Inline Conductivity Meter – OQ Protocol

Document Control

Document Number: OQ-ICM-001

Version: 1.0

Effective Date: [Insert Date]

Prepared By: [Insert Name]

Approved By: [Insert Name]

Operational Qualification Protocol for Inline Conductivity Meter

Objective: To validate the operational performance of the Inline Conductivity Meter used in the production of IV Infusions to ensure it meets specified requirements.

Scope: This protocol applies to the Inline Conductivity Meter utilized in the production area for monitoring solution conductivity of LVP/SVP bags and bottles.

Responsibilities:

  • Validation Team: Execute the protocol and document results.
  • Quality Assurance: Review and approve the OQ protocol and results.
  • Production Personnel: Assist in the execution of the tests as needed.

Prerequisites:

  • Completion of Installation Qualification (IQ).
  • Training of personnel on the use of the Inline Conductivity Meter.
  • Availability of all necessary equipment and materials for testing.

Equipment Description:

The Inline Conductivity Meter is a system used to monitor the conductivity of solutions during the production of IV infusions. It provides real-time data essential for ensuring product quality and compliance.

Test ID Procedure Acceptance Criteria Evidence
OQ-001 Verify accuracy of conductivity measurement at specified points. ±5% of target conductivity values. Calibration records, measurement logs.
OQ-002 Test alarm functionality for out-of-range conductivity. Alarm triggers within 5 seconds of out-of-range condition. Alarm test logs.
OQ-003 Audit trail verification for data integrity. Complete and accurate audit trail for all measurements. Audit trail report.
See also  LAF Workstation / Grade A Bench – Equipment Validation SOP

Detailed Test Cases:

Test Case OQ-001

Objective: Verify accuracy of conductivity measurement.

Procedure: Measure conductivity at three predetermined points using standard solutions.

Acceptance Criteria: Measurements must be within ±5% of target values.

Evidence: Documented calibration records and measurement logs.

Test Case OQ-002

Objective: Validate alarm functionality.

Procedure: Simulate out-of-range conductivity conditions and observe alarm response.

Acceptance Criteria: Alarm must trigger within 5 seconds.

Evidence: Logs of alarm tests conducted.

Test Case OQ-003

Objective: Ensure data integrity through audit trails.

Procedure: Review audit trail for completeness and accuracy of all measurements.

Acceptance Criteria: Audit trail must reflect all data entries accurately.

Evidence: Audit trail report.

Deviations:

Any deviations from the acceptance criteria must be documented and assessed for impact on product quality. A corrective action plan should be developed if necessary.

Approvals:

All results must be reviewed and approved by the Quality Assurance team to ensure compliance with regulatory standards.