Calibration Documents: Torque Tester (Cap torque) (CAL-0265)
| Asset ID | CAL-0265 |
|---|---|
| Equipment | Torque Tester (Cap torque) |
| Equipment Type | Torque Tester |
| Department | Packaging |
| Area/Location | Packing Line |
| Process Use | Cap torque |
| Parameter | Torque |
| Units | Ncm |
| Range Min | 0 |
| Range Max | 300 |
| Acceptance Criteria | ±2% |
| Calibration Frequency (days) | 180 |
| Calibration Type | External |
| Traceability | ISO17025 |
Table of Contents
Calibration SOP
1. Purpose
The calibration of the Torque Tester (Cap torque) is performed to ensure accurate measurement of torque applied to caps during packaging. This is critical for maintaining batch quality, ensuring patient safety, and preserving data integrity. Inaccurate torque measurements can lead to improper sealing, resulting in compromised product quality and potential risks to patient safety.
2. Scope
- Asset ID: CAL-0265
- Equipment Name/Type: Torque Tester (Cap torque)
- Department/Area: Packaging
- Process Use: Cap torque
- Parameter: Torque
- Range: 0 – 300 Ncm
- Units: Ncm
- Calibration Frequency: 180 days
3. Definitions
- Calibration: The process of comparing a measurement instrument to a standard to ensure accuracy.
- Verification: The process of confirming that a measurement instrument is functioning correctly.
- Adjustment: The act of correcting a measurement instrument to ensure it meets specified standards.
- As Found: The condition of the instrument prior to any adjustments made during calibration.
- As Left: The condition of the instrument after adjustments have been made and verified.
- OOT (Out of Tolerance): A condition where the measurement falls outside the specified acceptance criteria.
- Drift: The gradual change in a measurement instrument’s performance over time.
- Traceability: The ability to trace the measurement results back to national or international standards.
- Measurement Uncertainty: A parameter that characterizes the range of values within which the true value of a measurement lies.
- Intermediate Check: A periodic verification of the instrument’s performance between calibrations.
4. References
- Company Metrology SOP [TBD]
- ISO/IEC 17025 traceability via calibration certificates
- ALCOA+ principles
5. Responsibilities
- Technician/Metrology: Perform calibration and document results.
- QC/QA Reviewer: Review calibration documentation and results.
- Owner Department: Ensure equipment is available and maintained.
- External Service Provider: Conduct external calibration as per schedule.
6. Safety & Preconditions
- Ensure environmental controls are in place (temperature, humidity).
- Allow equipment to warm-up and stabilize as per manufacturer guidelines.
- Maintain cleanliness of the equipment and surrounding area.
- Check that the equipment is level and stable.
- Verify power supply is stable and within specifications.
7. Standards & Tools
- Reference standards required: [TBD]
- Traceability statement: ISO17025
- Certificate validity check: [TBD]
- Standard ID placeholders: [TBD]
8. Calibration Strategy
Calibration will follow the As Found and As Left logic. The minimum number of test points will be 5, including endpoints and midpoints, across the specified range of 0 to 300 Ncm. Each point will have 3 replicate readings to ensure accuracy and reliability.
9. Step-by-Step Procedure
- Preparation: Gather all necessary tools and reference standards.
- Stabilization: Allow the Torque Tester to stabilize in the operating environment.
- As Found checks and data capture: Measure torque at defined points and record raw observations contemporaneously.
- Adjustment rules: If OOT is detected, follow adjustment protocols; if not allowed, escalate to QA.
- As Left verification: Re-measure to confirm adjustments are within specifications.
- Labeling + documentation completion: Complete all required documentation and label the equipment with calibration details.
10. Acceptance Criteria & Calculations
Acceptance criteria: ±2% of the specified torque value. The error calculation formula is:
Error (%) = [(Observed Value – True Value) / True Value] x 100
Pass/Fail logic: If the error is within ±2%, the result is a pass; otherwise, it is a fail.
Rounding rule: [TBD]
11. Handling OOT / Failures
- Immediate containment: Tag the equipment and stop its use.
- Notification & deviation initiation: Notify relevant personnel and initiate a deviation report.
- Impact assessment trigger: Assess the impact on results/batches produced since the last calibration.
- Recalibration after repair/adjustment: Ensure recalibration is performed post-adjustment.
- QA disposition placeholders: [TBD]
12. Records & Retention
- Logbook entries: Document all calibration activities in the logbook.
- Calibration report/certificate filing: Maintain records of calibration certificates.
- e-Records if applicable: [TBD]
- Label content requirements: Include calibrated date, due date, technician name, and status.
13. Annexures
Annexure A: Calibration Data Sheet
| Test Point | Reference | Observed | Error | Acceptance | Pass/Fail | As Found | As Left |
|---|---|---|---|---|---|---|---|
| [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] |
Annexure B: Quick Checklist
- Environmental controls verified
- Equipment warm-up completed
- Cleanliness checked
- Leveling confirmed
- Power supply stable
Annexure C: Example Label Text
Calibration Date: [TBD]
Due Date: [TBD]
Calibrated By: [TBD]
Status: [TBD]
Calibration Protocol
Calibration Protocol
| Protocol ID | Asset ID | Equipment Name | Equipment Type | Dept | Area | Process Use | Parameter | Units | Range | Acceptance Criteria | Frequency | Calibration Type | Traceability |
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| [TBD] | CAL-0265 | Torque Tester (Cap torque) | Torque Tester | Packaging | Packing Line | Cap torque | Torque | Ncm | 0 – 300 | ±2% | 180 days | External | ISO17025 |
Environmental Conditions
Temperature: [TBD]
Relative Humidity: [TBD]
Vibration: [TBD]
Drafts: [TBD]
Test Point Selection
- Minimum: 0 Ncm
- 25% of Range: 75 Ncm
- 50% of Range: 150 Ncm
- 75% of Range: 225 Ncm
- Maximum: 300 Ncm
Results Table
| Test Point | Ref Value | Observed (As Found) | Error | Limit | Pass/Fail | Observed (As Left) | Error | Pass/Fail |
|---|---|---|---|---|---|---|---|---|
| 0 Ncm | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] |
| 75 Ncm | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] |
| 150 Ncm | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] |
| 225 Ncm | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] |
| 300 Ncm | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] | [TBD] |
Stepwise Instructions
- Preparation: Ensure the Torque Tester is clean and free from any obstructions.
- Stabilization: Allow the equipment to stabilize for [TBD] before taking measurements.
- Performing Readings: Take three replicate measurements at each test point.
- Calculating Error: Calculate the error for each observed value against the reference value.
- Evaluation: Compare the calculated error to the acceptance criteria of ±2%.
- Documentation: Record all observations in the results table and ensure data integrity (ALCOA+).
Deviation Instruction for OOT
If any observed value falls outside the acceptance criteria:
- Stop use of the Torque Tester immediately.
- Notify the supervisor and quality assurance personnel.
- Open a deviation report.
- Quarantine impacted results window: [TBD]
- Await QA decision on further actions.
Approvals
| Prepared | Reviewed | Approved |
|---|---|---|
| [Signature/Date] | [Signature/Date] | [Signature/Date] |
Calibration Checklist
| Asset ID | Equipment | Dept | Area | Parameter | Range | Acceptance | Frequency | Traceability |
|---|---|---|---|---|---|---|---|---|
| CAL-0265 | Torque Tester (Cap torque) | Packaging | Packing Line | Torque | 0 – 300 Ncm | ±2% | 180 days | ISO17025 |
Pre-calibration
- [ ] Verify environmental conditions are within specified limits
- [ ] Ensure equipment has warmed up adequately
- [ ] Clean the equipment as per SOP
- [ ] Level the equipment if necessary
- [ ] Check validity of reference standards
- [ ] Ensure calibration logbook is ready for entries
As Found Execution
- [ ] Set the torque tester to specified setpoints
- [ ] Perform repeat measurements as required
- [ ] Record all raw data accurately
- [ ] Perform necessary calculations for results
OOT Immediate Actions
- [ ] Tag out the equipment if out of tolerance
- [ ] Stop use of the equipment immediately
- [ ] Notify relevant personnel
- [ ] Document deviation and root cause analysis
- [ ] Trigger impact review process
As Left Verification
- [ ] Conduct post-adjustment checks
- [ ] Confirm all measurements are within acceptance limits
Documentation & Closure
- [ ] Update equipment label with new calibration date
- [ ] Attach calibration certificate to logbook
- [ ] Obtain reviewer sign-off
- [ ] Update next due date in the logbook
Data Integrity (ALCOA+): Ensure all data recorded is Attributable, Legible, Contemporaneous, Original, and Accurate.
Deviation Impact Assessment
| Event Summary |
|---|
|
Asset: Torque Tester (Cap torque) Date: [TBD] Discovered By: [TBD] Document References: [TBD] As Found Results Summary: [TBD] |
Immediate Containment Checklist
- Stop use of the equipment
- Label/tag the equipment as out of service
- Segregate impacted work
Define Impact Window
Last Known In-Tolerance Date: [TBD] to Discovery Date: [TBD]
Product/Batch/Test Impact Review
| Batch/Study/Report ID | Results Potentially Impacted | Re-test/Re-weigh Decision |
|---|---|---|
| [TBD] | [TBD] | [TBD] |
Data Integrity Review Prompts (ALCOA+)
- Attributable: [TBD]
- Legible: [TBD]
- Contemporaneous: [TBD]
- Original: [TBD]
- Accurate: [TBD]
- Complete: [TBD]
- Consistent: [TBD]
- Enduring: [TBD]
- Available: [TBD]
Risk Assessment
| Severity | Probability | Detectability | Rationale | Overall Risk |
|---|---|---|---|---|
| [TBD] | [TBD] | [TBD] | [TBD] | [TBD] |
Investigation Summary
Root Cause: [TBD]
CAPA Plan
| Action | Owner | Due Date | Effectiveness Check |
|---|---|---|---|
| [TBD] | [TBD] | [TBD] | [TBD] |
QA Disposition
Release/Hold/Retest/Recall Triggers: [TBD]
Calibration Certificate
Certificate No: [TBD]
Date: [TBD]
Next Due Date: [TBD]
Asset Details
| Asset ID | Equipment Name | Equipment Type | Dept | Area | Make/Model | Serial No | Location | Parameter/Range/Units | Acceptance Criteria |
|---|---|---|---|---|---|---|---|---|---|
| CAL-0265 | Torque Tester (Cap torque) | Torque Tester | Packaging | Packing Line | [TBD] | [TBD] | [TBD] | Torque: 0 – 300 Ncm | ±2% |
Conditions
Environment: [TBD]
Reference Standards
| Standard Name/ID | Certificate No | Due Date | Traceability Statement |
|---|---|---|---|
| [TBD] | [TBD] | [TBD] | Traceability to ISO17025 |
Results
As Found Results
| Test Point | Measurement (Ncm) | Pass/Fail |
|---|---|---|
| 1 | [TBD] | [TBD] |
| 2 | [TBD] | [TBD] |
| 3 | [TBD] | [TBD] |
| 4 | [TBD] | [TBD] |
| 5 | [TBD] | [TBD] |
As Left Results
| Test Point | Measurement (Ncm) | Pass/Fail |
|---|---|---|
| 1 | [TBD] | [TBD] |
| 2 | [TBD] | [TBD] |
| 3 | [TBD] | [TBD] |
| 4 | [TBD] | [TBD] |
| 5 | [TBD] | [TBD] |
Measurement Uncertainty Statement: [TBD]
Statement of Conformity: [Pass/Fail overall]
Remarks: [TBD]
Signatures
Performed by: ____________________ Date: __________
Checked by: ______________________ Date: __________
Approved by: _____________________ Date: __________