Water Bath Temp Sensor (CAL-0158) – Calibration Documents

Calibration Documents: Water Bath Temp Sensor (CAL-0158)

Asset ID CAL-0158
Equipment Water Bath Temp Sensor
Equipment Type Water Bath
Department QC
Area/Location Chemistry Lab
Process Use Heating
Parameter Temperature
Units °C
Range Min 20
Range Max 100
Acceptance Criteria ±0.5°C
Calibration Frequency (days) 180
Calibration Type External
Traceability ISO17025

Calibration SOP

1. Purpose

The calibration of the Water Bath Temp Sensor is performed to ensure accurate temperature readings during the heating process in the Chemistry Lab. This calibration controls risks related to accuracy, patient safety, batch quality, and data integrity, thereby supporting compliance with Good Manufacturing Practices (GMP).

2. Scope

  • Asset ID: CAL-0158
  • Equipment Name/Type: Water Bath Temp Sensor / Water Bath
  • Department/Area: QC / Chemistry Lab
  • Process Use: Heating
  • Parameter: Temperature
  • Range: 20°C to 100°C
  • Units: °C
  • Calibration Frequency: Every 180 days

3. Definitions

  • Calibration: The process of configuring an instrument to provide a result for a sample within an acceptable range.
  • Verification: The process of confirming that an instrument is functioning correctly.
  • Adjustment: The act of correcting an instrument to ensure it provides accurate readings.
  • As Found: The condition of the instrument prior to any adjustments.
  • As Left: The condition of the instrument after adjustments have been made.
  • OOT (Out of Tolerance): A condition where the instrument’s readings fall outside the specified acceptance criteria.
  • Drift: The gradual change in an instrument’s output over time.
  • Traceability: The ability to verify the history, location, or application of an item by means of documented recorded identification.
  • Measurement Uncertainty: A parameter associated with the result of a measurement that characterizes the range of values within which the true value is asserted to lie.
  • Intermediate Check: A periodic assessment of an instrument’s performance between calibrations.

4. References

  • Company Metrology SOP [TBD]
  • ISO/IEC 17025 traceability via calibration certificates
  • ALCOA+ principles

5. Responsibilities

  • Technician/Metrology: Perform calibration and maintain records.
  • QC/QA Reviewer: Review calibration results and documentation.
  • Owner Department: Ensure equipment is available and in good condition for calibration.
  • External Service Provider: Conduct external calibration as per schedule.

6. Safety & Preconditions

  • Ensure environmental controls are in place (temperature, humidity).
  • Allow equipment to warm up and stabilize before calibration.
  • Maintain cleanliness of the equipment and surrounding area.
  • Check that the equipment is level and stable.
  • Verify power supply is functioning correctly.
See also  Flow Hood / Airflow Capture Hood (CAL-0251) - Calibration Documents

7. Standards & Tools

  • Reference standards required: [TBD]
  • Traceability statement: ISO17025
  • Certificate validity check: [TBD]
  • Standard ID placeholders: [TBD]

8. Calibration Strategy

The calibration will follow the As Found and As Left logic:

  • Minimum number of points: At least 5 points including endpoints (20°C and 100°C) and midpoints (e.g., 40°C, 60°C, 80°C).
  • Replicate readings: Perform 3 repeats at each point as a best practice.

9. Step-by-Step Procedure

  1. Preparation: Gather all necessary tools and standards.
  2. Stabilization: Allow the water bath to reach the desired temperature range.
  3. As Found checks and data capture: Record initial temperature readings at specified points.
  4. Adjustment rules: If OOT is detected, follow escalation procedures.
  5. As Left verification: Confirm that adjustments bring the instrument within acceptance criteria.
  6. Labeling + documentation completion: Ensure all records are filled out accurately and contemporaneously (ALCOA+).

10. Acceptance Criteria & Calculations

Acceptance criteria: ±0.5°C

Error calculation formula: Error = Observed Value – True Value

Pass/Fail logic: If the absolute value of the error is within ±0.5°C, the calibration passes; otherwise, it fails.

Rounding rule placeholder: [TBD]

11. Handling OOT / Failures

  • Immediate containment: Tag equipment and stop use.
  • Notification & deviation initiation: Inform relevant personnel.
  • Impact assessment trigger: Define impact window and review of affected results/batches.
  • Recalibration after repair/adjustment: Schedule recalibration as necessary.
  • QA disposition placeholders: [TBD]

12. Records & Retention

  • Logbook entries: Document all calibration activities.
  • Calibration report/certificate filing: Maintain records as per company policy.
  • e-records if applicable: [TBD]
  • Label content requirements: Calibrated date, due date, by whom, status.

13. Annexures

Annexure A: Calibration Data Sheet

Test Points Reference Observed Error Acceptance Pass/Fail As Found As Left
[TBD] [TBD] [TBD] [TBD] [TBD] [TBD] [TBD] [TBD]

Annexure B: Quick Checklist

  • Environmental controls checked
  • Equipment warm-up completed
  • Cleanliness verified
  • Leveling confirmed
  • Power supply checked

Annexure C: Example Label Text

Calibration Date: [TBD]

Due Date: [TBD]

Calibrated By: [TBD]

Status: [TBD]

⬆ Back to TOC


Calibration Protocol

Calibration Protocol

Protocol ID Asset ID Equipment Name Equipment Type Dept Area Process Use Parameter Units Range Acceptance Criteria Frequency Calibration Type Traceability
[TBD] CAL-0158 Water Bath Temp Sensor Water Bath QC Chemistry Lab Heating Temperature °C 20 – 100 ±0.5°C 180 days External ISO17025
See also  O2 Analyzer (Gas line) (CAL-0100) - Calibration Documents

Environmental Conditions

Temperature: [TBD] °C

Relative Humidity: [TBD] %

Vibration: [TBD]

Drafts: [TBD]

Test Point Selection

  • Minimum: 20°C
  • 25%: 35°C
  • 50%: 60°C
  • 75%: 85°C
  • Maximum: 100°C

Results Table

Test Point Ref Value Observed (As Found) Error Limit Pass/Fail Observed (As Left) Error Pass/Fail
20°C [TBD] [TBD] [TBD] ±0.5°C [TBD] [TBD] [TBD] [TBD]
35°C [TBD] [TBD] [TBD] ±0.5°C [TBD] [TBD] [TBD] [TBD]
60°C [TBD] [TBD] [TBD] ±0.5°C [TBD] [TBD] [TBD] [TBD]
85°C [TBD] [TBD] [TBD] ±0.5°C [TBD] [TBD] [TBD] [TBD]
100°C [TBD] [TBD] [TBD] ±0.5°C [TBD] [TBD] [TBD] [TBD]

Stepwise Instructions

  1. Preparation: Ensure the water bath is clean and properly set up in the designated area.
  2. Stabilization: Allow the water bath to reach the desired temperature range for at least 30 minutes.
  3. Performing Readings: Take three replicate readings at each test point to ensure accuracy.
  4. Calculating Error: Calculate the error for each observed value against the reference value.
  5. Evaluation: Compare the calculated errors to the acceptance criteria of ±0.5°C.
  6. Documentation: Record all observations in the results table and ensure data integrity (ALCOA+).

Deviation Instruction for OOT

If any observed value is outside the acceptance criteria:

  • Stop use of the equipment immediately.
  • Notify the supervisor and QA.
  • Open a deviation report.
  • Quarantine impacted results window: [TBD]
  • Await QA decision on further actions.

Approvals

Prepared Reviewed Approved
[Signature/Date] [Signature/Date] [Signature/Date]

⬆ Back to TOC


Calibration Checklist

Asset ID Equipment Dept Area Parameter Range Acceptance Frequency Traceability
CAL-0158 Water Bath Temp Sensor QC Chemistry Lab Temperature 20 – 100 °C ±0.5°C 180 days ISO17025

Pre-calibration

  • [ ] Verify environmental conditions are within specified limits
  • [ ] Ensure water bath is warmed up to operating temperature
  • [ ] Clean the equipment as per SOP
  • [ ] Level the water bath if necessary
  • [ ] Check validity of reference standards
  • [ ] Ensure calibration logbook is ready for entries

As Found Execution

  • [ ] Set temperature setpoints as per calibration protocol
  • [ ] Perform multiple repeats for accuracy
  • [ ] Record raw data accurately
  • [ ] Perform necessary calculations for results

OOT Immediate Actions

  • [ ] Tag out the equipment if out of tolerance (OOT)
  • [ ] Stop use of the equipment immediately
  • [ ] Notify the supervisor and QA
  • [ ] Document deviation in the logbook
  • [ ] Trigger impact review as per SOP
See also  Dissolved Oxygen Sensor (Bioreactor) (CAL-0098) - Calibration Documents

As Left Verification

  • [ ] Perform post-adjustment checks to confirm accuracy
  • [ ] Ensure all readings are within acceptance limits

Documentation & Closure

  • [ ] Update equipment label with new calibration date
  • [ ] Attach calibration certificate to logbook
  • [ ] Obtain reviewer sign-off on calibration documentation
  • [ ] Update next due date for calibration

Data Integrity (ALCOA+): Ensure all data is Attributable, Legible, Contemporaneous, Original, Accurate, and Complete.

⬆ Back to TOC


Deviation Impact Assessment

Event Summary
Asset: Water Bath Temp Sensor
Date: [TBD]
Discovered By: [TBD]
Document References: [TBD]
As Found Results Summary: [TBD]

Immediate Containment Checklist

  • Stop use of the equipment
  • Label/Tag the equipment as out of service
  • Segregate impacted work

Define Impact Window

Last known in-tolerance date: [TBD] to discovery date: [TBD]

Product/Batch/Test Impact Review

Batch/Study/Report ID Results Potentially Impacted Re-test/Re-weigh Decision
[TBD] [TBD] [TBD]

Data Integrity Review Prompts (ALCOA+)

  • Attributable: [TBD]
  • Legible: [TBD]
  • Contemporaneous: [TBD]
  • Original: [TBD]
  • Accurate: [TBD]
  • Complete: [TBD]
  • Consistent: [TBD]
  • Enduring: [TBD]
  • Available: [TBD]

Risk Assessment

Severity Probability Detectability Rationale Overall Risk
[TBD] [TBD] [TBD] [TBD] [TBD]

Investigation Summary

Root Cause: [TBD]

CAPA Plan

Action Owner Due Date Effectiveness Check
[TBD] [TBD] [TBD] [TBD]

QA Disposition

Release/Hold/Retest/Recall Triggers: [TBD]

⬆ Back to TOC


Calibration Certificate

Certificate No: [TBD]

Date: [TBD]

Next Due Date: [TBD]

Asset Details

Asset ID Equipment Name Equipment Type Dept Area Make/Model Serial No Location Parameter/Range/Units Acceptance Criteria
CAL-0158 Water Bath Temp Sensor Water Bath QC Chemistry Lab [TBD] [TBD] [TBD] Temperature/20-100/°C ±0.5°C

Conditions

Environment: [TBD]

Reference Standards

Standard Name/ID Certificate No Due Date Traceability Statement
[TBD] [TBD] [TBD] ISO17025

Results

As Found Results

Test Point Measured Value (°C) Acceptance Criteria
1 [TBD] ±0.5°C
2 [TBD] ±0.5°C
3 [TBD] ±0.5°C
4 [TBD] ±0.5°C
5 [TBD] ±0.5°C

As Left Results

Test Point Measured Value (°C) Acceptance Criteria
1 [TBD] ±0.5°C
2 [TBD] ±0.5°C
3 [TBD] ±0.5°C
4 [TBD] ±0.5°C
5 [TBD] ±0.5°C

Measurement Uncertainty Statement: [TBD]

Statement of Conformity: [Pass/Fail overall]

Remarks: [TBD]

Signatures

Performed by: ______________________ Date: ___________

Checked by: ______________________ Date: ___________

Approved by: ______________________ Date: ___________

⬆ Back to TOC