Calibration Documents: Temperature/RH Data Logger (CAL-0056)
| Asset ID | CAL-0056 |
|---|---|
| Equipment | Temperature/RH Data Logger |
| Equipment Type | TempRH Data Logger |
| Department | Validation |
| Area/Location | Stability/Storage |
| Process Use | Mapping |
| Parameter | Temperature/RH |
| Units | °C/%RH |
| Range Min | 0 |
| Range Max | 60 |
| Acceptance Criteria | ±0.5°C/±3%RH |
| Calibration Frequency (days) | 365 |
| Calibration Type | External |
| Traceability | ISO17025 |
Table of Contents
Calibration SOP
1. Purpose
The calibration of the Temperature/RH Data Logger is performed to ensure accurate measurement of temperature and relative humidity within the stability/storage area. This calibration controls risks related to data integrity, patient safety, and batch quality by ensuring that the equipment operates within specified limits, thereby maintaining compliance with regulatory standards.
2. Scope
- Asset ID: CAL-0056
- Equipment Name/Type: Temperature/RH Data Logger
- Department/Area: Validation / Stability/Storage
- Process Use: Mapping
- Parameter: Temperature/RH
- Range: 0 to 60
- Units: °C/%RH
- Calibration Frequency: 365 days
3. Definitions
- Calibration: The process of comparing a measurement instrument to a standard of known accuracy.
- Verification: The process of confirming that a measurement instrument meets specified requirements.
- Adjustment: The process of modifying a measurement instrument to bring it into alignment with a standard.
- As Found: The condition of the equipment prior to any adjustments made during calibration.
- As Left: The condition of the equipment after calibration adjustments have been made.
- OOT (Out of Tolerance): A condition where the measurement falls outside the specified acceptance criteria.
- Drift: The gradual change in the output of a measurement instrument over time.
- Traceability: The ability to verify the history, location, or application of an item by means of documented recorded identification.
- Measurement Uncertainty: A parameter that characterizes the dispersion of the values attributed to a measured quantity.
- Intermediate Check: A periodic check of the measurement instrument to ensure it remains within specified limits between calibrations.
4. References
- Company Metrology SOP [TBD]
- ISO/IEC 17025 traceability via calibration certificates
- ALCOA+ principles
5. Responsibilities
- Technician/Metrology: Perform calibration and document results.
- QC/QA Reviewer: Review calibration documentation for compliance.
- Owner Department: Ensure equipment is available and operational for calibration.
- External Service Provider: Conduct external calibration as per the calibration type.
6. Safety & Preconditions
- Ensure environmental controls are in place (temperature and humidity within specified limits).
- Allow equipment to warm up and stabilize as per manufacturer specifications.
- Maintain cleanliness of the equipment and surrounding area.
- Ensure equipment is level and properly positioned.
- Check power supply and connections prior to calibration.
7. Standards & Tools
- Reference standards required: [TBD]
- Traceability: ISO17025
- Certificate validity check: [TBD]
- Standard ID placeholders: [TBD]
8. Calibration Strategy
The calibration will follow an As Found and As Left logic:
- Minimum number of points: At least 5 points including endpoints and midpoints within the range of 0 to 60 °C/%RH.
- Replicate readings at each point: 3 repeats as a best practice.
9. Step-by-Step Procedure
- Preparation: Gather all necessary tools and reference standards.
- Stabilization: Allow the Temperature/RH Data Logger to stabilize in the calibration environment.
- As Found checks and data capture: Record initial readings at specified points.
- Adjustment rules: If OOT is found, escalate for adjustment if allowed; otherwise, document and notify QA.
- As Left verification: Record final readings after any adjustments.
- Labeling + documentation completion: Complete all required documentation and label the equipment accordingly.
Explicit instruction to record raw observations contemporaneously (ALCOA+).
10. Acceptance Criteria & Calculations
Acceptance criteria: ±0.5°C/±3%RH.
Error calculation formula: Error = Observed Value – Reference Value.
Pass/Fail logic: If the error falls within the acceptance criteria, the calibration passes; otherwise, it fails.
Rounding rule placeholder: [TBD]
11. Handling OOT / Failures
- Immediate containment: Tag equipment and stop use.
- Notification & deviation initiation: Notify relevant personnel and initiate deviation process.
- Impact assessment trigger: Define impact window and review of affected results/batches.
- Recalibration after repair/adjustment: Ensure recalibration is performed post-adjustment.
- QA disposition placeholders: [TBD]
12. Records & Retention
- Logbook entries: Document all calibration activities.
- Calibration report/certificate filing: Maintain records in accordance with company policy.
- e-Records if applicable: [TBD]
- Label content requirements: Calibrated date, due date, by whom, status.
13. Annexures
Annexure A: Calibration Data Sheet
| Test Points | Reference | Observed | Error | Acceptance | Pass/Fail | As Found | As Left |
|---|---|---|---|---|---|---|---|
| [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] |
Annexure B: Quick Checklist
- Environmental controls verified
- Equipment warm-up completed
- Cleanliness checked
- Leveling confirmed
- Power supply checked
Annexure C: Example Label Text
Calibration Date: [TBD]
Due Date: [TBD]
Calibrated By: [TBD]
Status: [TBD]
Calibration Protocol
Calibration Protocol
| Protocol ID | Asset ID | Equipment Name | Equipment Type | Dept | Area | Process Use | Parameter | Units | Range | Acceptance Criteria | Frequency | Calibration Type | Traceability |
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| [TBD] | CAL-0056 | Temperature/RH Data Logger | TempRH Data Logger | Validation | Stability/Storage | Mapping | Temperature/RH | °C/%RH | 0 to 60 | ±0.5°C/±3%RH | 365 days | External | ISO17025 |
Environmental Conditions
Temperature: [TBD]
Relative Humidity: [TBD]
Vibration: [TBD]
Drafts: [TBD]
Test Point Selection
- Minimum: 0°C/%RH
- 25%: 15°C/%RH
- 50%: 30°C/%RH
- 75%: 45°C/%RH
- Maximum: 60°C/%RH
Results Table
| Test Point | Ref Value | Observed (As Found) | Error | Limit | Pass/Fail | Observed (As Left) | Error | Pass/Fail |
|---|---|---|---|---|---|---|---|---|
| 0°C/%RH | [TBD] | [TBD] | [TBD] | ±0.5°C/±3%RH | [TBD] | [TBD] | [TBD] | [TBD] |
| 15°C/%RH | [TBD] | [TBD] | [TBD] | ±0.5°C/±3%RH | [TBD] | [TBD] | [TBD] | [TBD] |
| 30°C/%RH | [TBD] | [TBD] | [TBD] | ±0.5°C/±3%RH | [TBD] | [TBD] | [TBD] | [TBD] |
| 45°C/%RH | [TBD] | [TBD] | [TBD] | ±0.5°C/±3%RH | [TBD] | [TBD] | [TBD] | [TBD] |
| 60°C/%RH | [TBD] | [TBD] | [TBD] | ±0.5°C/±3%RH | [TBD] | [TBD] | [TBD] | [TBD] |
Stepwise Instructions
- Preparation: Ensure the Temperature/RH Data Logger is clean and calibrated prior to use.
- Stabilization: Allow the equipment to stabilize in the calibration environment for at least 30 minutes.
- Performing Readings: Take three replicate readings at each test point to ensure accuracy.
- Calculating Error: Calculate the error for each observed value against the reference value.
- Evaluation: Compare the calculated errors to the acceptance criteria.
- Documentation: Record all findings in the results table and ensure data integrity (ALCOA+).
Deviation Instruction for OOT
- Stop use of the equipment immediately.
- Notify the QA department of the out-of-tolerance condition.
- Open a deviation report for the incident.
- Quarantine the impacted results window: [TBD].
- Await QA decision on further actions.
Approvals
| Prepared By | Reviewed By | Approved By |
|---|---|---|
| [Signature/Date] | [Signature/Date] | [Signature/Date] |
Calibration Checklist
| Asset ID | Equipment | Dept | Area | Parameter | Range | Acceptance | Frequency | Traceability |
|---|---|---|---|---|---|---|---|---|
| CAL-0056 | Temperature/RH Data Logger | Validation | Stability/Storage | Temperature/RH | 0 to 60 °C/%RH | ±0.5°C/±3%RH | 365 days | ISO17025 |
Pre-calibration
- [ ] Verify environmental conditions are within specified limits
- [ ] Allow equipment to warm up as per manufacturer guidelines
- [ ] Clean the equipment as per SOP
- [ ] Level the equipment if applicable
- [ ] Confirm reference standards are valid and calibrated
- [ ] Ensure logbook is ready for data entry
As Found Execution
- [ ] Set the required setpoints for calibration
- [ ] Perform multiple repeats for accuracy
- [ ] Record all raw data accurately
- [ ] Perform necessary calculations for results
OOT Immediate Actions
- [ ] Tag out the equipment
- [ ] Stop use of the equipment immediately
- [ ] Notify the appropriate personnel
- [ ] Document deviation in the logbook
- [ ] Trigger impact review as per SOP
As Left Verification
- [ ] Conduct post-adjustment checks
- [ ] Confirm all results are within acceptance limits
Documentation & Closure
- [ ] Update equipment label with new calibration date
- [ ] Attach calibration certificate to documentation
- [ ] Obtain reviewer sign-off
- [ ] Update next due date in logbook
Deviation Impact Assessment
| Event Summary |
|---|
|
Asset: Temperature/RH Data Logger Date: [TBD] Discovered By: [TBD] Document References: [TBD] As Found Results Summary: [TBD] |
Immediate Containment Checklist
- Stop use of the asset
- Label/tag the asset as out of tolerance
- Segregate impacted work
Define Impact Window
Last known in-tolerance date: [TBD] to discovery date: [TBD]
Product/Batch/Test Impact Review
| Batch/Study/Report ID | Results Potentially Impacted | Re-test/Re-weigh Decision |
|---|---|---|
| [TBD] | [TBD] | [TBD] |
Data Integrity Review Prompts (ALCOA+)
- Attributable: [TBD]
- Legible: [TBD]
- Contemporaneous: [TBD]
- Original: [TBD]
- Accurate: [TBD]
- Complete: [TBD]
- Consistent: [TBD]
- Enduring: [TBD]
- Available: [TBD]
Risk Assessment
| Severity | Probability | Detectability | Rationale | Overall Risk |
|---|---|---|---|---|
| [TBD] | [TBD] | [TBD] | [TBD] | [TBD] |
Investigation Summary
Root Cause: [TBD]
CAPA Plan
| Action | Owner | Due Date | Effectiveness Check |
|---|---|---|---|
| [TBD] | [TBD] | [TBD] | [TBD] |
QA Disposition
Release/Hold/Retest/Recall Triggers: [TBD]
Calibration Certificate
Certificate No: [TBD]
Date: [TBD]
Next Due Date: [TBD]
Asset Details
| Asset ID | Equipment Name | Equipment Type | Dept | Area | Make/Model | Serial No | Location | Parameter/Range/Units | Acceptance Criteria |
|---|---|---|---|---|---|---|---|---|---|
| CAL-0056 | Temperature/RH Data Logger | TempRH Data Logger | Validation | Stability/Storage | [TBD] | [TBD] | [TBD] | Temperature: 0-60 °C, RH: 0-100 %RH | ±0.5°C/±3%RH |
Conditions
Environment: [TBD]
Reference Standards
| Standard Name/ID | Certificate No | Due Date | Traceability Statement |
|---|---|---|---|
| [TBD] | [TBD] | [TBD] | Traceability to ISO17025 |
Results
As Found Results
| Test Point | Measured Value | Acceptance Criteria | Pass/Fail |
|---|---|---|---|
| Point 1 | [TBD] | ±0.5°C/±3%RH | [TBD] |
| Point 2 | [TBD] | ±0.5°C/±3%RH | [TBD] |
| Point 3 | [TBD] | ±0.5°C/±3%RH | [TBD] |
| Point 4 | [TBD] | ±0.5°C/±3%RH | [TBD] |
| Point 5 | [TBD] | ±0.5°C/±3%RH | [TBD] |
As Left Results
| Test Point | Measured Value | Acceptance Criteria | Pass/Fail |
|---|---|---|---|
| Point 1 | [TBD] | ±0.5°C/±3%RH | [TBD] |
| Point 2 | [TBD] | ±0.5°C/±3%RH | [TBD] |
| Point 3 | [TBD] | ±0.5°C/±3%RH | [TBD] |
| Point 4 | [TBD] | ±0.5°C/±3%RH | [TBD] |
| Point 5 | [TBD] | ±0.5°C/±3%RH | [TBD] |
Measurement Uncertainty Statement: [TBD]
Statement of Conformity: [TBD]
Remarks: [TBD]
Signatures
Performed by: ______________________ Date: __________
Checked by: ______________________ Date: __________
Approved by: ______________________ Date: __________