Document Control
Document Number: IQ-WFI-001
Version: 1.0
Effective Date: [Insert Date]
Prepared by: [Insert Name]
Approved by: [Insert Name]
Installation Qualification Protocol for Water for Injection System
Meta Description: This document outlines the Installation Qualification (IQ) Protocol for the Water for Injection (WFI) System, ensuring compliance with critical parameters and acceptance criteria.
Objective
The objective of this Installation Qualification (IQ) Protocol is to verify that the Water for Injection (WFI) System is installed according to the design specifications and operational requirements, ensuring it meets the necessary standards for generating WFI.
Scope
This protocol applies to the Installation Qualification of the Water for Injection (WFI) System located in the Utilities area. It covers all aspects of installation, including equipment setup, configuration, and initial performance verification.
Responsibilities
Validation Team: Responsible for executing the IQ protocol and documenting results.
Quality Assurance: Responsible for reviewing and approving the IQ protocol and results.
Maintenance Team: Responsible for ensuring the equipment is maintained and operational prior to IQ execution.
Prerequisites
All installation activities must be completed, and the system should be ready for qualification. Relevant documentation, including User Requirements Specifications (URS), should be reviewed prior to the initiation of the IQ protocol.
Equipment Description
The Water for Injection (WFI) System is designed to generate purified water that meets the standards for pharmaceutical manufacturing. Key features include:
- Conductivity measurement
- Total Organic Carbon (TOC) monitoring
- Microbial control
- Temperature regulation
- Audit trail capabilities
Test Plan
| Test ID | Procedure | Acceptance Criteria | Evidence |
|---|---|---|---|
| IQ-001 | Verify installation against design specifications | All components installed as per specifications | Installation checklist |
| IQ-002 | Check functionality of conductivity meter | Conductivity < 1.3 µS/cm | Calibration certificate |
| IQ-003 | Verify TOC levels | TOC < 500 ppb | TOC measurement report |
| IQ-004 | Microbial testing | No microbial growth detected | Microbial test results |
| IQ-005 | Temperature verification | Temperature maintained at 20-25°C | Temperature log |
| IQ-006 | Audit trail verification | Audit trail functionality confirmed | Audit trail report |
Detailed Test Cases
Test ID: IQ-001
Procedure: Confirm all installation points are completed as per the design specifications.
Acceptance Criteria: All components must be installed correctly.
Evidence: Installation checklist signed off by the validation team.
Test ID: IQ-002
Procedure: Test the conductivity meter with a known standard.
Acceptance Criteria: Conductivity reading must be less than 1.3 µS/cm.
Evidence: Calibration certificate showing the meter is within specification.
Test ID: IQ-003
Procedure: Measure TOC levels using the designated TOC analyzer.
Acceptance Criteria: TOC levels must be below 500 ppb.
Evidence: TOC measurement report from the analyzer.
Test ID: IQ-004
Procedure: Conduct microbial testing on the output water.
Acceptance Criteria: No microbial growth should be detected.
Evidence: Microbial test results report.
Test ID: IQ-005
Procedure: Monitor and record the temperature of the system.
Acceptance Criteria: Temperature must be maintained between 20-25°C.
Evidence: Temperature log documenting the readings.
Test ID: IQ-006
Procedure: Verify that the audit trail is functioning correctly.
Acceptance Criteria: Audit trail must document all system interactions.
Evidence: Audit trail report confirming functionality.
Deviations
Any deviations from the acceptance criteria must be documented and reviewed. A deviation report should be generated for any test that does not meet the specified criteria, outlining the cause and corrective actions taken.
Approvals
All results must be reviewed and approved by the Quality Assurance team before the Water for Injection System is deemed qualified for use.