: [TBD]
Status: [TBD]
Calibration Protocol
Calibration Protocol
| Protocol ID | Asset ID | Equipment Name | Equipment Type | Dept | Area | Process Use | Parameter | Units | Range | Acceptance Criteria | Frequency | Calibration Type | Traceability |
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| [TBD] | CAL-0180 | Stability Chamber Temp Probe | Temperature Probe | Stability | Stability Room | ICH stability | Temperature | °C | 0 – 60 | ±0.5°C | 180 days | External | ISO17025 |
Environmental Conditions
Temperature: [TBD] °C
Relative Humidity: [TBD] %
Vibration: [TBD]
Drafts: [TBD]
Test Point Selection
Test points will be selected based on the following range:
- Min: 0°C
- 25%: 15°C
- 50%: 30°C
- 75%: 45°C
- Max: 60°C
Results Table
| Test Point | Ref Value | Observed (As Found) | Error | Limit | Pass/Fail | Observed (As Left) | Error | Pass/Fail |
|---|---|---|---|---|---|---|---|---|
| 0°C | [TBD] | [TBD] | [TBD] | ±0.5°C | [TBD] | [TBD] | [TBD] | [TBD] |
| 15°C | [TBD] | [TBD] | [TBD] | ±0.5°C | [TBD] | [TBD] | [TBD] | [TBD] |
| 30°C | [TBD] | [TBD] | [TBD] | ±0.5°C | [TBD] | [TBD] | [TBD] | [TBD] |
| 45°C | [TBD] | [TBD] | [TBD] | ±0.5°C | [TBD] | [TBD] | [TBD] | [TBD] |
| 60°C | [TBD] | [TBD] | [TBD] | ±0.5°C | [TBD] | [TBD] | [TBD] | [TBD] |
Stepwise Instructions
- Preparation: Ensure the stability chamber is clean and free from any obstructions.
- Stabilization: Allow the stability chamber to stabilize at the test points for a minimum of [TBD] hours.
- Performing Readings: Take at least three replicate readings at each test point.
- Calculating Error: Calculate the error for each reading using the formula: Error = Observed Value – Reference Value.
- Evaluation: Compare the calculated errors against the acceptance criteria of ±0.5°C.
- Documentation: Record all observations, calculations, and results in the calibration log.
Deviation Instruction for OOT
If any reading is outside the acceptance criteria:
- Stop use of the equipment immediately.
- Notify the supervisor and quality assurance (QA) team.
- Open a deviation report.
- Quarantine impacted results window: [TBD]
- Await QA decision on further actions.
Approvals
| Prepared By | Reviewed By | Approved By |
|---|---|---|
| [Signature/Date] | [Signature/Date] | [Signature/Date] |
Calibration Checklist
| Asset ID | Equipment | Dept | Area | Parameter | Range | Acceptance | Frequency | Traceability |
|---|---|---|---|---|---|---|---|---|
| CAL-0180 | Stability Chamber Temp Probe | Stability | Stability Room | Temperature | 0 to 60 °C | ±0.5°C | 180 days | ISO17025 |
Pre-calibration
- [ ] Verify environmental conditions are within specified limits
- [ ] Ensure temperature probe has warmed up adequately
- [ ] Clean the equipment as per SOP
- [ ] Level the temperature probe if necessary
- [ ] Confirm reference standards are valid and calibrated
- [ ] Ensure logbook is ready for documentation
As Found Execution
- [ ] Set temperature probe to specified setpoints
- [ ] Perform repeat measurements as required
- [ ] Record raw data accurately
- [ ] Perform necessary calculations for analysis
OOT Immediate Actions
- [ ] Tag out the equipment if out of tolerance (OOT)
- [ ] Stop use of the temperature probe immediately
- [ ] Notify QA and relevant personnel
- [ ] Document deviation in the logbook
- [ ] Trigger impact review as per SOP
As Left Verification
- [ ] Conduct post-adjustment checks to confirm accuracy
- [ ] Ensure all measurements are within acceptance limits
Documentation & Closure
- [ ] Update equipment label with new calibration date
- [ ] Attach calibration certificate to logbook
- [ ] Obtain reviewer sign-off on calibration documentation
- [ ] Update next due date for calibration
Data Integrity (ALCOA+): Ensure all data recorded is Attributable, Legible, Contemporaneous, Original, Accurate, and Complete.
Deviation Impact Assessment
| Event Summary | Details |
|---|---|
| Asset | Stability Chamber Temp Probe |
| Date | [TBD] |
| Discovered By | [TBD] |
| Document References | [TBD] |
| As Found Results Summary | [TBD] |
Immediate Containment Checklist
- Stop use of the equipment
- Label/Tag the equipment as out of service
- Segregate impacted work
Define Impact Window
| Last Known In-Tolerance Date | Discovery Date |
|---|---|
| [TBD] | [TBD] |
Product/Batch/Test Impact Review
| Batch/Study/Report ID | Results Potentially Impacted | Re-test/Re-weigh Decision |
|---|---|---|
| [TBD] | [TBD] | [TBD] |
Data Integrity Review Prompts (ALCOA+)
- Attributable: [TBD]
- Legible: [TBD]
- Contemporaneous: [TBD]
- Original: [TBD]
- Accurate: [TBD]
- Complete: [TBD]
- Consistent: [TBD]
- Enduring: [TBD]
- Available: [TBD]
Risk Assessment
| Severity | Probability | Detectability | Rationale | Overall Risk |
|---|---|---|---|---|
| [TBD] | [TBD] | [TBD] | [TBD] | [TBD] |
Investigation Summary
Root Cause: [TBD]
CAPA Plan
| Action | Owner | Due Date | Effectiveness Check |
|---|---|---|---|
| [TBD] | [TBD] | [TBD] | [TBD] |
QA Disposition
Release/Hold/Retest/Recall Triggers: [TBD]
Calibration Certificate
Certificate No: [TBD]
Date: [TBD]
Next Due Date: [Calculated based on calibration_frequency_days]
Asset Details
| Asset ID | Equipment Name | Equipment Type | Dept | Area | Make/Model | Serial No | Location | Parameter/Range/Units | Acceptance Criteria |
|---|---|---|---|---|---|---|---|---|---|
| CAL-0180 | Stability Chamber Temp Probe | Temperature Probe | Stability | Stability Room | [TBD] | [TBD] | [TBD] | Temperature: 0 – 60 °C | ±0.5°C |
Conditions
Environment: [TBD]
Reference Standards
| Standard Name/ID | Certificate No | Due Date | Traceability Statement |
|---|---|---|---|
| [TBD] | [TBD] | [TBD] | Traceability to ISO17025 |
Results
As Found Results
| Test Point | Measured Value (°C) |
|---|---|
| 1 | [TBD] |
| 2 | [TBD] |
| 3 | [TBD] |
| 4 | [TBD] |
| 5 | [TBD] |
As Left Results
| Test Point | Measured Value (°C) |
|---|---|
| 1 | [TBD] |
| 2 | [TBD] |
| 3 | [TBD] |
| 4 | [TBD] |
| 5 | [TBD] |
Measurement Uncertainty Statement: [TBD]
Statement of Conformity: [Pass/Fail overall]
Remarks: [TBD]
Signatures
Performed by: ______________________ Date: __________
Checked by: ______________________ Date: __________
Approved by: ______________________ Date: __________