| URS ID | URS Statement | Risk (H/M/L) | DQ Test Ref | IQ Test Ref | OQ Test Ref | PQ Test Ref | Evidence/Record | Result |
|---|---|---|---|---|---|---|---|---|
| URS-001 | The Fluid Bed Dryer must maintain the specified inlet temperature within ±2°C. | H | DQ-001 | IQ-001 | OQ-001 | PQ-001 | Temperature Log | Pass/Fail |
| URS-002 | The Fluid Bed Dryer must maintain the specified outlet temperature within ±2°C. | H | DQ-002 | IQ-002 | OQ-002 | PQ-002 | Temperature Log | Pass/Fail |
| URS-003 | The Fluid Bed Dryer must provide adequate airflow to ensure proper drying. | M | DQ-003 | IQ-003 | OQ-003 | PQ-003 | Airflow Measurement Report | Pass/Fail |
| URS-004 | The Fluid Bed Dryer must operate within the specified drying time. | M | DQ-004 | IQ-004 | OQ-004 | PQ-004 | Drying Time Log | Pass/Fail |
| URS-005 | The Fluid Bed Dryer must monitor and control filter differential pressure (DP). | M | DQ-005 | IQ-005 | OQ-005 | PQ-005 | DP Monitoring Report | Pass/Fail |
| URS-006 | The Fluid Bed Dryer must be controlled by a PLC system for automation. | L | DQ-006 | IQ-006 | OQ-006 | PQ-006 | PLC Configuration Document | Pass/Fail |