Temperature/RH Data Logger (CAL-0056) – Calibration Documents

Calibration Documents: Temperature/RH Data Logger (CAL-0056)

Asset ID CAL-0056
Equipment Temperature/RH Data Logger
Equipment Type TempRH Data Logger
Department Validation
Area/Location Stability/Storage
Process Use Mapping
Parameter Temperature/RH
Units °C/%RH
Range Min 0
Range Max 60
Acceptance Criteria ±0.5°C/±3%RH
Calibration Frequency (days) 365
Calibration Type External
Traceability ISO17025

Calibration SOP

1. Purpose

The calibration of the Temperature/RH Data Logger is performed to ensure accurate measurement of temperature and relative humidity within the stability/storage area. This calibration controls risks related to data integrity, patient safety, and batch quality by ensuring that the equipment operates within specified limits, thereby maintaining compliance with regulatory standards.

2. Scope

  • Asset ID: CAL-0056
  • Equipment Name/Type: Temperature/RH Data Logger
  • Department/Area: Validation / Stability/Storage
  • Process Use: Mapping
  • Parameter: Temperature/RH
  • Range: 0 to 60
  • Units: °C/%RH
  • Calibration Frequency: 365 days

3. Definitions

  • Calibration: The process of comparing a measurement instrument to a standard of known accuracy.
  • Verification: The process of confirming that a measurement instrument meets specified requirements.
  • Adjustment: The process of modifying a measurement instrument to bring it into alignment with a standard.
  • As Found: The condition of the equipment prior to any adjustments made during calibration.
  • As Left: The condition of the equipment after calibration adjustments have been made.
  • OOT (Out of Tolerance): A condition where the measurement falls outside the specified acceptance criteria.
  • Drift: The gradual change in the output of a measurement instrument over time.
  • Traceability: The ability to verify the history, location, or application of an item by means of documented recorded identification.
  • Measurement Uncertainty: A parameter that characterizes the dispersion of the values attributed to a measured quantity.
  • Intermediate Check: A periodic check of the measurement instrument to ensure it remains within specified limits between calibrations.

4. References

  • Company Metrology SOP [TBD]
  • ISO/IEC 17025 traceability via calibration certificates
  • ALCOA+ principles

5. Responsibilities

  • Technician/Metrology: Perform calibration and document results.
  • QC/QA Reviewer: Review calibration documentation for compliance.
  • Owner Department: Ensure equipment is available and operational for calibration.
  • External Service Provider: Conduct external calibration as per the calibration type.

6. Safety & Preconditions

  • Ensure environmental controls are in place (temperature and humidity within specified limits).
  • Allow equipment to warm up and stabilize as per manufacturer specifications.
  • Maintain cleanliness of the equipment and surrounding area.
  • Ensure equipment is level and properly positioned.
  • Check power supply and connections prior to calibration.
See also  Volumetric Flask 200 mL (CAL-0036) - Calibration Documents

7. Standards & Tools

  • Reference standards required: [TBD]
  • Traceability: ISO17025
  • Certificate validity check: [TBD]
  • Standard ID placeholders: [TBD]

8. Calibration Strategy

The calibration will follow an As Found and As Left logic:

  • Minimum number of points: At least 5 points including endpoints and midpoints within the range of 0 to 60 °C/%RH.
  • Replicate readings at each point: 3 repeats as a best practice.

9. Step-by-Step Procedure

  1. Preparation: Gather all necessary tools and reference standards.
  2. Stabilization: Allow the Temperature/RH Data Logger to stabilize in the calibration environment.
  3. As Found checks and data capture: Record initial readings at specified points.
  4. Adjustment rules: If OOT is found, escalate for adjustment if allowed; otherwise, document and notify QA.
  5. As Left verification: Record final readings after any adjustments.
  6. Labeling + documentation completion: Complete all required documentation and label the equipment accordingly.

Explicit instruction to record raw observations contemporaneously (ALCOA+).

10. Acceptance Criteria & Calculations

Acceptance criteria: ±0.5°C/±3%RH.

Error calculation formula: Error = Observed Value – Reference Value.

Pass/Fail logic: If the error falls within the acceptance criteria, the calibration passes; otherwise, it fails.

Rounding rule placeholder: [TBD]

11. Handling OOT / Failures

  • Immediate containment: Tag equipment and stop use.
  • Notification & deviation initiation: Notify relevant personnel and initiate deviation process.
  • Impact assessment trigger: Define impact window and review of affected results/batches.
  • Recalibration after repair/adjustment: Ensure recalibration is performed post-adjustment.
  • QA disposition placeholders: [TBD]

12. Records & Retention

  • Logbook entries: Document all calibration activities.
  • Calibration report/certificate filing: Maintain records in accordance with company policy.
  • e-Records if applicable: [TBD]
  • Label content requirements: Calibrated date, due date, by whom, status.

13. Annexures

Annexure A: Calibration Data Sheet

Test Points Reference Observed Error Acceptance Pass/Fail As Found As Left
[TBD] [TBD] [TBD] [TBD] [TBD] [TBD] [TBD] [TBD]

Annexure B: Quick Checklist

  • Environmental controls verified
  • Equipment warm-up completed
  • Cleanliness checked
  • Leveling confirmed
  • Power supply checked

Annexure C: Example Label Text

Calibration Date: [TBD]

Due Date: [TBD]

Calibrated By: [TBD]

Status: [TBD]

⬆ Back to TOC

See also  Microbial Air Sampler Flow (CAL-0174) - Calibration Documents

Calibration Protocol

Calibration Protocol

Protocol ID Asset ID Equipment Name Equipment Type Dept Area Process Use Parameter Units Range Acceptance Criteria Frequency Calibration Type Traceability
[TBD] CAL-0056 Temperature/RH Data Logger TempRH Data Logger Validation Stability/Storage Mapping Temperature/RH °C/%RH 0 to 60 ±0.5°C/±3%RH 365 days External ISO17025

Environmental Conditions

Temperature: [TBD]

Relative Humidity: [TBD]

Vibration: [TBD]

Drafts: [TBD]

Test Point Selection

  • Minimum: 0°C/%RH
  • 25%: 15°C/%RH
  • 50%: 30°C/%RH
  • 75%: 45°C/%RH
  • Maximum: 60°C/%RH

Results Table

Test Point Ref Value Observed (As Found) Error Limit Pass/Fail Observed (As Left) Error Pass/Fail
0°C/%RH [TBD] [TBD] [TBD] ±0.5°C/±3%RH [TBD] [TBD] [TBD] [TBD]
15°C/%RH [TBD] [TBD] [TBD] ±0.5°C/±3%RH [TBD] [TBD] [TBD] [TBD]
30°C/%RH [TBD] [TBD] [TBD] ±0.5°C/±3%RH [TBD] [TBD] [TBD] [TBD]
45°C/%RH [TBD] [TBD] [TBD] ±0.5°C/±3%RH [TBD] [TBD] [TBD] [TBD]
60°C/%RH [TBD] [TBD] [TBD] ±0.5°C/±3%RH [TBD] [TBD] [TBD] [TBD]

Stepwise Instructions

  1. Preparation: Ensure the Temperature/RH Data Logger is clean and calibrated prior to use.
  2. Stabilization: Allow the equipment to stabilize in the calibration environment for at least 30 minutes.
  3. Performing Readings: Take three replicate readings at each test point to ensure accuracy.
  4. Calculating Error: Calculate the error for each observed value against the reference value.
  5. Evaluation: Compare the calculated errors to the acceptance criteria.
  6. Documentation: Record all findings in the results table and ensure data integrity (ALCOA+).

Deviation Instruction for OOT

  • Stop use of the equipment immediately.
  • Notify the QA department of the out-of-tolerance condition.
  • Open a deviation report for the incident.
  • Quarantine the impacted results window: [TBD].
  • Await QA decision on further actions.

Approvals

Prepared By Reviewed By Approved By
[Signature/Date] [Signature/Date] [Signature/Date]

⬆ Back to TOC


Calibration Checklist

Asset ID Equipment Dept Area Parameter Range Acceptance Frequency Traceability
CAL-0056 Temperature/RH Data Logger Validation Stability/Storage Temperature/RH 0 to 60 °C/%RH ±0.5°C/±3%RH 365 days ISO17025

Pre-calibration

  • [ ] Verify environmental conditions are within specified limits
  • [ ] Allow equipment to warm up as per manufacturer guidelines
  • [ ] Clean the equipment as per SOP
  • [ ] Level the equipment if applicable
  • [ ] Confirm reference standards are valid and calibrated
  • [ ] Ensure logbook is ready for data entry

As Found Execution

  • [ ] Set the required setpoints for calibration
  • [ ] Perform multiple repeats for accuracy
  • [ ] Record all raw data accurately
  • [ ] Perform necessary calculations for results
See also  Inline Flow Meter (Solvent) (CAL-0203) - Calibration Documents

OOT Immediate Actions

  • [ ] Tag out the equipment
  • [ ] Stop use of the equipment immediately
  • [ ] Notify the appropriate personnel
  • [ ] Document deviation in the logbook
  • [ ] Trigger impact review as per SOP

As Left Verification

  • [ ] Conduct post-adjustment checks
  • [ ] Confirm all results are within acceptance limits

Documentation & Closure

  • [ ] Update equipment label with new calibration date
  • [ ] Attach calibration certificate to documentation
  • [ ] Obtain reviewer sign-off
  • [ ] Update next due date in logbook

⬆ Back to TOC


Deviation Impact Assessment

Event Summary
Asset: Temperature/RH Data Logger
Date: [TBD]
Discovered By: [TBD]
Document References: [TBD]
As Found Results Summary: [TBD]

Immediate Containment Checklist

  • Stop use of the asset
  • Label/tag the asset as out of tolerance
  • Segregate impacted work

Define Impact Window

Last known in-tolerance date: [TBD] to discovery date: [TBD]

Product/Batch/Test Impact Review

Batch/Study/Report ID Results Potentially Impacted Re-test/Re-weigh Decision
[TBD] [TBD] [TBD]

Data Integrity Review Prompts (ALCOA+)

  • Attributable: [TBD]
  • Legible: [TBD]
  • Contemporaneous: [TBD]
  • Original: [TBD]
  • Accurate: [TBD]
  • Complete: [TBD]
  • Consistent: [TBD]
  • Enduring: [TBD]
  • Available: [TBD]

Risk Assessment

Severity Probability Detectability Rationale Overall Risk
[TBD] [TBD] [TBD] [TBD] [TBD]

Investigation Summary

Root Cause: [TBD]

CAPA Plan

Action Owner Due Date Effectiveness Check
[TBD] [TBD] [TBD] [TBD]

QA Disposition

Release/Hold/Retest/Recall Triggers: [TBD]

⬆ Back to TOC


Calibration Certificate

Certificate No: [TBD]

Date: [TBD]

Next Due Date: [TBD]

Asset Details

Asset ID Equipment Name Equipment Type Dept Area Make/Model Serial No Location Parameter/Range/Units Acceptance Criteria
CAL-0056 Temperature/RH Data Logger TempRH Data Logger Validation Stability/Storage [TBD] [TBD] [TBD] Temperature: 0-60 °C, RH: 0-100 %RH ±0.5°C/±3%RH

Conditions

Environment: [TBD]

Reference Standards

Standard Name/ID Certificate No Due Date Traceability Statement
[TBD] [TBD] [TBD] Traceability to ISO17025

Results

As Found Results

Test Point Measured Value Acceptance Criteria Pass/Fail
Point 1 [TBD] ±0.5°C/±3%RH [TBD]
Point 2 [TBD] ±0.5°C/±3%RH [TBD]
Point 3 [TBD] ±0.5°C/±3%RH [TBD]
Point 4 [TBD] ±0.5°C/±3%RH [TBD]
Point 5 [TBD] ±0.5°C/±3%RH [TBD]

As Left Results

Test Point Measured Value Acceptance Criteria Pass/Fail
Point 1 [TBD] ±0.5°C/±3%RH [TBD]
Point 2 [TBD] ±0.5°C/±3%RH [TBD]
Point 3 [TBD] ±0.5°C/±3%RH [TBD]
Point 4 [TBD] ±0.5°C/±3%RH [TBD]
Point 5 [TBD] ±0.5°C/±3%RH [TBD]

Measurement Uncertainty Statement: [TBD]

Statement of Conformity: [TBD]

Remarks: [TBD]

Signatures

Performed by: ______________________ Date: __________

Checked by: ______________________ Date: __________

Approved by: ______________________ Date: __________

⬆ Back to TOC